K182625 · DiaDent Group International · KIF · Jun 3, 2019 · Dental
Device Facts
Record ID
K182625
Device Name
Diapex Plus
Applicant
DiaDent Group International
Product Code
KIF · Dental
Decision Date
Jun 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
DIAPEX PLUS is a calcium hydroxide paste with iodoform, used as a temporary root canal filling material. Application : Root canal filling material / Apexification and hard tissue formation/ Apexogensis
Device Story
Diapex Plus is a premixed calcium hydroxide and iodoform paste used as a temporary root canal filling material. It is supplied in a syringe with disposable tips and a silicon cap. The device is used by dental professionals in a clinical setting. The clinician inserts the syringe tip into the prepared root canal, extrudes the paste by pressing the piston, and removes the tip. The paste serves as a temporary filling to facilitate apexification and hard tissue formation. After the temporary period, the material is removed from the canal. The device benefits patients by providing a temporary barrier and therapeutic agent during endodontic procedures.
Clinical Evidence
No clinical data was collected or provided. Substantial equivalence is supported by non-clinical bench testing, including flowability, film thickness, and radio-opacity per ISO 6876, and comprehensive biocompatibility testing per ISO 10993 series (cytotoxicity, sensitization, systemic toxicity, genotoxicity).
Technological Characteristics
Premixed paste containing calcium hydroxide, iodoform, polydimethylsiloxane, and olive oil. Delivered via syringe with disposable tips. Conforms to ISO 6876 for root canal sealing materials and ISO 10993 for biocompatibility. Non-sterile delivery; requires disinfection before use.
Indications for Use
Indicated for use as a temporary root canal filling material, apexification, and hard tissue formation/apexogenesis in patients requiring endodontic treatment.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 3, 2019
Diadent Group International Myung Sub Kim Quality Assurance Manager 16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu Cheongji-si, Chungcheongbuk-do, 28161 Republic of Korea
Re: K182625
Trade/Device Name: Diapex Plus Regulation Number: 21 CFR 872.3820 Regulation Name: Root canal filling resin Regulatory Class: Class II Product Code: KIF Dated: April 15, 2019 Received: April 19, 2019
Dear Myung Sub Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina B Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name DIAPEX PLUS
Indications for Use (Describe)
DIAPEX PLUS is a calcium hydroxide paste with iodoform, used as a temporary root canal filling material. Application : Root canal filling material/Apexification and hard tissue formation/Apexogensis
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Product Name : DIAPEX PLUS
# 5.0 510(k) Summary
This summary of 510(k) information is submitted in accordance with the requirements of 21 CFR 807.92.
# 5.1 Application Information
| Date Prepared: | 21st September, 2018 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Name and Address: | DiaDent Group International<br>16, Osongsaengmyeong 4-ro, Osong-eup, Heungdeok-gu, Cheongju-<br>si, Chungcheongbuk-do, 28161, Republic of Korea |
| Contact Person: | Myung Sub Kim<br>Quality Assurance Manager<br>Phone: +82-43-266-2315<br>FAX: +82-43-235-2315<br>Email: diadent32@diadent.co.kr |
# 5.2 Device Information
| Device Type: | Root Canal Filling Resin |
|-------------------------|--------------------------|
| Regulation Description: | Root Canal Filling Resin |
| Review Panel: | Dental |
| Regulation Number: | 21 CFR 872.3820 |
| Product Code: | KIF |
| Device Class: | II |
| Device Name: | DIAPEX PLUS |
## 5.3 Predicate Devices
The legally marketed devices to which substantial equivalence is being claimed are:
| 510(k) Number: | K973667 | K032603 |
|--------------------|------------------------------------------|-----------------------|
| Applicant: | Neo Dental Chemical Products<br>Co, Ltd. | Meta Biomed Co., Ltd. |
| Device Name: | Vitapex | Metapex |
| Regulation Number: | 21 CFR 872.3820 | 21 CFR 872.3820 |
| Product Code: | KIF | KIF |
| Device Class: | II | II |
# 5.4 Device Description
The subject device is packaged with the following:
- 0 Syringe
- Disposable Tip .
