← Product Code [NLH](/productcode/NLH) · K182488

# Reprocessed Response Diagnostic Electrophysiology Catheter (K182488)

_Innovative Health, LLC · NLH · Feb 22, 2019 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182488

## Device Facts

- **Applicant:** Innovative Health, LLC
- **Product Code:** [NLH](/productcode/NLH.md)
- **Decision Date:** Feb 22, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1220
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Reprocessed Response Diagnostic Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

## Device Story

Reprocessed diagnostic electrophysiology catheter used for endocardial and intravascular mapping of cardiac arrhythmias. Device consists of a catheter body with multiple electrodes; inserted by a physician into the heart to record electrical signals. Signals are transmitted to external electrophysiology recording equipment to assist in arrhythmia diagnosis. Device is a reprocessed version of a single-use predicate; undergoes cleaning, decontamination, inspection, and functional testing. Used in clinical electrophysiology labs. Benefits include providing diagnostic data for arrhythmia management while reducing medical waste through reprocessing.

## Clinical Evidence

Bench testing only. Testing included biocompatibility, cleaning validation, sterilization validation, functional testing (visual inspection, dimensional verification, electrical continuity/resistance, simulated use, mechanical characteristics), and electrical safety testing (dielectric and current leakage).

## Technological Characteristics

Diagnostic EP catheter with multiple electrodes (2-10), various French sizes (5-6F), and fixed curve types. Materials, design, and function are identical to predicate. Reprocessed via cleaning and decontamination; limited to maximum 3 reprocessing cycles. Tracked via marking system. No software or active electronic components; passive electrode recording.

## Regulatory Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

## Predicate Devices

- Response Electrophysiology Catheter ([K151622](/device/K151622.md))
- Reprocessed Response Diagnostic Electrophysiology Catheter ([K161827](/device/K161827.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 22, 2019

Innovative Health, LLC. Amanda Babcock Principal Regulatory Affairs Specialist 1435 North Havden Road. Suite 100 Scottsdale, Arizona 85257

Re: K182488

Trade/Device Name: Reprocessed Response Diagnostic Electrophysiology Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: February 1, 2019 Received: February 4, 2019

# Dear Amanda Babcock:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice

(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Minde Yellman
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| Description                                             | Item<br>Number | Number of<br>Electrodes | French<br>Size | Electrode<br>Spacing (mm) | Curve<br>Type | Usable<br>Length (cm) |
|---------------------------------------------------------|----------------|-------------------------|----------------|---------------------------|---------------|-----------------------|
| Response<br>Diagnostic<br>Electrophysiology<br>Catheter | 401381         | 10                      | 6              | 2-8-2                     | CSL           | 120                   |
|                                                         | 401353         | 10                      | 6              | 2-8-2                     | CSL           | 65                    |
|                                                         | 401392         | 10                      | 6              | 2-2-2                     | CSL           | 120                   |
|                                                         | 401399         | 10                      | 6              | 5-5-5                     | CSL           | 120                   |
|                                                         | 401400         | 10                      | 6              | 5-5-5                     | CSL           | 65                    |
|                                                         | 401360         | 6                       | 5              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401317         | 2                       | 6              | 10                        | CRD           | 120                   |
|                                                         | 401318         | 2                       | 6              | 10                        | JSN           | 120                   |
|                                                         | 401305         | 10                      | 6              | 2-5-2                     | CRD           | 120                   |
|                                                         | 401306         | 10                      | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401308         | 10                      | 6              | 2-5-2                     | CRD           | 65                    |
|                                                         | 401309         | 10                      | 6              | 2-5-2                     | JSN           | 65                    |
|                                                         | 401310         | 10                      | 6              | 2-5-2                     | DAO           | 65                    |
|                                                         | 401311         | 10                      | 6              | 2-2-2                     | CRD           | 120                   |
|                                                         | 401312         | 10                      | 6              | 2-2-2                     | JSN           | 120                   |
|                                                         | 401150         | 10                      | 6              | 2-1-1-10-1-1-1-1-1        | CRD           | 120                   |
|                                                         | 401271         | 6                       | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401275         | 6                       | 6              | 2-2-2                     | CRD           | 120                   |
|                                                         | 401276         | 6                       | 6              | 2-2-2                     | JSN           | 120                   |
|                                                         | 401278         | 6                       | 6              | 2-2-2                     | CRD           | 65                    |
|                                                         | 401281         | 6                       | 6              | 5-5-5                     | CRD           | 120                   |
|                                                         | 401282         | 6                       | 6              | 5-5-5                     | JSN           | 120                   |
|                                                         | 401425         | 6                       | 6              | 5-5-5-150-5               | CRD           | 120                   |
|                                                         | 401386         | 6                       | 6              | 5-5-5                     | CSL           | 120                   |
|                                                         | 401206         | 4                       | 5              | 10                        | CRD           | 120                   |
|                                                         | 401207         | 4                       | 5              | 10                        | JSN           | 120                   |
|                                                         | 401222         | 4                       | 5              | 5                         | CRD           | 120                   |
|                                                         | 401223         | 4                       | 5              | 5                         | JSN           | 120                   |
|                                                         | 401210         | 4                       | 6              | 10                        | CRD           | 120                   |
|                                                         | 401211         | 4                       | 6              | 10                        | JSN           | 120                   |
|                                                         | 401212         | 4                       | 6              | 10                        | DAO           | 120                   |
|                                                         | 401226         | 4                       | 6              | 5                         | CRD           | 120                   |
|                                                         | 401356         | 4                       | 6              | 5                         | CRD-1         | 120                   |
|                                                         | 401227         | 4                       | 6              | 5                         | JSN           | 120                   |
|                                                         | 401357         | 4                       | 6              | 5                         | JSN-1         | 120                   |
|                                                         | 401228         | 4                       | 6              | 5                         | DAO           | 120                   |
|                                                         | 401329         | 4                       | 6              | 5                         | DAO-1         | 120                   |
|                                                         | 401260         | 4                       | 6              | 2-5-2                     | CRD           | 120                   |
|                                                         | 401261         | 4                       | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401154         | 4                       | 6              | 5                         | CRD           | 120                   |
|                                                         | 401155         | 4                       | 6              | 5                         | JSN           | 120                   |
|                                                         | 401156         | 4                       | 6              | 5                         | DAO-1         | 120                   |
|                                                         | 401158         | 4                       | 6              | 10                        | CRD           | 120                   |

