COCET Digital Thermometer (Models KFT-01/KFT-02/KFT-03/KFT-04/KFT-05/KFT-06/KFT-07/KFT-08/KFT-09/KFT-10
K182652 · Kangfu Medical Equipment Factory · FLL · Nov 20, 2018 · General Hospital
Device Facts
Record ID
K182652
Device Name
COCET Digital Thermometer (Models KFT-01/KFT-02/KFT-03/KFT-04/KFT-05/KFT-06/KFT-07/KFT-08/KFT-09/KFT-10
Applicant
Kangfu Medical Equipment Factory
Product Code
FLL · General Hospital
Decision Date
Nov 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
COCET Digital Thermometer measures body temperature. This digital thermometer is used in the armpit (axillary) to take temperature measuring results. The device is for adult use.
Device Story
Hand-held electronic thermometer; utilizes NTC thermistor sensor to detect temperature; converts electrical signal via PCB to digital readout on LCD screen. Device used in home or clinical settings by adults for self-measurement or by clinicians. Powered by replaceable 1.5V LR41 battery. ABS plastic housing with stainless steel tip. Provides digital temperature display to user; assists in monitoring body temperature for health assessment.
Clinical Evidence
Bench testing only. Compliance with ISO 80601-2-56:2017 for clinical thermometer performance, IEC 60601-1 for electrical safety, and IEC 60601-1-2 for EMC. Biocompatibility testing performed per ISO 10993-1, including cytotoxicity (ISO 10993-5), skin irritation, and sensitization (ISO 10993-10).
Indicated for adult patients for the measurement of body temperature via the axillary (armpit) route.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
DIGITAL THERMOMETER (Models: YT301/ YT302/ YT303) (K120004)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
November 20, 2018
Kangfu Medical Equipment Factory Mr. Zelong Cao Quality Manager No. 380 Ningkang East Road Yueqing City, Zhejiang 325600 CHINA
Re: K182652
Trade/Device Name: COCET Digital Thermometer (Models: KFT-01/KFT-02/KFT-04/KFT-04/KFT-05/KFT-06/KFT-07/KFT-08/KFT-09/KFT-10) Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: September 24, 2018 Received: September 24, 2018
Dear Mr. Zelong Cao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, >Tina Kiang - S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182652
Device Name
COCET Digital Thermometer (Models: KFT-01/KFT-02/KFT-03/KFT-07/KFT-07/KFT-07/KFT-08/KFT-09/KFT-09/KFT-10)
Indications for Use (Describe)
COCET Digital Thermometer measures body temperature. This digital thermometer is used in the armpit (axillary) to take temperature measuring results. The device is for adult use.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
[Refer to 21 CFR §807.92]
## K182652
| 1. | Submitted by: | Kangfu Medical Equipment Factory<br>No.380 Ningkang East Road, Yueqing City 325600 Zhejiang Province<br>China<br>Phone: 086-577-55775583<br>E-mail: zelong_858@126.com<br>Contact Person: Zelong Cao<br>Date Prepared: November 16, 2018 |
|----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Trade Name: | COCET Digital Thermometer (Models: KFT-01/KFT-02/KFT-03/KFT-04/KFT-05/KFT-06/ KFT-07/ KFT-08/ KFT-09/KFT-10 )<br>Common Name: Clinical electronic thermometer<br>Regulation Number: 21 CFR 880.2910<br>Regulation Name: Clinical electronic thermometer<br>Classification: Class II<br>Product Code: FLL |
| 3. | Predicate Device: | K120004, DIGITAL THERMOMETER (Models: YT301/ YT302/ YT303)<br>Manufacturer: CHINA (Shenyang) Med-land Imp/Exp Corp., Ltd. |
| 4. | Device Description: | COCET Digital Thermometer (Models: KFT-01/KFT-02/KFT-03/KFT-04/ KFT-05/KFT-06/ KFT-07/KFT-08/ KFT-09/KFT-10) is a hand-held<br>electronic thermometer, it comprised of an electronic thermo resistor<br>(thermistor) sensor connected to printed circuit board(PCB) with liquid<br>crystal display(LCD) user readout. The outer material of thermometer is<br>made of ABS and the measuring tip is stainless steel. The thermometer<br>using a temperature sensor to output an electrical signal, and then<br>converting the current signal into a liquid crystal digital display<br>temperature. The thermometers are powered by LR41 battery. The<br>battery can be replaced. |
| 5. | Indications for Use: | COCET Digital Thermometer measures body temperature. This digital<br>thermometer is used in the armpit (axillary) to take temperature<br>measuring results. The device is for adult use. |
