PowerLook® Tomo Detection V2 software is a computer-assisted detection and diagnosis (CAD) software device intended to be used concurrently by interpreting physicians while reading digital breast tomosynthesis (DBT) exams from compatible DBT systems. The system detects soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in the 3D DBT slices. The detections and Certainty of Finding and Case Scores assist interpreting physicians in identifying soft tissue densities and calcifications that may be confirmed or dismissed by the interpreting physician.
Device Story
PowerLook® Tomo Detection V2 (PLTD V2) is a CAD software device for digital breast tomosynthesis (DBT) exams. It processes 3D DBT slices to detect soft tissue densities (masses, architectural distortions, asymmetries) and calcifications. Using deep learning, the algorithm identifies suspicious lesions and assigns a 'Certainty of Finding' score (0-100%) to each detection and a 'Case Score' (0-100%) to each case, representing relative confidence of malignancy. The software operates as a stand-alone executable within the PowerLook AMP framework. Interpreting physicians use the output as a concurrent read; detections appear as overlays on tomosynthesis images, and marks serve as navigation tools to specific image planes. The output informs clinical decision-making but does not replace physician judgment. The device is intended for use in clinical settings by radiologists to aid in identifying findings that may require further work-up, potentially improving diagnostic performance and reading efficiency.
Clinical Evidence
Clinical performance was evaluated via a retrospective, fully-crossed, multi-reader, multi-case (MRMC) pivotal study with 24 radiologists and 260 Hologic DBT cases (65 cancer). Co-primary endpoints were met: CAD with DBT was non-inferior and superior to DBT alone for case-level AUC (0.852 vs 0.795, p<0.01) and reading time was 52.7% shorter (p<0.01). Secondary endpoints showed superior case-level sensitivity (0.850 vs 0.770) and lesion-level sensitivity (0.853 vs 0.769), with non-inferior specificity and recall rates. Standalone performance was also validated on Hologic and GE DBT datasets, meeting design specifications for sensitivity, specificity, and AUC.
Technological Characteristics
Radiological CAD software; deep learning-based pattern recognition; operates on softcopy workstations; compatible with Hologic Selenia Dimensions and GE Pristina DBT systems; standalone executable within PowerLook AMP framework; software level of concern: Moderate.
Indications for Use
Indicated for symptomatic and asymptomatic women undergoing mammography to assist interpreting physicians in the detection, localization, and characterization of soft tissue densities (masses, architectural distortions, asymmetries) and calcifications in digital breast tomosynthesis (DBT) exams.
Regulatory Classification
Identification
A radiological computer-assisted detection and diagnostic software is an image processing device intended to aid in the detection, localization, and characterization of fracture, lesions, or other disease-specific findings on acquired medical images (e.g., radiography, magnetic resonance, computed tomography). The device detects, identifies, and characterizes findings based on features or information extracted from images, and provides information about the presence, location, and characteristics of the findings to the user. The analysis is intended to inform the primary diagnostic and patient management decisions that are made by the clinical user. The device is not intended as a replacement for a complete clinician's review or their clinical judgment that takes into account other relevant information from the image or patient history.
Special Controls
A radiological computer assisted detection and diagnosis software must comply with the following special controls: Design verification and validation must include: 1. i. A detailed description of the image analysis algorithm, including but not limited to a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable. iii. Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contain a sufficient number of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment. iv. Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of but not limited to clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (e.g., receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in 1(iii) above. v. Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures. 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output. iii. A detailed description of the intended user, and any user training materials as programs that addresses appropriate reading protocols for the device to ensure that the end user is fully aware of how to interpret and apply the device output. iv. A detailed description of the device inputs and outputs. v. A detailed description of compatible imaging hardware and imaging protocols. vi. Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality or for certain subpopulations), as applicable. vii. A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithm, including a description of the algorithm inputs and outputs, each major component or block, how the algorithm and output affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide improved assisted-read detection and diagnostic performance as intended in the indicated user population(s), and to characterize the standalone device performance for labeling. Performance testing includes standalone test(s), side-by-side comparison(s), and/or a reader study, as applicable.
(iii) Results from standalone performance testing used to characterize the independent performance of the device separate from aided user performance. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Devices with localization output must include localization accuracy testing as a component of standalone testing. The test dataset must be representative of the typical patient population with enrichment made only to ensure that the test dataset contains a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant disease, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Results from performance testing that demonstrate that the device provides improved assisted-read detection and/or diagnostic performance as intended in the indicated user population(s) when used in accordance with the instructions for use. The reader population must be comprised of the intended user population in terms of clinical training, certification, and years of experience. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, positive and negative predictive values, and diagnostic likelihood ratio). Test datasets must meet the requirements described in paragraph (b)(1)(iii) of this section.(v) Appropriate software documentation, including device hazard analysis, software requirements specification document, software design specification document, traceability analysis, system level test protocol, pass/fail criteria, testing results, and cybersecurity measures.
(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the device instructions for use, including the intended reading protocol and how the user should interpret the device output.
