Single-Use Non-Stick Bipolar forceps Bayonet, Single-Use Non-Stick Bipolar Irrigating forceps, Single Use Non-Stick Bipolar Suction forceps
K182773 · Faulhaber Pinzetten Ohg · GEI · Oct 8, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182773
Device Name
Single-Use Non-Stick Bipolar forceps Bayonet, Single-Use Non-Stick Bipolar Irrigating forceps, Single Use Non-Stick Bipolar Suction forceps
Applicant
Faulhaber Pinzetten Ohg
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 8, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Faulhaber Single Use Non-Stick Bipolar Forceps sterile and Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non-sterile are intended for use by a physician familar with electrosurgery for bipolar coagulation and irrigation of tissue for general surgery. The bipolar forceps are used with the bipolar output for standard electrosurgical generators. The products are intended for single use and are provided sterile as well as non-sterile. Products supplied non-sterile must be cleaned, disinfected and sterilized prior to their use by the validated cleaning, disinfection and sterilization process.
Device Story
Single-use bipolar forceps (with or without irrigation) used for tissue coagulation in general surgery. Device connects to standard electrosurgical generators via bipolar cable. Irrigation models feature a drain along the tines connected to an irrigation pump via Luer-Lock. Operated by physicians in clinics/hospitals. Provides mechanical tissue grasping and electrical coagulation. Benefits include non-stick tips to reduce tissue adhesion and irrigation to manage thermal effects. Substantially equivalent to reusable Egon Faulhaber forceps (K101080) in design/materials, adapted for single-use and irrigation.
Clinical Evidence
Bench testing only. No clinical data. Testing included drop tests (ISO 2248), electrical safety (IEC 60601-2-2), mechanical strength, and thermal effects on tissue (liver, kidney, muscle) compared to reusable predicate. All results met pre-defined acceptance criteria.
Technological Characteristics
Materials: Stainless steel (ISO 7153-1), sterling silver, Rilsan (nylon) coating. Bipolar electrosurgical principle. Bayonet form factor. Connectivity: US-Pin bipolar connector, Luer-Lock irrigation port. Sterilization: Gamma irradiation (sterile units) or steam sterilization (non-sterile units). Electrical safety per IEC 60601-2-2.
Indications for Use
Indicated for bipolar coagulation and irrigation of tissue in general surgery, including laryngeal, orthopedic, thoracic, neurosurgical, gynecological (excluding female sterilization), urological, and ENT procedures. Intended for use by physicians familiar with electrosurgery.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Egon Faulhaber Bipolar, Non-Stick Bipolar and Monopolar Forceps (K101080)
Reference Devices
Olsen Medical Single Use Bayonet Bipolar Irrigating Forceps (K080187)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the full name written out to the right of it.
October 8, 2019
Faulhaber Pinzetten OHG Ms. Dorothea Faulhaber General Manager Daimlerstr. 1 D-78665 Frittlingen Germany
Re: K182773
Trade/Device Name: Single-Use Non-Stick Bipolar Forceps sterile/non-sterile Single-Use Non-Stick Bipolar Irrigating forceps sterile/non-sterile
Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 5, 2019 Received: August 8, 2019
Dear Dorothea Faulhaber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182773
#### Device Name
Single Use Non-Stick Bipolar Forceps sterile/ non-sterile Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non-sterile
#### Indications for Use (Describe)
Faulhaber Single Use Non-Stick Bipolar Forceps sterile and Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non-sterile are intended for use by a physician familar with electrosurgery for bipolar coagulation and irrigation of tissue for general surgery. The bipolar forceps are used with the bipolar output for standard electrosurgical generators. The products are intended for single use and are provided sterile as well as non-sterile.
Products supplied non-sterile must be cleaned, disinfected and sterilized prior to their use by the validated cleaning, disinfection and sterilization process.
The bipolar forceps have not been shown to be effective for tubal coagulation for sterilization procedures and should not be used for these procedures.
The types of surgery intended are
- General surgery
- Laryngeal coagulation
- Orthopedic coagulation
- Thoracic coagulation
- Neurosurgical coagulation
- Gynecological coagulation (except for use in female sterilization)
- Urological coagulation
- Ear-, Nose-, Throat coagulation
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a golden emblem inside of a diamond shape. The emblem is symmetrical and appears to be made up of two stylized letters that are mirror images of each other. The emblem and the diamond shape have a thin golden outline.