#### Predicate Device Comparison 6.
Table 1 Predicate Product Comparisons - COCET Digital Thermometer Models: KFT-01/ KFT-02/ KFT-03/ KFT-04/ KFT-05/ KFT-06/ KFT-07/ KFT-08/ KFT-09/ KFT-10
{4}------------------------------------------------
| Feature | Kangfu Medical Equipment<br>Factory Models: KFT-01/ KFT-<br>02/ KFT-03/ KFT-04/ KFT-05/<br>KFT-06/ KFT-07/ KFT-08/ KFT-<br>09/ KFT-10 | CHINA (Shenyang) Med-<br>land Imp/Exp Corp.,Ltd<br>YT301/ YT302/ YT303 | Discussion |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| 510(k) Number | K182652 | K120004 | Different |
| Indications for Use | COCET Digital Thermometer<br>measures body temperature.<br>This digital thermometer is<br>used in the armpit (axillary) to<br>take temperature measuring<br>results. The device is for adult<br>use. | The Digital Thermometer is<br>used for the measurement<br>and monitoring of human<br>body temperature. Body<br>temperature can be<br>measured with the Digital<br>Thermometer by oral,<br>axillary (under the arm) and<br>rectal methods. The device<br>is for adult and pediatric<br>use. | Different |
| thermometer type | under arm | oral, axillary (under the<br>arm) and rectal. | Different |
| components | Electronic thermo<br>resistor(thermistor) sensor<br>Printed circuit board(PCB)<br>Liquid crystal display(LCD)<br>Outer<br>Tip | Electronic thermo<br>resistor(thermistor) sensor<br>Printed circuit board(PCB)<br>Liquid crystal display(LCD)<br>Outer<br>Tip | Same |
| principal of operation | Using a temperature sensor to<br>output an electrical signal, and<br>then converting the current<br>signal into a liquid crystal<br>digital display temperature | Using a temperature sensor<br>to output an electrical<br>signal, and then converting<br>the current signal into a<br>liquid crystal digital display<br>temperature | Same |
| Temperature Measurement<br>Technology | NTC Thermistor Resistance<br>Technique | NTC Thermistor Resistance<br>Technique | Same |
| Key Temperature<br>Sensor | NTC Thermistor | NTC Thermistor | Same |
| Power requirements | 1xLR41 battery<br>Standard voltage 1.5V,<br>termination voltage 0.9V.<br>Discharge with 1K discharge<br>resistor, sustainable discharge<br>for 670 hours. Working | 1xLR41 battery | Same |
| capacity 28mAh, weight 0.61g.<br>Load resistance 22K | | | |
| Materials | ABS plastic cabinet (White and blue), Stainless steel | ABS plastic cabinet (White), Stainless steel | Different |
| Temperature range | 32.0 °C ~ 42.9 °C (90°F-109.2 °F) | 32.0 °C ~ 42.9 °C (90°F - 109.2 °F) | Same |
| Accuracy | +/- 0.1 °C(±0.2°F) from 35.5 °C to 42 °C(95.9°F-107.6°F)<br>+/- 0.2 °C(± 0.4 °F) in the rest measuring range | +/- 0.1 °C( ± 0.2 °F) from 35.5 °C to 42 °C(95.9°F-107.6 °F)<br>+/- 0.2 °C(±0.4°F) in the rest measuring range | Same |
| Biocompatibility | Comply with ISO 10993-1<br>ISO10993-5<br>ISO 10993-10 | Comply with ISO 10993-1<br>ISO10993-5<br>ISO 10993-10 | Same |
| Voluntary standards for Clinical Electronic Thermometers | Comply with ISO 80601-2-56 | Comply with ASTME1112-00 | Different |
| Medical Electrical Safety and EMC | IEC 60601-1, IEC 60601-1-2 | IEC 60601-1, IEC 60601-1-2 | Same |
{5}------------------------------------------------
Discussion:
1. Indications for Use is different from the description of the contrasting devices are measuring human body temperature, the predicate device is used orally, axillary (under the arm) and rectal for adult and pediatric, our COCET Digital Thermometer is only used under arm for adult, the difference does not raise performance questions.
