← Product Code [KNT](/productcode/KNT) · K182804

# Traditional Length GJ Feeding Device (K182804)

_Applied Medical Technology, Inc. · KNT · Mar 1, 2019 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182804

## Device Facts

- **Applicant:** Applied Medical Technology, Inc.
- **Product Code:** [KNT](/productcode/KNT.md)
- **Decision Date:** Mar 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

Provide simultaneous gastric decompression/drainage and delivery of enteral nutrition into the distal duodenum or proximal jejunum.

## Device Story

Single-use feeding tube for simultaneous gastric decompression and enteral nutrition delivery. Device enters stomach via surgically-created gastric stoma. Features multi-lumen tubing with graduated markings, inflatable internal retention balloon, and sliding external bolster for adjustable gastric length. External tri-port overmold includes separate ports for jejunal feeding, gastric decompression, and balloon inflation. Available in legacy (funnel ports) or ENFit (ISO 80369-3 compliant) configurations. Distal portion of 16Fr/18Fr tubes includes metal spring anti-kink design. Used in clinical settings by healthcare providers. Provides nutrition and decompression; benefits patients unable to tolerate gastric feeding or requiring simultaneous drainage. MR conditional.

## Clinical Evidence

Bench testing only. No clinical or animal studies performed. Biocompatibility testing conducted per ISO 10993 series. Performance testing included balloon integrity (ASTM F2528-06), connector testing (ISO 80369-20), and MRI safety (ASTM F2052, F2213, F2119, F2182).

## Technological Characteristics

Medical grade silicone construction; inflatable silicone internal retention balloon; sliding external bolster; multi-lumen tubing; tri-port adapter. Sterilized via Ethylene Oxide. MR conditional. ENFit models compliant with ISO 80369-3. Anti-kink metal spring in 16Fr/18Fr jejunal segments.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- AMT Low-Profile Transgastric-Jejunal Feeding Tube Kit ([K123716](/device/K123716.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out as "U.S. Food & Drug Administration".

March 1, 2019

Applied Medical Technology, Inc. Joy Tubero Regulatory Affairs Specialist 8006 Katherine Boulevard Brecksville, OH 44141

Re: K182804

> Trade/Device Name: Traditional Length GJ Feeding Device Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: II Product Code: KNT, PIF Dated: January 23, 2019 Received: January 28, 2019

Dear Joy Tubero:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark J. Antonino -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182804

Device Name Traditional Length GJ Feeding Device

### Indications for Use (Describe)

The Traditional Length GJ Feeding Device is indicated for use in adult, adolescent, children, and infants over 10kg who cannot absorb adequate nutrition through the stomach, who have intestinal motility problems, gastric outlet obstruction, severe gastroesophageal reflux, are at risk of aspiration, or in those who have had previous esophagectorny or gastrectomy.

The use of this tube is also clinically indicated when simultaneous gastric decompression and jejunal feeding are needed. This includes patients in whom malnutrition already exists, or may result, secondary to concurrent conditions.

Type of Use (Select one or both, as applicable)

| <div> <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span>Prescription Use (Part 21 CFR 801 Subpart D)    </div> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span>Over-The-Counter Use (21 CFR 801 Subpart C)    </div> </div>                                                                                                                                                                                                       |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

Traditional Length GJ Feeding Device

#### I. SUBMITTER:

Applied Medical Technology, Inc. 8006 Katherine Boulevard Brecksville, OH 44141 Phone: 440-717-4000 Fax: 440-717-4200

Contact Person: Joy Tubero - Regulatory Affairs Specialist Email: joy.tubero@appliedmedical.net Date Prepared: January 23, 2019

#### II. DEVICE INFORMATION:

Trade/Device Name: Traditional Length GJ Feeding Device Common Name: Gastrointestinal Tube Classification Name: Gastrointestinal tubes and accessories. (21 CFR 876.5980) Product Code: KNT, PIF Regulatory Class: II Review Panel: Gastroenterology and Urology

#### III. PREDICATE INFORMATION:

Primary Predicate: AMT Low-Profile Transgastric-Jejunal Feeding Tube Kit (K123716)

- The predicate device has not been subject to design-related recalls.
#### INDICATIONS FOR USE: IV.