# 510 (k) Summary
### 1. Submitter's Contract Information
| FAULHABER PINZETTEN OHG | |
|-------------------------|----------------------------------------|
| Daimlerstr. 1 | |
| D – 78665 Frittlingen | |
| Telephone: | 00497426 - 96 38 53-0 |
| Fax: | 00497426 – 96 38 53-30 |
| Contact Person: | Ms. Dorothea Faulhaber, CEO |
| Contact Person: | Ms. Patricia Pfaff, Quality Management |
| Date prepared: | 06/Oct/2019 |
### 2. Device name and classification
| Trade name: | Single Use Non-Stick Bipolar Forceps sterile/ non sterile<br>Single Use Non-Stick Bipolar Irrigating Forceps sterile/ non sterile |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Common name: | Single Use Bipolar Forceps |
| Classification name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Product code: | GEI |
| Regulation number: | 878.4400 |
| Classification: | Class II |
| Predicate Device: | K101080 Egon Faulhaber Pinzetten |
| Reference Devices: | K080187 Olsen Medical |
#### 3. Predicate devices
#### 3.1 Primary predicate device
| | Device Owner/<br>Trade Name | 510(k) # | Product Code |
|-----------------------------|--------------------------------------------------------------------------|----------|-------------------------------------------------------------------------|
| Primary<br>Predicate Device | Egon Faulhaber<br>Bipolar, Non-Stick<br>Bipolar and Monopolar<br>Forceps | K101080 | GEI Electrosurgical, Cutting &<br>Coagulation & Accessories<br>Class II |
#### 3.2 Reference device
| | Device Owner/<br>Trade Name | 510(k) # | Product Code |
|---------------------|-------------------------------------------------------------------|----------|-------------------------------------------------------------------------|
| Reference<br>Device | Olsen Medical<br>Single Use Bayonet<br>Bipolar Irrigating Forceps | K080187 | GEI Electrosurgical, Cutting &<br>Coagulation & Accessories<br>Class II |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with a gold-colored design inside a diamond shape. The design appears to be two stylized letters, possibly 'T' and 'E', mirrored and joined together in the center. The logo has a symmetrical and geometric appearance, with sharp lines and angles that create a modern and abstract look.
#### 4. Device Description
The product family "Single Use Bipolar Forceps", including Faulhaber Single Use Non-Stick Bipolar Forceps and Single Use Non-Stick Bipolar Irrigating Forceps, are intended to be used for bipolar coagulation and irrigation of tissue by physicians familiar with bipolar coagulation in medical practices and clinics.
The Single Use Bipolar Forceps are single use products and must not be reused. They are provided sterile as well as non-sterile. Products delivered non sterile must be cleaned, disinfected and sterilized before use.
For the application the Single Use Bipolar Forceps have to be connected by appropriate bipolar cable to the bipolar output of an HF generator. Bipolar cables and ESU are not part of the subject device.
The Single Use Bipolar Forceps are provided in bayonet design with non-stick tips and are identical in design, construction, materials and manufacturing to the reusable device EGON FAULHABER Bipolar Non-Stick Forceps (K101080). The principles of operation and mechanism of action are identical as well.
In addition to the cleared and leqally marketed EGON FAULHABER devices, the products are with irrigation function available. The irrigation function works via a drain running along the forceps tines from tip to handle height, the drain is connected by Luer-Lock via an irrigation tubing with an irrigation pump.
#### 5. Intended Use/ Indications for use
Faulhaber Single Use Non-Stick Bipolar Forceps sterile and Single Use Non-Stick Irrigating Forceps sterile are intended for use by a physician familiar with electrosurgery for bipolar coaqulation and irrigation of tissue for general surgery. The bipolar forceps are used with the bipolar output for standard electrosurgical generators. The products are intended for single use and are provided sterile as well as non sterile. Products supplied non sterile must be cleaned, disinfected and sterilized prior to their use by the validated cleaning, disinfection and sterilization process.
The bipolar forceps have not been shown to be effective for tubal sterilization or tubal coaqulation for sterilization procedures and should not be used for these procedures.
