← Product Code [GKZ](/productcode/GKZ) · K182389

# Sysmex XN-L Automated Hematology Analyzer (K182389)

_Sysmex America, Inc. · GKZ · Jan 25, 2019 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182389

## Device Facts

- **Applicant:** Sysmex America, Inc.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** Jan 25, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology
- **Attributes:** Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K182389 · Jan 25, 2019 | Sysmex XN-L Automated Hematology Analyzer | Sysmex America, Inc. | Clinical samples from pediatric patients (less than 2 years of age) | Real-world data was collected to support the expansion of the intended use population to include pediatric patients under the age of 2 years. | Pediatric population; Age limitation removal; Clinical samples |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Retrospective/Routine clinical data collection | Pediatric patients less than 2 years of age | Not applicable for this study | Performance characteristics for pediatric population |

## Indications for Use

The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in venous and capillary whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal, peritoneal, pleural, and synovial fluids. Whole blood should be collected in K2 or K3EDTA anticoagulant and peritoneal, pleural, and synovial fluids in K2EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

## Device Story

Quantitative multi-parameter automated hematology analyzer; processes venous/capillary whole blood and body fluids (CSF, peritoneal, pleural, synovial). Uses electrical impedance, laser light scattering, and fluorescent labeling; performs DC sheath flow detection, flow cytometry (semiconductor laser), and SLS-hemoglobin method. Main unit aspirates, dilutes, mixes, and analyzes samples; pneumatic unit provides pressure/vacuum. External touchscreen monitor used for operation and data processing. Used in clinical laboratories by trained personnel. Automatically classifies cells from aspiration to result output. Provides clinicians with hematological parameters to assist in patient screening and clinical decision-making.

## Clinical Evidence

Bench testing only. Performance evaluated via accuracy, precision (repeatability/reproducibility), linearity, carryover, stability, and limits of blank/detection/quantitation for both whole blood and body fluid modes. Results met predetermined acceptance criteria, demonstrating equivalence to predicate XN-Series analyzers. Real-world data included to support use in pediatric patients under 2 years.

## Technological Characteristics

Automated multi-parameter hematology analyzer. Methods: DC Sheath Flow Detection, semiconductor laser flow cytometry, SLS hemoglobin method. Materials: K2/K3EDTA anticoagulated samples. Connectivity: Standalone, non-wireless data transmission. Software: Rule-based rerun/reflex. Sterilization: N/A (IVD instrument). Compliance: IEC 60825-1 (laser safety), IEC 61010-1/2-081/2-101 (electrical/IVD safety), IEC 61326-2-6 (EMC).

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzer ([K112605](/device/K112605.md))

## Reference Devices

- XN CAL ([K160585](/device/K160585.md))
- XN-L CHECK ([K160586](/device/K160586.md))
- XN CHECK ([K160590](/device/K160590.md))
- XN CHECK BF ([K160588](/device/K160588.md))
- Whole Blood Linearity Range Check X III ([K960557](/device/K960557.md))
- Retic Chex ([K000115](/device/K000115.md))
- XN CAL PF ([K120747](/device/K120747.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM

A. 510(k) Number:
K182389

B. Purpose for Submission:
Expand Intended Use to include pediatric subjects under the age of 2 years old.

C. Manufacturer and Instrument Name:
Sysmex America Inc., Sysmex® XN-L Automated Hematology Analyzer

D. Type of Test or Tests Performed:
The Sysmex XN-L Automated Hematology Analyzer (hereafter, the XN-L analyzer) classifies and enumerates the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal (CSF), peritoneal, pleural and synovial fluids.

E. System Descriptions:

1. Device Description:
The XN-L analyzer is a quantitative multi-parameter automated differential cell counter that classifies and enumerates whole blood and body fluid parameters by means of electrical impedance, laser light scattering, and fluorescent labeling. Cell counts and parameters are performed on whole blood samples collected in K₂EDTA or K₃EDTA anticoagulant, body fluids (peritoneal, pleural and synovial) collected in K₂EDTA anticoagulant and CSF collected without anticoagulant. The instrument consists of two principal units: (1) the Main Unit which will aspirate, dilute, mix, and analyze whole blood and body fluid samples and (2) the Pneumatic Unit which supplies pressure and vacuum to the analyzer.

The XN-L analyzer has an external monitor with touch screen capability that is used to operate the instrument and process data from the Main Unit. The monitor also allows for operator interfacing with the instrument by use of a panel keyboard.

