Cope Pediatric Gastrointestinal Suture Anchor Set, Enterostomy Suture Anchor Set
Applicant
Cook Incorporated
Product Code
KGC · Gastroenterology, Urology
Decision Date
Jun 26, 2019
Decision
SESK
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Cope Pediatic Gastrointestinal Suture Anchor Set is intended for anchoring the anterior wall of the abdominal wall prior to the introduction of interventional catheters and can stay in place for up to 14 days in infant, child, and adolescent populations. The Enterostomy Suture Anchor Set is intended for anchoring the wall of a hollow viscus to the abdominal wall prior to the introduction of interventional catheters and stay in place for up to 14 days in child, adolescent, and adult populations.
Device Story
Suture anchor sets used to secure hollow viscera to the abdominal wall. Components include stainless steel anchors (0.025-inch), Tevdek sutures, introducer needles (lancet-tip), backload needles (blunt-tip), wire guides, and (for Enterostomy set) a polyethylene dilator. Operated by physicians in clinical settings. Anchors are deployed into the viscus via the introducer needle; sutures are secured externally (e.g., around gauze rolls). After the procedure, anchors are removed by cutting the suture, allowing passage through the gastrointestinal system. Provides temporary fixation to stabilize the target site for subsequent interventional catheter placement.
Clinical Evidence
Bench testing only. Includes visual inspection, dimensional verification, tensile testing, MR compatibility, radiopacity, and biocompatibility (ISO 10993-1: cytotoxicity, sensitization, reactivity, toxicity, pyrogenicity, genotoxicity, implantation). Sterility and packaging validated via bioburden, endotoxin, EO residual, and simulated distribution testing.
Indicated for anchoring the anterior wall of the stomach (pediatric) or hollow viscus (child, adolescent, adult) to the abdominal wall to facilitate interventional catheter introduction; indwell time up to 14 days.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 26, 2019
Cook Incorporated Rohini Patel. Ph.D. Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K182832
> Trade/Device Name: Cope Pediatric Gastrointestinal Suture Anchor Set, Enterostomy Suture Anchor Set Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KGC Dated: May 28, 2019 Received: May 30, 2019
Dear Rohini Patel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the Cope Pediatric Gastrointestinal Suture Anchor Set and the Enterostomy Suture Anchor Set have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Shani P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182832
#### Device Name
Cope Pediatric Gastrointestinal Suture Anchor Set Enterostomy Suture Anchor Set
#### Indications for Use (Describe)
The Cope Pediatic Gastrointestinal Suture Anchor Set is intended for anchoring the anterior wall of the abdominal wall prior to the introduction of interventional catheters and can stay in place for up to 14 days in infant, child, and adolescent populations.
The Enterostomy Suture Anchor Set is intended for anchoring the wall of a hollow viscus to the abdominal wall prior to the introduction of interventional catheters and stay in place for up to 14 days in child, adolescent, and adult populations.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="True" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, smaller block letters. The red background is a rectangle with a diagonal cut on the bottom right.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA ONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
### 510(k) Summary
# K182832 Cope Pediatric Gastrointestinal Suture Anchor Set Enterostomy Suture Anchor Set 21 CFR §876.5010 Date Prepared: 24 June 2019
## Submitted By:
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | Rohini Patel |
| Email: | RegSubmission@CookMedical.com |
| Contact Phone Number: | (812) 335-3575 x104516 |
| Contact Fax number: | (812) 332-0281 |
### Device Information:
| Trade Name: | Cope Pediatric Gastrointestinal Suture Anchor Set<br>Enterostomy Suture Anchor Set |
|----------------------------|------------------------------------------------------------------------------------|
| Regulation Name: | Gastrointestinal tube and accessories |
| Classification Regulation: | 21 CFR §876.5980, Product Code KGC |
| Device Class: | Class II |
| Classification Panel: | Gastroenterology/Urology |
## Predicate Device:
The predicate device for the Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set is the Modified Cope Suture Anchor (K873606, Cook Incorporated) cleared on June 17, 1988.
## Device Description:
This bundled submission includes two devices: Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set. Both subject devices include two suture anchors that are identical.