The following item numbers are included in the scope of this submission (K182488):

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K182488

Device Name

Reprocessed Response Diagnostic Electrophysiology (EP) Catheter

Indications for Use (Describe)

The Reprocessed Response Diagnostic Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <div> <span>       Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> | <div> <span>       Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# SECTION 5: 510(k) SUMMARY

As required by 21 CFR 807.92(c)

#### Submitter's Name and Address:

Innovative Health, LLC. 1435 N. Hayden Road, Suite 100 Scottsdale, AZ 85257

#### Contact Name and Information:

Amanda Babcock Principal Regulatory Affairs Specialist Innovative Health, LLC. (480) 525-5911 (office) (888) 965-7705 (fax) ababcock@innovative-health.com

## Date prepared:

September 10, 2018

## Device Information:

Trade/Proprietary Name: Common Name: Classification Name: Classification Number: Product Code:

Reprocessed Response Diagnostic Electrophysiology Catheters Diagnostic Electrophysiology Catheter Catheter, Recording, Electrode, Reprocessed Class II, 21 CFR 870.1220 NLH

# Predicate Device:

| 510(k) Number | 510(k) Device                                                 | Manufacturer            |
|---------------|---------------------------------------------------------------|-------------------------|
| K151622       | Response Electrophysiology Catheter                           | St. Jude Medical        |
| K161827       | Reprocessed Response Diagnostic<br>Electrophysiology Catheter | Innovative Health, LLC. |

#### Device Description:

The Reprocessed Response Diagnostic Electrophysiology Catheters are manufactured in various fixed curves and electrode spacing for electrophysiological mapping for the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

The item numbers in scope of this submission are as follows (next page):

