The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature.
Device Story
CerusEndo Microcatheter is a percutaneous, intravascular catheter designed for delivery of therapeutic devices to peripheral, coronary, and neurovascular sites. Device features include a proximal Luer hub, hydrophilic coating, radiopaque marker, polymer tapered shaft, stainless steel reinforcement, and Teflon-lined inner lumen. Available in three configurations (150 cm or 160 cm lengths). Physician operates device in hospital interventional neuroradiology suites; access gained via introducer sheath, advanced over guidewire through guide catheter. Distal tip is steam-shapeable at point of use to navigate tortuous anatomy. Device facilitates precise placement of interventional tools; output is the successful delivery of therapeutic devices to target vascular locations. Benefits include improved access to complex vascular anatomy for minimally invasive procedures.
Clinical Evidence
Bench testing only. Testing included tensile strength, shaft flexibility, shape retention, kink resistance, static burst, simulated use, particulate assessment, coating durability, and sterilization validation. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility (hemolysis, complement activation, thrombogenicity), and genotoxicity (Ames assay, micronucleus). All tests met acceptance criteria.
Technological Characteristics
Percutaneous intravascular catheter. Materials: Polyether outer layer, stainless steel braid/coil, PTFE/polyether liner, platinum/iridium marker, nylon hub, Pebax strain relief. Hydrophilic coating. Dimensions: 150-160 cm length, 0.0205 in min ID. Steam-shapeable distal tip. Sterilization: Ethylene oxide. Single-use. No software/electronics.
Indications for Use
Indicated for patients undergoing vascular procedures requiring delivery of therapeutic devices to peripheral, coronary, or neurovasculature. No known contraindications.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 20, 2019
Cerus Endovascular, Inc. Theresa Brandner Vice President of Regulatory 47757 Fremont Boulevard Fremont, California 94538
Re: K182487
Trade/Device Name: CerusEndo Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: June 11, 2019 Received: June 19, 2019
Dear Theresa Brandner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182487
Device Name CerusEndo Microcatheter
### Indications for Use (Describe)
The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> | <input type="checkbox"/> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K182487
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 8807.92.
#### I. SUBMITTER
Submitter Name: Cerus Endovascular, Inc.
Address: 47757 Fremont Boulevard, Fremont, CA 94538
Phone Number: (510) 651-4000
Fax Number: (510) 405-8356
Contact Person: Theresa Brandner VP of Regulatory
Date Prepared: July 17, 2019
#### II. DEVICE
Name of Device: CerusEndo Microcatheter
Common Name: Catheter, Percutaneous
Classification Name: Percutaneous Catheter 21 CFR 870.1250
Regulatory Class: Class II
Product Code: DQY
#### III. PREDICATE DEVICE
K093160: Headway 21 Microcatheter manufactured by MicroVention Inc.
The predicate device has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
The subject device is a microcatheter available in three versions: a 150 cm length microcatheter with a 15 cm length distal segment; a 150 cm length microcatheter with a 34 cm length distal segment; and a 160 cm length microcatheter with a 42 cm length
{4}------------------------------------------------
distal segment. The distal segment of the catheter is flexible to facilitate access into tortuous anatomy, and the distal tip of the catheter is formable, allowing the physician to shape it according to the needs of the procedure at the point of use. The CerusEndo Microcatheter has a hydrophilic coating, radiopaque marker, Luer hub on the proximal end, polymer tapered shaft construction, stainless steel reinforced shaft, and Teflon lined inner lumen. The subject device is controlled by user manipulation to access discrete locations within the vascular anatomy. It is intended to deliver other interventional or therapeutic devices through its inner lumen. It is designed to be used in peripheral, coronary, and neurovascular locations.
#### V. INDICATIONS FOR USE
The CerusEndo Microcatheter is intended for general intravascular use to deliver therapeutic devices to the peripheral, coronary, and neurovasculature.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject device has similar design, dimensions, materials, intended use, and technological characteristics to the legally marketed predicate cleared under K093160. Non-clinical testing has been performed to demonstrate that any differences in technological characteristics do not raise new questions of safety and effectiveness. A table comparing the intended use and technical characteristics of the proposed device and the legally marketed predicate is provided in Table 1.
