← Product Code [GWO](/productcode/GWO) · K182711

# Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI) (K182711)

_Kelyniam Global, Inc. · GWO · May 21, 2019 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182711

## Device Facts

- **Applicant:** Kelyniam Global, Inc.
- **Product Code:** [GWO](/productcode/GWO.md)
- **Decision Date:** May 21, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5320
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).

## Device Story

Customized prosthetic plates fabricated from PEEK billet blocks using patient CT scan data; designed to fill cranial/craniofacial bony voids. Device features include perfusion holes, integrated fixation system (IFS) tabs, temporal cutbacks, and multi-part configurations. Used in hospital/surgical settings by neurosurgeons; implanted via standard cranioplasty hardware. Provides precise anatomical fit to support bone healing. Non-sterile; requires steam sterilization by user. MR safe.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), cleaning validation (ANSI/AAMI ST72, USP <161>, USP <85>), steam sterilization validation (ISO 17665-1, ISO 14937), mechanical testing, and shipping validation.

## Technological Characteristics

Material: Implant-grade PEEK (Invibio PEEK-Optima LT-1, Evonik Vestakeep i4) per ASTM F2026-16. Form factor: Patient-specific CNC-machined plates (2-10mm thick). Non-conductive, non-magnetic. Sterilization: Steam (user-performed).

## Regulatory Identification

A preformed alterable cranioplasty plate is a device that is implanted into a patient to repair a skull defect. It is constructed of a material, e.g., tantalum, that can be altered or reshaped at the time of surgery without changing the chemical behavior of the material.

## Predicate Devices

- Stryker PEEK Customized Cranial Implant Kit ([K121153](/device/K121153.md))
- Synthes Patient Specific Cranial/Craniofacial Implant (PSCI) ([K053199](/device/K053199.md))
- Kelyniam Global Inc. (KGI) Customized Skull Implant (CSI) ([K103582](/device/K103582.md))
- Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI) ([K121755](/device/K121755.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 21, 2019

Kelyniam Global Inc. Eric Boyea Quality Director 97 River Road. Suite A Canton, Connecticut 06019

Re: K182711

Trade/Device Name: Customized Craniofacial Implant (CCI), Customized Skull Implant (CSI) Regulation Number: 21 CFR 882.5320 Regulation Name: Preformed Alterable Cranioplasty Plate Regulatory Class: Class II Product Code: GWO Dated: April 15, 2019 Received: April 16, 2019

Dear Eric Boyea:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Matthew Krueger, M.S.E. Assistant Director THT5A1: Neurosurgical Devices DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K182711

Device Name Customized Craniofacial Implant (CCI) Customized Skull Implant (CSI)

#### Indications for Use (Describe)

The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="font-size:120%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:120%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### KELYNIAM

