The Lansinoh® Signature Pro™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users. The Lansinoh® Smartpump™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.
Device Story
Powered breast pump system; diaphragm-type vacuum pump driven by microprocessor-controlled DC electric motor; powered by 6 AA batteries or AC adapter. Inputs: user-selected cycle mode and vacuum level via front panel keypad. Outputs: vacuum suction for milk expression. Used in home or clinical settings by nursing women. Smartpump model includes Bluetooth connectivity for mobile app integration. Closed system design with diaphragm barrier prevents milk backflow into tubing/pump. Healthcare providers/users adjust settings based on comfort; output facilitates milk collection for infant feeding; benefits include support for breastfeeding and milk supply maintenance.
Clinical Evidence
Bench testing only. Performance evaluated via suction curves, battery life testing (approx. 90 mins), and 500-hour use-life testing. Backflow testing confirmed diaphragm efficacy in preventing milk contamination. Electrical safety and EMC testing performed per ANSI/AAMI ES60601-1 and IEC 60601-1-2. No clinical data presented.
Technological Characteristics
Diaphragm-type vacuum pump; microprocessor-controlled DC motor. Materials meet 21 CFR 175-179. Powered by 6 AA batteries or AC adapter. Connectivity: Bluetooth (Smartpump only). Standards: ANSI/AAMI ES60601-1:2005/(R)2012, A1:2012; IEC 60601-1-2:2007; IEC 60601-1:2005/(R)2012. Closed system design.
Indications for Use
Indicated for nursing women to express and collect breastmilk for infant feeding. Intended for both single-user and multiple-user settings.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
June 12, 2019
Lansinoh Laboratories Saglik Gerecleri Tasarim San. Tic. Ltd. Sti. % Lindsay Ewers Director of Quality Assurance Lansinoh Laboratories 333 N. Fairfax Street, Suite 400 Alexandria, VA 22314
Re: K182749
Trade/Device Name: Lansinoh® Signature Pro™ Double Electric Breast Pump (DEBP 2.1) Lansinoh® Smartpump™ Double Electric Breast Pump (DEPB 2.2) Regulation Number: 21 CFR§ 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: May 16, 2019 Received: May 17, 2019
Dear Lindsay Ewers:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sharon M. Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182749
#### Device Name
Lansinoh® Signature Pro™ Double Electric Breast Pump (DEBP 2.1) Lansinoh® Smartpump™ Double Electric Breast Pump (DEPB 2.2)
#### Indications for Use (Describe)
The Lansinoh® Signature Pro™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.
The Lansinoh® Smartpump™M Double Electric Breast Pump is intended to express and collect the breasmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;"><input type="checkbox"/></div> <span> Exempted New Part 613 Subchapter S Project</span> |
| <div style="display:inline-block;"><input checked="" type="checkbox"/></div> <span> One-Time Community SEP New Subchapter S Project</span> |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Device Common Name: | Powered Breast Pump |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | Lansinoh® Signature Pro™ Double Electric Breast Pump<br>(DEBP 2.1) and Lansinoh® Smartpump™ Double Electric<br>Breast Pump (DEBP 2.2) |
| Submitter: | Lansinoh Laboratories Saglık Gerecleri San. Tic. Ltd. Sti.<br>10006 sokak, No:64<br>A.O.S.B. Cigli - Izmir 35620<br>TURKEY<br>Tel: 0090 232 328 05 06<br>Fax: 0090 232 328 05 58<br>www.lansinoh.com.tr |
| Company Contact: | Barbaros Karadağ<br>General Manager<br>Lansinoh Laboratories Saglık Gerecleri San. Tic. Ltd. Sti.<br>Tel: +90 232 328 05 06<br>barbaroskaradag@lansinoh.com.tr |
| Date Prepared: | June 11, 2019 |
| Classification Regulation: | 21 CFR 884.5160, Class II, Powered Breast Pump |
| Panel: | Obstetrics/Gynecology |
| Product Code: | HGX - Powered Breast Pump |
# Indication for Use:
The Lansinoh® Signature Pro™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.
