K182589 · Becton, Dickinson and Company · FMF · Mar 14, 2019 · General Hospital
Device Facts
Record ID
K182589
Device Name
BD Plastipak Syringe
Applicant
Becton, Dickinson and Company
Product Code
FMF · General Hospital
Decision Date
Mar 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The BD Plastipak™ Syringe is intended for single use by health care professionals for general purpose fluid aspiration/injection.
Device Story
BD Plastipak™ Syringe is a three-piece, single-use, sterile or bulk non-sterile hypodermic syringe; available in 3mL and 20mL sizes. Device consists of a lubricated polypropylene barrel with graduated scale, a lubricated thermoplastic elastomer (TPE) stopper, and a polypropylene plunger rod. Operates via manual aspiration (pulling plunger) or injection/expulsion (depressing plunger). Connects to compatible female 6% (Luer) connectors via Luer Slip or male 6% (Luer Lock) connection. Used by healthcare professionals in clinical settings for general fluid management. Output is the delivery or aspiration of fluids; healthcare providers use the graduated scale for volume measurement. Benefits include precise fluid handling and delivery. Device is provided sterile (Irradiation or EtO) or bulk non-sterile.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical design verification and biocompatibility testing. Bench testing included volumetric accuracy, leakage, plunger force, and sterility/biocompatibility (cytotoxicity, hemolysis, systemic toxicity, sensitization, pyrogenicity) per ISO 10993 and USP standards.
Technological Characteristics
Materials: Polypropylene barrel/plunger, TPE stopper, silicone lubricant. Dimensions: 3mL and 20mL. Connectivity: Luer Slip or Luer Lock. Sterilization: Irradiation or EtO (SAL 10^-6). Standards: ISO 7886-1, ISO 7886-2, ISO 10993, USP <87>, <88>, <381>, <788>, <85>.
Indications for Use
Indicated for health care professionals for general purpose fluid aspiration and injection.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Becton Dickinson Single Use Hypodermic Syringes (K980987)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 14, 2019
Becton, Dickinson and Company % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. Nw Buffalo, Minnesota 55313
Re: K182589
Trade/Device Name: BD Plastipak Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: February 14, 2019 Received: February 15, 2019
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Expiration Date: 06/30/2020 | Food and Drug Administration |
| See PRA Statement below. | <b>Indications for Use</b> |
| 510(k) Number (if known) | K182589 |
| Device Name | BD Plastipak <sup>TM</sup> Syringe |
| Indications for Use (Describe) | The BD Plastipak <sup>TM</sup> Syringe is intended for single use by health care professionals for general purpose fluid aspiration/injection. |
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="float:left">☒</span> <span style="float:left">Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span style="float:left">☐</span> <span style="float:left">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|
PSC Publishing Services (301) 443-6740 EF
0881 to 48 app 1
{3}------------------------------------------------
bd.com
Image /page/3/Picture/2 description: The image shows the BD logo. The logo consists of an orange circle with a white starburst-like design inside it. To the right of the circle are the letters "BD" in blue. The letters are in a bold, sans-serif font.
# 510(k) Summary (21 CFR §807.92) BD Plastipak™ Syringe
| Submitter: | Leslie Robinson-Frye<br>Staff Regulatory Affairs Specialist<br>Becton, Dickinson and Company<br>1 Becton Drive<br>Franklin Lakes, NJ 07417<br>Phone: (201) 847-6891<br>Fax: (201) 847-4306 | |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:<br>Date Prepared: | Leslie Robinson-Frye<br>March 13, 2019 | |
| Subject Devices: | Name of Device:<br>Common Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code: | BD Plastipak™ Syringe<br>Piston Syringe<br>Piston Syringe, (21 CFR §880.5860)<br>II<br>FMF |
| Predicate Devices: | Name of Device:<br>510(k) Reference:<br>Common Name:<br>Classification:<br>Product Code: | Becton Dickinson Single Use Hypodermic Syringes<br>K980987<br>Piston Syringe<br>Piston Syringe, (21 CFR §880.5860, Class II device)<br>FMF |
This predicate has not been subject to a design-related recall.
