Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable PVC patient examination glove; acts as physical barrier to prevent contamination between patient and examiner. Device conforms to hand via tensile properties; water-tight under normal use. Used in clinical settings by healthcare providers. Provides protection against cross-contamination; benefits patient and examiner by maintaining hygiene during medical procedures.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via ASTM D5250 (dimensions, physical properties), ASTM D5151 (pinhole detection), and ASTM D6124 (residual powder). Biocompatibility verified via ISO 10993-10 (irritation/sensitization) and ISO 10993-5 (cytotoxicity).
Technological Characteristics
Material: Poly (vinyl chloride) (PVC). Form factor: Disposable glove. Standards: ASTM D5250, ASTM D5151, ASTM D6124. Biocompatibility: ISO 10993-5, ISO 10993-10. Non-sterile. Powder-free (residual powder < 2mg/glove).
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 12, 2019
Jiangxi Jaysun Medcare Co., Ltd. % Chu Xiaoan Official Correspondent Beijing Easylink CO., LTD Rm. F302 Bldg., 41, Jing Cheng Ya Ju, Courtyard 6 of Southern Dou Ge Zhuang, Beijing, 100021 Cn
Re: K182575
Trade/Device Name: Jaysun Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: April 11, 2019 Received: April 25, 2019
Dear Chu Xiaoan:
This letter corrects our substantially equivalent letter of May 24, 2019.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}------------------------------------------------
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elizabeth Claverie, M.S. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K182575
Device Name
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)
Indications for Use (Describe)
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
"The assigned 510(k) number is:___K182575
### Premarket Notification [510(k)] Summary
#### Submitter: 1.0
| Submitter's name : | Jiangxi Jaysun Medcare Co.,Ltd. |
|-------------------------|------------------------------------------------------------------------------------------------------|
| Submitter's address : | No.1 Jaysun Road,Penghuwan Industrial Park<br>Pengze City, Jiujiang Jiangxi<br>Province,332708,China |
| Phone number : | 0086-0792- 5866108 |
| Fax number : | 0086-0792- 5866108 |
| Name of contact person: | Mr. Zong Run |
| Date of preparation: | 2019-05-15 |
#### 2.0 Name of the Device
| Device Name: | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) |
|-------------------------|----------------------------------------------------------------------|
| Proprietary/Trade name: | Powder Free Vinyl Patient Examination<br>Gloves, Clear (non-colored) |
| Common Name: | Exam gloves |
| Classification Name: | Patient examination glove |
| Device Classification: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Panel: | General Hospital |
| Product Code: | LYZ |
### 3.0 Predicate device
| Device Name: | Powder Free Vinyl Patient Examination Gloves,<br>Clear (non-colored) |
|----------------|----------------------------------------------------------------------|
| Company name: | Zhang Jia Gang Fengyuan Plastic Product Co. Ltd. |
| 510(K) Number: | K091663 |
### 4.0 Device Description:
#### 4.1 How the device functions:
PVC films form a barrier to prevent contamination between patient and examiner.
#### 4.2 Scientific concepts that form the basis for the device
The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure.
{4}------------------------------------------------
### Physical and performance characteristics such as design, materials and 4.3 physical properties:
The Poly (vinyl chloride) glove acts as a barrier to prevent contamination between patient and examiner. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151.
### 5.0 Device Intended Use (Indication for use):
Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### 6.0 Technological Characteristics:
Shown below is a comparison of the technological characteristics of the subject device and the predicate device:
| Features & Description | Predicate Device<br>(K091663) | Subject Device<br>(K182575) | Result of<br>Comparison | | |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------|
| Company | Zhang Jia Gang<br>Fengyuan Plastic<br>Product Co. Ltd. | Jiangxi Jaysun<br>Medcare Co.,Ltd. | -- | | |
| 510(K) Number | K091663 | K182575 | -- | | |
| Product name | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored) | Powder Free Vinyl<br>Patient<br>Examination<br>Gloves, Clear<br>(non-colored) | Same | | |
| Product Code | LYZ | LYZ | Same | | |
| Size | Small/ Medium/<br>Large/X large | Small/ Medium/<br>Large/X large | Same | | |