The Lansinoh® Smartpump™ Double Electric Breast Pump is intended to express and collect the breastmilk of a nursing woman for the purpose of feeding collected breastmilk to a baby. The pump is intended for multiple users, and single users.
# Device Description:
The Lansinoh Powered Breast Pumps are intended to express the breast milk of a nursing woman. The pumping can be performed on one breast or on both breasts at the same time. The pump can be powered by 6 AA batteries or by an AC adaptor that is provided with the pump. The pumping system consists of a diaphragm-type vacuum pump which is driven by a microprocessor controlled DC electric motor. The user interface consists of a front panel keypad
{4}------------------------------------------------
and LCD display. The user is able to adjust cycle mode and vacuum level based on personal comfort and preference.
The purpose of this 510(k) is to update the indications for use statement to include use by multiple users. The labeling has been revised to clarify that the collection kits are for single users only and are not to be shared with other users.
# Device Comparison Table:
| | New Device | Predicate Device | Substantial<br>Equivalence<br>Comparison |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Lansinoh Laboratories<br>Saglik Gerecleri Tasarim<br>San. Tic. Ltd. Sti. | Lansinoh Laboratories<br>Saglik Gerecleri<br>Tasarim San. Tic. Ltd.<br>Sti. | N/A |
| Device Name | Lansinoh® Signature<br>Pro™ Double Electric<br>Breast Pump (DEBP 2.1)<br>and Lansinoh®<br>Smartpump™ Double<br>Electric Breast Pump<br>(DEBP 2.2) | Powered Breast Pump | N/A |
| 510(k) # | K182749 | K122474 | N/A |
| Intended Use | The Lansinoh® Signature<br>Pro™ Double Electric<br>Breast Pump is intended<br>to express and collect the<br>breastmilk of a nursing<br>woman for the purpose of<br>feeding collected<br>breastmilk to a baby. The<br>pump is intended for<br>multiple users, and single<br>users.<br>The Lansinoh®<br>Smartpump™ Double<br>Electric Breast Pump is<br>intended to express and<br>collect the breastmilk of a<br>nursing woman for the<br>purpose of feeding<br>collected breastmilk to a<br>baby. The pump is<br>intended for multiple | The Powered Breast<br>Pump is intended to<br>express and collect the<br>breast milk of a nursing<br>woman for the purpose<br>of feeding the collected<br>milk to a baby. The<br>Powered Breast Pump is<br>intended for a single<br>user. | Different; the<br>indications for use of<br>the subject devices are<br>different from the<br>predicate in that the<br>subject device is<br>intended for multiple<br>users. However, this<br>difference does not alter<br>the intended use of the<br>device as compared to<br>the predicate. |
| | New Device | Predicate Device | Substantial<br>Equivalence<br>Comparison |
| | users, and single users. | | |
| Suction Levels<br>(stimulation) | 46-140 mmHg | 55 - 140 mmHg | Similar; the subject<br>devices have a slightly<br>lower suction level.<br>However, this difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Cycles per Second<br>(stimulation) | 1.61-2.33 | 1.55 - 2.4 | Similar; the subject<br>devices have a slightly<br>different cycle rates.<br>However, this difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Suction Levels<br>(expression) | 95 - 280 mmHg | 80 – 220 mmHg | Different; the subject<br>device has higher<br>pressure levels for<br>expression than the<br>predicate. However,<br>this difference does not<br>raise different questions<br>of safety and<br>effectiveness. |
| Cycles per Second<br>(expression) | 0.58 - 1.69 | 0.61 - 1.52 | Similar; the subject<br>devices have a slightly<br>different cycle rate.<br>However, this difference<br>does not raise different<br>questions of safety and<br>effectiveness. |
| Suction Settings | 8 | 8 | Same |
| Expression modes | 3 | 3 | Same |
| Power Supply | a) 6 AA alkaline batteries<br>b) AC Adapter | a) 6 AA alkaline<br>batteries<br>b) AC Adapter | Same |
| Pumping Option | Single or Double | Single or Double | Same |
| Back Flow<br>Protection | Yes | Yes | Same |
| Let Down Function | Yes | Yes | Same |
| | New Device | Predicate Device | Substantial<br>Equivalence<br>Comparison |
| Cycling/Suction<br>Control Mechanism | Microprocessor | Microprocessor | Same |
| Communication<br>with mobile app<br>(Smartpump only) | Bluetooth | N/A | Different; the subject<br>device (SmartPump) has<br>Bluetooth capability.<br>The addition of<br>Bluetooth<br>communication does not<br>alter the intended use or<br>raise different questions<br>of safety and<br>effectiveness. |
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Data:
### Bench Performance
### Suction Curves
Devices were evaluated for suction performance at each of the available settings. The suction curves for each cycle mode and suction level demonstrated that the device meets its specifications and performs within the specified working ranges of pressure and cycle speed for each mode/level.
