alpha-AMYLASE DIRECT, alpha-AMYLASE EPS, alpha-AMYLASE PANCREATIC, BILIRUBIN DIRECT, BILIRUBIN TOTAL

K182474 · Biosystems S.A. · JFJ · May 2, 2019 · Clinical Chemistry

Device Facts

Record IDK182474
Device Namealpha-AMYLASE DIRECT, alpha-AMYLASE EPS, alpha-AMYLASE PANCREATIC, BILIRUBIN DIRECT, BILIRUBIN TOTAL
ApplicantBiosystems S.A.
Product CodeJFJ · Clinical Chemistry
Decision DateMay 2, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1070
Device ClassClass 2

Indications for Use

alpha-AMYLASE DIRECT: Reagent for the measurement of alpha-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of acute and chronic pancreatitis. This reagent institus of in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. alpha-AMYLASE EPS: Reagent for the measurement of alpha-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of acute and chronic pancreatitis. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. alpha-AMYLASE PANCREATIC: Reagent for the measurement of pancreatic c-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of acute and chronin pancreatitis. This reagentis for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. BILIRUBIN DIRECT: Reagent for the measurement of direct bilirubin concentration in human serum or plasma. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hematological and metabolic disorders, including hepatitis and gall bladder block. This reagent is for use in the BioSystems Blockers. Only for in vitro use in the clinical laboratory. BILIRUBIN TOTAL: Reagent for the measurement of total bilirubin concentration in human serum or plasma. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, herrytological and metabolic disorders, including hepatitis and gall bladder block. This reagent is for nuse in the BioSystems Blockers. Only for in vitro use in the clinical laboratory.

Device Story

Photometric assays for clinical chemistry analysis; reagents react with patient samples (serum, plasma, urine) to produce colored complexes or enzymatic rate changes; measured via spectrophotometry at 405 nm or 535 nm on BioSystems BA400 analyzers. Amylase assays use enzymatic hydrolysis (CNP-G3 or 4-nitrophenyl-maltoheptaoside-ethylidene); pancreatic amylase assay uses monoclonal antibodies to inhibit salivary isoenzyme. Bilirubin assays use 3,5-dichlorophenyl diazonium salt reaction. Used in clinical laboratories by trained personnel; results aid physicians in diagnosing pancreatitis, liver, and metabolic disorders.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP05-A3 (n=80 per level). Linearity evaluated per CLSI EP06-A. Detection limits (LoB, LoD, LoQ) evaluated per CLSI EP17-A2. Analytical specificity/interference evaluated per CLSI EP07-A2. Method comparison against predicate devices performed on BA400 analyzer showing high correlation (R > 0.99). No clinical studies performed.

Technological Characteristics

Photometric enzymatic and colorimetric assays. Reagents include buffers (MES, HEPES), substrates (CNP-G3, 4-Nitrophenyl-maltoheptaoside-ethylidene), enzymes (alpha-glucosidase), and diazonium salts. Analyzed on BioSystems BA400 automated clinical chemistry analyzer. Traceability to internal Master Calibrators or NIST SRM 916. Storage at 2-8°C.

Indications for Use

Indicated for measurement of alpha-amylase, pancreatic alpha-amylase, total bilirubin, or direct bilirubin in human serum, plasma, or urine to aid in diagnosis/treatment of pancreatitis, liver, hemolytic, hematological, or metabolic disorders. For prescription, in vitro diagnostic use in clinical laboratories using BioSystems BA400 analyzers.

Regulatory Classification

Identification

An amylase test system is a device intended to measure the activity of the enzyme amylase in serum and urine. Amylase measurements are used primarily for the diagnosis and treatment of pancreatitis (inflammation of the pancreas).