# K182711 510(k) Summary

| Date Prepared:<br>Purpose:                                                               | 20MAY2019                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Purpose of the<br>Submission:                                                            | The purpose for the Premarket Notification 510(k) is for authorization to add additional features to the<br>currently cleared (devices) base implant or base implant with Prefusion Holes. The additional features do not<br>alter the intended therapeutic use of the subject device, the technological characteristic nor do they affect the<br>safety and effectiveness of the subject devices relative to the predicates. The additional features are intended<br>to respond to requests by surgeons to facilitate the surgical application of the implant(s) and do not introduce<br>any new technological issues. The base implant, as well as the proposed new features is fabricated from a<br>billet block of natural implant grade Polyether ether ketone (PEEK) Thermoplastic Polymer. This is the same<br>material currently used by the listed predicate device organizations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Submitter Information:                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Submitter:                                                                               | Kelyniam Global Inc. (KGI), 97 River Road, Suite A, Canton, Connecticut 06019<br>Tel: (800) 280-8192<br>Fax: (501) 641-2000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Official<br>Correspondent:                                                               | Eric Boyea, Director of Quality & Regulatory Affairs<br>Toll Free: (800) 280-8192 x 701<br>Tel: (860) 352-5736<br>Email: eboyea@kelyniam.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Information                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name:                                                                              | • Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Common Name:<br>Classification<br>Information:                                           | • Customized Cranial Implant or Craniofacial/Skull Implant<br>• Panel: Neurology / Product Code: GWO / Classification: Class II in accordance with 21 CFR 882.5320,<br>Preformed Alterable Cranioplasty Plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate<br>Devices:                                                                    | • Primary Predicate - K121153 Stryker PEEK Customized Cranial Implant Kit<br>• Secondary Predicate - K053199 Synthes Patient Specific Cranial/Craniofacial Implant (PSCI)<br>• Tertiary Predicate - K103582 Kelyniam Global Inc. (KGI) Customized Skull Implant (CSI)<br>• Quaternary Predicate - K121755 Kelyniam Global Inc. (KGI) Customized Craniofacial Implant (CCI)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device<br>Description (with<br>included features<br>previously<br>cleared):              | Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is an individually sized and shaped<br>implantable prosthetic plates intended to fill a bony void or defect area in a specific patient's cranial and<br>craniofacial skeleton (orbital rim, zygoma, & adjacent bone), ranging from 2 mm to 10 mm thick (typically<br>4mm thick based on the patient CT Scan imaging data) x 25mm to 250 mm wide x 25 mm to 250 mm long. A<br>Customized Skull Implant (CSI) is intended to fill a bony void or defect area in a specific patient's skull whereas<br>a Customized Craniofacial Implants (CCI) is intended to fill a bony void or defect area in a specific patient's<br>facial region of the skull, excluding the Maxilla (upper jaw area surrounding the teeth only) and Mandible,<br>which are both considered load bearing areas of the facial region of the skull. The size, asymmetrical shape,<br>thickness, contour, and edge profile are design elements of the non-load bearing patient-specific Base Implant<br>that are used to support the base implant in the bony void or defect area while providing for a "Precise Fit".<br>The single patient use base implant is (1) fabricated from a billet block of natural implant grade Polyether<br>ether ketone (PEEK) Thermoplastic Polymer using the patient's CT Scan imaging data, (2) provided clean but<br>non-sterile for steam sterilization prior to implantation at a hospital or surgical site with neurosurgery<br>capabilities, and (3) attached to the native bone using commercially available cranioplasty hardware and<br>fasteners. The additional features listed below are additional design elements that may be added to the Base<br>Implant and Base Implant with Perfusion Holes, as requested by the Physician.<br>The Base Implant, as defined above (as previously cleared on K103582 and K121755). |
| Device Information (Continued)                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Additional<br>Features<br>(Subject of this<br>submission):                               | Image: Integrated Fixation System (IFS) or IFS Tabs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                                          | Image: A Temporal Cutback                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                                          | Image: A Multi-Part Implant                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for<br>Use:                                                                  | The Customized Craniofacial Implant (CCI) and Customized Skull Implant (CSI) is intended to fill a bony void or<br>defect area in a patient's specific cranial and craniofacial skeleton (orbital rim, zygoma, & adjacent bone).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Statement of<br>Substantial<br>Equivalence:                                              | The KGI subject devices are intended to fill a bony void or defect area in a patient's specific Cranial and<br>craniofacial skeleton and are substantially equivalent in indications for use, technology, and material to<br>K121153 Stryker PEEK Customized Cranial Implant Kit (Primary), K053199 Synthes Patient Specific<br>Cranial/Craniofacial Implant (PSCI) (Secondary), K103582 KGI Customized Skull Implant (CSI) (Tertiary), and<br>K121755 KGI Customized Craniofacial Implant (CCI) (Quaternary).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Comparison of<br>Technological<br>Characteristics<br>with the<br>Predicate<br>Device(s): | The KGI subject devices are manufactured from implant grade Polyether ether ketone Thermoplastic Polymer<br>material, known hereafter as PEEK, sold non-sterile, and is intended to fill a bony void or defect area in a patient's<br>specific Cranial and craniofacial skeleton. The additional features, as listed above, are incorporated in the Base<br>Implant during the design phase and CNC Machined from a billet block of PEEK. The additional features are<br>intended to respond to requests by surgeons to facilitate the surgical application of the implant(s) and do not                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

Perfusion Holes are a design element that allow the passage of fluid to an organ or tissue (as previously cleared on K103582 and K121755).

