1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, 1.5T 30ch AIR AA
K182590 · GE Healthcare · MOS · Oct 19, 2018 · Radiology
Device Facts
Record ID
K182590
Device Name
1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, 1.5T 30ch AIR AA
Applicant
GE Healthcare
Product Code
MOS · Radiology
Decision Date
Oct 19, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR AA are receive only RF coils designed for use with GE 1.5T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.
Device Story
Receive-only RF coils for GE 1.5T MRI systems; detect hydrogen nuclei to produce diagnostic images of general human anatomy and extremities. Coils feature 16, 20, 21, or 30 elements with integrated preamplifiers for improved signal-to-noise ratio and uniform coverage. Flexible design with soft materials conforms to patient anatomy to enhance comfort; optional coil holders assist in positioning. Operated by MRI technicians or clinicians in clinical imaging settings. Output is raw signal data processed by the MRI system into diagnostic images for physician interpretation. Benefits include high-quality imaging and patient comfort due to flexible form factor.
Clinical Evidence
No clinical studies required. Substantial equivalence supported by bench testing, risk management, and sample clinical images.
Technological Characteristics
Receive-only RF coils; 16, 20, 21, or 30 elements; integrated preamplifiers; P-connectors. Enclosures made of flame-rated V2/V0 co-polyester. Compliant with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, NEMA MS 6, and NEMA MS 9. Flexible form factor.
Indications for Use
Indicated for use with GE 1.5T MRI systems to produce diagnostic images of general human anatomy, including extremities, in patients requiring MRI examinations. No specific disease states or contraindications listed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
GE Healthcare Coils (USA Instruments, Inc.) Veronica Meridith Regulatory Affairs Leader 1515 Danner Drive AURORA OH 44202
October 19, 2018
### Re: K182590
Trade/Device Name: 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, 1.5T 30ch AIR AA Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: MOS Dated: September 19, 2018 Received: September 20, 2018
Dear Veronica Meridith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182590
Device Name 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, 1.5T 30ch AIR AA
Indications for Use (Describe)
The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR AA are receive only RF coils designed for use with GE 1.5T MRI systems to produce diagnostic images on general human anatomy including extremities. The nucleus detected is hydrogen.
Type of Use (Select one or both, as applicable)
| <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------|
| <span> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has a swirling design around the letters, giving it a dynamic and recognizable appearance.
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | Sept 19, 2018 |
|---------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare Coils (USA Instruments, Inc.)<br>1515 Danner Drive<br>Aurora, OH 44202<br>USA |
| Primary Contact Person: | Veronica Meridith<br>Regulatory Affairs Leader<br>GE Healthcare<br>Phone: 262-955-5427<br>Fax: 414-908-9585 |
| Secondary Contact Person: | Andrew Menden<br>Regulatory Affairs Manager<br>GE Healthcare<br>Phone: 262-521-6223<br>Fax: 414-908-9585 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the text "GE Healthcare" in bold font, followed by the text "510(k) Premarket Notification Submission" in a smaller font size. The text is left-aligned and appears to be the title or heading of a document. The text is black and the background is white.
Image /page/4/Picture/2 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized, cursive font in the center. The circle is surrounded by a series of swirling lines, which give the impression of movement or energy. The logo is simple, yet recognizable, and it is associated with a well-known and established company.
| Device Trade Name: | 1.5T AIR MP M<br>1.5T AIR MP L<br>1.5T 16ch AIR AA<br>1.5T 30ch AIR AA |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual<br>Name: | Coil, Magnetic Resonance, Specialty |
| Classification Names: | Magnetic Resonance Diagnostic Device per 21 CFR 892.1000 |
| Product Code: | MOS |
| Predicate Device(s): | 3.0T Anterior Array (K172695) |
| Device Description: | The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA and 1.5T 30ch AIR<br>AA are receive-only coils designed to provide optimum signal-to noise ratio and<br>uniform coverage of general human anatomy including extremities and are<br>designed for use with GE 1.5T MRI Systems. These are 20, 21, 16, and 30<br>element coils tuned to image proton nuclei. Each coil element has an integrated<br>preamplifier to improve image quality. These coils are provided with P-<br>connectors. The coils have a soft material to conform to the patient's anatomy<br>and maximize patient comfort. Due to the flexibility of the coils, a coil holder<br>may be used to assist in securing the coil in place. |
| Intended Use: | The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR<br>AA are receive only RF coils designed for use with GE 1.5T MRI systems to<br>produce diagnostic images on general human anatomy including extremities. The<br>nucleus detected is hydrogen. |
| Comparison of<br>Intended Use: | The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR<br>AA and the predicate device are classified as coils for magnetic resonance<br>imaging devices and are intended for diagnostic use. Both indications for use<br>statements are functional in nature, and do not list specific diseases or<br>conditions. The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and<br>1.5T 30ch AIR AA and the predicate device are indicated for the same patient<br>population, and for the same clinical setting. |
