R3 Anteverted Liners

K182535 · Smith & Nephew, Inc. · MBL · Nov 19, 2018 · Orthopedic

Device Facts

Record IDK182535
Device NameR3 Anteverted Liners
ApplicantSmith & Nephew, Inc.
Product CodeMBL · Orthopedic
Decision DateNov 19, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDID) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, atthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection: fracture-dislocation of the hin; and correction of deformity. The R3 Acetabular System is for single use only and is intended for cementless use.

Device Story

R3 Anteverted Liners are acetabular liner line extensions for R3 Acetabular System; used in primary/revision hip arthroplasty. Device features 20° anteverted entry angle and variable angle chamfer/bevel on rim; +4mm lateralized ID. Manufactured from 10MRAD cross-linked polyethylene (XLPE). Implanted by orthopedic surgeons in clinical settings. Device provides articulating surface for femoral head; restores hip joint function; addresses joint damage/deformity. Substantially equivalent to existing R3 XLPE liners; maintains identical locking mechanism and material properties.

Clinical Evidence

Bench testing only. No clinical data provided. Mechanical testing included push-in, push-out, lever-out, torque to failure, finite element analysis, range of motion analysis, and poly thickness comparison. Biocompatibility and bacterial endotoxin testing met acceptable limits per FDA guidance and ANSI/AAMI ST72.

Technological Characteristics

Material: 10MRAD Cross-linked Polyethylene (XLPE) per ASTM F648. Design: Acetabular liner with 20° anteverted entry angle, variable angle chamfer/bevel, +4mm lateralized ID. Sizes: ID 28-44mm, OD 44-80mm. Sterilization: Ethylene Oxide. Single-use, cementless. No software or electronics.

Indications for Use

Indicated for individuals undergoing primary/revision hip surgery due to trauma, noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant), inflammatory degenerative joint disease (rheumatoid arthritis, secondary arthritis, congenital dysplasia), or proximal femur fractures/nonunion unmanageable by other techniques.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ November 19, 2018 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. The logo is simple and professional, and it is easily recognizable. Smith & Nephew, Inc. Meenakshi Gupta Senior Regulatory Affairs Specialist 7135 Goodlett Farms Parkway Cordova, Tennessee 38016 #### Re: K182535 Trade/Device Name: R3 Anteverted Liners Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: MBL Dated: October 23, 2018 Received: October 24, 2018 ### Dear Meenakshi Gupta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Daniel S. Ramsey -S 2018.11.19 16:51:55 -05'00' For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182535 Device Name R3 Anteverted Liners #### Indications for Use (Describe) Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDID) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, atthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection: fracture-dislocation of the hin; and correction of deformity. The R3 Acetabular System is for single use only and is intended for cementless use. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------| | <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY | Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 | |----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Submission: | September 13, 2018 | | Contact Person: | Meenakshi Gupta, Sr. Regulatory Affairs Specialist<br>T (901) 399-6139<br>F (901) 721-2748 | | Name of Device: | R3 Anteverted Liners | | Common Name: | Acetabular Liners | | Device Classification Name<br>and Reference: | 21 CFR 888.3358 Hip joint metal/polymer/metal semi<br>constrained porous-coated uncemented prosthesis | | Device Class: | Class II | | Panel Code: | Orthopaedics/87 | | Product Code: | MBL | | Predicate Device: | Primary Predicate-R3 XLPE Liners-K113848<br>Predicate 2-R3 XLPE Liners-K093363<br>Predicate 3-REFLECTION Acetabular System-<br>K920430<br>Reference Predicate-R3 XLPE Anteverted Liners-<br>K102370 | {4}------------------------------------------------ #### K182535 # Device Description: The subject of this Special 510(k) is the R3® Anteverted Liners. The subject R3 Anteverted Liners are a line extension to the R3 XLPE Liners that were cleared under premarket notifications K113848 and K093363. The R3 Anteverted Liners have same overall design, indication for use and material as the predicate R3 XLPE liners. The subject R3 Anteverted Liners feature a 20° anteverted, or "face changing", entry angle to the articulating surface (Inner diameter). Also, a variable angle chamfer/bevel was added on the rim of the liner. The R3 Anteverted Liners have +4mm lateralized inside Diameter, 20° andle, and identical locking mechanism as the predicate R3 XLPE Liners, cleared under K113848 and K093363. The R3 Anteverted Liners are available with Inner diameter (ID) sizes 28, 32, 36, 40 and 44mm and the Outer Diameter (OD) sizes ranging from 44 to 76-80mm.The R3 Anteverted Liner is manufactured from 10MRAD Cross-linked Polyethylene (XLPE) confirming to ASTM F648, as the predicate R3 XLPE Liners cleared under a premarket notification K113848 and K093363. The R3 Anteverted Liners are provided sterile and is intended for single use only. # Indications for Use Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using {5}------------------------------------------------ ## K182535 other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracturedislocation of the hip; and correction of deformity. The R3 Acetabular System is for single use only and is intended for cementless use. The above indication for use for the subject device, is identical to the R3 XLPE Liner cleared under K113848 and K093363, the REFLETION Acetabular System (K920430) and R3 XLPE Anteverted Liners (K102370). ## Comparison of Technological Characteristics with the Predicate Device Intended Use: The proposed R3 Anterverted Liners have identical Intended use as the predicate R3 XLPE Liners Indication for Use: The proposed R3 Anterverted Liners have identical Indication for Use as the predicate R3 XLPE Liners Materials: The proposed R3 Anterverted Liners are manufactured by same material and manufacturing process as the predicate R3 XLPE Liners Design Features: The proposed R3 Anterverted Liners have overall same design features as the predicate R3 XLPE Liners. Sterilization: The proposed R3 Anterverted Liners and predicate R3 XLPE Liners are provided sterile by Ethylene Oxide for single-use. Therefore, the technological characteristics of the subject device are the similar to the technological characteristics of the predicate device R3 XLPE Liners cleared under K113848 and K093363. # Substantial Equivalence Information The overall design, materials, and indications for use for R3 Anteverted Liners are substantially equivalent to the following commercially available predicate devices. {6}------------------------------------------------ # Table: Predicate Devices | Manufacturer | Description | Submission<br>Number | Clearance<br>Date | |----------------------|------------------------------|----------------------|-------------------| | Smith & Nephew, Inc. | R3 XLPE Liner | K113848<br>(Primary) | 04/27/2012 | | Smith & Nephew, Inc. | R3 XLPE Liner | K093363 | 01/26/2010 | | Smith & Nephew, Inc. | Reflection Acetabular System | K920430 | 07/21/1992 | | Smith & Nephew, Inc. | R3 XLPE Anteverted Liners | K102370 | 01/19/2011 | # Performance Testing To further support a determination of substantial equivalence, non-clinical bench (mechanical) testing were conducted on the R3 Anteverted Liners. A review of the mechanical data and the technical memo, indicates that the R3 Anteverted Liners are substantially equivalent to predicate devices listed in the Table above. The following tests were used as a basis for the determination of substantial equivalence. - Push-in . - Push-out - Lever-out ● - Torque to Failure ● - Finite Element Analysis . - Range of Motion Analysis ● - Poly thickness comparison . - o Biocompatibility Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72. {7}------------------------------------------------ # K182535 # Conclusion As previously noted, this special 510(k) Premarket Notification is being submitted to request clearance for the R3 Anteverted Liners. Based on the similarities to the predicate devices and a review of the mechanical testing performed, the subject device is substantially equivalent to the commercially available predicate devices listed above.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...