Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications
Device Story
PIXX series are digital radiography systems using flat panel detectors (FPD) to capture X-ray images; replaces conventional film/screen systems. Input: X-ray radiation; Output: digital radiographic images. Used in clinical settings by healthcare professionals. Device integrates with PACS for filmless environment. Modification from predicate (K180976) adds rechargeable lithium battery power capability for wireless operation. Healthcare providers view images on connected workstations to support diagnostic decision-making. Benefits include improved cost-effectiveness and workflow efficiency.
Clinical Evidence
No clinical data. Bench testing only. Performance metrics (MTF, DQE) were identical to the predicate device. Battery life testing confirmed operational duration (8 hours/480 images for PIXX 1212; 5 hours/300 images for PIXX 1717/1417).
Technological Characteristics
Flat panel digital detector (FPD) with CsI or GOS scintillator. 16-bit A/D conversion. Connectivity via Gigabit Ethernet (1000BASE-T) or IEEE 802.11ac wireless. Power: AC line or rechargeable lithium battery. Standards: IEC 60601-1 (Electrical Safety), IEC 60601-1-2 (EMC), IEC 62133 (Battery Safety). Software remains identical to predicate.
Indications for Use
Indicated for general radiographic imaging of human anatomy. Intended to replace film/screen systems in general-purpose diagnostic procedures. Excludes fluoroscopic, angiographic, and mammographic applications.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 4, 2018
Pixxgen Corporation % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K182533
Trade/Device Name: PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-ray Receptor Panels Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MOB Dated: September 12, 2018 Received: September 14, 2018
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hol 2. Nils
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K182533
Device Name
PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels
Indications for Use (Describe)
Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, angiographic, and mammographic applications
Type of Use (Select one or both, as applicable)
| <div> <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9307 2.66663 1.33325 3.26407 1.33325 3.99997V12C1.33325 12.7359 1.9307 13.3333 2.66663 13.3333H13.3333C14.0692 13.3333 14.6666 12.7359 14.6666 12V3.99997C14.6666 3.26407 14.0692 2.66663 13.3333 2.66663Z" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.33333"></path> <path d="M5.33325 8L7.33325 10L10.6666 6" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.33333"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> </div> |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9307 2.66663 1.33325 3.26407 1.33325 3.99997V12C1.33325 12.7359 1.9307 13.3333 2.66663 13.3333H13.3333C14.0692 13.3333 14.6666 12.7359 14.6666 12V3.99997C14.6666 3.26407 14.0692 2.66663 13.3333 2.66663Z" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.33333"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary 510(k) Number K182533 PIXXGEN Corporation 5F, SMART BAY, 123, Beolmal-ro, Dongan-gu, Anyang-si, Gyeonggi-do, 14056, Korea Tel +82.70.4846.8888 Date Prepared: September 28, 2018 Contact: Young Kim, President
- 1. Identification of the Device:
2
Device Class:
Proprietary-Trade Names: PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels Device: Solid State X-Ray Imager (Flat Panel/Digital Imager) Regulation Name: Stationary x-ray system. Product Code: MQB Regulation Number: 892.1680
- 2. Equivalent legally marketed device: PIXX 1717; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels.
| Manufacturer: | PIXXGEN Corporation |
|--------------------|------------------------------------------------------|
| 510(k) Number: | K180976 |
| Device: | Solid State X-Ray Imager (Flat Panel/Digital Imager) |
| Regulation Name: | Stationary x-ray system. |
| Product Code: | MQB |
| Regulation Number: | 892.1680 |
| Device Class: | 2 |
- 3. Indications for Use (intended use) Indicated for use in general radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general-purpose diagnostic procedures, excluding fluoroscopic, and mammographic applications. (Rx Only)
### 4. Description of the Device:
The PIXX series are digital radiography systems, featuring an integrated flat panel digital detector (FPD). They are designed to perform digital radiographic examinations as a replacement for conventional film. This integrated platform provides the benefits of PACS with the advantages of digital radiography for a filmless environment and improves cost effectiveness. The major functions and principle of operation of the updated panels are the same as our previous panel retaining the Wi-Fi wireless features. The modification involves adding battery power capability to the devices in our previous submission K180976.
- 5. Safety and Effectiveness, comparison to predicate device. The results of bench and test laboratory results indicates that the new devices are as safe and effective as the predicate devices.