- Silicon Cap .
# 5.5 Intended Use/Indications for Use
DIAPEX PLUS is a calcium hydroxide paste with iodoform, used as a temporary root canal filling material. Application : Root canal filling material / Apexification and hard tissue formation/ Apexogensis
{4}------------------------------------------------
# Product Name : DIAPEX PLUS
# 5.6 Technological Characteristics
This device compares to the legally marketed devices as follows:
| | Proposed Device | Primary predicate<br>device | Reference<br>predicate device | Discussion |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DiaPex Plus | Vitapex | Metapex | |
| 510(k)<br>Number | | K973667 | K032603 | - |
| Intended<br>Use/Indic<br>ations for<br>Use | DIAPEX PLUS is a<br>calcium hydroxide<br>paste with iodoform,<br>used as a temporary<br>root canal filling<br>material<br>Application : Root<br>canal filling material /<br>Apexification and<br>hard<br>tissue<br>formation/<br>Apexogensis | VitaPex can be used<br>as a temporary or<br>permanent root canal<br>filling material. It is<br>ideal for the treatment<br>of infected root canals<br>and for vital<br>pulpotomies In<br>deciduous teeth. | •Exposed pulp in<br>capping and<br>pulpectomy<br>•Leakage canal<br>•Apexification<br>•Formation of hard<br>tissue barrier<br>•Root canal filling<br>material | equivalent |
| Direction<br>s for Use | •Before using Diapex<br>Plus, ensure that<br>the syringe and tip<br>are free of<br>damage(e.g. chips,<br>cracks, deformation,<br>and/or other<br>defects)<br>•Wipe and disinfect<br>the outer syringe<br>and tip with alcohol<br>as tips come<br>unsterilized.<br>•After root canal<br>preparation, the<br>canal is irrigated<br>and dried.<br>•Insert the tip of the<br>Diapex Plus syringe<br>to one fifth of foot<br>canal length from<br>the apex. the tip is<br>bent to obtain<br>easier access to<br>canal.<br>•Slowly and gently<br>press the piston of<br>the syringe to<br>extrude the paste.<br>the paste should<br>reach the periapical<br>tissue and the tip<br>should be pulled out | Extrude VitaPex into<br>the canal.<br>place temporary<br>restoration.<br>At subsequent visits<br>remove VitaPex with<br>sodium hydrochloride<br>and mechanical<br>means. | •Clean and dry it<br>after preparing the<br>root canal<br>•Set up a disposable<br>tip in syringe and<br>insert the ring<br>provided for and<br>easy re-direction of<br>the tip<br>•Disinfect the<br>disposable tip with<br>ethanol<br>•Fill the canal<br>completely by<br>pushing the syringe<br>plunger while<br>withdrawing the tip<br>slowly<br>•Clean any excess<br>paste with a<br>sterilized cotton<br>pellet<br>•Remove the used<br>tip, fix a new one and<br>cover up | equivalent |
| | Proposed Device | Primary predicate<br>device | Reference<br>predicate device | Discussion |
| | DiaPex Plus | Vitapex | Metapex | |
| | slowly. the paste<br>should reverse back<br>to the pulp<br>chamber.<br>•After placing the<br>paste in the canal<br>and fully removing<br>the syringe, clean<br>off any excess<br>paste.<br>•After each<br>application, the<br>used tip should be<br>discarded and a<br>new tip or the<br>silicon cap put back<br>on | | | |
| Package<br>Contents | •Syringe<br>•Disposable Tip<br>•Silicon Cap | •Syringe<br>•Disposable Tip | •Syringe<br>•Disposable Tip<br>•One ring rotator for<br>the direction control<br>of the tip | |
| Period of<br>Use | Temporary (remains<br>in the body for 29<br>days or less) | Temporary | Temporary | |