#### Reason for Submission
The purpose of this submission is to modify the stopper material, barrel/stopper lubricant, stopper dimensions, stopper manufacturing location, product name, and label statements (pump claim and do not re-sterilize) of the predicate Becton Dickinson Single Use Hypodermic Syringes.
### Device Description
The 3mL and 20mL BD Plastipak™ Syringe is a three-piece, single use, sterile or bulk non-sterile (BNS) hypodermic syringe with Luer Slip or male 6% (Luer Lock) connection, which are connectable to a compatible female 6% (Luer) connector. The syringe assemblies consist of a lubricated polypropylene barrel with a graduated scale, a lubricated thermoplastic elastomer stopper and a polypropylene plunger rod. The plunger rod is pulled back to aspirate fluids or depressed to inject or expel fluids. The barrel scale of the BD Plastipak™ Syringe incorporates a scale graduated in units of milliliters.
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image contains the logo for BD, a global medical technology company. The logo consists of two parts: a stylized sun-like symbol in orange on the left and the letters "BD" in blue on the right. The sun-like symbol has a central circle with radiating lines, and the letters "BD" are in a bold, sans-serif font.
The BD Plastipak™ Syringe is provided sterile by Irradiation or Ethylene Oxide Gas (EtO) sterilization methods in a syringe only configuration or with a BD Hypodermic Needle, BD SafetyGlide™ Hypodermic Needle or BD Eclipse™ Hypodermic Needle. The BD Plastipak™ Syringe is also provided in a bulk nonsterile, syringe only configuration.
#### Intended Use
The BD Plastipak™ Syringe is intended for single use by health care professionals for general purpose fluid aspiration/injection.
### Comparison of Technological Characteristics
The subject BD Plastipak ™ Syringe is equivalent to that of the predicate BD Hypodermic Syringe in intended use, materials and performance characteristics.
The stopper material, lubricant and dimensions are different between the subject and predicate devices However, the above differences are not critical in the safety and effectiveness of the device when used as labeled as demonstrated through functional testing.
| Element of Comparison | Subject Device | Predicate Device | |
|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
| Indications for Use/Intended Use | The BD Plastipak™ Syringe is intended for single use by health care professionals for general purpose fluid aspiration/injection. | These syringes are intended for use by health care professionals for general purpose fluid aspiration/injection. | |
| Syringe Materials | Plunger | Polypropylene | Polypropylene |
| | Barrel | Polypropylene | Polypropylene |
| | Barrel Silicone | Silicone | Silicone |
| | Stopper | Thermo Plastic Elastomer (TPE) Copolymer | Polyisoprene Rubber |
| | Stopper Silicone | Silicone | Silicone |
| Syringe Dimensions | Stopper | - Stopper Outer Diameter<br>-Stopper Inner Diameter | - Stopper Outer Diameter<br>-Stopper Inner Diameter |
| | | | |
| Sterilization Methods | EtO<br>Irradiation | EtO<br>Irradiation | |
| SAL | 10-6 | 10-6 | |
| Shelf Life | 5 Years | 5 Years | |
| Functional Testing: | | | |
| Startup Time | ISO7886-2 | ISO7886-2 | |
| Volumetric Accuracy<br>(Manual Use, tolerance on graduated capacity) | ISO7886-1 and ISO 7886-2 | ISO7886-1 and ISO 7886-2 | |
| Sterility Product<br>(N/A* for BNS) | ISO 11135 , ISO 11137 and<br>PH Eur. 5.0 | ISO 11135 , ISO 11137 and<br>PH Eur. 5.0 | |
