← Product Code [MBI](/productcode/MBI) · K182446

# TruShot with Y-Knot All-Suture Anchor (K182446)

_Conmed Corporation · MBI · Dec 19, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182446

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** Dec 19, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures. The system may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

## Device Story

TruShot™ with Y-Knot® All-Suture Anchor is a non-absorbable orthopedic fixation device. It consists of a high-strength flat-braided suture anchor preloaded on a disposable inserter with a drill bit and guide handle. Used in arthroscopic or open surgery by orthopedic surgeons to reattach soft tissue (ligaments, tendons, joint capsules) to bone. The device is inserted into a drilled pilot hole; the anchor expands to provide fixation. The surgeon uses the attached sutures to secure the soft tissue, stabilizing it during the healing period. The device is provided sterile for single-use, benefiting patients by providing a secure anchor for tissue repair.

## Clinical Evidence

Bench testing only. No clinical data presented. Verification testing included reliability, ultimate fixation strength, cyclic testing, sterilization, pyrogen testing, biocompatibility, shelf-life, post-aging functional testing, and MR safety. Validation testing included user validation, packaging, and transportation. Bacterial endotoxin testing met established limits.

## Technological Characteristics

Non-absorbable flat-braid suture anchor system. Components: All-Suture Anchor, disposable guide handle, #2 Hi-Fi® suture strands. Principle: Press-fit insertion into a drilled pilot hole with an expandable anchor design. Sterilization: Ethylene Oxide (EO). Single-use, disposable.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- ConMed Linvatec Y-Knot™ All-Suture Anchor ([K111779](/device/K111779.md))

## Reference Devices

- Y-Knot Flex All-Suture Anchor, w/Two #2 (5 Metric) Hi-Fi Sutures, 1.8mm ([K131035](/device/K131035.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

December 19, 2018

CONMED Corporation Diana L. Nader-Martone, M.S. Regulatory Affairs Specialist 525 French Road Utica, New York 13502

Re: K182446

Trade/Device Name: TruShot™ with Y-Knot® All-Suture Anchor Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: November 16, 2018 Received: November 20, 2018

Dear Ms. Nader-Martone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurence D. Coyne -S

For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

# Indications for Use

510(k) Number (if known)

## K182446

Device Name TruShot™ with Y-Knot® All-Suture Anchor

Indications for Use (Describe) Intended Use The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures.

Indications for Use

The device may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the suture anchor systems thereby stabilize the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white abstract design on the left. To the right of the square, the word "CONMED" is written in bold, black letters. Below "CONMED", the word "CORPORATION" is written in smaller, black letters.

## 510(k) SUMMARY

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K182446.

#### I. SUBMITTER

CONMED Corporation 11311 Concept Blvd Largo, Florida 33773

Phone: 727-399-5425 Fax: 727-399-5264

Contact Persons: Diana L. Nader-Martone or Kathy Reddig Date Prepared: September 6, 2018

#### II. DEVICE NAME

| Device Name:         | TruShot™ with Y-Knot® All-Suture Anchors       |
|----------------------|------------------------------------------------|
| Common Name:         | Nonabsorbable Suture Anchor System             |
| Classification Name: | Fastener, fixation, nondegradable, soft tissue |
| Regulatory Class:    | Class II, per 21 CFR Part 888. 3040            |
| Product Codes:       | MBI                                            |

#### III. PREDICATE/ LEGALLY MARKET DEVICE

| Device Name:  | ConMed Linvatec Y-Knot™ All-Suture Anchor |
|---------------|-------------------------------------------|
| Company Name: | ConMed Linvatec                           |
| 510(k) #:     | K111779                                   |

#### IV. REFERENCE DEVICE

| Device Name:   | Y-Knot Flex All-Suture Anchor, w/Two #2 (5 Metric) Hi-Fi |
|----------------|----------------------------------------------------------|
|                |                                                          |
| Sutures, 1.8mm |                                                          |
| Company Name:  | ConMed Linvatec                                          |
| 510(k) #:      | K131035                                                  |

#### V. DEVICE DESCRIPTION

The CONMED TruShot™ with Y-Knot® All-Suture Anchor is manufactured from High Strength Flat Braided Suture threaded with either one or two #2 (5 metric) Hi-Fi® suture strands with needles. Each All-Suture Anchor is preloaded on a disposable inserter, and pre-assembled with a disposable drill bit and guide handle. The anchor, suture(s) with

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white abstract shape inside, resembling a stylized "C". To the right of the square is the word "CONMED" in bold, black letters. Below "CONMED" is the word "CORPORATION" in smaller, lighter letters, underlined with a blue line.

needles, inserter, drill bit, and guide handle are pre-assembled and provided sterile, intended for single-use. The device is EO Sterilized.

#### VI. INTENDED USE/ INDICATIONS FOR USE

The non-absorbable suture anchors are intended to reattach soft tissue to bone in orthopedic surgical procedures.

The system may be used in either arthroscopic or open surgical procedures. After the suture is anchored to the bone, it may be used to reattach soft tissue, such as ligaments, tendons, or joint capsules to the bone. The suture anchor system thereby stabilizes the damaged soft tissue, in conjunction with appropriate postoperative immobilization, throughout the healing period.