2. Principles of Operation:
The XN-L analyzer analyzes samples using the following methods: DC Sheath Flow Detection method, Flow Cytometry method using a semiconductor laser, and SLS (cyanide-free sodium lauryl sulfate) hemoglobin method. Particle characterization and identification is based on detection of forward scatter, fluorescence, and adaptive cluster analysis. The XN-L analyzer automatically classifies cells from whole blood and body

{1}

fluids and carries out all processes automatically from aspiration of the sample to result output.

3. **Modes of Operation:**
- Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
- Yes ☐ X or No ☐
- Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
- Yes ☐ or No ☐ X

4. **Specimen Identification:**
Specimen identification input is manual (by operator) or by barcode reader.

5. **Specimen Sampling and Handling:**
There are two modes of sample introduction: (1) Sampler Mode; (2) Manual Mode. In the Sampler Mode the operator loads the sample tubes into a rack, which is then automatically transported and analyzed by the instrument. This mode automatically mixes, aspirates, and analyzes samples without removing their caps. The Sampler Mode is used for processing of whole blood samples. In the Manual Mode, there are two sample tube holders: (1) Normal sample tube holder; (2) Micro collection tube holder. In this mode the operator loads and mixes the samples tubes individually by hand. The samples in the Manual Mode can be analyzed with the cap on or off. The Manual Mode is used for processing whole blood and body fluid samples.

6. **Calibration:**
The XN CAL calibrator (K160585) is used for calibration of the WBC, RBC, HGB, HCT, PLT and RET parameters. XN CAL is used for the calibration and calibration verification of Sysmex XN series (XN-10, XN-11, XN-20, XN-21, XN-L) analyzers. Calibration is performed as needed (e.g., when QC data is fluctuating) to ensure accuracy of the system.

7. **Quality Control:**
The XN-L CHECK (K160586) is used as quality control (three levels) for Sysmex XN-L analyzers. XN-L CHECK™ is an in-vitro diagnostic product that contains the following: stabilized red blood cell component(s), stabilized white blood cell component(s), and stabilized platelet component(s) in a preservative medium.

XN CHECK (K160590) is used as quality control (three levels) for Sysmex XN series (XN-10, XN-11, XN-20, XN-21, XN-L) analyzers. XN CHECK is an in-vitro diagnostic product that contains the following: stabilized red blood cell component(s), stabilized white blood cell component(s), stabilized platelet component(s), and stabilized nucleated red blood cell component(s) in a preservative medium.

{2}

XN CHECK BF (K160588) is used as quality control (two levels) for Sysmex XN series (XN-10, XN-11, XN-20, XN-21, XN-L) analyzers. Assayed parameters include: WBC-BF, RBC-BF, MN%, PMN%, TC-BF#.

8. Software:
FDA has reviewed applicant’s Hazard Analysis and Software Development processes for this line of product types:
Yes ☐ X ☐ or No ☐

F. Regulatory Information:

1. Regulation section:
21 CFR 864.5220, Automated differential cell counter

2. Classification:
Class II

3. Product code:
GKZ, Counter, Differential Cell

4. Panel:
Hematology (81)

G. Intended Use:

1. Indication(s) for Use:
The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in venous and capillary whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal, peritoneal, pleural, and synovial fluids. Whole blood should be collected in K2 or K3EDTA anticoagulant and peritoneal, pleural, and synovial fluids in K₂EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.