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k182474 B. Purpose for Submission: New devices C. Measurand: α-Amylase Pancreatic α-Amylase Bilirubin Total Bilirubin Direct D. Type of Test: α-Amylase assays: Quantitative, enzymatic assays Bilirubin (Total and Direct) assays: Colorimetric, photometric assays E. Applicant: BioSystems S.A. F. Proprietary and Established Names: alpha-AMYLASE DIRECT alpha-AMYLASE EPS alpha-AMYLASE PANCREATIC BILIRUBIN TOTAL BILIRUBIN DIRECT G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | JFJ | II | 21 CFR § 862.1070 Amylase test system | Clinical Chemistry (75) | | CIG | II | 21 CFR § 862.1110 Bilirubin (total or direct) test system | | {1} H. Intended Use: 1. Intended use(s): See Indications for use below 2. Indication(s) for use: alpha-AMYLASE DIRECT: Reagent for the measurement of alpha-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis and of acute and chronic pancreatitis. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. alpha-AMYLASE EPS: Reagent for the measurement of alpha-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of acute and chronic pancreatitis. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. alpha-AMYLASE PANCREATIC: Reagent for the measurement of pancreatic $\alpha$-amylase concentration in human serum, plasma or urine. The obtained values are useful as an aid in the diagnosis of acute and chronic pancreatitis. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. BILIRUBIN TOTAL: Reagent for the measurement of total bilirubin concentration in human serum or plasma. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. BILIRUBIN DIRECT: Reagent for the measurement of direct bilirubin concentration in human serum or plasma. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block. This reagent is for use in the BioSystems BA analyzers. Only for in vitro use in the clinical laboratory. 3. Special conditions for use statement(s): These devices are intended for prescription use and in vitro diagnostic use only. 4. Special instrument requirements: BioSystems BA400 {2} I. Device Description: α-AMYLASE DIRECT: MES 50 mmol/L, calcium chloride 5 mmol/L, sodium chloride 300 mmol/L, sodium thiocyanate 450 mmol/L, CNP-G3 2.25 mmol/L, pH 6.1. α-AMYLASE EPS: A: 2 x 60 mL. HEPES 50 mmol/L, calcium chloride 0.075 mmol/L, sodium chloride 90 mmol/L, magnesium chloride 13 mmol/L, α-glucosidase > 4 U/mL, pH 7.1. B: 2 x 15 mL. HEPES 50 mmol/L, 4-Nitrophenyl-maltoheptaoside-ethylidene 18 mmol/L, pH 7.1. α-AMYLASE PANCREATIC: A: 2 x 60 mL. HEPES 50 mmol/L, calcium chloride 0.075 mmol/L, sodium chloride 90 mmol/L, magnesium chloride 13 mmol/L, α-glucosidase > 4 U/mL, pH 7.1, monoclonal antibodies (mouse) 50 mg/L. B: 2 x 15 mL. HEPES 50 mmol/L, 4-Nitrophenyl-maltoheptaoside-ethylidene 18 mmol/L, pH 7.1. BILIRUBIN DIRECT: A: Phosphoric acid 90 mmol/L, HEDTA 4.5 mmol/L, sodium chloride 50 mmol/L, pH 1.5. B: 3,5-dichlorophenyl diazonium 1.5 mmol/L. BILIRUBIN TOTAL: A: Hydrochloric acid 170 mmol/L, cetrimide 40 mmol/L, pH 0.9. B: 3,5-dichlorophenyl diazonium 1.5 mmol/L. J. Substantial Equivalence Information: 1. Predicate device name(s): AMYL-α-Amylase EPS, BILD2- Bilirubin Direct Gen.2, BILT2- Bilirubin Total Gen.2 2. Predicate 510(k) number(s): k100853 {3} 3. Comparison with predicate: | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate Device alpha-AMYLASE DIRECT (k182474) | Predicate Device AMYL-α-Amylase EPS (k100853) | | Intended Use | Reagent for the measurement of α-amylase concentration in human serum, plasma or urine. | Same | | Measurand | α-amylase | Same | | Sample Type | Serum, plasma or urine | Same | | Assay Technology | Direct Substrate | IFCC | | Assay Range | 8.20-1300 U/L | 3-1500 U/L | | Reagent storage | Store at 2-8 °C until the expiry date marked in the label. | Same | | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate Device alpha-AMYLASE EPS (k182474) | Predicate Device AMYL-α-Amylase EPS (k100853) | | Intended Use | Reagent for the measurement of α-amylase concentration in human serum, plasma or urine. | Same | | Measurand | α-amylase | Same | | Sample Type | Serum, plasma or urine | Same | | Assay Technology | IFCC | Same | | Assay Range | 14.3-1300 U/L | 3-1500 U/L | | Reagent storage | Store at 2-8 °C until the expiry date | Same | | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate Device alpha-AMYLASE PANCREATIC (k182474) | Predicate Device AMYL-α-Amylase EPS Pancreatic (k100853) | | Intended Use | Reagent for the measurement of pancreatic α-amylase concentration in human serum, plasma or urine. | Same | | Measurand | Pancreatic α-amylase | Same | | Sample Type | Serum, plasma or urine | Same | | Assay Technology | Immuno-inhibition | Same | | Assay Range | 7.49-1300 U/L | 3-1500 U/L | | Reagent storage | Store at 2-8 °C until the expiry date | Same | {4} | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate Device BILIRUBIN DIRECT (k182474) | Predicate Device BILD2- Bilirubin Direct Gen.2 (k100853) | | Intended Use | Reagent for the measurement of direct bilirubin concentration in human serum or plasma. | Same | | Measurand | Bilirubin | Same | | Sample Type | Serum, plasma | Same | | Assay Technology | Dichlorophenyl diazonium | Same | | Assay Range | 0.14-15 mg/dL | 0.09-17 mg/dL | | Reagent storage | Store at 2-8 °C until the expiry date | Same | | Similarities and Differences | | | | --- | --- | --- | | Item | Candidate Device BILIRUBIN TOTAL (k182474) | Predicate Device BILT2- Bilirubin Total Gen.2 (k100853) | | Intended Use | Reagent for the measurement of total bilirubin concentration in human serum or plasma. | Same | | Measurand | Bilirubin | Same | | Sample Type | Serum, plasma | Same | | Assay Technology | Dichlorophenyl diazonium | Same | | Assay Range | 0.26-38 mg/dL | 0.1-38 mg/dL | | Reagent storage | Store at 2-8 °C until the expiry date | Same | # K. Standard/Guidance Document Referenced (if applicable): CLSI - EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Third Edition. CLSI - EP06-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline. CLSI - EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition. # L. Test Principle: The alpha-AMYLASE DIRECT reagent is a photometric assay in which the $\alpha$ -amylase catalyzes the hydrolysis of 2-chloro-4-nitrophenyl-malto-trioside (CNPG3) to 2-chloro-4-nitrophenol (CNP). The catalytic concentration is determined from the rate of 2-chloro-4-nitrophenol formation, measured at $405~\mathrm{nm}$ . {5} The alpha-AMYLASE EPS reagent is a photometric assay in which the $\alpha$ -amylase catalyzes the hydrolysis of 4-nitrophenyl-maltoheptaoside-ethylidene to smaller oligosaccharides which are hydrolyzed by $\alpha$ -glucosidase liberating 4-nitrophenol. The catalytic concentration is determined from the rate of 4-nitrophenol formation, measured at $405~\mathrm{nm}$ . The alpha-AMYLASE PANCREATIC reagent is a photometric assay in which pancreatic $\alpha$ -amylase catalyzes the hydrolysis of 4-nitrophenyl-maltoheptaoside-ethylidene to smaller oligosaccharides which are hydrolyzed by $\alpha$ -glucosidase liberating 4-nitrophenol. The catalytic concentration is determined from the rate of 4-nitrophenol formation, measured at $405~\mathrm{nm}$ . Specific antibodies inhibit the salivary isoenzyme and thus allow the measurement of pancreatic $\alpha$ -amylase. The BILIRUBIN DIRECT reagent is a photometric assay in which the direct bilirubin in the sample reacts with 3,5-dichlorophenyl diazonium salt forming a colored complex that can be measured by spectrophotometry at $535~\mathrm{nm}$ . The BILIRUBIN TOTAL reagent is a photometric assay in which both the direct and indirect bilirubin in the sample reacts with 3,5-dichlorophenyl diazonium salt in the presence of cetrimide forming a colored complex that can be measured by spectrophotometry at $535~\mathrm{nm}$ . # M. Performance Characteristics (if/when applicable): # 1. Analytical performance: # a. Precision/Reproducibility: Precision studies were performed according to the CLSI EP5-A3 guideline. For alpha-AMYLASE DIRECT, alpha-AMYLASE EPS and alpha-AMYLASE PANCREATIC, three levels of commercial biochemistry control serum (human serum matrix) or Urine, Level I, II and III, and for BILIRUBIN DIRECT and BILIRUBIN TOTAL, four levels of commercial biochemistry control serum (human serum matrix) Level I, II, III and IV were tested on two runs per day, two measurements per run, for twenty days with total of 80 observations. The results of the precision studies are summarized in the tables below: alpha-AMYLASE DIRECT: | Matrix | Samples | n | Mean (U/L) | Repeatability | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | CV% | SD | CV% | | Serum | QC 1 | 80 | 53.3 | 0.8 | 1.6 | 1.6 | 2.2 | | | QC 2 | 80 | 201 | 1.0 | 0.5 | 0.5 | 0.9 | | | QC 3 | 80 | 1061 | 6.7 | 0.6 | 0.6 | 1.5 | | Urine | QC 1 | 80 | 92 | 2.3 | 2.5 | 2.3 | 2.5 | | | QC 2 | 80 | 407 | 10.1 | 2.5 | 10.2 | 2.5 | | | QC 3 | 80 | 2078 | 12.6 | 0.6 | 35.4 | 1.7 | {6} alpha-AMYLASE EPS: | Matrix | Samples | n | Mean (U/L) | Repeatability | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | CV% | SD | CV% | | Serum | QC 1 | 80 | 54.3 | 0.9 | 1.7 | 1.9 | 3.5 | | | QC 2 | 80 | 207 | 4.5 | 2.2 | 4.9 | 2.4 | | | QC 3 | 80 | 1090 | 4.3 | 0.4 | 16.9 | 1.6 | | Urine | QC 1 | 80 | 103 | 2.3 | 2.2 | 2.8 | 2.7 | | | QC 2 | 80 | 384 | 4.7 | 1.2 | 7.8 | 2.0 | | | QC 3 | 80 | 2161 | 26.1 | 1.2 | 36.6 | 1.7 | alpha-AMYLASE PANCREATIC: | Matrix | Samples | n | Mean (U/L) | Repeatability | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | CV% | SD | CV% | | Serum | QC 1 | 80 | 52.7 | 0.6 | 1.0 | 1.2 | 2.2 | | | QC 2 | 80 | 156 | 2.1 | 1.4 | 2.2 | 1.4 | | | QC 3 | 80 | 1058 | 6.1 | 0.6 | 11.7 | 1.1 | | Urine | QC 1 | 80 | 66.2 | 1.4 | 2.1 | 1.6 | 2.5 | | | QC 2 | 80 | 366 | 5.4 | 1.5 | 6.2 | 1.7 | | | QC 3 | 80 | 2259 | 43.8 | 1.9 | 45.2 | 2.0 | BILIRUBIN DIRECT: | Matrix | Samples | n | Mean (mg/dL) | Repeatability | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | CV% | SD | CV% | | Serum | QC 1 | 80 | 