### VII. COMPARISON OF THE TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The following table represents a summary of the technological characteristics between the proposed and the predicate device.

|                       | TruShotTM with Y-Knot® All-<br>Suture Anchor<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                            | ConMed Linvatec Y-Knot® All-<br>Suture Anchor<br>Predicate Device | Y-Knot Flex All-Suture Anchor,<br>w/Two #2 (5 Metric) Hi-Fi<br>Sutures, 1.8mm<br>Reference Device |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Device<br>Description | The non-absorbable suture anchors<br>are intended to reattach soft tissue<br>to bone in orthopedic surgical<br>procedures.                                                                                                                                                                                                                                                                                                 | Same                                                              | Same                                                                                              |
| Intended Use          | The non-absorbable suture anchors<br>are intended to reattach soft tissue<br>to bone in orthopedic surgical<br>procedures.                                                                                                                                                                                                                                                                                                 | Same                                                              | Same                                                                                              |
| Indication for Use    | The system may be used in either<br>arthroscopic or open surgical<br>procedures. After the suture is<br>anchored to the bone, it may be<br>used to reattach soft tissue, such as<br>ligaments, tendons, or joint<br>capsules to the bone. The suture<br>anchor system thereby stabilizes<br>the damaged soft tissue, in<br>conjunction with appropriate<br>postoperative immobilization,<br>throughout the healing period. | Same                                                              | Same                                                                                              |
| Contraindications     | 1. Pathological conditions of bone<br>which would adversely affect<br>the Y-Knot anchor.<br>2. Pathological conditions in the<br>soft tissue to be repaired or<br>reconstructed which would<br>adversely affect suture fixation.<br>3. Physical conditions that would<br>eliminate, or tend to eliminate,                                                                                                                  | Same                                                              | Same                                                                                              |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a blue square with a white, stylized "C" shape inside. To the right of the square, the word "CONMED" is written in a bold, black font. Below the word "CONMED", the word "CORPORATION" is written in a smaller, black font.

|                                  | TruShot™ with Y-Knot® All-<br>Suture Anchor<br>Proposed Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | ConMed Linvatec Y-Knot® All-<br>Suture Anchor<br>Predicate Device | Y-Knot Flex All-Suture Anchor,<br>w/Two #2 (5 Metric) Hi-Fi<br>Sutures, 1.8mm<br>Reference Device |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|                                  | adequate implant support or<br>retard healing.<br>4. Conditions which tend to limit<br>the patient's ability or<br>willingness to restrict activities<br>or follow directions during the<br>healing period.<br>5. Attachment of artificial<br>ligaments or other implants.<br>6. Foreign body sensitivity, known<br>or suspected allergies to implant<br>and/or instrument materials.<br>7. This device is not approved for<br>screw attachment or fixation to<br>the posterior elements (pedicles)<br>of the cervical, thoracic or<br>lumbar spine.<br>8. Patients with active sepsis or<br>infection. |                                                                   |                                                                                                   |
| Components                       | All-Suture Anchor<br>Disposable Guide Handle<br>Suture                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | All-Suture Anchor<br>Disposable Driver<br>Suture                  | All-Suture Anchor<br>Disposable Driver<br>Suture                                                  |
| Technological<br>Characteristics | Flat-braid suture anchor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Flat-braid suture anchor                                          | Flat-braid suture anchor                                                                          |
|                                  | One or two #2 sutures<br>Drilled pilot hole                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | One #2 suture<br>Drilled pilot hole                               | Two #2 sutures<br>Drilled pilot hole                                                              |
|                                  | Press-fit insertion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Press-fit insertion                                               | Press-fit insertion                                                                               |
|                                  | Expandable design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Expandable design                                                 | Expandable design                                                                                 |

#### VIII. PERFORMANCE DATA

Testing has been completed to demonstrate that the TruShot™ with Y-Knot® All-Suture Anchor performs as intended and is substantially equivalent to the predicate device. Bacterial endotoxin testing was conducted and met the endotoxin limits.

Completed testing includes the following:

Verification Testing

- Reliability ●
- Ultimate Fixation Strength ●
- Cyclic ●
- Sterilization ●
- Pyrogen
- Biocompatibility ●
- Shelf-life ●
- Post Aging Functional Testing
- MR Safety Testing ●

Validation Testing

- . User Validation
- Packaging .
- Transportation .

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a blue square with a white curved shape inside on the left. To the right of the square is the word "CONMED" in a bold, sans-serif font, with a horizontal line underneath. Below the line is the word "CORPORATION" in a smaller, sans-serif font.

#### IX. CONCLUSION

The TruShot™ with Y-Knot® All-Suture Anchor is either substantially equivalent or identical in design, materials, intended use, principles of operation, and technical characteristics to the predicate ConMed Linvatec Y-Knot™ All-Suture Anchor. Based upon the findings of our performance testing, the differences present no new issues of safety and efficacy, and the TruShot™ with Y-Knot® All-Suture Anchor is substantially equivalent to the ConMed Linvatec Y-Knot™ All-Suture Anchor (K111779).

---

**Source:** [https://fda-staging.innolitics.com/device/K182446](https://fda-staging.innolitics.com/device/K182446)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).