{5}------------------------------------------------

| Description                                             | Item Number | Number of<br>Electrodes | French<br>Size | Electrode Spacing<br>(mm) | Curve<br>Type | Usable<br>Length (cm) |
|---------------------------------------------------------|-------------|-------------------------|----------------|---------------------------|---------------|-----------------------|
| Response<br>Diagnostic<br>Electrophysiology<br>Catheter | 401381      | 10                      | 6              | 2-8-2                     | CSL           | 120                   |
|                                                         | 401353      | 10                      | 6              | 2-8-2                     | CSL           | 65                    |
|                                                         | 401392      | 10                      | 6              | 2-2-2                     | CSL           | 120                   |
|                                                         | 401399      | 10                      | 6              | 5-5-5                     | CSL           | 120                   |
|                                                         | 401400      | 10                      | 6              | 5-5-5                     | CSL           | 65                    |
|                                                         | 401360      | 6                       | 5              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401317      | 2                       | 6              | 10                        | CRD           | 120                   |
|                                                         | 401318      | 2                       | 6              | 10                        | JSN           | 120                   |
|                                                         | 401305      | 10                      | 6              | 2-5-2                     | CRD           | 120                   |
|                                                         | 401306      | 10                      | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401308      | 10                      | 6              | 2-5-2                     | CRD           | 65                    |
|                                                         | 401309      | 10                      | 6              | 2-5-2                     | JSN           | 65                    |
|                                                         | 401310      | 10                      | 6              | 2-5-2                     | DAO           | 65                    |
|                                                         | 401311      | 10                      | 6              | 2-2-2                     | CRD           | 120                   |
|                                                         | 401312      | 10                      | 6              | 2-2-2                     | JSN           | 120                   |
|                                                         | 401150      | 10                      | 6              | 2-1-1-10-1-1-1-1-1        | CRD           | 120                   |
|                                                         | 401271      | 6                       | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401275      | 6                       | 6              | 2-2-2                     | CRD           | 120                   |
|                                                         | 401276      | 6                       | 6              | 2-2-2                     | JSN           | 120                   |
|                                                         | 401278      | 6                       | 6              | 2-2-2                     | CRD           | 65                    |
|                                                         | 401281      | 6                       | 6              | 5-5-5                     | CRD           | 120                   |
|                                                         | 401282      | 6                       | 6              | 5-5-5                     | JSN           | 120                   |
|                                                         | 401425      | 6                       | 6              | 5-5-5-150-5               | CRD           | 120                   |
|                                                         | 401386      | 6                       | 6              | 5-5-5                     | CSL           | 120                   |
|                                                         | 401206      | 4                       | 5              | 10                        | CRD           | 120                   |
|                                                         | 401207      | 4                       | 5              | 10                        | JSN           | 120                   |
|                                                         | 401222      | 4                       | 5              | 5                         | CRD           | 120                   |
|                                                         | 401223      | 4                       | 5              | 5                         | JSN           | 120                   |
|                                                         | 401210      | 4                       | 6              | 10                        | CRD           | 120                   |
|                                                         | 401211      | 4                       | 6              | 10                        | JSN           | 120                   |
|                                                         | 401212      | 4                       | 6              | 10                        | DAO           | 120                   |
|                                                         | 401226      | 4                       | 6              | 5                         | CRD           | 120                   |
|                                                         | 401356      | 4                       | 6              | 5                         | CRD-1         | 120                   |
|                                                         | 401227      | 4                       | 6              | 5                         | JSN           | 120                   |
|                                                         | 401357      | 4                       | 6              | 5                         | JSN-1         | 120                   |
|                                                         | 401228      | 4                       | 6              | 5                         | DAO           | 120                   |
|                                                         | 401329      | 4                       | 6              | 5                         | DAO-1         | 120                   |
|                                                         | 401260      | 4                       | 6              | 2-5-2                     | CRD           | 120                   |
|                                                         | 401261      | 4                       | 6              | 2-5-2                     | JSN           | 120                   |
|                                                         | 401154      | 4                       | 6              | 5                         | CRD           | 120                   |
|                                                         | 401155      | 4                       | 6              | 5                         | JSN           | 120                   |
|                                                         | 401156      | 4                       | 6              | 5                         | DAO-1         | 120                   |
|                                                         | 401158      | 4                       | 6              | 10                        | CRD           | 120                   |

Table 5.1: Device Scope

{6}------------------------------------------------

# Indications for Use:

The Reprocessed Response Diagnostic Electrophysiology Catheters can be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites.

## Technological Characteristics:

The purpose, design, materials, function, and intended use of the Reprocessed Response Diagnostic Electrophysiology (EP) Catheters are identical to the predicate devices. There are no changes to the claims, clinical applications, patient population, performance specifications, or method of operation. In addition, Innovative Health's reprocessing of these devices includes removal of visible soil and decontamination. Each device is inspected and function tested prior to packaging and labeling.

## Functional and Safety Testing:

Bench and laboratory testing was conducted to demonstrate performance (safety and effectiveness) of the Reprocessed Response Diagnostic EP Catheter. This included the following:

- Biocompatibilitv ●
- Cleaning Validation ●
- Sterilization Validation
- Functional Testing ●
	- Visual Inspection .
	- Dimensional Verification ●
	- Electrical Continuity and Resistance .
	- Simulated Use .
	- . Mechanical Characteristics
- Electrical Safety Testing .
	- Dielectric and Current Leakage .
- . Packaging Validation

The Reprocessed Response Diagnostic EP Catheters are reprocessed no more than three (3) times. Each device is marked and tracked. After the device has reached the maximum number of reprocessing cycles, the device is rejected from further reprocessing. Reprocessing is performed only by Innovative Health. Innovative Health restricts its reprocessing to exclude devices previously reprocessed by other reprocessors.

#### Conclusion:

Innovative Health concludes that the Reprocessed Response Diagnostic EP Catheters are as safe and effective as the predicate devices described herein.

---

**Source:** [https://fda-staging.innolitics.com/device/K182488](https://fda-staging.innolitics.com/device/K182488)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).