| | Subject Device | Predicate Device | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Manufacturer | CerusEndo Microcatheter<br>(Cerus Endovascular) | Headway Microcatheter<br>(MicroVention) | Comparison |
| Clinical Attributes | | | |
| Indications for Use | The CerusEndo Microcatheter is<br>intended for general intravascular<br>use to deliver therapeutic devices to<br>the peripheral, coronary, and<br>neurovasculature. | The Headway Microcatheter is<br>intended for general intravascular<br>use, including the peripheral,<br>coronary and neuro vasculature for<br>the infusion of diagnostic agents,<br>such as contrast media, and<br>therapeutic agents, such as<br>occlusion coils. | Same |
| Environments of Use | Hospital interventional<br>neuroradiology suites | Hospital interventional<br>neuroradiology suites | Identical |
| Patient Population | Patients undergoing vascular<br>procedures | Patients undergoing vascular<br>procedures | Identical |
| Contraindications | No known contraindications | No known contraindications | Identical |
| Functions | To facilitate introduction of<br>therapeutic devices into the<br>vasculature | To facilitate introduction of<br>diagnostic and therapeutic<br>devices into the vasculature | Same |
| | Subject Device | Predicate Device | |
| Manufacturer | CerusEndo Microcatheter | Headway Microcatheter | Comparison |
| | (Cerus Endovascular) | (MicroVention) | |
| Patient Access | Device access is gained using an | Device access is gained using an | Identical |
| | introducer sheath inserted into | introducer sheath inserted into | |
| | the vasculature and is advanced | the vasculature and is advanced | |
| | over a guidewire and through a | over a guidewire and through a | |
| | guide catheter | guide catheter | |
| Intraoperative | Yes | Yes | Identical |
| Use | | | |
| Technological Attributes | | | |
| General | Percutaneous Catheter | Percutaneous Catheter | Same |
| Description | Intravascular Catheter | Diagnostic Intravascular Catheter | |
| Device | Proximal Luer hub, hydrophilic | Proximal Luer hub, hydrophilic | Same |
| Configuration | coating, radiopaque marker, | coating, radiopaque marker, | |
| | polymer tapered shaft | polymer tapered shaft | |
| | construction, stainless steel | construction, stainless steel | |
| | reinforced shaft, and lined inner | reinforced shaft, and lined inner | |
| | lumen | lumen | |
| Catheter Body | Polyether outer layer, stainless steel | Polyurethane outer layer, | Similar |
| Materials | polyether inner layer, braid/coil, | polyether/polyamide inner layer, | |
| | PTFE/polyether liner | stainless steel braid/coil,<br>PTFE/polyolefin liner | |
| Marker | Platinum/iridium | Platinum/iridium | Same |
| Hub | Nylon | Nylon | Same |
| Strain Relief | Pebax | Pebax | Same |
| Introducer | Not applicable | Pebax | The CerusEndo |
| | | | Microcatheter is not |
| | | | provided with an |
| | | | introducer |
| Shaping Mandrel<br>Proximal ID/OD | Stainless Steel<br>0.0205 in min / 0.036 in | Stainless steel<br>0.021 in min / 0.033 in (2.5 Fr) | Same<br>Similar |
| | | | |
| Distal ID/OD | 0.0205 in min / 0.031 in | 0.021 in min / 0.026 in (2.0 Fr) | Similar |
| | | | |
| Effective Length | 150 cm, 160 cm | 150 cm, 156 cm | Similar |
| Coating | Hydrophilic coating | Hydrophilic coating | Similar |
| Tip Configuration | Straight - Steam Shapeable by | Straight - Steam Shapeable by | Same |
| Guidewire | physician prior to use<br>≤ 0.018 in | physician prior to use<br>≤ 0.018 in | Same |
| Compatibility | | | |
| Accessories | Shaping mandrel | Shaping mandrel, introducer | The CerusEndo |
| | | | Microcatheter is |
| | | | provided with a |
| | | | shaping mandrel but is |
| | | | not provided with an |
| | | | introducer |
| Method of Supply | Sterile, single use | Sterile, single use | Same |
| Sterilization | Ethylene oxide,<br>single patient use | Ethylene oxide,<br>single patient use | Same |
| Biocompatibility | Biocompatible per ISO 10993-1 | Biocompatible per ISO 10993-1 | Same |
Table 1 - Comparison Table of Proposed Device and Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
There are no significant technological differences between the subject device and predicate device. The technological differences between the subject and predicate device are indicated in the table above.
#### VII. PERFORMANCE DATA
Non-clinical tests were performed to demonstrate safety and substantial equivalence. Bench testing performed is summarized in Table 2. Biocompatibility testing performed is summarized in Table 3.