The types of surgery intended are
- General surgery -
- -Laryngeal coagulation
- -Orthopedic coaqulation
- Thoracic coagulation -
- -Neurosurgical coagulation
- -Gynecological coaqulation (except for use in female sterilization)
- -Urological coaqulation
- Ear-, Nose- and Throat coagulation -
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a gold-colored emblem. The emblem is diamond-shaped and contains a stylized design that resembles two mirrored letter 'T's. The design is symmetrical and has a sharp, modern look.
## 6. Substantial Equivalence discussion
| | Subject Device<br>K182773<br>Single Use Bipolar<br>Forceps (sterile and non-<br>sterile)<br>Faulhaber Pinzetten<br>OHG | Primary Predicate<br>Device K101080<br>Bipolar, Non-Stick<br>Bipolar and Monopolar<br>Forceps<br>EGON FAULHABER<br>Pinzetten | Reference Device<br>K080187<br>Single Use Bayonet<br>Bipolar Irrigating Forceps<br>Olsen Medical | |
|----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k)<br>number | K182773 | K101080 | K080187 | |
| 510(k)<br>submitter/hold<br>er | Faulhaber Pinzetten<br>OHG | EGON FAULHABER<br>Pinzetten | Olsen Medical | |
| Product Code | GEI | GEI | GEI | |
| Indications for<br>Use/ Intended<br>Use | Faulhaber Single Use<br>Non-Stick Bipolar<br>Forceps and Single Use<br>Non-Stick Bipolar<br>Irrigating Forceps are<br>intended for use by a<br>physician familar with<br>electrosurgery for bipolar<br>coagulation and irrigation<br>of tissue for general<br>surgery. The bipolar<br>forceps are used with the<br>bipolar output for<br>standard electrosurgical<br>generators.<br>The products are<br>intended for single use<br>and are provided sterile<br>as well as non-sterile.<br>Products supplied non-<br>sterile must be cleaned,<br>disinfected and sterilized<br>prior to their use by the<br>validated cleaning,<br>disinfection and<br>sterilization process.<br>The bipolar forceps have<br>not been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization procedures<br>and should not be used | The EGON<br>FAULHABER Bipolar,<br>Non-Stick Bipolar and<br>Monopolar Forceps are<br>intended for use by a<br>physician familar with<br>electrosurgery in bipolar<br>and monopolar<br>coagulation for general<br>surgery where<br>coagulation of soft tissue<br>is needed. The Bipolar<br>Forceps are used with<br>the bipolar and the<br>Monopolar Forceps are<br>used with the<br>monopolar output for<br>standard electrosurgical<br>generators.<br>The EGON<br>FAULHABER Bipolar,<br>Non-Stick Bipolar and<br>Monopolar Forceps have<br>not been shown to be<br>effective for tubal<br>sterilization or tubal<br>coagulation for<br>sterilization procedures<br>and should not be used<br>for this procedures.<br>The types of surgery<br>intended are | The Olsen Medical<br>Single Use Bayonet<br>Bipolar Irrigating<br>Forceps is a single use<br>product sold sterile and<br>is intended for use in<br>electro surgery for<br>coagulation and irrigation<br>of tissue. This device is<br>intended for use with the<br>OLSEN MEDICAL<br>Integrated Irrigation<br>Tubing and Bipolar Cord<br>Set or similar design of<br>Bipolar Cord and<br>Irrigation Tubing. | |
| | Subject Device<br>K182773<br>Single Use Bipolar<br>Forceps (sterile and non-<br>sterile)<br>Faulhaber Pinzetten | Primary Predicate<br>Device K101080<br>Bipolar, Non-Stick<br>Bipolar and Monopolar<br>Forceps<br>EGON FAULHABER | Reference Device<br>K080187<br>Single Use Bayonet<br>Bipolar Irrigating Forceps<br>Olsen Medical | |
| | OHG | Pinzetten | | |