0.35 | 0.03 | 7.7 | 0.03 | 8.0 | | | QC 2 | 80 | 0.61 | 0.03 | 4.1 | 0.03 | 5.0 | | | QC 3 | 80 | 1.68 | 0.03 | 2.0 | 0.05 | 2.9 | | | QC 4 | 80 | 15.1 | 0.09 | 0.6 | 0.19 | 1.3 | BILIRUBIN TOTAL: | Matrix | Samples | n | Mean (mg/dL) | Repeatability | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | | | SD | CV% | SD | CV% | | Serum | QC 1 | 80 | 1.03 | 0.09 | 8.5 | 0.09 | 8.7 | | | QC 2 | 80 | 2.08 | 0.07 | 3.3 | 0.07 | 3.3 | | | QC 3 | 80 | 4.90 | 0.05 | 0.9 | 0.09 | 1.8 | | | QC 4 | 80 | 30.9 | 0.13 | 0.4 | 0.33 | 1.1 | b. Linearity/assay reportable range: Linearity studies were conducted according to the CLSI EP06-A guideline. Eleven levels of dilutions were prepared by mixing different proportions of samples with high and low concentration of each analyte. {7} Samples were analyzed on one BA-400 instrument in duplicate. The results of the regression analysis are shown below: | Analyte | Matrix | Slope | Intercept | r2 | Range Conc. tested | Claimed range | | --- | --- | --- | --- | --- | --- | --- | | alpha-AMYLASE DIRECT (U/L) | Serum | 1.011 | -3.50 | 1.000 | 0.0 - 1400 | 8.18 - 1300 | | | Urine | 1.006 | 1.58 | 1.000 | 0.0 - 2700 | 19.2 - 2600 | | alpha-AMYLASE EPS (U/L) | Serum | 0.981 | 3.48 | 1.000 | 0.0 - 1400 | 14.3 - 1300 | | | Urine | 0.982 | 9.72 | 1.000 | 0.0 - 2700 | 19.1 - 2600 | | alpha-AMYLASE PANCREATIC (U/L) | Serum | 0.982 | 2.13 | 1.000 | 0.0 - 1300 | 8.18 - 1300 | | | Urine | 0.990 | 4.46 | 1.000 | 0.0 - 2700 | 19.6 - 2600 | | BILIRUBIN DIRECT (mg/dL) | Serum | 0.999 | 0.003 | 1.000 | 0.08 - 22.80 | 0.14 - 15.00 | | BILIRUBIN TOTAL (mg/dL) | Serum | 0.972 | 0.327 | 0.999 | 0.00 - 38.00 | 0.32 - 38.00 | The linearity studies support the sponsor's claimed measuring ranges as described in the table above. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability: alpha-AMYLASE DIRECT reagent: Values assigned and traceable to an internal BioSystems Master Calibrator (BMC). alpha-AMYLASE-EPS reagent: Values assigned and traceable to C-SER/IFCC; International Federation of Clinical Chemistry and Laboratory Medicine. alpha-AMYLASE-PANCREATIC reagent: Values assigned and traceable to C-SER/IFCC; International Federation of Clinical Chemistry and Laboratory Medicine. BILIRUBIN DIRECT Reagent: Values assigned and traceable to an internal BioSystems Master Calibrator (BMC). BILIRUBIN TOTAL Reagent: Values assigned and traceable to SRM 916 (NIST) Standard Reference Materials. {8} d. Detection limit: Detection capabilities studies for each analyte were evaluated based upon the CLSI EP17-A2 guideline. Limit of blank (LoB) studies were performed by testing 5 blank samples in replicates of 4 over 3 days, on two lots for a total of 120 observations. LoB was defined as the highest result that can reasonably be expected from a blank sample for a given error probability with $\alpha = 0.05$ . Limit of detection (LoD) studies were performed by testing 5 pool of human serum or urine with analyte concentrations close to expected detection limit for each analyte. Samples were tested in replicates of 4 over 3 days, on two lots for a total of 120 observations. The LoD was estimated based on a probability of error of $5\%$ ( $\beta = 0.05$ ). Limit of quantitation (LoQ) studies were performed using 4 to 5 pools of human serum or urine with analytes concentrations closed to the expected LoQ of the corresponding assay. Samples were tested in replicates of 4 over 3 days, on two lots for a total of 120 observations. The sponsor defines LoQ as the lowest concentration of a measurand with an imprecision $\leq 10\%$ . The results of the detection limit studies are presented in the table below: | Assays | Matrix | LoB | LoD | LoQ | Claimed Range | | --- | --- | --- | --- | --- | --- | | alpha-AMYLASE DIRECT (U/L) | Serum | 0.50 | 1.64 | 8.18 | 8.18 - 1300 | | | Urine | 1.36 | 5.14 | 19.2 | 19.2 - 2600 | | alpha-AMYLASE EPS (U/L) | Serum | 1.53 | 4.63 | 14.3 | 14.3 - 1300 | | | Urine | 4.48 | 8.90 | 19.1 | 19.1 - 2600 | | alpha-AMYLASE PANCREATIC (U/L) | Serum | 2.26 | 4.95 | 8.14 | 8.18 - 1300 | | | Urine | 3.06 | 6.42 | 19.6 | 19.6 - 2600 | | BILIRUBIN DIRECT (mg/dL) | Serum | 0.07 | 0.10 | 0.14 | 0.14 - 15.00 | | BILIRUBIN TOTAL (mg/dL) | Serum | 0.14 | 0.21 | 0.32 | 0.32 - 38.00 | e. Analytical specificity: Interference studies were performed in accordance with the CLSI EP07-A2 guideline. For alpha-AMYLASE DIRECT, alpha-AMYLASE EPS, alpha-AMYLASE PANCREATIC, specimens consist on human serum and urine representing low and high concentrations of the analyte, with and without added increasing concentration of the substance tested for interference. Interference studies for BILIRUBIN TOTAL and BILIRUBIN DIRECT were conducted using serum samples. Samples were spiked {9} with increasing concentrations of the potential interferents and analyzed in duplicate. Interference was calculated as: (b-a)/a, where a and b are the concentration of measurand found without (a) and with (b) interferent. The sponsor states that interference is considered to be non-significant if the difference between the samples with and without interferent are within $10\%$ . The results of the interference studies are summarized in the table below. | Analyte | Interferent | Highest concentration tested that did not show significant interference | | --- | --- | --- | | alpha-AMYLASE DIRECT | Bilirubin | 30 mg/dL | | | Hemoglobin | 200 mg/dL | | | Triglycerides | 1756 mg/dL | | alpha-AMYLASE EPS | Bilirubin | 30 mg/dL | | | Hemoglobin | 604mg/dL | | | Triglycerides | 659 mg/dL | | alpha-AMYLASE PANCREATIC | Bilirubin | 30 mg/dL | | | Hemoglobin | 500 mg/dL | | | Triglycerides | 1625 mg/dL | | BILIRUBIN DIRECT | Hemoglobin | 25 mg/dL | | | Triglycerides | 975 mg/dL | | BILIRUBIN TOTAL | Hemoglobin | 500 mg/dL | | | Triglycerides | 512 mg/dL | Exogenous interferences: | Analyte | Matrix | Interferent | Highest concentration tested that did not show significant interference | | --- | --- | --- | --- | | alpha-AMYLASE DIRECT | Serum | Acetaminophen | 20 mg/dL | | | | Ascorbic Acid | 30 mg/dL | | | | Acetylsalicylic Acid | 70 mg/dL | | | Urine | Ascorbic Acid | 30 mg/dL | | alpha-AMYLASE EPS | Serum | Acetaminophen | 20 mg/dL | | | | Ascorbic Acid | 30 mg/dL | | | | Acetylsalicylic Acid | 70 mg/dL | | | Urine | Ascorbic Acid | 30 mg/dL | | alpha-AMYLASE PANCREATIC | Serum | Acetaminophen | 20 mg/dL | | | | Ascorbic Acid | 30 mg/dL | | | | Acetylsalicylic Acid | 70 mg/dL | | | Urine | Ascorbic Acid | 30 mg/dL | | BILIRUBIN DIRECT | Serum | Acetaminophen | 20 mg/dL | | | | Ascorbic Acid | 30 mg/dL | | BILIRUBIN TOTAL | Serum | Acetaminophen | 20 mg/dL | | | | Ascorbic Acid | 30 mg/dL | {10} f. Assay cut-off: Not applicable ## 