{4}------------------------------------------------

### KELYNIAM

## K182711 510(k) Summary

introduce any new technological issues. See Device Comparison Chart below for details. 1

### Subject Devices vs Predicate Devices Comparison Chart with included Additional Features

|                            | Subject Device                                                                                                                                           | Subject Device                    | Primary Predicate                                                                       | Secondary Predicate                                                           | Tertiary Predicate                                                                                             | Quaternary Predicate                     |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Manufacturer               | Kelyniam Global Inc. (KGI)                                                                                                                               |                                   | Stryker                                                                                 | Synthes                                                                       | Kelyniam Global Inc. (KGI)                                                                                     |                                          |
| 510(k) #                   | K182711                                                                                                                                                  |                                   | K121153                                                                                 | K053199                                                                       | K103582                                                                                                        | K121755                                  |
| Description                | Customized Craniofacial<br>Implant (CCI)                                                                                                                 | Customized Skull<br>Implant (CCI) | PEEK Customized Craniofacial<br>Implant Kit                                             | Patient Specific Cranial/<br>Craniofacial Implant (PSCI)                      | Customized Skull<br>Implant (CSI)                                                                              | Customized Craniofacial<br>Implant (CCI) |
| Indications for Use        | Is intended to fill a bony void or defect<br>area in a patient's specific cranial and<br>craniofacial skeleton (orbital rim, zygoma,<br>& adjacent bone) |                                   | Is intended to be used to<br>replace bony voids in the<br>cranial/craniofacial skeleton | Is intended to replace<br>bony voids in the Cranial/<br>craniofacial skeleton | Is intended to fill a bony void or defect<br>area in a patient's specific cranial and<br>craniofacial skeleton |                                          |
| Material                   | Invibio Inc PEEK-Optima LT-1® and Evonik<br>Vestakeep i4®                                                                                                |                                   | Invibio Inc PEEK-Optima LT-1®                                                           | CP Titanium and PEEK                                                          | Invibio Inc PEEK-Optima LT-1®                                                                                  |                                          |
| Technical<br>Specification | Plate - Custom sized to each patient using<br>CT scan data                                                                                               |                                   | Customized patient-specific<br>implant based on CT scan data                            | Implants are<br>preformed/pre-shaped to<br>fit the anatomy of the<br>patient  | Custom sized to each patient using CT scan<br>data                                                             |                                          |
| Sterilization              | Provided Non-sterile                                                                                                                                     |                                   | Provided Non-sterile                                                                    | Provided Non-sterile                                                          | Provided Non-sterile                                                                                           |                                          |
| Product Code               | GWO (882.5320, Preformed Alterable<br>Cranioplasty Plate)                                                                                                |                                   | GWO                                                                                     | GXN                                                                           | GXN (882.5330, Preformed Non-alterable<br>Cranioplasty Plate)                                                  |                                          |
| Classification             | Class II                                                                                                                                                 |                                   | Class II                                                                                | Class II                                                                      | Class II                                                                                                       |                                          |
| Perfusion Holes            | X                                                                                                                                                        | X                                 | X                                                                                       | -                                                                             | X                                                                                                              | X                                        |
| IFS                        | X                                                                                                                                                        | X                                 | -                                                                                       | -                                                                             | -                                                                                                              | -                                        |
| Temporal Cutback           | X                                                                                                                                                        | X                                 | -                                                                                       | -                                                                             | -                                                                                                              | -                                        |
| Multi-Part Implant         | X                                                                                                                                                        | X                                 | -                                                                                       | X                                                                             | -                                                                                                              | -                                        |