The Traditional Length GJ Feeding Device is indicated for use in adult, adolescent, child, and infant patients over 10kg who cannot absorb adequate nutrition through the stomach, who have intestinal motility problems, gastric outlet obstruction, severe gastroesophageal reflux, are at risk of aspiration, or in those who have had previous esophagectomy or gastrectomy.

The use of this tube is also clinically indicated when simultaneous gastric decompression and jejunal feeding are needed. This includes patients in whom malnutrition already exists, or may result, secondary to concurrent conditions.

> Applied Medical Technology, Inc. - 510(k) Submission Traditional Length GJ Feeding Device Section 5 - Page 5.1

{4}------------------------------------------------

#### V. DEVICE DESCRIPTION:

The Traditional Length GJ Feeding Device is a single-use feeding tube that provides for simultaneous gastric decompression / drainage and delivery of enteral nutrition into the distal duodenum or proximal jejunum. It enters the stomach through a surgically-created gastric stoma. The multi-lumen tubing is marked with a graduated scale and held in place by means of an inflatable internal retention balloon and a sliding external bolster. The external tri-port is overmolded onto the tubing and contains three ports: one large port for administering nutrition/medication in the jejunum labeled "JEJUNAL," one large port for gastric decompression/drainage labeled "GASTRIC," and one smaller port housing the balloon fill valve labeled "BAL."

The tri-port component of the Traditional Length GJ Feeding Device is offered in two variations: a legacy model and an ENFit® model designed to ISO 80369-3. Both variations of the device are identical except for the design of the tri-port component configurations.

- The tri-port of the legacy model has two large silicone funnel ports, and a small port to fill 1. the retention balloon. The JEJUNAL and the GASTRIC ports have individual attached silicone straps and plugs intended to independently close each funnel when not in use. The BAL port houses a small, MR conditional balloon fill valve.
- 2. The tri-port of the ENFit® model has 2 large ports and one small port akin to the legacy model, but with a twist-lock thermoplastic male ENFit® connector housed in each the JEJUNAL and the GASTRIC funnel ports. The ENFit® connectors are compliant to ISO 80369-3. The JEJUNAL and the GASTRIC ENFit® ports have individual attached tethers and plugs intended to independently close each ENFit® port when not in use. The BAL port houses an MR safe balloon fill valve.

Both the legacy and the ENFit® variations of the Traditional Length GJ Feeding Device are offered in different French sizes. The gastric length of both variations of the Traditional Length GJ Feeding Device is adjustable, depending on where the sliding external bolster is placed along the multi-lumen tubing. Various jejunal lengths are available. The distal portion of only the 16Fr and 18Fr jejunal tubing contains an anti-kink design using a metal spring. At the distal end of every tube, a silicone tip is attached to the device through which a non-absorbable suture is attached. The Traditional Length GJ Feeding Device is provided sterile (Ethylene Oxide) for prescription use only.

| TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE AND THE PREDICATE<br>DEVICE |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|---------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                 | Subject Device: Traditional Length GJ Feeding<br>Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Predicate: AMT Low-Profile Transgastric-<br>Jejunal Feeding Tube Kit (K123716)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Class and<br>Product Code:                                               | II, KNT, PIF                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | II, KNT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE AND THE PREDICATE<br>DEVICE |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterilization:                                                                  | Sterile (Ethylene Oxide), single use only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Sterile (Ethylene Oxide), single use only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Prescription                                                                    | Single use: Prescription Only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Single use: Prescription Only                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| MRI Safety<br>Information                                                       | Labeled MR Conditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Labeled MR Conditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use:                                                         | The Traditional Length GJ Feeding Device is indicated<br>for use in adult, adolescent, child, and infant patients<br>over 10kg who cannot absorb adequate nutrition<br>through the stomach, who have intestinal motility<br>problems, gastric outlet obstruction, severe<br>gastroesophageal reflux, are at risk of aspiration, or in<br>those who have had previous esophagectomy or<br>gastrectomy.<br>The use of this tube is also clinically indicated when<br>simultaneous gastric decompression and jejunal feeding<br>are needed. This includes patients in whom malnutrition<br>already exists, or may result, secondary to concurrent<br>conditions. | The AMT Transgastric-Jejunal feeding tube is<br>indicated for use in patients who cannot<br>absorb adequate nutrition through the<br>stomach, who have intestinal motility<br>problems, gastric outlet obstruction, severe<br>gastroesophageal reflux, are at risk of<br>aspiration, or in those who have had previous<br>esophagectomy or gastrectomy.<br>The use of this tube is also clinically indicated<br>when simultaneous gastric decompression and<br>jejunal feeding are needed. This includes<br>patients in whom malnutrition already exists,<br>or may result, secondary to concurrent<br>conditions. |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended Use:                                                                   | Provide simultaneous gastric decompression/drainage<br>and delivery of enteral nutrition into the distal<br>duodenum or proximal jejunum.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Provide simultaneous gastric<br>decompression/drainage and delivery of<br>enteral nutrition into the distal duodenum or<br>proximal jejunum.                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Principles of<br>Operation                                                      | • Deliver enteral nutrition directly into the small<br>intestine by enteral feeding pump.<br>• Allow gastric decompression by gravity (leaving<br>the G plug open), or connect to low intermittent<br>suction.                                                                                                                                                                                                                                                                                                                                                                                                                                                | • Deliver enteral nutrition directly into the<br>small intestine by enteral feeding pump.<br>• Allow gastric decompression by gravity<br>(connect extension set to gastric port,<br>leaving the plug open), or connect to low<br>intermittent suction.                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE AND THE PREDICATE<br>DEVICE |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Major Design<br>Characteristics                                                 | Medical grade silicone construction Medical grade silicone tubing with graduated centimeter markings Sliding external bolster External tri-port adapter:  Jejunal feeding port with strap/plug Gastric decompression port with strap/plug Balloon fill valve   Inflatable silicone internal retention balloon Multiple Gastric and Jejunal exit ports Anti-kink enforced Jejunal portion (16Fr & 18Fr only) Tapered distal tip with suture MR Conditional                                                                                                                                                                                                     | Medical grade silicone construction Medical grade silicone tubing External low-profile bolster:  Jejunal feeding port with interlock and strap/plug Gastric decompression port with interlock and strap/plug Balloon fill valve   Inflatable silicone internal retention balloon Multiple Gastric and Jejunal exit ports Anti-kink enforced Jejunal portion (16Fr & 18Fr only) Tapered distal tip with suture MR Conditional                                                                                                                                                                                       |
|                                                                                 | Substantial Equivalence:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE: VI.

{5}------------------------------------------------

Applied Medical Technology, Inc. – 510(k) Submission Traditional Length GJ Feeding Device

{6}------------------------------------------------

As indicated above, minor differences exist between the Traditional Length GJ Feeding Device and the predicate device. These differences include:

- External device profile/size: The subject device has a tri-port (overmold) attached to a standard length of adjustable gastric tubing available in various jejunal lengths while the primary predicate is not adjustable, but rather has a low-profile (overmold) bolster attached to various gastric and jejunal length configurations.
These minor design differences between the Traditional Length GJ Feeding Device (subject device) and the primary predicate (K123716) device do not raise different questions of safety and/or effectiveness from the predicate device. The intended use and indications for use remain the same between the Traditional Length GJ Feeding Device and the predicate device.