# Battery Life Testing
Battery life was measured with the pumps at the highest level ((Expression Mode 3, Level 8). The "fail time" was defined as the number of minutes the pump could run at this level before dropping below specifications. Testing supports that the battery life of the device, as noted in the labeling, is approximately 90 minutes.
### Sterilization, Cleaning and Shelf Life
### Sterilization and Cleaning
The device is reusable, provided non-sterile, and is not sterile when used. Cleaning instructions are provided in the labeling.
### Shelf-life
Shelf-life is not applicable due to the low likelihood of time-dependent product degradation. However, in accordance with IEC 60601-1:2005/(R)2012 the subject devices expected use-life is 500 hours. In testing, the devices were demonstrated to operate within specifications for up to 500 hours of continuous operation at the maximum setting (Expression mode 3, Level 8).
### Backflow Test
The Lansinoh pumps are designed as a closed milk collection system. The diaphragm provides a physical barrier, preventing breastmilk from flowing into the tubing or pump body. The purpose
{7}------------------------------------------------
of the backflow test is to demonstrate that the design prevents backflow into the tubing and pump. Devices were tested at maximum pressure/cycle settings (Expression Mode 3, Level 8) at various orientations to simulate worst-case conditions. The testing demonstrated that no milk was present in test devices' tubing during and following the test; therefore the diaphragm was demonstrated to prevent back flow of milk into the tubing and pump.
### Biocompatibility
There have been no changes to the patient contacting materials used in the subject device since its previous clearance in K122474. Therefore, biocompatibility data from K122474 can be leveraged to support the biocompatibility of the subject devices.
The milk contacting components are those components that are provided in the pump kit. All materials in contact with milk have been tested to meet FDA's Food Additive criteria (21 CFR 175-179).
### Electrical Safety and EMC
The device was tested for electrical safety in accordance with (ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012) and EMC in accordance with (IEC 60601-1-2:2007). The device was demonstrated to meet applicable test limits for electrical safety and EMC per the cited standards.
### Software and Cybersecurity
The software was evaluated as a minor level of concern per FDA guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005. Documentation consistent with a minor level of concern demonstrated the software functions as intended.
Cybersecurity for the subject devices was evaluated through a risk-based approach. Appropriate mitigations were implemented based upon the level of risk of the devices.
### Substantial Equivalence Conclusion:
The purpose of this 510(k) is to establish substantial equivalence of the Lansinoh® Signature Pro™ Double Electric Breast Pump (DEBP 2.1) and Lansinoh® Smartpump™ Double Electric Breast Pump (DEBP 2.2) to the previously cleared Lansinoh Double Electric Breast Pump (K122474).
The subject device indications are different from that of the predicate in that they allow for multiple users. This difference does not alter the intended use of the subject device from that of the predicate.
The subject devices have slightly different technological characteristics as described in the table above. However, the technological differences do not alter the intended use of the subject device as compared to the predicate or raise different questions of safety and effectiveness. The technological differences were evaluated through performance testing that demonstrated that the subject device is as safe and effective as the predicate.
Therefore, the subject devices are substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.