| Intend for use | Powder Free Vinyl<br>Patient Examination<br>Gloves, Clear<br>(non-colored)<br>is a disposable<br>device<br>intended for medical<br>purposes that is<br>worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient and<br>examiner. | Powder Free<br>Vinyl<br>Patient<br>Examination<br>Gloves, Clear<br>(non-colored) is a<br>disposable<br>device<br>intended for<br>medical purposes<br>that is worn on the<br>examiner's hand<br>or finger to<br>prevent<br>contamination<br>between patient<br>and examiner. | Same | | |
| Device Description and<br>Specifications | Meets ASTM<br>D5250-06 | Meets ASTM<br>D5250-06<br>(Reapproved 2011) | Same | | |
| Dimensions<br>Length<br>(mm )<br>ILS-2<br>AQL4.0 | ≥230mm | 231-241mm | 232-243mm | Same | |
| Dimensions<br>Width<br>(mm)<br>IL S-2<br>AQL4.0 | Small | 80-90 | 81-89 | 82-88 | Same |
| | Medium | 90-100 | 93-99 | 93-98 | |
| | Large | 100-110 | 102-110 | 102-109 | |
| | X large | 110-120 | 111-1119 | 113-118 | |
| Dimensions<br>Thickness<br>(mm )<br>IL S-2<br>AQL4.0 | Finger | $\ge$ 0.05 | 0.05-0.10 | 0.08-0.11 | Same |
| | Palm | $\ge$ 0.08 | 0.09-0.13 | 0.10-0.11 | Same |
| Physical<br>Properties<br>IL S-2<br>AQL4.0 | Before aging/after aging | | | | Same |
| | Elongation $\ge$ 300% | | 340-410% | 350-420% | |
| | Tensile Strength<br>$\ge$ 14MPa | | 15-25 MPa | 15-20 MPa | |
| Freedom<br>from<br>Pinholes<br>Residual<br>Powder | Holes at<br>Inspection Level<br>LAOL2.5<br>below 2mg of<br>residual powder | | Holes at<br>Inspection<br>Level I | Holes at<br>Inspection Level I<br>AOL2.5 | Same |
| | | 0.3mg | | 0.1mg | Same |
| Materials used to fabricate<br>the devices | | | PVC | PVC | Same |
| Dusting or Donning<br>Powder: name | | | Surface Coating<br>Agent | Surface Coating<br>Agent | Same |
| Compare performance data<br>supporting substantial<br>equivalence | | | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation 2011) | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D5250-06<br>(Reapproved 2011)<br>ASTM D6124-06<br>(Reaffirmation | Same |
| Single Patient Use | | | Single Patient Use | Single Patient Use | Same |
| Biocompatibility<br>Irritation and Sensitization<br>(ISO 10993-10) | | | Under the conditions<br>of this study, not an<br>irritant and Under<br>the conditions of<br>this study, not a<br>sensitizer. | Under the<br>conditions<br>of this study, not<br>an irritant and<br>Under the<br>conditions of this<br>study, not a<br>sensitizer. | Same |
| Cytotoxicity<br>(ISO 10993-5) | | | / | Under the<br>conditions of this<br>study, the test<br>article was<br>non-cytotoxicity<br>to L-929 cells. | |
| Labeling. | | | -Powder Free<br>-Patient<br>Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | -Powder Free<br>-Patient<br>Examination<br>Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot | Same |
{5}------------------------------------------------
{6}------------------------------------------------
## 7.0 Summary of Non-Clinical Performance Data:
Non-clinical tests were conducted to verify that the proposed device will meet acceptance criteria for each test. The test results demonstrated that the proposed device met the acceptance criteria found in the following standards below:
| Characteristics | Standard |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Dimension | ASTM D 5250-06(Reapproved 2011). |
| | Length |
| | Width |
| | Thickness |
| | Physical<br>Properties |
| | |
| | |
| | |
| Powder Residual | ASTM standard D 5250-06 (Reapproved<br>2011).and D6124-06(Reaffirmation 2011) Meets<br><2mg/glove |
| Biocompatibility | Primary Skin Irritation - ISO 10993-10:<br>2010 Under the conditions of<br>the study, the subject<br>device is not a primary<br>skin irritant. |
| | Dermal Sensitization - ISO 10993-10:<br>2010 Under the conditions<br>of the study, the<br>subject device is not a<br>skin sensitizer. |
| | Cytotoxicity - ISO 10993-5: 2009 Under the conditions<br>of this study, the test<br>article was<br>non-cytotoxicity to<br>L-929 cells. |
{7}------------------------------------------------
| ISO 10993-5:2009 | Biological Evaluation of Medical Devices - Part 5: Tests<br>For In Vitro Cytotoxicity |
|-------------------------------------|-----------------------------------------------------------------------------------------------------|
| ISO 10993-10:2010 | Biological Evaluation of Medical Devices - Part 10: Tests<br>For Irritation And Skin Sensitization. |
| ASTM D5151-06<br>(Reapproved 2015), | Standard Test Method for Detection of Holes in Medical<br>Gloves. |
| ASTM D5250-06<br>(Reapproved 2015), | Standard specification for poly (vinyl chloride) gloves for<br>medical application. |
| ASTM D6124-06<br>(Reapproved 2017) | Standard Test Method for Residual Powder on Medical<br>Gloves |
#### Summary of Clinical Performance Data: 8.0
Clinical testing is not needed for this device.
#### 9.0 Conclusion:
The conclusions drawn from the nonclinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the legally marketed predicated device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.