| Not Manufactured with BPA | Manufacturing process does not<br>include BPA | Manufacturing process does<br>not include BPA | |
| Not made with natural rubber latex | Not made with natural rubber<br>latex | Not made with natural<br>rubber latex | |
| Stopper Separation from plunger | ISO7886-1 | ISO7886-1 | |
| Liquid leakage past stopper under<br>pressure (back rib) | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Liquid leakage past stopper under<br>pressure (front rib) | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Air Leakage past Stopper in<br>aspiration | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Dead space | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Maximum Usable Capacity | ISO7886-1 | ISO7886-1 | |
| Fiducial Line | ISO7886-1 | ISO7886-1 | |
| Hard Height | ISO7886-2 | ISO7886-2 | |
| Short-Term Flow Rate Accuracy<br>(incl Stiction) | ISO 7886-2 | ISO 7886-2 | |
| Overall Percentage Error | ISO7886-2 | ISO7886-2 | |
| Max Variation in Flow Rate | ISO7886-2 | ISO7886-2 | |
| Plunger Movement Forces (Pump) | ISO7886-2 | ISO7886-2 | |
| Syringe Compliance | ISO7886-2 | ISO7886-2 | |
| Uniform Stopper Appearance | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Scale -<br>Graduation Line Intervals | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Scale -<br>Unit of measure | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Scale Lines | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Scale Numbering | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| | | | |
| Scale Length | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Scale Position | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Marking - Syringe Barrel | ISO7886-1 and ISO7886-2 | ISO7886-1 and ISO7886-2 | |
| Air leakage past stopper in<br>aspiration (vacuum) | ISO7886-1 | ISO 7886-1 | |
| Stopper separation from plunger | ISO7886-1 | ISO 7886-1 | |
| | Biocompatibility Testing: | | |
| Cytotoxicity | ISO 10993-5, 10993-12 and<br>USP <87> | ISO 10993-5, 10993-12 and<br>USP <87> | |
| Hemolysis | Per ISO 10993-4 and 10993-12 | Per ISO 10993-4 and 10993-12 | |
| Acute Systemic Toxicity | Per ISO 10993-11 and 10993-12 | Per ISO 10993-11 and 10993-12 | |
| Intracutaneous Reactivity | Per ISO 10993-10, 10993-12 USP<br>and <88> | Per ISO 10993-10, 10993-12<br>USP and <88> | |
| Sensitization | Per ISO 10993-10 | Per ISO 10993-10 | |
| Pyrogenicity | Per ISO 10993-11, 10993-12 and<br>USP 151 | Per ISO 10993-11, 10993-12<br>and USP 151 | |
| Elastomeric Closures for Injections | USP <381> | USP <381> | |
| ISO 7886-1: Acidity/Alkalinity and<br>Extractable Metals | ISO 7886-1 | ISO 7886-1 | |
| Extractables/Leachables Assessment | ISO 10993-17 and ISO 10993-18 | ISO 10993-17 and<br>ISO 10993-18 | |
| Particulate Matter | USP <788> | USP <788> | |
| Endotoxin, LAL | USP <85> | USP <85> | |
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows the BD logo. The logo consists of an orange circle with white rays emanating from the center, next to the letters "BD" in blue. The letters are bold and sans-serif.
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the BD logo. The logo consists of two parts: a stylized orange sun-like symbol on the left and the letters "BD" in blue on the right. The sun-like symbol is a circle with rays emanating from it, and the letters "BD" are in a bold, sans-serif font.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the BD logo. The logo consists of an orange circle with a white starburst inside, followed by the letters "BD" in blue. The starburst has 10 points, and the letters "BD" are in a bold, sans-serif font.