| Composit<br>ion | Calcium Hydroxide<br>IODOFORM<br>Polydimethylsiloxane<br>Olive Oil | Calcium Hydroxide<br>IODOFORM<br>Silicone Oil<br>Inert | Calcium Hydroxide<br>IODOFORM<br>Polydimethylsiloxane<br>Silicone Oil | The main<br>ingredients are<br>similar and<br>have some<br>other<br>ingredients.<br>However,<br>biocompatibilit<br>y and<br>performance<br>tests confirm<br>that DiaPex<br>Plus and<br>similar<br>products are<br>substantially<br>equivalent. |
| Biocomp<br>atibility | Biocompatible | Biocompatible | Biocompatible | equivalent |
| Delivery<br>forms | Premixed paste | Premixed paste | Premixed paste | equivalent |
| Standard<br>S | ISO 6876 | ISO 6876 | ISO 6876 | equivalent |
{5}------------------------------------------------
## 510k DRAFT SUBMISSION
## Diadent Group International
## Product Name : DIAPEX PLUS
As demonstrated in the above comparison table, the subject and predicate devices have simil ar intended uses, contraindications, composition and contents, Base and Catalyst part mixture
{6}------------------------------------------------
#### Product Name : DIAPEX PLUS
design, shelf life, biocompatibility, and conformance with standards.
#### 5.7 Non-Clinical Performance Data
This device has demonstrated conformance with non-clinical performance requirements through evaluation and testing in accordance with the following harmonized standards:
| Standard | contents | Relevant test | Relevant File |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------------------------------|
| ISO 6876:2012 | Dentistry -- Root canal<br>sealing materials | •Flowability<br>•Film Thickness<br>•Radio-opacity | 18.0 Performance<br>Testing- Bench |
| ISO 7405:2008 | Dentistry -- Evaluation of biocompatibility of medical<br>devices used in dentistry | | 15.0<br>Biocompatibility |
| ISO 10993-1:2009 | Biological evaluation of medical devices -- Part 1:<br>Evaluation and testing within a risk management process | | 15.0<br>Biocompatibility |
| ISO 10993-3:2014 | Biological evaluation of<br>medical devices -- Part 3:<br>Tests for genotoxicity,<br>carcinogenicity and<br>reproductive toxicity | •Genotoxicity Bacterial<br>Reverse Mutation<br>•Genotoxicity Mouse<br>Lymphoma Assay | 15.0<br>Biocompatibility |
| ISO 10993-5:2009 | Biological evaluation of<br>medical devices -- Part 5:<br>Tests for in vitro cytotoxicity | •Cytotoxicity | 15.0<br>Biocompatibility |
| ISO 10993-6: 2016 | Biological evaluation of<br>medical devices -- Part 6:<br>Tests for local effects after<br>implantation | •4 Week Systemic Toxicity | 15.0<br>Biocompatibility |
| ISO 10993-10:2010 | Biological evaluation of<br>medical devices -- Part 10:<br>Tests for irritation and skin<br>sensitization | •Sensitization | 15.0<br>Biocompatibility |
| ISO 10993-11:2017 | Biological evaluation of<br>medical devices -- Part 11:<br>Tests for systemic toxicity | •Acute Systemic Toxicity<br>•Pyrogenicity<br>•4 Week Systemic Toxicity | 15.0<br>Biocompatibility |
Additional non-clinical bench testing demonstrates the substantial equivalence of the subject device.
#### 5.8 Clinical Performance Data
No clinical data was collected or provided to support substantial equivalence between the subject and predicate devices.
#### 5.9 Conclusions
Based on the above information and all data provided in this submission, the comparison of intended uses, technological characteristics, and non-clinical performance testing demonstrates that the subject device is substantially equivalent to the legally marketed devices identified in this submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.