The Venture 0.014" catheters are indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the coronary and peripheral vasculature. The OTW version (Model 5821) may also be used for manual delivery of saline solution or diagnostic contrast agents.
Device Story
Venture 0.014" catheter is a single-lumen device with a radiopaque, deflectable distal tip; used to support guidewires during coronary and peripheral vascular interventional procedures. Device features a proximal control handle with a deflection knob allowing 90° tip articulation; hydrophilic coating on distal end enhances deliverability. Available in Rapid Exchange (RX) and Over-The-Wire (OTW) configurations. Operated by physicians in clinical settings. OTW version allows manual delivery of saline or contrast agents. Device provides mechanical support and steering for guidewires, facilitating access to target vasculature. Benefits include improved navigation and support in complex vascular anatomy.
Clinical Evidence
Bench testing only. No clinical data presented. Verification testing included deliverability, tip deflection, tensile strength, torque robustness, pressure testing, and coating durability. Biocompatibility testing included cytotoxicity, sensitization, irritation, pyrogenicity, hemolysis, complement activation, and acute systemic toxicity. All tests met acceptance criteria.
Indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the coronary and peripheral vasculature in patients requiring interventional procedures. OTW version also indicated for manual delivery of saline or diagnostic contrast agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 13, 2019
Vascular Solutions, LLC Becky Astrup Regulatory Affairs Specialist 6464 Sycamore Court North Minneapolis. Minnesota 55369
Re: K182570
Trade/Device Name: Venture 0.014" Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: February 8, 2019 Received: February 11, 2019
Dear Becky Astrup:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn
(http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Lydia S. | Digitally signed by<br>Lydia S. Glaw -S |
|----------|-----------------------------------------|
| Glaw -S | Date: 2019.03.13<br>15:26:36 -04'00' |
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K182570
Device Name Venture 0.014" Catheter
Indications for Use (Describe)
The Venture 0.014" catheters are indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the coronary and peripheral vasculature. The OTW version (Model 5821) may also be used for manual delivery of saline solution or diagnostic contrast agents.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Vascular Solutions. The logo features the word "Vascular" in a bold, serif font, with the word "SOLUTIONS" in a smaller, sans-serif font underneath. Above the text is a square shape in a light brown color, with a pattern of black squares that fade out towards the top, creating a visual effect of movement or flow.
### 510(k) Summary
[As required by 21 CFR 807.92]
510(k) Number: K182570
## SUBMITTER AND DEVICE
| Submitter: | Vascular Solutions, LLC<br>6464 Sycamore Court North<br>Minneapolis, MN 55369 USA |
|-----------------------------|-----------------------------------------------------------------------------------|
| Establishment Registration: | 2134812 |
| Phone: | 763-656-4300 |
| Fax: | 763-656-4253 |
| Contact Person: | Becky Astrup,<br>Regulatory Product Specialist |
| Date Prepared: | September 14th, 2018 |
| Name of Device: | Venture 0.014" Catheter |
| Common or Usual Name: | Catheter |
| Classification Name: | Catheter, Percutaneous<br>(21 CFR 870.1250) |
| Regulatory Class: | II |
| Product Code: | DQY |
### PREDICATE DEVICES
The legally marketed devices to which substantial equivalence is claimed is:
Primary Predicate Device: Vascular Solutions, Venture Wire Control Catheter, K042910 - cleared November 22, 2004.
Predicate Device: Vascular Solutions, Venture Wire Control Catheter, K061843 - cleared August 1, 2006.
The Vascular Solutions Turnpike Catheter, K142065- cleared November 25, 2014 and the Vascular Solutions Venture 038 Catheter, K171335- cleared June 7, 2017 are included as reference devices for this submission.