2. Comparison studies: a. Method comparison with predicate device: Method comparison studies were conducted by testing native human serum, plasma and urine samples (as applicable) on the $\alpha$-AMYLASE DIRECT, $\alpha$-AMYLASE EPS, $\alpha$-AMYLASE PANCREATIC, BILIRUBIN DIRECT and BILIRUBIN TOTAL candidate assays on the BA400 analyzer and the predicate devices. The results of the regression analysis are summarized below: | Analyte | Specimen Type | N | Slope | Intercept | R | Test range* | Claimed Measuring Range | | --- | --- | --- | --- | --- | --- | --- | --- | | alpha-AMYLASE DIRECT (U/L) | Serum | 83 | 0.994 | 4.74 | 0.999 | 17.5-1285 | 8.18-1300 | | | Li Heparin Plasma | 110 | 0.951 | 2.75 | 0.995 | 19.3-1295 | | | | Urine | 136 | 1.029 | 21.65 | 0.998 | 79-2572 | 19.2-2600 | | alpha-AMYLASE EPS (U/L) | Serum | 136 | 1.075 | 0.04 | 0.997 | 13.9-1287 | 14.3-1300 | | | Li Heparin Plasma | 111 | 1.109 | -5.317 | 0.999 | 19.3-1217 | | | | Urine | 136 | 0.932 | 2.77 | 0.993 | 86.7-2363 | 19.1-2600 | | alpha-AMYLASE PANCREATIC (U/L) | Serum | 110 | 1.000 | -1.20 | 0.999 | 12.0-1289 | 8.18-1300 | | | Li Heparin Plasma | 115 | 1.011 | -3.176 | 0.998 | 12.0-1283 | | | | Urine | 81 | 1.063 | -1.30 | 1.000 | 12.0-2410 | 19.6-2600 | | BILIRUBIN DIRECT (mg/dL) | Serum | 125 | 1.005 | 0.02 | 0.997 | 0.33-15.0 | 0.14-15 | | | Li Heparin Plasma | 142 | 1.029 | 0.01 | 0.997 | 0.12-14.0 | | | BILIRUBIN TOTAL (mg/dL) | Serum | 108 | 1.015 | 0.09 | 0.997 | 0.55-35.1 | 0.32-38 | | | K₃ EDTA plasma | 106 | 1.02 | 0.02 | 0.998 | 0.28-28.8 | | *Concentrations tested based on the results reported by the predicate devices. Method comparison data supports that serum and lithium heparin plasma are acceptable samples types for the BILIRUBIN DIRECT assay. Studies also support that serum and $\mathrm{K}_3$ EDTA plasma samples are acceptable sample types for the {11} BILIRUBIN TOTAL assay. The alpha-AMYLASE DIRECT, alpha-AMYLASE EPS, alpha-AMYLASE PANCREATIC assays can be used with serum, lithium heparin plasma and urine specimens. b. Matrix comparison: Not applicable. Different sample types were assessed in other analytical studies described above. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: The following reference ranges are cited from the literature: alpha-AMYLASE DIRECT¹: Serum and plasma: 22 - 80 U/L Urine: <321 U/L alpha-AMYLASE EPS²,³: Serum and plasma: 28 - 100 U/L Urine: 16 - 491 U/L alpha-AMYLASE PANCREATIC²,³: Serum and Plasma: 13-53 U/L Urine: 7 - 356 U/L BILIRUBIN DIRECT¹,⁴: Adults: 0 - 0.3 mg/dL 12 {12} BILIRUBIN TOTAL¹: Adults: 0 - 2.0 mg/dL 1- Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 5th ed. Burtis CA, Ashwood ER, Bruns DE. WB Saunders Co, 2012. 2- IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37°C. Part 8. Reference procedure for the measurement of catalytic concentration of α-amylase. Clin Chem Lab Med 2006; 44: 1146-1155. 3- Junge W, Werner W, Wilke B et al. Development and evaluation of assays for the determination of total and pancreatic amylase at 37°C according to the principle recommended by the IFCC. Clin Biochem 2001;34:607-615. 4- McPherson RA, Pincus MR. Henry's Clinical Diagnosis and Management by Laboratory Methods. 20st ed. Saunders Elsevier, 2001:1427. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 13
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...