2. Special Conditions for Use Statement(s):
For prescription use only.

H. Substantial Equivalence Information:

1. Predicate Device Name(s) and 510(k) numbers:
Sysmex XN-Series (XN-10, XN-20) Automated Hematology Analyzer, K112605

{3}

2. Comparison with Predicate Device:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Candidate
Sysmex XN-L analyzer
K182389 | Predicate
Sysmex XN-Series (XN-10)^{a}
K112605  |
|  Intended Use | The Sysmex XN-L analyzer is a quantitative multi-parameter automated hematology analyzer intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-L analyzer classifies and enumerates the following parameters in venous and capillary whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBC-BF, RBC-BF, MN%/#, PMN%/#, and TC-BF# parameters in cerebrospinal, peritoneal, pleural, and synovial fluids. Whole blood should be collected in K_{2} or K_{3}EDTA anticoagulant and peritoneal, pleural, and synovial fluids in K_{2}EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended. | The XN-Series modules (XN-10, XN-20) are quantitative multi-parameter automated hematology analyzers intended for in vitro diagnostic use in screening patient populations found in clinical laboratories. The XN-Series modules classify and enumerate the following parameters in whole blood: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, EO%/#, BASO%/#, IG%/#, RDW-CV, RDW-SD, MPV, NRBC%/#, RET%/#, IPF, IRF, RET-He and has a Body Fluid mode for body fluids. The Body Fluid mode enumerates the WBCBF, RBC-BF, MN%/#, PMN%/# and TC-BF parameters in cerebrospinal fluid (CSF), serous fluids (peritoneal, pleural) and synovial fluids. Whole blood should be collected in K_{2} or K_{3}EDTA anticoagulant and, Serous and Synovial fluids in K_{2}EDTA anticoagulant to prevent clotting of fluid. The use of anticoagulants with CSF specimens is neither required nor recommended.  |
|  Specimen Type | Whole Blood and Body Fluids (CSF and peritoneal, pleural, synovial fluids) | Same  |
|  Test Principle | Hydro Dynamic Focusing (DC Detection), flow cytometry method using a semiconductor laser and SLS hemoglobin method. | Same  |
|  Parameters | Whole Blood Mode: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, PLT, NEUT%/#, LYMPH%/#, MONO%/#, | Same  |

4

{4}

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Candidate
Sysmex XN-L analyzer
K182389 | Predicate
Sysmex XN-Series (XN-10)^{a}
K112605  |
|   | EO%/#, BASO%/#, RDW-CV, RDW-SD, MPV, RET%/#, IRF, IG%/#, RET-He# Body Fluid
Mode: WBC-BF, RBC-BF, MN%/#, PMN%/#,TC-BF# |   |
|  Reagents | CELLPACK DCL (Diluent)
CELLPACK DFL (Diluent)
Lysercell WDF (Lyse)
Fluorocell WDF (Stain)
Fluorocell RET (Stain)
SULFOLYSER (Lyse) | Same  |
|  Analysis Modes | Sampler Analysis Mode
(rack autoloader)
Whole Blood Mode

Manual Analysis Mode
Whole Blood Mode;
LWBC Analysis Mode;
Pre-Dilute Analysis Mode;
Body Fluid Mode | Same  |
|  Sample Aspiration/
Fluidic Pathway | Single Pathway | Same  |
|  Measuring Channels | RBC/PLT, HGB, RET, WDF | Same  |
|  Controls/Calibrators/
Linearity Material | Whole Blood
XN CHECK 3 Levels
(K160590);
XN CAL (K160585);
Body Fluid
XN CHECK BF 2 Levels
(K160588)

Whole Blood Linearity
Range Check X III (K960557);
Retic Chex (K000115) | Same  |
|  Cleaning Detergent | CELLCLEAN AUTO | Same  |
|  Software/ Hardware | Rule based rerun/reflex | Same  |

{5}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Candidate
Sysmex XN-L analyzer
K182389 | Predicate
Sysmex XN-Series (XN-10)^{a}
K112605  |
|  Parameters | Not Available | PLT (PLT-F), NRBC%/#, IPF  |
|  Reagents | Not Available | K112605
Lysercell WNR (Lyse)
Fluorocell WNR (Stain)
Fluorocell PLT (Stain)  |
|  Measuring Channels | Not Available | WNR, PLT-F  |
|  Controls/Calibrators | Not Available
XN-L CHECK^{b} | XN CAL PF – (K120747)
Not Available  |
|  Throughput | Whole Blood Mode
60 samples/hour maximum
depending on mode used.
Body Fluid Mode
30 samples/hour maximum | Whole Blood Mode
100 samples/hour maximum
depending on mode used.
Body Fluid Mode
40 samples/hour maximum  |
|  Sample Aspiration
Volumes | Sampler Mode - 25 μL
Manual (Closed Cap) Mode - 25 μL
Manual (Open Cap) Mode - 25 μL
Dilution Mode - 70 μL
Body Fluid Mode - 70 μL | Sampler Mode - 88 μL
Manual (Closed Cap) Mode - 88 μL
Manual (Open Cap) Mode - 88 μL
Dilution Mode - 70 μL
Body Fluid Mode - 88 μL  |

a Intended use for the predicate analyzer was cleared in submission K112605. All information listed for the predicate analyzer refers to the XN-10 module.<br>
b Control material specific for the XN-L analyzer.