### VII. PERFORMANCE DATA:

- A. Biocompatibility Testing: Following a Biological Evaluation Plan, the Traditional Length GJ Feeding Device has been tested for biocompatibility based on the applicable sections of the following standards:
- . ISO 10993-1: 2009 Biological Evaluation of Medical Devices - Part 1: Evaluation and testing within a risk management process
- ISO 10993-3: 2014 Biological Evaluation of Medical Devices - Part 3: Tests for genotoxicity, carcinogenicity, and reproductive toxicity.
- ISO 10993-5: 2009 Biological Evaluation of Medical Devices - Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-6: 2016 Biological Evaluation of Medical Devices . - Part 6: Test for local effects after implantation.
- ISO 10993-7: 2008 Biological Evaluation of Medical Devices

Applied Medical Technology, Inc. - 510(k) Submission Traditional Length GJ Feeding Device Section 5 - Page 5.4

{7}------------------------------------------------

- Part 7: Ethylene Oxide Sterilization Residuals.

- ISO 10993-10: 2010 Biological Evaluation of Medical Devices ● - Part 10: Tests for irritation and skin sensitization.
- ISO 10993-11:2006 Biological Evaluation of Medical Devices . - Part 11: Tests for systemic toxicity
- ISO 10993-12:2012 Biological Evaluation of Medical Devices ● - Part 12: Sample preparation and reference materials
- ISO 10993-17:2002 Biological Evaluation of Medical Device ● - Part 17: Sample preparation and reference materials

In accordance with a Biological Evaluation Plan, a Biological Risk Assessment was completed on the patient contacting materials and it was determined that the Traditional Length GJ Feeding Device met the acceptance criteria for permanent contact (greater than 30 days) with mucosal membrane and breached/compromised surfaces.

- B. Software: There are no software components related in any way to this device. Therefore, Software Validation is not applicable to this device.
- C. Electromagnetic Compatibility & Electrical Safety: There are NO electronic components related in any way to this device.

### D. Performance Testing:

AMT conducted various performance tests on the components contained within the Traditional Length GJ Feeding Device. Testing found that all components and materials met or exceeded design specifications established by AMT.

Bench tests have been carried out to demonstrate conformance to applicable recognized standards, as well as to compare performance to the predicate. The tests carried out are outlined below.

- Testing per AMT Design Specifications: .
	- Balloon Assembly Bond Peel/Tear Strength O
	- Balloon Burst O
	- Fill Valve Blow Out O
	- Fill Valve Pullout O
	- Flow Rate O
	- Leak Test O
	- Tensile Test to tear tubing at jejunal holes O
	- Tensile Test to tear tubing at gastric holes O
	- Minimum Overmold Tri-Port Bond Strength O
	- Stoma Pullout O
	- Tubing Tensile Test O
	- Suture Pullout Tensile Test O
	- Gastric Strap Tensile Test O
	- Jejunal Strap Tensile Test O
- Testing per ASTM F2528-06:
	- Balloon Integrity in Simulated Gastric Fluid O

Applied Medical Technology, Inc. - 510(k) Submission Traditional Length GJ Feeding Device

### Section 5 - Page 5.5

{8}------------------------------------------------

- Balloon volume maintenance O
- Balloon size and shaft size O
- Balloon concentricity O
- Balloon integrity O
- Testing per ISO 80369-20: ●
	- O Fluid leakage
	- Stress cracking O
	- Resistance to separation from axial load O
	- Resistance to separation from unscrewing O
	- Resistance to overriding O
	- Disconnection by unscrewing O
- Testing per ASTM F2052:
	- Magnetically induced displacement force O
- Testing per ASTM F2213:
	- Magnetically induced torque O
- Testing per ASTM F2119:
	- MR image artifact o
- Testing per ASTM F2182:
	- Radio frequency induced heating O

The Traditional Length GJ Feeding Device meets all the acceptance criteria and performed comparable to or better than the primary predicate.

- E. Animal Study: Animal testing was NOT performed.
- F. Clinical Study: Clinical testing was NOT performed. Clinical evidence gathered from independent organizations has been included in the submission.

### VIII. Conclusion:

The Traditional Length GJ Feeding Device can be found substantially equivalent to the primary predicate device cleared under K123716 in intended use, performance, and principles of operation. The minor design differences between the subject device and the legally marketed predicate do not raise different questions of safety and/or efficacy, and the information submitted in this application demonstrates that the subject device is at least as safe and effective as the current legally marketed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K182804](https://fda-staging.innolitics.com/device/K182804)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).