{5}------------------------------------------------

## KELYNIAM

# K182711 510(k) Summary

| Performance Data                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Data<br>Biocompatibility<br>Testing:<br>Performance<br>data has been<br>provided in<br>support of the<br>substantial<br>equivalence<br>determination                                                                                                                                                                 | • Biocompatibility was leveraged from previously cleared devices. A risk assessment was conducted to evaluate<br>the impact of design changes on the biocompatibility of subject device. The results demonstrate that the<br>subject device meets biological safety requirements per ISO 10993-1 for permanently implanted devices that<br>have tissue/bone and cerebrospinal fluid contact.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Performance<br>Testing Bench:<br>Performance<br>data has been<br>provided in<br>support of the<br>verification &<br>validation testing<br>conducted for a<br>determination of<br>substantially<br>equivalence. The<br>subject devices<br>are safe,<br>effective, and<br>substantially<br>equivalent to the<br>predicate<br>devices. | • Invibio PEEK Optima LT-1® and Evonik Vestakeep i4® Material Specifications and Certification of<br>Analysis/Certification of Compliance<br>o ASTM F2026-16 Standard Specification for Polyether ether ketone (PEEK) Polymers for Surgical Implant<br>Applications.<br>o PEEK Certification of Analysis/Certification of Compliance are reviewed and accepted against approved<br>specifications at each receipt as part of receiving inspection activities.<br>• Cleaning Validation testing was conducted in accordance with the following:<br>o ANSI/AAMI ST72:2011 Bacterial endotoxins, Test methods, routine monitoring, and alternatives to batch<br>testing.<br>o USP <161>, USP <85>, EP 2.6.14, and JP 4.01. USP <85> Bacterial Endotoxin Test (BET).<br>• Steam Sterilization Validation testing was conducted in accordance with the following:<br>o ANSI/AAMI/ISO 17665-1:2006, Sterilization of health care products - Moist heat - Part 1 - Requirements for<br>the development, validation, routine control of a sterilization process for medical devices, Annex D.<br>o ANSI/AAMI/ISO 14937:2009, Sterilization of health care products, general requirements for<br>characterization of a sterilizing agent & development, validation, and routine control of a sterilization<br>process for medical devices, Annex D.<br>• Mechanical Testing Validation<br>o There is no industry accepted standard governing mechanical testing for non-load bearing implantable<br>prosthetic plates. A Mechanical Testing Validation Protocol was developed by KGI to validate the subject<br>devices.<br>• Ship Testing Validation testing was conducted in accordance with the following: |
| Summary on<br>Non-Clinical<br>Testing:                                                                                                                                                                                                                                                                                              | o FedEx Packaging Testing Under 150Lb (Same as ITSA-2A-2011 Packaged-Products 150 lb (68 kg) or Less)<br>The safety and compatibility of passive implants in the Magnetic Resonance (MR) environment concern<br>magnetically induced displacement force and torque, radio frequency (RF) heating, and image artifacts. The<br>static magnetic field of the MR system induces displacement forces and torques on magnetic materials: The<br>Kelyniam Global Inc. (KGI) implants are electrically nonconductive or a nonmagnetic item and poses no known<br>hazards in all MR environments. The KGI Customized Craniofacial (CCI) and Customized Skull (CSI) Implants are<br>MR Safe.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Summary on<br>Clinical Testing:                                                                                                                                                                                                                                                                                                     | Clinical Testing was determined not applicable for Kelyniam Global Inc. Customized Craniofacial Implant (CCI)<br>and Customized Skull Implant (CSI).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Conclusion:                                                                                                                                                                                                                                                                                                                         | The biocompatibility testing, steam sterilization testing, cleaning testing, mechanical testing, and ship testing of<br>the subject devices with and without the proposed features was performed and the results of these tests as<br>well as additional product related validations (all samples passed the acceptance criteria) support the device is<br>safe and effective and substantially equivalent to the predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

---

**Source:** [https://fda-staging.innolitics.com/device/K182711](https://fda-staging.innolitics.com/device/K182711)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com