# I. Special Control/Guidance Document Referenced (if applicable):

- CLSI C28-A3c Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline - Third Edition
- CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: Statistical Approach; Approved Guideline
- CLSI EP12-A2, User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second Edition
- CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI H20-A2 Reference Leukocyte (WBC) Differential Count (Proportional) And Evaluation Of Instrumental Methods; Approved Standard-Second Edition
- CLSI H26-A2, Validation, Verification, and Quality Assurance of Automated Hematology Analyzer; Approved Guideline - Second Edition
- IEC 60825-1:2007 Safety of laser products – Part 1: Equipment classification and requirements
- IEC 61010-1:2001 Safety requirements for electrical equipment for measurement, control

{6}

and laboratory use – Part 1: General requirements

- IEC 61010-2-081:2001+A1 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-081: Particular requirements for automatic and semi-automatic laboratory equipment for analysis and other purposes
- IEC 61010-2-101:2002 Safety requirements for electrical equipment for measurement, control and laboratory use – Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
- IEC 61326-2-6:2005 Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 2-6: Particular requirements – In vitro diagnostic (IVD) medical equipment

## J. Performance Characteristics:

### 1. Analytical Performance:

#### a. Method comparison:

i. Method comparison - &gt; 2 years of age: Refer to K160538

ii. Method comparison - ≤ 2 years of age:

Whole Blood Studies (excluding reticulocyte parameters - RET, IRF, Ret-He): A total of 52 pediatric samples under the age of 2 years old were collected across three clinical pediatric sites. The residual specimens were collected in K₂EDTA according to the manufacturer's recommendations. All samples were tested within 8 hours of collection. Samples were analyzed on both the XN-10 analyzer and the XN-L in singlet by the labeled procedures within 2 hours of each other.

Deming regression analyses were used to estimate the parameters of the regression model (slope, intercept, 95% confidence intervals (CI) and correlation coefficient). The 95% CI and estimates of the bias/difference were determined for each parameter. All results were within the pre-defined acceptance criteria.

All sites combined:

|  Parameter (unit) | Result Range | Correlation Coefficient | Slope (95% CI) | Intercept (95% CI)  |
| --- | --- | --- | --- | --- |
|  WBC (10³/μL) | 2.64 – 23.85 | 0.9982 | 0.925 (0.909, 0.941) | 0.199 (0.017, 0.380)  |
|  RBC (10⁶/μL) | 2.74 – 5.78 | 0.994 | 1.015 (0.984, 1.047) | -0.064 (-0.200, 0.071)  |
|  HGB (g/dL) | 7.1 – 15.9 | 0.9981 | 1.032 (1.014, 1.050) | -0.11 (-0.32, 0.10)  |
|  HCT | 22.0 – 45.8 | 0.9833 | 0.975 | 0.92  |

