The WOLF Thrombectomy Device is indicated for the nonsurgical removal of emboli and thrombi from arterial blood vessels in the peripheral vasculature.
Device Story
WOLF Thrombectomy Device is a coaxial system for removing clots from peripheral arterial vessels. It consists of an inner and outer catheter with a nitinol weave/clot-grabbing structure at the distal end of the inner catheter. The device is delivered through a guide catheter (min 0.068" ID) via the femoral artery. Operation involves pulling the inner catheter to retract the weave, which ingests the clot into the outer catheter. A syringe-based aspiration source is used. Radiopaque markers on the catheters and a radiopaque polymer cuff on the weave allow for visualization under fluoroscopy. Used by physicians in clinical settings. The device benefits patients by enabling minimally invasive mechanical clot removal, reducing the need for surgical intervention.
Clinical Evidence
No human clinical data. Evidence includes bench testing (dimensional verification, durability, kink radius, tensile strength, leak testing, clot retrievability, particulate testing, corrosion resistance), biocompatibility (ISO 10993-1), sterilization validation (EO), and package integrity. Animal testing (acute and chronic, 28-day) compared the device to a control, assessing angiographic and histopathological vessel damage. Results showed minimal acute injury, full re-endothelialization by day 28, and equivalent or less endothelial erosion/mural injury compared to the control.
Technological Characteristics
Coaxial system; inner and outer catheters composed of metallic/polymer composites with hydrophilic coating. Distal nitinol wire weave structure. Radiopaque marker bands. Aspiration via syringe. Sterilized by Ethylene Oxide (EO). Compatible with guide catheters/sheaths and guidewires.
Indications for Use
Indicated for nonsurgical removal of emboli and thrombi from arterial blood vessels in the peripheral vasculature.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Inari Medical Infusion Aspiration Catheter System, also known as the FlowTriever (K143563)
Reference Devices
Penumbra Embolectomy Aspiration System (INDIGO Aspiration System) (K142870)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 27, 2019
880 Medical, LLC % Ms. Roberta Hines Regulatory Consultant Northwest Clinical Research Group, Inc. 19836 NE 125th Place Woodinville, WA 98077
Re: K182835
Trade/Device Name: WOLF Thrombectomy Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: DXE Dated: February 19, 2019 Received: February 22, 2019
Dear Ms. Hines:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed Eleni by Eleni Whatley Date: 2019.03.27 Whatley 13:17:55 -04'00' For
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182835
Device Name WOLF Thrombectomy Device
Indications for Use (Describe)
The WOLF Thrombectorny Device is indicated for the nonsurgical removal of emboli and thrombi from arterial blood vessels in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 201 Subpart D)</span> | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 201 Subpart C)</span> |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for 880 Medical LLC. The logo consists of a circular graphic on the left and the text "880 Medical LLC" on the right. The circular graphic is a spiral design with a dark red center and lighter red lines spiraling outward.
# 510(k) SUMMARY 880 Medical, LLC WOLF Thrombectomy Device
# K182835
## General Company Information
| Name: | 880 Medical, LLC |
|------------|-----------------------------------------------|
| Address: | 46714 Fremont Blvd.<br>Fremont, CA 94538, USA |
| Contact: | Roberta Hines, Regulatory Consultant |
| Telephone: | 425-766-0308 |
| Email: | rhines@nwcrq.com |
#### Date Prepared
February 19, 2019
#### General Device Information
Product Name: WOLF Thrombectomy Device Common Name: Thrombectomy Catheter Classification: Catheter, Embolectorny (21 CFR 870.5150, Product Code: DXE)
#### Predicate Devices
| Inari Medical, Inc. | Inari Medical Infusion Aspiration Catheter System, also known<br>as the FlowTriever (K143563) (Primary) |
|---------------------|---------------------------------------------------------------------------------------------------------|
|---------------------|---------------------------------------------------------------------------------------------------------|
## Reference Device
Penumbra, Inc. Penumbra Embolectomy Aspiration System (INDIGO Aspiration System) (K142870)
## Description
The WOLF Thrombectomy Device is comprised of two single lumen variable stiffness catheters (WOLF Inner & Outer) designed for use in removing clot from arterial peripheral vessels. The WOLF inner catheter is attached to the weave which when pulled, inqests the clot into the WOLF outer catheter. The WOLF device is delivered through a quide catheter that has an inner diameter of at least 0.068″. Both WOLF catheters have hydrophilic coating to facilitate tracking and reduce friction during ingestion of the clot.
The WOLF inner catheter has two radiopague marker bands. The distal marker is used to indicate the tip of the catheter during tracking. The proximal marker is used to indicate how far the physician should pull back the GC prior to pulling. The outer catheter has a distal radiopaque marker band to indicate its distal tip during tracking and pulling. The weave also has a radiopaque polymer cuff that allows for visualization of the weave movement.
## Indication for Use
The WOLF Thrombectomy Device is indicated for the nonsurgical removal of emboli and thrombi from arterial blood vessels in the peripheral vasculature.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo for 880 Medical LLC. The logo consists of a circular design on the left and the text "880 Medical LLC" on the right. The circular design features a series of curved lines that radiate outward from the center, creating a spiral effect.
#### Substantial Equivalence
The data presented in this submission demonstrates the technological similarity of the WOLF Thrombectory Device compared with the primary predicate device, the Inari Medical Infusion Aspiration Catheter System.
The devices have the same intended use (peripheral vascular), use the same mechanism of action, incorporate similar components, use similar construction and material, are compatible with a quide catheter/sheath and are packaged and sterilized using the same processes.
Both devices are coaxial systems consisting of an Inner catheter with a metallic/polymer composite, an Outer catheter with a metallic/polymer composite, and a nitinol Weave/Clot Grabbing structure attached to the distal end of the Inner catheter.
In addition to the primary predicate device, a reference device has been included in this submission to support substantial equivalence to a marketed device. The INDIGO Aspiration Catheter and Separator is considered as a reference device because it has the same intended use, same mechanisms of actions and its 5F device is used in similar sized vessels as the WOLF Thrombectomy Device.
Biocompatibility, performance testing, simulated use testing, and animal testing demonstrate that the device has appropriate properties for its intended use.
#### Performance Data
Bench studies indicate that the 880 Medical WOLF Thrombectomy Device performs as intended. The following testing was performed in conformance with design inputs, including performance standards for peripheral vascular embolectomy devices. Testing included dimensional design verification/validation (durability and integrity, kink radius, torsion and tensile strength, air and liquid leak testing, clot retrievability, surface condition, coating integrity, particulate testing, corrosion resistance), sterilization validation, transit and package integrity testing, chronic animal safety testing, biocompatibility testing, simulated use testing, and comparative performance analysis with the predicate device. In addition, comparative testing of the INDIGO Aspiration Catheter was conducted in the simulated use test.
#### Technical Comparison
The technical features of the WOLF Thrombectomy Device and the Inari FlowTriever device are the same or similar for both the design components and the mechanism of action. Both devices are also provided sterile and are sterilized by the same method (EO).
Both the WOLF Thrombectomy and Inari FlowTriever Device have four main design components, highlighting their similarities in design:
- 1. Weave/Clot Grabber – attached to distal end of Inner Catheter
- 2. Inner Catheter
- 3. Outer Catheter
- 4. Aspiration Source
Both devices are made of similar materials and come in similar configurations (shape, diameter and lengths). The Weave/Clot Grabber component in both devices is a tubular weave structure formed from nitinol wire and is attached to the distal end of the Inner Catheter. The Weave/Clot Grabber is the component that integrates into the clot and aids in clot removal. Also, for both devices the Inner and Outer Catheters are coaxial single lumen metallic/polymeric compositic coating. The aspiration source for both the WOLF Thrombectomy Device and the Inari FlowTriever is a syringe-based design.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for 880 Medical LLC. The logo consists of a circular design with swirling lines on the left and the text "880 Medical LLC" on the right. The logo is in a maroon color.
Both the WOLF Thrombectomy Device and Inari FlowTriever Device are delivered through the femoral artery, provide delivery of contrast, and contain radiopague markers for visualization under fluoroscopy. Both devices are used with introducer sheaths/quide catheters and quidewires. The WOLF Thrombectorny Device has been tested for compatibility with the appropriate accessories in preclinical (animal) testing and simulated use design validation testing.
## Biocompatibility Testing
The WOLF Thrombectomy Device was subjected to the following biocompatibility testing per the ISO-10993-1 standard: cytotoxicity, sensitization, irritation/intracutaneous reactivity, acute systemic toxicity, pyrogenicity, hemocompatibility testing (partial thromboplastin time, assay for coagulation and hematology parameters, and hemolysis) and endotoxin testing. The results showed that the WOLF Thrombectomy Device meets biocompatibility requirements of the ISO standard.
## Sterilization Validation
Sterilization validation testing verified with a high degree of assurance that Ethylene Oxide sterilization is effective in achieving sterility of the 880 WOLF Thrombectorny Device at a sterility assurance level of 10°.
## Package Integrity After Aging and Distribution
Packaging was verified to protect the WOLF Thrombectomy Device adequately to ensure product function throughout the claimed shelf life and after exposure to the storage and distribution environment.
#### Animal Testing
The WOLF Thrombectomy Device was subjected to acute and chronic animal study to compare the safety of the WOLF Device to a control device. The study was conducted per the US Food and Drug Administration, 21 CFR, Part 58 Good Laboratory Practice for Nonclinical Laboratory Studies. The purpose of this study was to assess the angiographic damage to the target vessel on Day 28 and histopathological assessment of damage to the target vessel on Day 0 and Dav 28. The objectives, as well as the acceptance criteria for success, were met. No evidence of vessel dissection was seen acutely or through histopathology at the acute and chronic time points. When analyzed histologically, the treatment with the WOLF device resulted in equivalent or less endothelial erosion and mural injury as the control device. Acute injury was minimal and consistent with catheterization procedures. Vessel injuries were found to have healed completely and resulted in full re-endothelialization by the chronic time point (28 days).
## Conclusions
The WOLF Thrombectomy Device and its predicate device have the same intended use and similar technological characteristics. The differences do not raise different questions of safety or effectiveness. Performance testing further demonstrates that the device is substantially equivalent to the predicate for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.