The Cope Pediatric Gastrointestinal Suture Anchor Set is available in two set configurations. Both set configurations are supplied with two suture anchors that are identical. The first suture anchor is pre-loaded in the lancet-tip introducer needle. The
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Cook Medical. The word "COOK" is in white, bold letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, bold letters on a red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
second suture anchor may come pre-loaded in the blunt-tip backload needle. A wire guide is supplied if the second suture anchor comes pre-loaded in the backload needle. The suture anchor is made of a 0.025-inch stainless steel anchor and a 30-cm suture. The anchor is either 2 cm or 1.3 cm long. The suture is the 5-0 size Tevdek® and one end of the suture is bonded at the center of the anchor. The introducer needle is made of 19gauge stainless steel cannula in the length of 7 cm. The distal tip of the introducer needle is lancet bevel. Similarly, the backload needle is made of 19-gauge stainless steel cannula and 3 cm long. The distal tip of the backload needle is blunt. The wire guide is made of 0.025-inch stainless steel coil and 80 cm in length. Again, the wire guide is only supplied in the set configuration that the suture anchor is pre-loaded in the introducer needle.
The Enterostomy Suture Anchor Set is supplied with two suture anchors, the lancet-tip introducer needle, the blunt-tip backload needle, the wire guide, and the dilator. The suture anchors are pre-loaded in the backload needle. The suture anchor is made of a 0.025-inch stainless steel anchor and a 30-cm suture. The suture is the 5-0 size Tevdek® and one end of the suture is bonded at the center of the anchor. The introducer needle is made of 18-gauge stainless steel cannula in the length of 7 cm. The distal tip of the introducer needle is lancet bevel. Similarly, the backload needle is made of 18-gauge stainless steel cannula and 5 cm long. The distal tip of the backload needle is blunt. The wire guide is made of 0.035-inch stainless steel coil and 80 cm in length. The wire guide is coated in polytetrafluoroethylene. The dilator is made of 8 French polyethylene tubing and 20 cm in length. The distal end of the dilator is tapered to 0.035 inches in diameter.
Where Chait Enterostomy Suture Anchor Set is mentioned instead of Enterostomy Suture Anchor Set in the submission, the references are made to the identical product. The "Chait" naming convention is used in the internal purposes.
## Indications for Use:
The Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set are used for anchoring the anterior wall of the hollow viscus to the abdominal walls for up to 14 days. The exact verbiage of the intended use can be found in Table 1.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters, stacked on top of the word "MEDICAL" in white, block letters. The words are set against a red background. The logo is simple and modern.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA HONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
| | Intended Use |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cope Pediatric Gastrointestinal<br>Suture Anchor Set | Intended for anchoring the anterior wall of the stomach to the abdominal<br>wall prior to the introduction of interventional catheters and can stay in<br>place for up to 14 days in infant, child, and adolescent populations |
| Enterostomy Suture Anchor Set | Intended for anchoring the wall of a hollow viscus to the abdominal<br>wall prior to the introduction of interventional catheters and stay in<br>place for up to 14 days in child, adolescent, and adult populations |
#### Table 1 Subject Device Intended Use
# Comparison to Predicate Device:
The subject devices, Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set, and the predicate device, the Modified Cope Suture Anchor (K873606), are substantially equivalent in that these devices have similar intended use and technological characteristics and are identical in method of placement.
The anchor material, introducer needle cannula material, wire guide material, number of sutures per anchor, and mode of operation are identical between the subject device Cope Pediatric Gastrointestinal Suture Anchor Set and the predicate device Modified Cope Suture Anchor (K873606). The differences between the subject device and predicate device, including target patient population, suture materials and dimensions, anchor dimensions, number of anchors, inclusion of a backload needle, suture securement method, and indications for use verbiage, do not raise any new questions of safety and/or effectiveness. The substantial equivalence comparison of the subject device, Cope Pediatric Gastrointestinal Suture Anchor Set, and the predicate device, Modified Cope Suture Anchor Set, is provided in Table 2 below.