| | for these procedures.<br>The types of surgery<br>intended are<br>- General surgery<br>- Laryngeal coagulation<br>- Orthopedic coagulation<br>- Thoracic coagulation<br>- Neurosurgical<br>coagulation<br>- Gynecological<br>coagulation (except for<br>use in female<br>sterilization)<br>- Urological coagulation<br>- Ear-, Nose-, and Throat<br>coagulation | - Laryngeal coagulation<br>- Orthopedic coagulation<br>- Thoracic coagulation<br>- Neurosurgical<br>coagulation<br>- Gynecological<br>coagulation (except for<br>use in female<br>sterilization)<br>- Urological coagulation<br>- Ear-, Nose-, and Throat<br>coagulation | | |
| Product life | Single use | Reusable | Single use | |
| Application<br>Technology | Bipolar | Bipolar and monopolar | Bipolar | |
| Tip<br>dimensions/<br>Tip material | 0.5 mm – 1.5 mm<br>Sterling silver | 0.25 mm – 2.0 mm<br>Sterling silver | 0.5 mm – 1.5 mm<br>Non-Stick Cermet<br>(Ceramic Metal) Coating | |
| Shaft style/<br>Shaft | Bayonet | Straight/ Angled/<br>Bayonet | Bayonet | |
| dimensions/<br>Shaft<br>materials | 203 mm – 300 mm<br>Coated Stainless steel | 110 mm – 250 mm<br>Coated Stainless steel | 180 mm – 267 mm<br>Insulated stainless steel<br>body | |
| Cable<br>connector<br>type | US-Pin | US-Pin | Twin Pin | |
| Irrigation<br>connector<br>type | Female luer to be<br>connected with male luer<br>of irrigation tubing | not applicable | Female luer for use with<br>the Olsen medical<br>Integrated Irrigation<br>Tubing and Bipolar Cord<br>Set or similar design of<br>Bipolar Cord and<br>Irrigation Tubing | |
| Body contact<br>material | - Stainless steel acc. ISO<br>7153-1<br>- Sterling silver | - Stainless steel acc.<br>ISO 7153-1<br>- Sterling silver | - Stainless steel body<br>with nylon insulation<br>- Non-stick Cermet | |
| | Subject Device<br>K182773<br>Single Use Bipolar<br>Forceps (sterile and non-<br>sterile)<br>Faulhaber Pinzetten<br>OHG | Primary Predicate<br>Device K101080<br>Bipolar, Non-Stick<br>Bipolar and Monopolar<br>Forceps<br>EGON FAULHABER<br>Pinzetten | Reference Device<br>K080187<br>Single Use Bayonet<br>Bipolar Irrigating Forceps<br>Olsen Medical | |
| | - Rilsan® (Nylon)<br>Coating | - Rilsan® (Nylon)<br>Coating | (Ceramic Metal) coating | |
| Delivery<br>status | Sterile and non-sterile | Non sterile | Sterile | |
| Sterilization<br>method for<br>products<br>provided<br>sterile | Gamma irradiation | not applicable | Ethylene Oxide (EtO) | |
| Packaging for<br>products<br>provided<br>sterile | Cleerpeel® Foil Pouch | not applicable | Tyvek® Pouch | |
| Sterilization<br>method for<br>initial<br>treatment/<br>reprocessing | Steam sterilization | Steam sterilization | not applicable | |
| Packaging for<br>products<br>provided non-<br>sterile | foil bag with cardboard<br>box | foil bag with cardboard<br>box | not applicable | |
| Maximum<br>Peak Voltage | ≤ 500 Vp | ≤ 500 Vp | Information not available | |
| Tissue<br>temperature<br>range | ~ 60°C - ~ 100°C | ~ 60°C - ~ 100°C | Information not available | |
| Method of<br>operation | mechanical activation,<br>no switch | both, mechanical<br>activation (bipolar) and<br>handle with switch<br>(monopolar) | mechanical activation,<br>no switch | |
| Accessories | No | No | Olsen medical Integrated<br>Irrigation Tubing and<br>Bipolar Cord Set or<br>similar design of Bipolar<br>Cord and Irrigation<br>Tubing | |
| Safety<br>features | Dielectric strength<br>insulation acc. IEC<br>60601-2-2; insulated<br>safety plug | Dielectric strength<br>insulation acc. IEC<br>60601-2-2; insulated<br>safety plug | Insulated | |
{6}------------------------------------------------
# 510 (k) Summary
K182773
Image /page/6/Picture/1 description: The image shows a logo with a diamond shape as the background. Inside the diamond, there is a stylized design that resembles the letters 'AE' or a symmetrical arrow-like figure. The color of the logo and the design is a golden hue, which gives it a premium or elegant feel. The design is simple, yet distinctive, making it easily recognizable.