| Test | Test Method Summary | Results |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tensile Strength | This test measures the tensile strength<br>of the catheter bonds using a tensile<br>tester and are pulled to failure. | Device met acceptance<br>criteria |
| Shaft Flexibility<br>(stiffness) | This test measures the bending stiffness<br>of the distal and proximal catheter shaft<br>segments. | Device met acceptance<br>criteria |
| Shape Retention | This test measures the ability of the<br>catheter tip to form and retain a steam<br>shape using conventional catheter lab<br>shaping techniques. | Device met acceptance<br>criteria |
| Kink Resistance | This test measures the distal and<br>proximal catheter shaft resistance to<br>kinking. | Device met acceptance<br>criteria |
| Static Burst | This test measures the catheters<br>resistance to burst failure by using a<br>high pressure injector to pressurize the<br>lumen while the distal tip is occluded. | Device met acceptance<br>criteria |
| Simulated Use | The device was used in accordance to<br>the Instructions for Use. | Device met acceptance<br>criteria |
| Particulate | This test assesses the coating integrity<br>by measuring the quantity and size of<br>particles generated during simulated<br>use of the device in an anatomical<br>model. | Device met acceptance<br>criteria |
| Coating Friction and<br>Durability | This test measures the lubricity of the<br>coating and the durability after repeated<br>abrasion cycles. | Device met acceptance<br>criteria |
| Sterilization<br>Validation | These tests confirms a minimum 6 log<br>sterility assurance level, assesses<br>ethylene oxide sterilant residual levels,<br>and detects pyrogens | Device met acceptance<br>criteria |
| Test | Extract/Test System | Results |
| Cytotoxicity (MEM<br>Elution) | The test extract, a positive control, and<br>a negative control were extracted at<br>37°C for 24 hours in MEM solution<br>(5% serum supplemented cell culture<br>medium) and exposed to mouse<br>fibroblast cells. | Non-cytotoxic.<br>The test article is<br>considered non-<br>cytotoxic<br>to cells. |
| Sensitization<br>(ISO Kligman<br>Guinea Pig<br>Maximization Test) | Test samples and control blanks were<br>extracted in normal saline and in cotton<br>seed oil at 37°C for 72 hours. | Non-sensitizing.<br>The test article did not<br>elicit a sensitization<br>response. |
| Irritation/<br>ISO Intracutaneous<br>Toxicity in<br>Rabbits | Test samples were extracted in normal<br>saline and in cotton seed oil at 37°C for<br>72 hours. | Non-irritant.<br>No evidence of<br>irritation. |
| Systemic Toxicity<br>(ISO Systemic<br>Injection) | Test samples and negative control<br>blanks were extracted and were<br>prepared in normal saline and in cotton<br>seed oil at 37°C for 72 hours. | Non-cytotoxic.<br>No weight loss,<br>mortality, or evidence of<br>systemic toxicity from<br>the extract exposure to<br>the mice. |
| Systemic Toxicity<br>(Rabbit Material-<br>Mediated<br>Pyrogenicity) | Test samples were extracted and<br>negative control blanks were prepared<br>in normal saline and in cotton seed oil<br>at 37°C for 72 hours. | Non-pyrogenic.<br>All individual rabbits for<br>both the test article and<br>negative control showed<br>a total rise in<br>temperature of < 0.5°C<br>and were determined to<br>be nonpyrogenic. |
| Hemocompatibility<br>(ISO In Vitro<br>Hemocompatibility<br>- Direct Contact) | Blood samples from three human<br>donors were pooled and diluted. The<br>test article is added to aliquots of<br>human blood and incubated at 37 °C<br>for a minimum of 3 hours. | Non-hemolytic. There<br>were no differences<br>between the hemolytic<br>index of the test article<br>and the negative control. |
| Hemocompatibility<br>(Hemolysis -<br>Indirect) | Test samples were extracted in<br>phosphate buffered saline at 37°C for<br>72 hours. The test article extract was<br>incubated at 37°C for a minimum of 3<br>hours. | Non-hemolytic.<br>There were no<br>significant differences<br>between the test article<br>extract/solid and<br>negative control article<br>results. |
| Hemocompatibility<br>(ISO Complement<br>Activation C3 and<br>SC5b-9 Test –<br>Direct Contact) | The test article, predicate, negative<br>control, and positive control (latex)<br>were added to the serum pooled from<br>three human blood samples. All were<br>incubated at 37 °C for 30, 60, and 90<br>minutes. | C3a and SC5b-9<br>complement proteins<br>were considered to be<br>non-activated as<br>compared to the<br>negative control. |
| Genotoxicity – Gene<br>Mutation<br>(Ames Assay,<br>S-9 Activation) | Test samples and negative control<br>blanks were extracted in normal saline<br>and PEG 400 at 37°C for 72 hours. | Non-mutagenic.<br>Based on the acceptance<br>criteria under the<br>experimental conditions<br>utilized, the test article<br>extracts were both<br>deemed non-mutagenic<br>in all strains under both<br>non-activated and<br>activated conditions. |
| Genotoxicity –<br>Micronucleus | Test samples and negative control<br>blanks were extracted in polar (saline)<br>and non-polar solvents at 37 °C for 72<br>hours. The extracts are placed on<br>Chinese Hamster Ovary (CHO) cells<br>with and without metabolic activations<br>(S-9). The cells are then incubated at<br>37 °C, with 5% CO2, for 15 – 21 hours.<br>The cells are harvested and the<br>chromosomes stained and examined for<br>aberrations. | Not genotoxic.<br>No significant increase<br>of aberrations when<br>compared to the<br>negative controls |
| Hemocompatibility<br>(Thrombogenicity) | Devices were placed in a canine carotid<br>vessel. | Non-thrombogenic. No<br>significant thrombus<br>was observed on any of<br>the subject catheters,<br>and the device was<br>determined to not show<br>thrombogenic potential |
## Table 2 - Bench Testing Summary
{7}------------------------------------------------
# Table 3 – Biocompatibility Testing Summary
{8}------------------------------------------------
## VIII. CONCLUSION
Differences between the subject device and the predicate device do not raise different questions of safety and effectiveness. After a comparison of the subject CerusEndo Microcatheter intended use, technological characteristics, and expected performance to the legally marketed predicate, the Headway 21 Microcatheter (K093160), Cerus Endovascular concludes that the CerusEndo Microcatheter performs as safely and effectively as the predicate device for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.