| | Modified Cope Suture Anchor<br>K873606 | Cope Pediatric Gastrointestinal<br>Suture Anchor Set<br>Subject Device | |
|-------------------|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| | | Pre-Loaded | Un-Loaded |
| Regulation Number | 876.5010 | 876.5980 | |
| Product Code | FGE | KGC | |
| Classification | II | II | |
| Intended Use | Used for retaining hollow viscera<br>prior to introduction of interventional<br>catheters | Intended for anchoring the anterior wall of the<br>stomach to the abdominal wall prior to the<br>introduction of interventional catheters and can<br>stay in place for up to 14 days in infant, child, and<br>adolescent populations | |
| Table 2 Substantial Equivalence Comparison Table - Cope Pediatric Gastrointestinal Suture Anchor Set | | | | |
|------------------------------------------------------------------------------------------------------|--|--|--|--|
{7}------------------------------------------------
MEDICAL
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
| | | Modified Cope Suture Anchor<br>K873606 | | Cope Pediatric Gastrointestinal<br>Suture Anchor Set | | | |
|-----------------------------|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-----------------------------------------------------------------------|--------|---------------------------------------------------------------|--------|
| | | | | Pre-Loaded | | Un-Loaded | |
| Device Picture | | | | | | | |
| | Material | Stainless Steel (T-302 or T-304) | | Stainless Steel (T-304) | | | |
| | Dimensions | Diameter | Length | Diameter | Length | Diameter | Length |
| Anchor | | 0.038 in | 2 cm | 0.025 in | 2 cm | 0.025 in | 1.3 cm |
| | Count | 1 | | 2 (second anchor loaded<br>in backload needle) | | 2 (1 anchor loaded,<br>1 anchor un-loaded) | |
| | Material | Polydek® suture | | Tevdek® suture | | | |
| | USP Size | 4-0 | | 5-0 | | | |
| Suture | Length | 40 cm | | 30 cm | | | |
| | Count | 1 | | Identical (1 per anchor) | | | |
| | Proximal<br>suture needle | Yes | | No | | | |
| | Material,<br>needle | Stainless Steel | | Identical | | | |
| | Material, hub | Polycarbonate | | Polypropylene | | | |
| Introducer<br>needle | Distal tip | Lancet | | Identical | | | |
| | Dimensions | Diameter | Length | Diameter | Length | Diameter | Length |
| | | 18 G | 12 cm | 19 G | 7 cm | 19 G | 7 cm |
| | Material | Not included | | Stainless Steel | | Identical | |
| Backload<br>needle | Distal tip | | | Blunt | | (the second anchor is<br>provided without<br>backload needle) | |
| | Dimensions | | | Diameter | Length | | |
| | | | | 19 G | 3 cm | | |
| | Material | Stainless Steel (T-302 or T-304) | | Stainless Steel (T-304) | | | |
| Wire guide | Distal tip | Flexible | | Flexible | | | |
| | Coating | Not coated | | Not coated | | Not included | |
| | Dimensions | Diameter | Length | Diameter | Length | | |
| | | 0.038 in | 80 cm | 0.025 in | 80 cm | | |
| | Indwell period | 10-14 days | | Up to 14 days | | | |
| Suture securement<br>method | | Secured to skin | | Wrap the ends of the suture around gauze rolls and<br>secure in place | | | |
| Mode of operation | | Loaded anchor is deployed into the<br>stomach as the wire guide pushes it<br>from the proximal end of the<br>introducer needle. After the tract<br>formation, the anchor is removed by<br>cutting the suture and allow their<br>passage via the gastrointestinal<br>system. | | Identical | | | |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in white, block letters on a darker red background.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
| | Modified Cope Suture Anchor<br>K873606 | Cope Pediatric Gastrointestinal<br>Suture Anchor Set<br>Subject Device | |
|--------------------|----------------------------------------|------------------------------------------------------------------------|-----------|
| | | Pre-Loaded | Un-Loaded |
| Sterilization, SAL | ETO, 10-6 | Identical | Identical |
| Packaging | Tyvek peel-open pouch | Identical | Identical |
The anchor material, introducer needle material, wire guide material, number of sutures per anchor, and mode of operation are identical between the subject device Enterostomy Suture Anchor Set and the predicate device Modified Cope Suture Anchor (K873606). The differences between the subject device and the predicate device, including target patient population, suture materials and dimensions, anchor dimensions, number of anchors, introducer needle dimensions, inclusion of a backload needle, wire guide dimensions and coating, inclusion of a dilator, suture securement method, and indications for use verbiage do not raise any new questions of safety and/or effectiveness. The substantial equivalence comparison of the subject device, Enterostomy Suture Anchor Set, and the predicate device, Modified Cope Suture Anchor Set, is provided in Table 3 below.
