The O2 Concepts Oxlife Independence is indicated on a prescription basis for the administration of supplemental oxygen. It is not intended for life support, nor does it provide any patient monitoring capabilities.
Device Story
Portable oxygen concentrator for home, institutional, or mobile use; provides supplemental oxygen to patients. Uses Pressure Swing Adsorption (PSA) with molecular sieve to convert ambient air to 87%-95% pure oxygen. Delivers oxygen via standard single lumen nasal cannula in either pulse mode (detects patient breath) or continuous flow mode. Includes Dynamic Network Analysis (DNA) technology for remote tracking of device location, performance, and usage data via cellular/GPS. Operated by patient or caregiver; settings adjusted via buttons/LCD. Benefits patients by providing portable oxygen therapy, improving mobility and quality of life. Does not provide life support or patient monitoring.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601 series), and usability validation with 10 users. Particulate matter and gas emissions (CO, CO2, ozone) were within safety limits.
Technological Characteristics
Portable oxygen concentrator; Pressure Swing Adsorption (PSA) with molecular sieve. Power: 100-240VAC, 12-15VDC, or rechargeable batteries. Connectivity: Cellular/GPS for DNA technology. Standards: IEC 60601-1, IEC 60601-1-2, ISO 80601-2-69, ISO 8359, ISO 10993. Biocompatibility: ISO 10993-5/10/12. Software: Embedded, moderate level of concern.
Indications for Use
Indicated for adult patients with chronic pulmonary diseases (e.g., chronic bronchitis, emphysema, asthma, lung cancer), terminal cancer, or any patient requiring supplemental oxygen. Prescription use only.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
O2 Concepts LLC % Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street. Nw Buffalo, Minnesota 55313
Re: K182454
Trade/Device Name: O2 Concepts Oxlife Independence Model 301-0001 Regulation Number: 21 CFR 868.5440 Regulation Name: Portable oxygen generator Regulatory Class: Class II Product Code: CAW Dated: November 2, 2018 Received: November 5, 2018
Dear Mark Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Todd D. Courtney -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182454
Device Name Oxlife Independence
Indications for Use (Describe)
The O2 Concepts Oxlife Independence is indicated on a prescription basis for the administration of supplemental oxygen. It is not intended for life support, nor does it provide any patient monitoring capabilities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for O2 Concepts. The logo features a blue circle with a white swoosh across the top. The number "2" is inside the circle. Below the circle, the word "Concepts" is written in a sans-serif font.
# 5. 510 k) Summary
#### Submitter I.
O2 Concepts LLC 199 Park Road Extension Middlebury, CT 06762 Phone: 877-867-4008 Contact Person: Kate Forgione, Director of Quality Date Prepared: December 17, 2018
#### II. Name of Device
Name of Device: O2 Concepts Oxlife Independence Model 301-0001
Common or Usual Name: Portable Oxygen Concentrator
CFR Regulation Number: 21 CFR 868.5440
Classification Panel: Anesthesiology
Regulatory Class: II
Product Code: CAW
- III. Predicate Device: Sequal Eclipse Portable Oxygen Concentrator (K013931) This predicate has not been subject to a design related recall. No reference devices were used in this submission.
#### IV. Device Description
The Oxlife Independence is a portable oxygen concentrator that enables patients requiring supplemental oxygen to be treated in a home environment, in an institutional environment or in a vehicle/mobile environment. The Oxlife Independence utilizes O2 Concepts' Dynamic Network Analysis (DNA) technology to provide equipment performance, location and usage data to O2 Concepts and its customers via cellular and GPS connection.
The device delivers 87%-95% pure oxygen to a patient through a standard single lumen nasal cannula. The Independence detects a patient breath and delivers a bolus of oxygen during the inhalation period in pulse mode or delivers a continuous flow of oxygen in continuous mode.
{4}------------------------------------------------
The Oxlife Independence can be set to deliver flowrates in pulse mode settings of 0.5 to 6.0 of 8ml - 96ml and in continuous mode of 0.5 to 3.0 liters per minute (LPM). Setting can be adjusted in increments of 0.5 for both modes.
Device standard power options include a 100-240VAC Power Supply, 10-15VDC Power Supply and rechargeable batteries.
#### V. Indications for Use
The O2 Concepts Oxlife Independence is indicated on a prescription basis for the administration of supplemental oxygen. It is not intended for life support, nor does it provide any patient monitoring capabilities.
#### Patient Population
Adult patients with chronic pulmonary diseases such as chronic bronchitis, emphysema, asthma, or lung cancer, those in the terminal stages of cancer or any patient requiring supplemental oxygen. It is a prescription device designed for indoor and outdoor use.
#### Environment of Use
Home, Travel or Institution
#### VI. Comparison of Technological Characteristics with the Predicate Device
Pressure Swing Absorption (PSA) is the technological principle for both the subject and predicate devices. It is based on molecular sieve/pressure swing absorption technology which draws in ambient air, pushes it through a sieve bed and then utilizes pressure swing absorption to convert the ambient air to pure oxygen.