## Non-Clinical Performance Data
BD has performed the following non-clinical/design verification testing, including shelf life testing and the results of these tests demonstrate that the BD Plastipak™ Syringe performed in an equivalent manner to the predicate devices. As there are no packaging changes occurring as part of this submission, all packaging data was leveraged from predicate devices. Labeling requiremented for BNS sterilization instructions.
| | Startup Time | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. |
|----------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Syringe Performance | Volumetric Accuracy<br>(Manual Use, tolerance on<br>graduated capacity) | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. |
| | Sterility Product<br>(N/A* for BNS) | Subject device met the pre-established acceptance criteria per<br>ISO 11135, ISO 11137 and PH Eur. 5.0 |
| | Stopper Separation from<br>plunger | Subject device met the pre-established acceptance criteria per<br>ISO7886-1. |
| | Liquid leakage past stopper<br>under pressure (back rib) | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. |
| | Liquid leakage past stopper<br>under pressure (front rib) | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. |
| | Air Leakage past Stopper in<br>aspiration | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. |
| | Dead space | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. |
| | Maximum Usable Capacity | Subject device met the pre-established acceptance criteria per<br>ISO7886-1. |
| | Fiducial Line | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 |
| | Hard Height | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. |
| Short-Term Flow Rate<br>Accuracy (incl Stiction) | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. | |
| Overall Percentage Error | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. | |
| Max Variation in Flow Rate | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. | |
| Plunger Movement<br>Forces (Pump) | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. | |
| Syringe Compliance | Subject device met the pre-established acceptance criteria per<br>ISO7886-2. | |
| Uniform<br>Stopper<br>Appearance | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale - Graduation<br>Line Intervals | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale - Unit of measure | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale Lines | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale Numbering | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale Length | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Scale Position | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Marking - Syringe Barrel | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 and ISO7886-2. | |
| Air leakage past stopper in<br>aspiration (vacuum) | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 | |
| Stopper separation<br>from plunger | Subject device met the pre-established acceptance criteria per<br>ISO7886-1 | |
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular orange symbol with white rays emanating from the center, and the letters "BD" in blue. The symbol is on the left, and the letters are on the right. The logo is simple and modern, and it is easily recognizable.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows the logo for BD, a global medical technology company. The logo consists of two parts: a circular symbol on the left and the letters "BD" on the right. The circular symbol is orange and features a stylized sun-like design with rays emanating from a central point. The letters "BD" are in blue and have a rounded, sans-serif font.
| Sterilization Test | Irradiation | Subject device met the pre-established acceptance criteria<br>per ISO 11137-1 and ISO 11137-2 |
|-----------------------|----------------------------------------------|-----------------------------------------------------------------------------------------------|
| | Ethylene Oxide (Eto) | Subject device met the pre-established acceptance criteria<br>per ISO 11135 and ISO 10993-7 |
| Biocompatibility Test | Cytotoxicity | Per ISO 10993-5 and 10993-12, USP <87><br>Non-cytotoxic |
| | Hemolysis | Per ISO 10993-4 and 10993-12,<br>Non-hemolytic |
| | Acute Systemic Toxicity | Per ISO 10993-11 and 10993-12,<br>Non-toxic |
| | Intracutaneous Reactivity | Per ISO 10993-10 and 10993-12, USP <88><br>Non-Irritant |
| | Sensitization | Per ISO 10993-10,<br>Non-Sensitizer |
| | Pyrogenicity | Per ISO 10993-11, 10993-12 and USP 151,<br>Non-Pyrogenic |
| | Elastomeric Closures for<br>Injections | USP <381><br>Met Type I and II closures |
| | ISO 7886-1: | ISO 7886-1 |
| | Acidity/Alkalinity and<br>Extractable Metals | Pass |
| | Extractables/Leachables<br>Assessment | ISO 10993-17 and 18<br>Pass |
| | Particulate Matter | USP <788> Pass |
| | Endotoxin LAL | USP <85> Pass |
### Clinical Testing
Clinical testing was not required for this submission.
### Substantial Equivalence
The BD Plastipak™ Syringe is substantially equivalent to the predicate devices in intended use, operating principle, technology, design, materials and performance.
### Conclusion
The BD Plastipak™ Syringe has been verified to meet the established performance criteria above. The results of the non-clinical/design verification testing demonstrate that the BD Plastipak™ Syringe perform as intended and perform as well as the legally marketed predicate devices for the same intended use. Therefore the devices are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.