← Product Code [OMP](/productcode/OMP) · K182409

# Applied Tissue Technologies PWD™ Platform Wound Dressing (K182409)

_Applied Tissue Technologies, LLC · OMP · Jan 11, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182409

## Device Facts

- **Applicant:** Applied Tissue Technologies, LLC
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Jan 11, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Applied Tissue Technologies PWD™ Platform Wound Dressing is intended to be used in conjunction with the Invia Motion Negative Pressure Wound Therapy (NPWT) system and is indicated in patients who would benefit from a suction device (NPWT) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as: diabetic, pressure and venous insufficiencies), flaps and grafts.

## Device Story

Single-use wound dressing; used with Invia Motion NPWT system; creates environment promoting healing via exudate removal. Input: negative pressure from pump; Output: wound exudate collected in external canister. Used in hospital/homecare settings; applied by healthcare practitioners. Transparent polyurethane embossed drape allows wound observation without removal; access port connects to pump. Benefits: promotes healing; enables monitoring. Maximum 3-day wear time.

## Clinical Evidence

Bench testing only. Evaluated vapor transmission, negative pressure maintenance, and exudate removal capacity. Biocompatibility testing performed per ISO 10993 series. All results acceptable.

## Technological Characteristics

Single-use sterile dressing; clear polyurethane embossed occlusive drape with tubing and access port. Materials: polymers and adhesive. Sterilization: Ethylene Oxide (EtO). Connectivity: connects to external NPWT pump system. No electronics or software.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Medela Ag Invia Foam Dressing Kit ([K170088](/device/K170088.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2019

Applied Tissue Technologies LLC % Michele Lucey Regulatory Affairs Manager Lakeshore Medical Device Consulting LLC 128 Blye Hill Landing Newbury, New Hampshire 03255

Re: K182409

Trade/Device Name: Applied Tissue Technologies PWD™ Platform Wound Dressing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: December 7, 2018 Received: December 12, 2018

Dear Michele Lucey:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Kimberly Ferlin -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182409

#### Device Name

Applied Tissue Technologies PWD™ Platform Wound Dressing

#### Indications for Use (Describe)

The Applied Tissue Technologies PWD™ Platform Wound Dressing is intended to be used in conjunction with the Invia Motion Negative Pressure Wound Therapy (NPWT) system and is indicated in patients who would benefit from a suction device (NPWT) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as: diabetic, pressure and venous insufficiencies), flaps and grafts.

| Type of Use (Select one or both, as applicable)                                |                                                                               |
|--------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D)       </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)       </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY

#### Submitter Information

| Submitters' Name:     | Applied Tissue Technologies LLC                             |
|-----------------------|-------------------------------------------------------------|
| Address:              | 99 Derby Street, Suite 200<br>Hingham, MA 02043             |
| Telephone:            | 781-366-3848                                                |
| Contact Person:       | Michele Lucey                                               |
| Telephone:            | 603-748-1374                                                |
| Date Prepared:        | January 11, 2019                                            |
| Device Trade Name:    | Applied Tissue Technologies<br>PWD™ Platform Wound Dressing |
| Classification:       | Class II                                                    |
| Product Code(s):      | OMP                                                         |
| Regulation Number(s): | 21 CFR 878-4780 Powered Suction Pump                        |
| Predicate Device:     | Medela Ag Invia Foam Dressing Kit,<br>K170088               |

## Intended Use:

The Applied Tissue Technologies PWD™ Platform Wound Dressing is intended to be used in conjunction with the Invia Motion Negative Pressure Wound Therapy (NPWT) system and is indicated in patients who would benefit from a suction device (NPWT) as it creates an environment that may promote wound healing by removing excess wound exudate. The Applied Tissue Technologies Platform Wound Dressing is appropriate for use on the following wounds: exudating, chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as: diabetic, pressure and venous insufficiencies), flaps and grafts.

#### Device Description/Technological Characteristics:

The Applied Tissue Technologies PWD™ Platform Wound Dressing is a single use device. Its transparency permits the healthcare practitioner to observe the progress of the wound healing without removing the wound dressing. There is an access port to administer negative pressure wound therapy. The PWD is intended to be used for a maximum of 3 days. Therapy duration may be less than indicated if clinical practice or other factors require more frequent dressing changes.

{4}------------------------------------------------

The PWD will be marketed in four different shapes and sizes, all having the same intended use. The four sizes are identified as follows.

| PWDTM Platform Wound Dressing |               |
|-------------------------------|---------------|
| Description                   | Catalog/REF # |
| 1" Round                      | AT1070-01     |
| 2" Round                      | AT1071-01     |
| 3" Round                      | AT1072-01     |
| 1" x 3" Oblong                | AT1073-01     |

## Performance Data:

Bench testing of the PWD™ Platform Wound Dressing was performed to evaluate device vapor transmission characteristics, ability to maintain negative pressure, and the ability to remove exudate. Biocompatibility testing was conducted according to the ISO 10993 standard series. Test results of all nonclinical testing were acceptable and demonstrate that the subject device is substantially equivalent to the predicate device