2. Thermometer type: Our product COCET Digital Thermometer is only used in the armpit, the predicated device can be used orally , axillary (under the arm) and rectal, the difference does not raise performance questions.
3. Materials: The ABS plastic cabinet of predicate device only has white color, COCET Digital Thermometer has two color white and blue. However, it is conformed with ISO 10993 series standard. It can be proved in the biocompatibility testing report. The difference does not raise performance questions.
4. Voluntary standards for Clinical Electronic Thermometers: Electronic Thermometer comply with ASTME1112-00 and COCET Digital Thermometer comply with The ISO 80601-2-56, the two standards were meets FDA guidelines. The difference does not raise performance questions.
| Name | Model | principal of operation | Dimension | Materials |
|------------------------------|--------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------|
| COCET DIGITAL<br>THERMOMETER | KFT-01 | This digital thermometer is used in<br>the armpit (axillary) and using a<br>temperature sensor to output an<br>electrical signal, and then converting<br>the current signal into a liquid crystal<br>digital display temperature | 124.5mm ×<br>18mm×<br>9mm | ABS plastic<br>cabinet<br>(White<br>code:WH),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-02 | same | 120mm ×<br>16mm ×<br>9.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-03 | same | 127mm×<br>19mm ×<br>11.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-04 | same | 127.5mm ×<br>19mm ×<br>11.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-05 | same | 133mm ×<br>27.5mm ×<br>10mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-06 | same | 132mm ×<br>20mm ×<br>7.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU), |
| | | | | |
| COCET DIGITAL<br>THERMOME | KFT-07 | same | 127mm×<br>19mm ×<br>11.5mm | Stainless<br>steel<br>ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU) |
| COCET DIGITAL<br>THERMOME | KFT-08 | same | 127.5mm ×<br>19mm ×<br>11.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-09 | same | 127mm ×<br>19mm ×<br>11.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
| COCET DIGITAL<br>THERMOME | KFT-10 | same | 127.5mm ×<br>19mm ×<br>11.5mm | ABS plastic<br>cabinet<br>(White<br>code:WH<br>Blue<br>code:BU),<br>Stainless<br>steel |
{6}------------------------------------------------
{7}------------------------------------------------
#### 7. Non-clinical Tests
Safety and EMC:
Safety and EMC test was performed in according to the
- IEC 60601-1:2005+CORR.1:2006+CORR.2.2007+a1:2012 . Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 :2014 -
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential
{8}------------------------------------------------
performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests Biocompatibility:
Biocompatibility testing was performed to evaluate the biocompatibility of the contact materials in accordance with ISO 10993-1:2009, the device passed each biocompatibility test identified below:
- । Cytotoxicity testing in according to ISO 10993-5:2009 Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
- Skin irritation testing in according to ISO 10993-10:2010 Biological evaluation of medical devices--Part 5: Tests for in vitro cytotoxicity
- Sensitization testing in according to ISO 10993-10:2010 Biological evaluation of medical devices- । Part 5: Tests for in vitro cytotoxicity
#### Performance Data:
The following bench testing was conducted in order to support substantial equivalence:
- । ISO 80601-2-56:2017 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.
### 8. Conclusion
Based on performance testing and compliance with voluntary standards demonstrate that the proposed device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.