{7}------------------------------------------------
# 510 (k) Summary
K182773
Image /page/7/Picture/1 description: The image shows a golden emblem. The emblem is diamond-shaped with a stylized design inside. The design appears to be a symmetrical arrangement of lines and shapes, possibly resembling letters or abstract symbols. The color is a consistent gold tone, giving the emblem a sense of value and importance.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo with a gold-colored design inside a diamond shape. The design appears to be an abstract representation of the letters 'T' and 'E' intertwined. The logo has a symmetrical appearance, with the 'T' shape in the center and the 'E' shape forming the outer edges of the design.
This submission supports the position that the Faulhaber Pinzetten OHG forceps is substantially equivalent to a number of previously cleared devices, including Olsen Medical Single Use Bayonet Irrigating Bipolar Forceps.
#### 7. Performance data:
Performance testing of the product family Single Use Bipolar Forceps was determined by the application of a risk management process acc. ISO 14971:2013-04.
As a result of the risk analysis and regarding the applicable FDA quidance documents, the following tests have been identified as essential.
The indicated performance tests were derived and have been projected. The performance data were provided in support of the substantial equivalence determination.
#### 7.1 Drop Test acc. DIN EN 22248:1993-02 (ISO 2248:1985) "Packaging; complete, filled transport packages; vertical impact test by dropping"
The vertical impact test was conducted on Faulhaber Single Use Bipolar Forceps acc. to DIN EN 22248:1993-02 (ISO 2248:1985).
The products in their primary and outer packaging have passed the tests without damage. The outer packaging is in a transportable condition.
The test of the packaging system has been successfully completed, the vertical impact test demonstrated the suitability of the packaging system successfully.
The test results are sufficient to support substantial equivalence of the subject device with the predicate devices.
#### 7.2 Electrical safety acc. DIN EN IEC 60601-2-2:2018-12 (IEC 60601-2-2:2017)
Electrical safety testing was conducted on Faulhaber Single Use Bipolar Forceps. The instruments have completed the entire test cycles without changes or impairments, in particular to insulation, but also in terms of mechanics and function.
The testing of the instruments for dielectric strength therefore have been completed successfully.
The system complies with the DIN EN IEC 60601-2-2:2018-12 (IEC 60601-2-2:2017) standards for electrical safety.
The test results are sufficient to support substantial equivalence of the subject device with the predicate devices.
#### 7.3 Performance Tests "Mechanical strength/ Electrical performance with all components and accessories as a system/ Thermal effects on tissue"
Performance tests regarding mechanical strength, electrical performance with all components and accessories as a system and thermal effects on tissue have been conducted on Faulhaber Single Use Bipolar Forceps.
The test procedures correspond to the FDA Guidance document "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery - Guidance for Industry and Food and Drug Administration Staff", issued August 15, 2016.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a golden emblem. The emblem is a diamond shape with a stylized design inside. The design appears to be a combination of geometric shapes and lines, possibly representing a stylized letter or symbol. The color of the emblem is a warm, golden hue, which contrasts with the white background.
For the test performance regarding thermal effects on tissue, three different types of tissue (liver, kidney and muscle) were coagulated with the Single Use Non-Stick Bipolar Forceps with irrigation function and without irrigation function at different settings. For comparison, all tests were carried out with the product "Faulhaber Pinzetten OHG Non-Stick Bipolar Forceps bavonet, reusable".
The tests were carried out in accordance with the Guidance for industry and food and drug administration staff "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" from August 15, 2016.
All test results met the pre-defined acceptance criteria.
The evaluation of the results also included a comparison between products with irrigation function and products without irrigation function. This comparison allows the conclusion that products with irrigation function achieve the same results, with maximum power setting even better result than the products without irrigation function.
The performance test showed that the products of the product family "Single Use Bipolar Forceps" are able to provide, with or without irrigation of tissue during coagulation, safe and appropriate results for the intended use.
The test results are sufficient to support substantial equivalence of the subject device with the predicate devices.
#### 8. Biocompatibility:
The Biological Evaluation of the Single Use Bipolar Forceps was conducted in accordance with use of International Standard ISO-10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA.
Based on the risk management it was proven, that the Single Use Bipolar Forceps are manufactured from identical material using identical processes and procedures as the primary predicate device K101080. Identical machines and equipment as well as identical operating- and auxiliary materials are used in production processes. Therefore, additional clinical tests are not necessary, because the safety for use of the materials has been proved in the past and will continue to be monitored by Post Marke…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.