{7}

|  Parameter (unit) | Result Range | Correlation Coefficient | Slope (95% CI) | Intercept (95% CI)  |
| --- | --- | --- | --- | --- |
|  (%) |  |  | (0.925, 1.026) | (-0.88, 2.73)  |
|  MCV (fL) | 72.8 – 103.1 | 0.922 | 1.041
(0.924, 1.157) | -3.2
(-12.91, 6.50)  |
|  MCH (pg) | 23.6 – 33.0 | 0.968 | 1.003
(0.931, 1.075) | 0.53
(-1.41, 2.47)  |
|  MCHC (g/dL) | 29.5 – 34.7 | 0.7726 | 0.878
(0.707, 1.049) | 4.6
(-0.93, 10.13)  |
|  PLT (10³/μL) | 135 – 888 | 0.9924 | 1.119
(1.080, 1.158) | -19.6
(-35.4, -3.8)  |
|  RDW-SD (fL) | 34.5 – 58.2 | 0.956 | 0.992
(0.909, 1.076) | 1.57
(-1.88, 5.02)  |
|  RDW-CV (%) | 11.7 – 17.4 | 0.983 | 0.996
(0.944, 1.048) | 0.29
(-0.42, 1.00)  |
|  MPV (fL) | 8.1 – 12.5 | 0.934 | 1.019
(0.914, 1.124) | 0.1
(-0.94, 1.14)  |
|  NEUT# (10³/μL) | 0.04 – 19.46 | 0.9992 | 0.93
(0.920, 0.941) | 0.036
(-0.027, 0.099)  |
|  LYMPH# (10³/μL) | 0.79 – 12.93 | 0.9966 | 0.93
(0.908, 0.952) | 0.039
(-0.079, 0.156)  |
|  MONO# (10³/μL) | 0.21 – 3.16 | 0.9729 | 0.933
(0.871, 0.995) | 0.05
(-0.017, 0.117)  |
|  EO# (10³/μL) | 0.00 – 2.24 | 0.9977 | 0.88
(0.863, 0.897) | 0.007
(0.000, 0.013)  |
|  BASO# (10³/μL) | 0.00 – 0.16 | 0.6237 | 1.348
(1.031, 1.664) | 0.011
(-0.004, 0.026)  |
|  IG# (10³/μL) | 0.00 – 0.20 | 0.8999 | 0.738
(0.643, 0.834) | -0.002
(-0.008, 0.003)  |
|  NEUT% (%) | 0.7 – 81.5 | 0.9994 | 0.996
(0.986, 1.006) | -0.05
(-0.51, 0.41)  |
|  LYMPH% (%) | 10.9 – 87.8 | 0.9973 | 0.977
(0.957, 0.997) | 0.65
(-0.43, 1.72)  |
|  MONO% (%) | 4.1 – 17.4 | 0.9236 | 1.013
(0.900, 1.125) | 0.31
(-0.78, 1.41)  |
|  EO% (%) | 0.0 – 11.6 | 0.9959 | 0.964
(0.939, 0.988) | 0.01
(-0.06, 0.08)  |
|  BASO% (%) | 0.0 – 1.2 | 0.3733 | 2.258
(1.643, 2.873) | -0.15
(-0.42, 0.12)  |
|  IG % (%) | 0.0 – 1.2 | 0.8464 | 1.011
(0.852, 1.171) | -0.09
(-0.16, -0.03)  |

Reticulocyte parameters (RET, IRF, Ret-He):
A total of 37 pediatric samples (1 month – 2.7 years) for the reticulocyte parameters were collected at one clinical site. The residual de-identified specimens were collected in K₂EDTA following manufacturer’s recommendations. All samples

{8}

were tested within 8 hours of collection. Samples were analyzed on both the XN-10 analyzer and the XN-L in singlet by the labeled procedures within 2 hours of each other.

Deming regression analyses were used to estimate the parameters of the regression model (slope, intercept, 95% confidence intervals (CI) and correlation coefficient). The 95% CI and estimates of the bias/difference were determined for each parameter. All results were within the pre-defined acceptance criteria.

All sites combined:

|  Parameter | Result Range | Correlation Coefficient | Slope (95% CI) | Intercept (95% CI)  |
| --- | --- | --- | --- | --- |
|  RET (10³/μL) | 0.0040 – 0.2154 | 0.9062 | 0.977 (0.831, 1.122) | -0.009 (-0.01994, 0.00126)  |
|  RET (%) | 0.16 – 6.45 | 0.9060 | 0.946 (0.805, 1.087) | -0.206 (-0.487, 0.076)  |
|  IRF (%) | 1.5 – 42.9 | 0.9041 | 0.971 (0.824, 1.117) | -0.200 (-2.78, 2.38)  |
|  RET-He (pg) | 22.6 – 35.5 | 0.9552 | 1.045 (0.938, 1.152) | -4.100 (-7.26, -0.94)  |

b. Precision/Reproducibility:
Refer to K160538

c. Linearity:
Refer to K160538

d. Carryover:
Refer to K160538

e. Interfering Substances:
Refer to K160538

2. Other Supportive Instrument Performance Data Not Covered Above:

a. Sample Stability
Refer to K160538

b. Verification of Reference Intervals

i. Body Fluids
Refer to K160538

ii. Reference range - pediatrics &gt; 2 years old:
Refer to K160538

{9}

iii. Reference range - pediatric birth to ≤2 years

Two peer-reviewed references were cited to substantiate the reference intervals for pediatrics.

Reference Ranges for WBC, RBC, HGB, HCT, MCV, MCH, PLT, RDW-SD, RDW-CV, MPV, NEUT#, NEUT%, LYMPH#, LMPH%, MONO#, MONO%, EO#, EO%, BASO%, RET#, RET%, IRF%, IG#:

Soldin, S.J., Brugnara, C., and Wong, E.C. 2005. Pediatric Reference Intervals, Fifth Edition, AACC Press, Washington, DC.