{4}------------------------------------------------
# DEVICE DESCRIPTION
The Venture 0.014" catheters are single lumen catheters with a radiopaque deflectable distal tip designed to provide guidewire support during interventional procedures in the coronary and peripheral vasculature. The Venture 0.014" catheters are compatible with 0.014" guidewires and are available in two configurations. The rapid exchange configuration consists of a hypotube shaft with a rapid exchange (RX) lumen and deflectable tip on the distal segment. The over the wire configuration consists of an over-the-wire (OTW) lumen that runs the entire length of the catheter with a deflectable tip. The Venture 0.014" catheter tips are deflectable 90° by rotating a deflection knob on the control handle of the proximal end and a hydrophilic coating on the distal end of the catheter enhances deliverability to the target vasculature. The Venture 0.014" catheters are sterilized with ethylene oxide.
### INDICATIONS FOR USE
The Venture 0.014" catheters are indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the coronary and peripheral vasculature. The OTW version (Model 5821) may also be used for manual delivery of saline solution or diagnostic contrast agents.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The Venture 0.014" catheter included in this Special 510(k) submission is a modification of the current Venture Wire Control Catheter. Modifications include a change in distal tip jacket material and, applicable to the OTW configuration only, an additional material processing step for the manufacturing of the liner component and removal of the proximal positioning marks. A comparison of technological differences between the subject and predicate devices is provided below.
| Characteristic | Predicate Devices:<br>Venture Wire Control Catheter<br>K042910, K061843 | Subject Device:<br>Venture 0.014" Catheter |
|-------------------------|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Principles of Operation | Deflectable tip is actuated by turning the knob on the proximal handle. | |
| Anatomical Location | Coronary and peripheral vasculature | |
| Guidewire Compatibility | 0.014" | |
| Working Length | K061843 (RX)<br>145 cm<br>K042910 (OTW)<br>140 cm | Model 5820 (RX):<br>144.5 cm<br>Model 5821 (OTW):<br>139.5 cm |
| Proximal Markers | Markers on the shaft for 90 cm and<br>100 cm guides. | Model 5820 (RX):<br>Markers on the shaft for 90 cm and<br>100 cm guides.<br>Model 5821 (OTW):<br>NA- No positioning marks |
| Tip Deflection | 90° from catheter axis. | |
| Lubricious Coating | Hydrophilic Coating | |
| Sterility | Ethylene Oxide (EO) – Sterility Assurance Level (SAL) 10-6 | |
| Biocompatibility | Meets ISO 10993-1 | |
{5}------------------------------------------------
# PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Performance Testing - Bench
The technological differences between the subject and predicate devices have been evaluated through bench tests to provide evidence that the Venture 0.014" catheter is substantially equivalent to the predicate devices. The device design was verified through the following tests:
- Deliver, Deflect, and Retract ●
- Guidewire Steerability
- Tip Deflection Angle ●
- Tip Rebound ●
- Tip Rotation ●
- Tip Tensile
- Surface Defects ●
- Guidewire Movement
- Torque Robustness ●
- Luer to Extension Tensile
- Extension to Back Handle Tensile ●
- Liquid Leakage from Fitting ● Assembly under Pressure
- Static Pressure
- Coating Lubricity/Durability ●
- Throw Length ●
- Particulates ●
- HPC Fluid Drops ●
- Therapeutic Agent Conditioning ●
- Dynamic Pressure ●
The results of the verification tests met the specified acceptance criteria and did not raise different questions of safety and effectiveness.
### Biocompatibility Testing
The biocompatibility evaluation for the Venture 0.014" catheter was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Venture 0.014" catheter is considered an externally communicating device in contact with circulating blood and tissue for a limited period of time (<24 hours) during use. The battery of tests included the following:
- Cytotoxicity ●
- Sensitization ●
- Irritation
- Material Mediated Pyrogenicity ●
- Hemolysis ●
- Complement Activation
- Acute Systemic Toxicity ●
- Thrombogenicity
Passing results from biomaterial tests demonstrate that the Venture 0.014" catheter is noncytotoxic, non-sensitizing, non-irritating, non-systemically toxic, non-hemolytic, not an activator of the complement system, and thromboresistant.
{6}------------------------------------------------
# CONCLUSION
The subject Venture 0.014" catheter is substantially equivalent to the Venture Wire Control Catheter predicate devices based on comparisons of the device functionality, technological characteristics, and indications for use. The results of design verification tests do not raise new or different questions of safety and effectiveness; the Venture 0.014" catheter is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.