#### Substantial Equivalence:

The following chart provides the evidence to support the claim for substantial equivalence:

| Substantial Equivalence Comparison Chart                                    |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                           |
|-----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature / Specification                                                     | PWDTM<br>Platform Wound<br>Dressing                                                                                                                                                                                                                 | Invia Foam Dressing Kit<br>(Medela Ag,)                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                                |
| Regulatory Clearance/<br>Approval Reference                                 | K182409                                                                                                                                                                                                                                             | K170088                                                                                                                                                                                                                                                            | N/A                                                                                                                                                                                                                       |
| Product Code                                                                | OMP                                                                                                                                                                                                                                                 | OMP                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                      |
| Regulation Number                                                           | 21 CFR 878.4780                                                                                                                                                                                                                                     | 21 CFR 878.4780                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                      |
| Regulation Name                                                             | Powered Suction Pump<br>(accessory)                                                                                                                                                                                                                 | Powered Suction Pump<br>(accessory)                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                      |
| Where used<br>(environment)                                                 | The Negative Pressure<br>Wound Therapy System<br>is suitable for use in both<br>hospital and homecare<br>settings.                                                                                                                                  | The Negative Pressure<br>Wound Therapy System is<br>suitable for use in<br>both hospital and<br>homecare settings.                                                                                                                                                 | Same                                                                                                                                                                                                                      |
| Maximum Duration of<br>a single dressing                                    | 72 Hours                                                                                                                                                                                                                                            | 72 Hours                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                      |
| Negative Pressure to<br>the wound surface<br>requirements in mmHg           | 80 mmHg (nominal)<br>+/- 20mmHG to the<br>wound surface                                                                                                                                                                                             | 125 mmHg to the wound<br>surface                                                                                                                                                                                                                                   | This constitutes a<br>nominal<br>recommended<br>setting, pump<br>settings range<br>from -40 to -175.<br>This difference<br>in recommended<br>nominal setting.                                                             |
| Substantial Equivalence Comparison Chart                                    |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                           |
| Feature / Specification                                                     | PWDTM<br>Platform Wound<br>Dressing                                                                                                                                                                                                                 | Invia Foam Dressing Kit<br>(Medela Ag,)                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                                |
|                                                                             |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                    | does not range<br>new questions of<br>safety.                                                                                                                                                                             |
| Multiple Shapes for<br>Wound Sizes up to<br>400cc (Surface Area x<br>depth) | 400cc                                                                                                                                                                                                                                               | 240 - 2700c                                                                                                                                                                                                                                                        | PWD is within<br>the range of the<br>predicate<br>device.                                                                                                                                                                 |
| Dressing<br>Configuration                                                   | Clear polyurethane<br>embossed occlusive drape<br>with tubing                                                                                                                                                                                       | Clear polyurethane<br>occlusive drape dressing<br>with foam insert negative<br>pressure/instillation tubing –<br>dressing is cut to size                                                                                                                           | Similar<br>materials and<br>configuration,<br>the PWD uses a<br>embossed layer<br>to allow for<br>fluid transfer,<br>the predicate<br>uses a<br>foam. This<br>difference does<br>not raise new<br>questions of<br>safety. |
| Access Port                                                                 | For attachment to the<br>negative pressure pump.<br>For removal of fluids.                                                                                                                                                                          | For attachment to the<br>negative pressure pump.<br>For removal of fluids                                                                                                                                                                                          | Same.                                                                                                                                                                                                                     |
| Collection Pack                                                             | The PWD tubing is<br>connected to the Medela<br>Invia Motion<br>canister/tubing set during<br>negative pressure wound<br>therapy for collection of<br>fluids. The fluids are<br>removed from the wound<br>site and secured within<br>the cannister. | The Invia Foam Dressing<br>Tubing is connected to the<br>Medela Invia Motion<br>canister/tubing set during<br>negative pressure wound<br>therapy for collection of<br>fluids. The fluids are<br>removed from the wound<br>site and secured within the<br>cannister | Same.                                                                                                                                                                                                                     |
| Pump                                                                        | To be used with the<br>Medela Invia®<br>Motion™ Negative<br>Pressure Wound<br>Therapy System                                                                                                                                                        | To be used with the Medela<br>Invia® Motion™ Negative<br>Pressure Wound Therapy<br>System                                                                                                                                                                          | Same                                                                                                                                                                                                                      |
| How Supplied                                                                | PWD is Sterile Single<br>Use Only                                                                                                                                                                                                                   | Dressing is Sterile<br>Single Use Only                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                      |
| Substantial Equivalence Comparison Chart                                    |                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                           |
| Feature / Specification                                                     | PWDTM<br>Platform Wound<br>Dressing                                                                                                                                                                                                                 | Invia Foam Dressing Kit<br>(Medela Ag,)                                                                                                                                                                                                                            | Comparison                                                                                                                                                                                                                |
| Tissue Contact<br>Materials                                                 | Polymers and Adhesive                                                                                                                                                                                                                               | Polymers Adhesive and<br>dressing                                                                                                                                                                                                                                  | Similar,<br>differences do<br>not introduce<br>new questions of<br>safety                                                                                                                                                 |
| Additives - antimicrobial,<br>animal origin                                 | No                                                                                                                                                                                                                                                  | No                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                      |
| Sterilization                                                               | Eto                                                                                                                                                                                                                                                 | Eto                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                      |

{5}------------------------------------------------

{6}------------------------------------------------

The PWD™ Platform Wound Dressing is substantially equivalent to the predicate device, the Invia Foam Dressing Kit (K170088) and does not raise different questions of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K182409](https://fda-staging.innolitics.com/device/K182409)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