Reference Ranges for BASO#, IG#, IG%, MCHC:

Soldin, S. J., Brugnara, C., and Wong, E.C. 2007. Pediatric Reference Intervals, Sixth Edition, AACC Press, Washington, DC. c. Matrix Studies

|  Measurand | Units | Females | Males | Combined^{1} RI Male/Females  |
| --- | --- | --- | --- | --- |
|  WBC | x 10^{3}/μL | 5.9 – 15.8 | 6.5 – 16.7 | 5.9 – 16.7  |
|  RBC | x 10^{6}/μL | 3.55 – 4.83 | 3.24 – 5.08 | 3.24 – 5.08  |
|  HGB | g/dL | 10.7 – 16.4 | 10.2 – 16.6 | 10.2 – 16.6  |
|  HCT | % | 30.5 – 47.7 | 29.1 – 47.4 | 29.1 – 47.7  |
|  MCV | fL | 76.6 – 105.4 | 75.6 – 106.3 | 75.6 – 106.3  |
|  MCH | pg | 26.5 – 36.3 | 26.0 – 36.4 | 26.0 – 36.4  |
|  MCHC | g/dL | 33.7 – 35.7 | 33.6 – 35.7 | 33.6 – 35.7  |
|  PLT | x 10^{3}/μL | 95.0 – 430 | 120 – 471 | 95.0 – 471  |
|  RDW-SD | fL | 34.9 – 65.7 | 35.3 – 61.7 | 34.9 – 65.7  |
|  RDW-CV | % | 13.3 – 17.8 | 13.5 – 18.2 | 13.3 – 18.2  |
|  MPV | fL | 7.3 – 9.9 | 7.3 – 9.3 | 7.3 – 9.9  |
|  NEUT | x 10^{3}/μL | 2.2 – 11.4 | 2.2 – 9.4 | 2.2 – 11.4  |
|  NEUT | % | 15.7 – 69.3 | 14.6 – 69.2 | 14.6 – 69.3  |
|  LYMPH | x 10^{3}/μL | 1.2 – 5.7 | 1.4 – 5.6 | 1.2 – 5.7  |
|  LYMPH | % | 8.0 – 70.0 | 9.0 – 68.0 | 8.0 – 70.0  |
|  MONO | x 10^{3}/μL | 0.1 – 5.0 | 0.2 – 3.5 | 0.1 – 5.0  |
|  MONO | % | 4.0 – 19.0 | 4.0 – 18.0 | 4.0 – 19.0  |
|  EO | x 10^{3}/μL | 0.0 – 0.4 | 0.0 – 0.5 | 0.0 – 0.5  |
|  EO | % | 1.0 – 6.0 | 1.0 – 7.0 | 1.0 – 7.0  |
|  BASO | x 10^{3}/μL | 0.0 – 0.1 | 0.0 – 0.1 | 0.0 – 0.1  |
|  BASO | % | 0.0 – 1.0 | 0.0 – 1.0 | 0.0 – 1.0  |
|  RET | % | 0.4 – 3.7 | 0.4 – 4.8 | 0.4 – 4.8  |
|  RET | x 10^{3}/μL | 35.0 – 120.0 | 29.0 – 104.0 | 29.0 – 120.0  |
|  RET-He^{2} | pg | 23.9 – 30.9 | 22.5 – 31.8 | 22.5 – 31.8  |
|  IRF | % | 11.4 – 35.1 | 11.4 – 35.1 | 11.4 – 35.1  |
|  IG | % | 0.0 – 1.7 | 0.0 – 1.7 | 0.0 – 1.7  |
|  IG | x 10^{3}/μL | 0.00 – 0.28 | 0.00 – 0.28 | 0.0 – 0.28  |

1 Combined Reference Intervals (RI) - The lowest and highest value of the above female and male ranges were used to define the lower and upper range for the combined RI for pediatric subgroup birth to &lt;2 years listed in the above table.

{10}

d. Bridging Studies
Refer to K160538

e. Determination of limit of Blank, lower limits of detection and quantitation:
Refer to K160538

K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

11

---

**Source:** [https://fda-staging.innolitics.com/device/K182389](https://fda-staging.innolitics.com/device/K182389)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
