K182831 · Coloplast Corp. · GEX · Dec 10, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182831
Device Name
SabreLine and SabreGuard Laser Fibers
Applicant
Coloplast Corp.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 10, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SabreLine™ and SabreGuard™ Laser Fibers are in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200mm have received regulatory clearance. Coloplast surgical fiber optic laser delivery devices are intended for use with any cleared surgical laser with an SMA 905 connector.
Device Story
Fiber optic laser delivery devices; transmit laser energy to soft and hard tissue in contact or non-contact modes. Device consists of silica inner core, primary/secondary cladding, and polymeric jacket; 3.0 meters long; SMA-905 proximal connector. Available with round, radius, or straight tips. Used in surgical settings; operated by surgeons. Connects to cleared surgical laser systems (500nm-2200nm). Output is laser energy delivered to tissue for surgical procedures. Benefits include precise tissue interaction for various surgical specialties.
Clinical Evidence
No clinical data or animal studies provided. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity) and mechanical verification (tip diameter, bend radius, ureteroscope navigation, energy transmission, connector bond strength).
Indicated for use in all surgical specialties utilizing compatible laser systems with operational wavelengths between 500nm - 2200nm. Compatible with any cleared surgical laser featuring an SMA 905 connector.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Laser Peripherals, LLC family of Bare Laser Fibers (K170366)
Submission Summary (Full Text)
{0}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182831
Device Name SabreLine™ Laser Fibers SabreGuard™ Laser Fibers
### Indications for Use (Describe)
The SabreLine™ and SabreGuard™ Laser Fibers are in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200mm have received regulatory clearance. Coloplast surgical fiber optic laser delivery devices are intended for use with any cleared surgical laser with an SMA 905 connector.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{1}------------------------------------------------
# 510(k) SUMMARY
#### SUBMITTER I.
| 510(K) Owner's Name: | Coloplast A/S |
|----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | Holtedam 1<br>3050 Humlebaek, Denmark |
| Phone/Fax/Email: | Office: 612.597.5106<br>Email: usclr@coloplast.com |
| Name of Contact Person: | Cori Ragan<br>Regulatory Affairs Manager |
| Address/Contact: | 1601 West River Road<br>Minneapolis, MN 55411 |
| Date Prepared: | October 4, 2018 |
| II.<br>DEVICE | |
| Trade or Proprietary Name: | SabreLine™ and SabreGuard™ Laser Fibers |
| Common or Usual Name: | Bare Laser Fibers |
| Classification Name: | Laser Surgical Instrument for Use in General and Plastic<br>Surgery and in Dermatology<br>(21CFR section 878.4810)<br>(Product Code: GEX)<br>Device Class: 2 |
#### III. PREDICATE DEVICE
The SabreLine™ and SabreGuard™ Laser Fibers are substantially equivalent in performance, indication, design and materials to the Laser Peripherals, LLC family of Bare Laser Fibers (K170366).
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
{2}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The Coloplast SabreGuard™ and SabreLine™ Laser Fibers are fiber optic laser delivery devices that transmit laser energy to soft and hard tissue in contact and non-contact mode during surgical laser procedures. The devices are 3.0 meters in length and are terminated with a SMA-905 connector on the proximal end.
A laser fiber is made of silica inner core, with a primary and secondary cladding surrounded by a polymeric jacket. The laser fibers are compatible with laser systems operating in wavelengths between 500 nm - 2200 nm. The laser delivery devices are intended for use with any cleared surgical laser with a SMA-905 connector.
Coloplast reference SUNXXX are equipped with a round tip, Coloplast reference SUEXXX are equipped with a radius tip; Coloplast reference SUDXXXX and RUSXXXX are equipped with a straight tip.
Single use laser fibers are delivered sterile in a retail box of 3, with an instruction for use in paper version. Reusable laser fibers are delivered sterile in a retail box of 1, with an instruction for use in paper version.
#### V. INDICATIONS FOR USE
The SabreLine™ and SabreGuard™ Laser Fibers are indicated for use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance. Coloplast surgical fiber optic laser delivery devices are intended for use with any cleared surgical laser with an SMA 905 connector.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The SabreLine™ and SabreGuard™ Laser Fibers are substantially equivalent in performance, design and materials to the predicate, Laser Peripherals, LLC family of Bare Laser Fibers, cleared under premarket notification number K170366.
The SabreLine™ and SabreGuard™ Laser Fibers are a standard silica core fiber, silica cladding, fluoropolymer coating, and Tefzel buffer. The technological characteristics of the proposed device, which allow use in all surgical specialties in which compatible laser systems with operational wavelengths between 500nm - 2200nm have received regulatory clearance, is substantially equivalent to the predicate device.
{3}------------------------------------------------
#### PERFORMANCE DATA VII.
The following performance data was provided in support of the substantial equivalence determination.
## Biocompatibility Testing
Biocompatibility testing was conducted based upon ISO 10993-1 (2009): Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management processand FDA Guidance for Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process - Guidance for Industry and Food and Drug Administration Staff - June 16, 2016. A comprehensive regimen of testing for the SabreLine™ and SabreGuard™ Laser Fibers included:
- o Cytotoxicity
- o Sensitization
- Irritation ●
- Acute Systemic toxicity ●
- Material-mediated pyrogenicity o
## Mechanical Testing
Mechanical testing was completed as follows to verify the product design performance.
- Tip Diameter
- o Bend Radius
- Deflected Ureteroscope Bend Navigation Test
- Energy Transfer/Transmission Testing ●
- o Spot Check
- Connector Bond Strength ●
## Sterilization
The SabreLine™ and SabreGuard™ Laser Fibers are sterilized using ethylene oxide in a validated cycle demonstrating a sterility assurance level (SAL) of 10-9.
## Packaging and Distribution
The SabreLine™ and SabreGuard™ Laser Fibers were subjected to distribution testing and verification testing to demonstrate that the product and package would be undamaged throughout the product life and maintain the device sterility.
No animal studies or clinical testing were provided to support substantial equivalence between the subject and predicate devices.
# VIII. CONCLUSIONS
The SabreLine™ and SabreGuard™ Laser Fibers have been demonstrated to be substantially equivalent to the predicate, Laser Peripherals, LLC family of Bare Laser Fibers based on the non-clinical data provided, similar intended use, materials, design and construction, biocompatibility, and technological characteristics.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Coloplast Corp Cori Ragan Regulatory Affairs Manager 1601 West River Road North Minneapolis, Minnesota 55411
December 10, 2018
Re: K182831
Trade/Device Name: SabreLine and SabreGuard Laser Fibers Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: Class II Product Code: GEX Dated: October 4, 2018 Received: October 9, 2018
Dear Cori Ragan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{5}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.