Table 1 compares the key features of the proposed O2 Concepts Oxlife Independence with the identified predicate, the Sequal Eclipse.
{5}------------------------------------------------
| Attribute | Predicate Device:<br>Sequal Eclipse K013931 | Proposed Device:<br>O2 Concepts Oxlife<br>Independence |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Image: Sequal Eclipse K013931 | Image: O2 Concepts Oxlife Independence |
| Indications for Use | The Eclipse Oxygen System is<br>intended for administration of<br>supplemental oxygen. The<br>Eclipse is not intended for life<br>supporting or life sustaining<br>applications nor does it provide<br>any patient monitoring<br>capabilities. | The O2 Concepts Oxlife<br>Independence is indicated on a<br>prescription basis for the<br>administration of supplemental<br>oxygen. It is not intended for<br>life support, nor does it provide<br>any patient monitoring<br>capabilities. |
| Environments of Use | Travel, Home or Institution | Travel, Home or Institution |
| Prescription Use | Yes | Yes |
| Patient Population | Adult | Adult |
| Single Patient<br>Multi Use | Yes | Yes |
| Patient Interface | Cannula port | Cannula port |
| Technology | Pressure Swing Adsorption<br>with molecular sieve | Pressure Swing Adsorption<br>with molecular sieve |
| Dimensions (HxWxD) | 19.3" x 12.3" x 7.1" | 20.29" x 10.85" x 9.45" |
| Weight | 17.9 lbs. | 16.7lbs. |
| Oxygen Concentration | 87-95.6% | 87-95% |
| Equivalent Flow Rates | Continuous 0.5 – 3.0LPM in<br>0.5 LPM increments<br>Pulse Dose Settings 1-6 | Continuous 0.5 – 3.0LPM in<br>0.5 LPM increments<br>Pulse Dose Settings 0.5-6 |
| Dose at Specified Flow | 8 mL | 8 mL |
| Filters | Air Inlet Filter | Air Intake Filter |
| User Interface | Buttons/LCD Display | Buttons/LCD Display |
| Electrical | 100-240 VAC, 50/60 Hz<br>11.5-16V DC | 100-240 VAC, 50/60 Hz<br>12-15V DC |
| Software | Embedded | Embedded |
| Acoustic Noise | 48 dBA | 56 dBA |
| Alarms | Unit Malfunction | Unauthorized battery |
| | Loss of Power | No external power |
| | | High external power |
| | | Low external power |
| | Low Battery | Low battery |
| | Low Oxygen Output | Low purity |
| | Flow Restricted | No Flow |
| | Power Cartridge Temperature | Over Temperature |
| | | Invalid Motor Temperature |
| | | Invalid Box Temperature |
| | No Inspiration in Pulse Mode | No Breath in Pulse Mode |
| Status Indicators | Battery/Power Condition | Battery/Power Condition |
| | Delivery Mode | Delivery Mode |
| | Flow Setting | Flow Setting |
| | Pulse Dose Breath Indicator | Pulse Dose Breath Indicator |
| | Alerts/Alarms | Alerts/Alarms |
| | | History Log |
| | | Diagnostics |
| Battery Duration | Approximately 2.4 hours at 2.0<br>Continuous | Approximately 2.5 hours at 2.0<br>Continuous |
| Operating | Temperature 50°F to 104°F<br>Humidity 10% to 95% @ 82.4°<br>F | Temperature 50°F to 104°F<br>Humidity 10% to 95% @ 82.4°<br>F |
| Shipping/Storage | Temperature -4°F to 140°F<br>Humidity Up to 95% non-<br>condensing | Temperature -4°F to 140°F<br>Humidity 0-95% non-<br>condensing |
| Electrical Safety | AAMI IEC 60601-1-2:2007 | IEC 60601-1<br>IEC 60601-1-2 |
| Mechanical Safety | AAMI ANSI ES60601-1:2005 | IEC 60601-1 |
| Chemical Safety | N/A | N/A |
| Thermal Safety | AAMI ANSI ES60601-1:2005 | IEC 60601-1 |
| Type of Protection<br>again electric shock | Type BF - Not for cardiac | Type BF - Not for cardiac |
| Biocompatibility | 4 VOC's less than ambient<br>Pass ISO 10993-5 | 4 VOC's less than ambient |
# TABLE 1 – COMPARISON OF PROPOSED DEVICE TO PREDICATE
{6}------------------------------------------------
{7}------------------------------------------------
| Standards Met | | |
|---------------|---------------------------------------------------------------------------------------------|---------------|
| | AAMI ANSI ES60601-<br>1:2005/(R)2012 and A1:2012,<br>C1:2009/(R)2012 and<br>A2:2010/(R)2012 | IEC 60601-1 |
| | ANSI AAMI IEC 60601-1-<br>2:2007/(R)2012 | IEC 60601-1-2 |
| | ANSI AAMI IEC 60601-1-<br>8:2006 and A1:2012 | |
| | ANSI AAMI HA60601-1-<br>11:2015 | |
| | ANSI AAMI IEC 62366-1:2015 | |
| | ISO 80601-2-69 1st ed. | |
| | ISO 8359:1996 | |
| | ISO 10993-2, 2006 | |
| | ISO 10993-5, 2014 | |
| | ISO 10993-10, 2010 | |
| | ISO 10993-12, 2012 | |
| | ISO 18562-2, 2017 | |
| | ISO 18562-3, 2017 | |
### Comparison Discussion
### Indications for Use
The indications for use are equivalent when compared to the predicate device.