| | | Modified Cope Suture Anchor<br>K873606 | Enterostomy Suture Anchor Set<br>Subject Device | | |
|--------|-------------------|----------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | Regulation Number | 876.5010 | 876.5980 | | |
| | Product Code | FGE | KGC | | |
| | Classification | II | II | | |
| | Intended Use | | Used for retaining hollow viscera prior to<br>introduction of interventional catheters | Intended for anchoring the wall of a hollow<br>viscus to the abdominal wall prior to the<br>introduction of interventional catheters and<br>stay in place for up to 14 days in child,<br>adolescent, and adult populations | |
| | Device Picture | | Image: Modified Cope Suture Anchor | Image: Enterostomy Suture Anchor Set | |
| Anchor | Material | Stainless Steel (T-302 or T-304) | | Stainless Steel (T-304) | |
| | Dimensions | Diameter | Length | Diameter | Length |
| | | 0.038 in | 2 cm | 0.025 in | 1.3 cm |
| | Count | 1 | | 2 (both anchors loaded in backload needle) | |
Table 3 Substantial Equivalence Comparison - Enterostomy Suture Anchor Set
{9}------------------------------------------------
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA PHONE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
| | | Modified Cope Suture Anchor<br>K873606 | | Enterostomy Suture Anchor Set<br>Subject Device | |
|-----------------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|-----------------------------------------------------------------------|--------|
| | Material | Polydek® suture | | Tevdek® suture | |
| | USP Size | 4-0 | | 5-0 | |
| | Length | 40 cm | | 30 cm | |
| | Count | 1 | | Identical (1 per anchor) | |
| Suture | Proximal<br>suture needle | Yes | | No | |
| | Material | Stainless Steel | | Stainless Steel | |
| | Material, hub | Polycarbonate | | Polycarbonate | |
| | Distal tip | Lancet | | Identical | |
| Introducer<br>needle | Dimensions | Diameter | Length | Diameter | Length |
| | | 18 G | 12 cm | 18 G | 7 cm |
| | Material | Not included | | Stainless Steel | |
| | Distal tip | Not included | | Blunt | |
| Backload<br>needle | Dimensions | Not included | | Diameter | Length |
| | | | | 18 G | 5 cm |
| | Material | Stainless Steel (T-302 or T-304) | | Stainless Steel (T-304) | |
| Wire guide | Distal Tip | Flexible | | Identical | |
| | Coating | Not coated | | Polytetrafluoroethylene | |
| | Dimensions | Diameter | Length | Diameter | Length |
| | | 0.038 in | 80 cm | 0.035 in | 80 cm |
| | Material | Not included | | Polyethylene | |
| | Distal Tip | Not included | | Tapered | |
| Dilator | Dimensions | Not included | | Diameter | Length |
| | | | | 8 Fr | 20 cm |
| | Indwell period | 10-14 days | | Up to 14 days | |
| Suture securement<br>method | | Secured to skin | | Wrap the ends of the suture around gauze<br>rolls and secure in place | |
| Mode of operation | | Loaded anchor is deployed into the<br>stomach with the wire guide from the<br>proximal end of the introducer needle.<br>After the tract formation, the anchor is<br>removed by cutting the suture and allow | | Identical | |
| | | their passage via the gastrointestinal<br>system. | | | |
| | One-time Use | Yes | | Identical | |
| | Sterilization, SAL | ETO, 10-6 | | Identical | |
| Packaging | | Tyvek peel-open pouch | | Identical | |
Image /page/9/Picture/3 description: The image shows the logo for Cook Medical. The word "COOK" is in white, sans-serif, uppercase letters on a red background. Below "COOK" is the word "MEDICAL" in white, sans-serif, uppercase letters on a darker red background.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, with a registered trademark symbol next to it. Below "COOK" is the word "MEDICAL" in white. Both words are set against a red background, with the bottom right corner of the red background cut off at an angle.
COOK INCORPORATED 750 DANIELS WAY BLOOMINGTON, IN 47404 USA NE: 812.339.2235 TOLL FREE: 800.457.4500 WWW.COOKMEDICAL.COM
## Technological Characteristics:
The subject devices, Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set, were subjected to the following applicable testing to assure reliable design and performance under the specified testing parameters:
Bench Testing (including time zero and applicable three year accelerated aged testing)
- Visual Inspection, Compatibility, and Dimensional Verification Testing (Zero -Time and Accelerated Aged)
- Tensile Testing (Time Zero and Accelerated Aged) -
- -MR Testing (Time Zero)
- Radiopacity Testing (Time Zero) -
# Biocompatibility Testing:
Per ISO 10993-1 and FDA guidance, testing for cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity, were performed on the subject device to ensure the biocompatibility of the subject device. Additional testing was performed on the suture anchors: testing for subacute/subchronic toxicity, genotoxicity, and implantation.
# Sterility and Packaging
- Finished Product Qualification Sterility, Bioburden, Endotoxin, and EO -Residual
- Simulated Distribution Testing -
# Conclusion:
The results of these tests confirm that the Cope Pediatric Gastrointestinal Suture Anchor Set and Enterostomy Suture Anchor Set meet the design input requirements based on the intended use and support the conclusion that this device does not raise new questions of safety or effectiveness and is substantially equivalent to the predicate device, the Modified Cope Suture Anchor (K873606, Cook Incorporated).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.