| | The proposed Indications for use are simplified to reflect current coil intent,<br>similar to the Siemens Contour 24 coil K173446, 510k cleared 11/17/2017. The<br>Siemens Contour 24 coil Indications for Use is: "The Contour 24 is intended for<br>use with Siemens 3.0T MR systems to produce diagnostic images of general<br>human anatomy that can be interpreted by a trained physician." |
| | Therefore, GE Healthcare believes that the 1.5T AIR MP M, 1.5T AIR MP L,<br>1.5T 16ch AIR AA, and 1.5T 30ch AIR AA has the same intended use as the<br>predicate device in accordance with the FDA's guidance document "The 510(k)<br>Program: Evaluating Substantial Equivalence in Premarket Notifications<br>[510(k)]", dated 28 July 2014. |
| Comparison of<br>Technological<br>Characteristics: | The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR<br>AA and the predicate device share many common technological characteristics.<br>Details can be found in the table below, which shows the high-level device<br>description for all four new coils as compared to the predicate device:<br><br>The most notable technological differences between the new coils and the<br>predicate device is that the new devices contain (respectively):<br><br>20 elements<br>21 elements<br>16 elements<br>30 elements<br>Predicate: 30 elements<br><br>The predicate is used on GE 3.0T MRI Systems while the new devices are for<br>use with GE 1.5T MRI systems.<br><br>These technological differences do not raise any different questions of safety and<br>effectiveness. All devices must address questions of whether they provide an<br>adequate level of image quality appropriate for diagnostic use. The performance<br>data described in this submission include results of both bench testing and<br>clinical testing that show the image quality performance of the 1.5T AIR MP M,<br>1.5T AIR MP L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR AA compared to the<br>predicate device.<br><br>Operating Principles: The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR<br>AA, and 1.5T 30ch AIR AA operate on the same principles as the predicate<br>devices.<br><br>Materials: The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA, and<br>1.5T 30ch AIR AA electrical elements and associated circuitry are supported<br>throughout the coil by flame rated V2 co-polyester injection mold enclosure and<br>V0 equivalent flame rated enclosures.<br>Both coils and the predicate device are certified to AAMI/ANSI ES60601-1,<br>which includes flammability requirements.<br><br>Safety and Performance Testing: The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T<br>16ch AIR AA, and 1.5T 30ch AIR AA and the predicate devices comply with<br>the same safety and performance testing (see Determination of Substantial<br>Equivalence, below). |
| Determination of<br>Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The predicate and modified devices have been subject to similar risk<br>management testing to demonstrate substantial equivalence of safety and<br>performance. Testing included: |
| | AAMI/ANSI ES60601-1 IEC 60601-1-2 IEC 60601-2-33 NEMA MS 6 NEMA MS 9 Maximum B1 peak Network blocking analysis Heat Testing Additionally, both predicate and modified devices have a successful<br>biocompatibility track record, as demonstrated by ISO 10993 testing and by their<br>history of use in previously cleared devices. |
| | The following quality assurance measures were applied to the development of<br>the devices: |
| | Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) Simulated use testing (Validation) |
| | Summary of Clinical Tests:<br>The subjects of this premarket submission, the 1.5T AIR MP M, 1.5T AIR MP<br>L, 1.5T 16ch AIR AA, and 1.5T 30ch AIR AA, do not require clinical studies to<br>support substantial equivalence. Sample clinical images have been included in<br>this submission. |
| | Substantial Equivalence Conclusion:<br>The indications for use of the proposed devices are comparable to the claimed<br>predicate devices. The 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR AA,<br>and 1.5T 30ch AIR AA employs equivalent technology to the claimed predicate<br>devices. Additionally, the results from the above non-clinical tests demonstrate<br>that the devices perform as intended. Thus, the 1.5T AIR MP M and 1.5T AIR<br>MP L are substantially equivalent to the predicate devices to which they have<br>been compared. |
| Conclusion: | GE Healthcare considers the 1.5T AIR MP M, 1.5T AIR MP L, 1.5T 16ch AIR<br>AA, and 1.5T 30ch AIR AA to be as safe, as effective, and performance is<br>substantially equivalent to the predicate devices. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the text "GE Healthcare" in bold font, followed by the text "510(k) Premarket Notification Submission" in a regular font. The text is left-aligned and appears to be the title or heading of a document. The text is black and the background is white.
Image /page/5/Picture/2 description: The image shows the logo of General Electric (GE). The logo consists of the letters 'GE' intertwined within a blue circle. The letters are stylized and appear to be handwritten. The blue circle has a white border, and the overall design is simple and recognizable.
{6}------------------------------------------------
# GE Healthcare 510(k) Premarket Notification Submission
Image /page/6/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by a series of curved, white lines that resemble water droplets or stylized flames, adding a dynamic element to the design.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.