### Patient Population
The patient populations are equivalent when compared to the predicate device. Adult patients with chronic pulmonary diseases such as chronic bronchitis, emphysema, asthma or lung cancer or those in the terminal stages of cancer, or any patient requiring supplemental oxygen.
### Environment of Use
The environment of use is equivalent when compared to the predicate device. Both devices are used in the home, for travel or in an institution.
{8}------------------------------------------------
### Principle of Operation & Technology
The essential design, components and principle operation of the Oxlife Independence based on molecular sieve/pressure swing absorption technology, to draw ambient air, push it through a sieve bed and then utilize pressure swing absorption to convert the ambient air to pure oxygen is equivalent to the predicate device.
The devices offer equivalent flow rates of oxygen and have equivalent alerts and alarms.
The difference between the device and the predicate is that the Oxlife Independence offers an optional technology, Dynamic Network Analysis (DNA), that providers can use to track device location, performance/usage data and alarms, via a cell and GPS module.
This technology does not change any performance characteristics of the device.
Both devices meet the requirements related to performance applicable to the portable oxygen concentrator and the differences noted in the comparison do not raise different questions of safety or effectiveness and therefore the subject device is substantially equivalent to the predicate device.
#### VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Guidance Document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" as recognized by FDA.
Particulate Matter 2.5 micron (PM2.5 testing was completed and the POC emitted a mean PM 2.5 level less than the EPA PM2.5 level of 15 µg/m3.
Inorganic gases (ozone. CO2, and CO) were evaluated with low levels detected of CO2; no level of CO and an ozone concentration less than the EPA allowed 0.050pm.
Patient contact with the device is limited to:
- Handle
- Device Enclosure
- Battery ●
- . Interface panel
{9}------------------------------------------------
The route of exposure to these materials would be through incidental contact with the patient's skin and does not contact the patient during normal device use. However, the oxygen that passes through the subject device per Annex A of ISO 10993-1:2009(E) is:
- . Category: External communicating device
- Contact: Tissue
- . Duration: C - Permanent (>30 days)
The POC is considered permanent contacting because the oxygen that passes through the device is inhaled by the patients through nasal tissue.
#### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing was conducted on the O2 Concepts Oxlife Independence. The system complies with the following:
AAMI ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 AAMI IEC 60601-1-2:2007/(R)2012 ANSI AAMI IEC 60601-1-8:2006 and A1:2012 ANSI AAMI HA60601-1-11:2015 ANSI AAMI IEC 62366-1:2015 ISO 80601-2-69 1ª ed. IEC 62304 Edition 1.1 2015-06
### Verification and Validation Testing
Verification and Validation testing was conducted in accordance with ISO 80601-2-69 1st edition and documented which included device software, acoustic testing and packaging testing. The software for this device was considered a "moderate" level of concern, since a failure or latent flaw in the software could result in Minor Injury, either to a patient or to a user of the device. Mechanical and electrical safety testing was performed according to the performance standards listed above and internal test protocols to identify test methods.
The O2 Concepts Oxlife Independence was evaluated in accordance with ISO 80601-2-69 1 st edition to assure it performs as intended and met user needs including the following:
Verification Testing - Testing of prototype devices to confirm that design outputs meet the design input requirements.
Validation Testing - Testing of production units to ensure that devices conform to user needs and intended uses.
Software Validation Testing - Testing of device software to ensure it meets the required specifications.
Usability Validation - Testing to ensure that users can be successfully operate the device using the supplied collateral.
{10}------------------------------------------------
# Usability Study
Usability testing was conducted with 10 users and included appropriate tasks based on the intended patient population in an environment representative of the intended conditions of use. Testing found that the design of the device and the instructions for use were appropriate for the intended user population.
### Animal Studies
Not applicable for this device
### Clinical Studies
No clinical studies were performed.
{11}------------------------------------------------
#### VIII. Conclusions
As detailed, the indications for use, patient population, environment of use, technology and principles of operation and performance are substantially equivalent to the predicate. The identified differences between the subject device and the predicate device conclude that there are no new risks and thus the subject device can be determined to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.