{4}------------------------------------------------
# 6. Substantial Equivalence Chart
| | K180976, PIXX 1717,<br>PIXX 1417, PIXX 1212 | PIXX 1717 | PIXX 1417 | PIXX 1212 |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| Intended Use | Indicated for use in general<br>radiographic images of human<br>anatomy. It is intended to<br>replace radiographic film/screen<br>systems in all general-purpose<br>diagnostic procedures, excluding<br>fluoroscopic, angiographic, and<br>mammographic applications | UNCHANGED | UNCHANGED | UNCHANGED |
| Configuration | This submission is for the<br>Digital Panel and Software<br>only, no generator or stand<br>provided. | UNCHANGED | UNCHANGED | UNCHANGED |
| Pixel Pitch | 140um | SAME | SAME | SAME |
| Limiting<br>Resolution | Over 3 lp/mm | SAME | SAME | SAME |
| DQE(CSI) | At 2 lp/mm 26.5% | SAME | SAME | SAME |
| MTF(CSI) | At 2 lp/mm 44% | SAME | SAME | SAME |
| DQE(GOS) | At 2 lp/mm 21 | SAME | SAME | SAME |
| MTF(GOS) | At 2 lp/mm 35% | SAME | SAME | SAME |
| DQE and MTF<br>Comments | Adding battery power capability did not change the panel performance. | | | |
| A/D Conversion | 16 bits | SAME | SAME | SAME |
| Active Area | 17 x 17 inch<br>14 x 17 inch<br>12 x 12 inch | 17 x 17 inch | 14 x 17 inch | 12 x 12 inch |
| Dimensions(m<br>m)/<br>Weights(Kg) | 460(W)x461(L)x15(H)/3.0Kg<br>385(W) x 460(L) x 15(H)/<br>2.8Kg<br>308.5(W) x 319.5(L) x 15(H)/<br>1.9Kg | SAME | SAME | SAME |
| Pixels | 3,072x3,072<br>2,560 x 3,072<br>2,048 x 2,048 | 3,072 x 3,072 | 2,560 x 3,072 | 2,048 x 2,048 |
| Software | Outputs a DICOM image | SAME | SAME | SAME |
| DICOM | Yes | Yes | Yes | Yes |
| Scintillator | Csl orGOS | UNCHANGED | UNCHANGED | UNCHANGED |
| Interface | Wired: Gigabit Ethernet<br>(1000BASE-T)<br>Wireless:IEEE802.11ac,<br>backward compatible | SAME | SAME | SAME |
| Power source | AC Line only | AC Line and/or<br>Rechargeable Lithium<br>Battery;<br>5 hours/300 images | AC Line and/or<br>Rechargeable<br>Lithium Battery;<br>5 hours/300 images | AC Line and/or<br>Rechargeable Lithium<br>Battery 8 hours/480<br>images |
| | K180976, PIXX 1717,<br>PIXX 1417, PIXX 1212 | PIXX 1717 | PIXX 1417 | PIXX 1212 |
| Standards | Electrical Safety per IEC<br>60601-1:2012 and EMC per<br>IEC 60601-1-22007+AC:2010<br>as well as IEEE802.11ac.<br>Meets FCC requirements. | SAME plus IEC 62133<br>Battery safety | SAME plus IEC 62133<br>Battery safety | SAME plus IEC 62133<br>Battery safety |
{5}------------------------------------------------
- 7. Summary of Bench Testing Conducted: IEC Standards were employed for: Electrical Safety and Electromagnetic Compatibility, and Battery Safety Tests. Standards met:
Electrical safety per: IEC/UL 60601-1, Medical Electrical Equipment – Part 1: General Requirements for Safety. (General)
Electromagnetic Compatibility per IEC 60601-1-2, Collateral Standard: Electromagnetic compatibility Requirements and tests.
IEC 62133 Edition 2.0 2012-12 Secondary cells and batteries containing alkaline or other nonacid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications
Risk Analysis was conducted in accordance with ISO 14971:2012 and EN 62304. The software remains the same as in K180976. MTF and DQE measurements were not needed because the panel characteristics are identical to our predicate cleared earlier this year.
Battery life testing was conducted to confirm that the PIXX 1212 runs for 8 hours/480 images and for the PIXX 1717/1417 for 5 hours/300 images.
Cybersecurity precautionary labeling was added per the FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, Guidance for Industry and Food and Drug Administration Staff, Document Issued on: October 2, 2014.
- 8. Summary of Clinical Testing: Clinical images were not required because the panel technology did not change from K180976.
- 9. Conclusion: After analyzing bench, clinical image, and external laboratory testing to applicable standards, it is the conclusion of Pixxgen that the PIXX 1717; PIXX 1417; PIXX 1212; Digital Diagnostic X-Ray Receptor Panels are as safe and effective as the predicate device, have few technological differences, and has the identical indications for use, thus rendering them substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.