(iii) A detailed description of the intended user, and any user training materials or programs that address appropriate reading protocols for the device, to ensure that the end user is fully aware of how to interpret and apply the device output.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) A detailed summary of the performance testing, including test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as anatomical characteristics, patient demographics and medical history, user experience, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
December 6, 2018
iCAD Inc. John Delucia VP, Regulatory Affairs, Clinical Affairs and Quality Assurance 98 Spit Brook Rd. Suite 100 NASHUA, NH 03062
Re: K182373
Trade/Device Name: PowerLook® Tomo Detection V2 Software Regulation Number: 21 CFR 892.2090 Regulation Name: Radiological Computer Assisted Detection and Diagnosis Software Regulatory Class: Class II Product Code: ODO Dated: November 6, 2018 Received: November 7, 2018
Dear Mr. Delucia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
Hse 2. Nils
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182373
Device Name PowerLook® Tomo Detection V2 Software
#### Indications for Use (Describe)
PowerLook® Tomo Detection V2 Software is a computer-assisted detection and diagnosis (CAD) software device intended to be used concurrently by interpreting physicians while reading digital breast tomosynthesis (DBT) exams from compatible DBT systems. The system detects soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in the 3D DBT slices. The detections and Certainty of Finding and Case Scores assist interpreting physicians in identifying soft tissue densities and calcifications that may be confirmed or dismissed by the interpreting physician .
Type of Use (Select one or both, as applicable)
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 12px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 1: 510(K) Summary
#### Date Prepared: December 5, 2018
#### Submitter:
iCAD, Inc. 98 Spit Brook Road Suite 100 Nashua, NH 03062
#### Contact Person:
John DeLucia VP, Regulatory Affairs, Clinical Affairs and Quality Assurance Email: jdelucia@icadmed.com Phone: (603) 309-1945 Fax: (603) 880-3043
#### Submission Date:
August 30, 2018
#### Trade Name:
PowerLook® Tomo Detection V2 software
#### Common Name:
Medical Imaging Software
#### Classification Name/Regulation
Radiological Computer Assisted Detection and Diagnosis Software 21 CFR 892.2090
#### Product Code
QDQ
# Device Classification
Class II
{4}------------------------------------------------
#### Legally Marketed Devices to Which Substantial Equivalence is Claimed
PowerLook Tomo Detection (PLTD) V2 Software is substantially equivalent to the following legally marketed predicate device.
| Device Name | Manufacturer | FDA DEN<br>Reference # | Decision Date |
|-------------|--------------|------------------------|---------------|
| OsteoDetect | Imagen | DEN180005 | May 24, 2018 |
#### Device Description
PLTD V2 detects malignant soft-tissue densities and calcifications in digital breast tomosynthesis (DBT) image. The PLTD V2 software allows a interpreting physician to quickly identify suspicious soft tissue densities and calcifications by marking the detected areas in the tomosynthesis images. When the PLTD V2 marks are displayed by a user, the marks will appear as overlays on the tomosynthesis images. The PLTD V2 marks also serve as a navigation tool for users, because each mark is linked to the tomosynthesis plane where the detection was identified. Users can navigate to the plane associated with each mark by clicking on the detection mark. Each detected region will also be assigned a "score" that corresponds to the PLTD V2 algorithm's confidence that the detected region is a cancer (Certainty of Finding Score). Certainty of Finding scores are relative scores assigned to each detected region and a Case Score is assigned to each case regardless of the number of detected regions. Certainty of Finding and Case Scores are computed by the PLTD V2 algorithm and represent the algorithm's confidence that a specific finding or case is malignant. The scores are represented on a 0% to 100% scale. Higher scores represent a higher algorithm confidence that a finding or case is malignant. Lower scores represent a lower algorithm confidence that a finding or case is malignant.
#### Key Outputs of PowerLook® Tomo Detection V2 Software
#### a. Lesion Detection
PowerLook® Tomo Detection V2 software detects soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in the 3D digital breast tomosynthesis images. The PowerLook Tomo Detection V2 algorithm uses deep learning technology to process feature computations and uses pattern recognition to identify suspicious breast lesions appearing as soft tissue densities or clusters of calcifications. Each detected region in the tomosynthesis data is identified or represented by marking the contour of the lesion in the tomosynthesis slice where it was detected.
{5}------------------------------------------------
#### b. Certainty of Finding and Case Scores
Certainty of Finding scores are relative scores assigned to each detected region and a Case Score is assigned to each case regardless of the number of detected regions. Certainty of Finding and Case Scores are computed by the PowerLook Tomo Detection V2 algorithm and represent the algorithm's confidence that a specific finding or case is malignant. The scores are represented on a 0% to 100% scale. Higher scores represent a higher algorithm confidence that a finding or case is malignant. Lower scores represent a lower algorithm confidence that a finding or case is malignant. The scores are based on a population with 50% prevalence of cancer and should be interpreted as the probability of the finding or case correctly being identified as malignant in a population of 50% cancers and 50% non-cancers. The scores serve as a guide to interpreting physicians to aid in determining if a suspicious finding or case needs further work-up. These scores are not intended to be the clinically used "probability of malignancy". Certainty of Finding and Case Scores are not calibrated to the prevalence in the intended use population or to the prevalence in the pivotal reader study outlined in the Assessment of Clinical Performance Data section, and consequently, the Certainty of Finding and Case Scores are in general higher than the actual probability of malignancy in an intended use population with less than 50% prevalence. These scores represent a relative level of concern or level of suspicion because they do not represent an absolute clinical probability of malignancy.
#### Supported Digital Breast Tomosynthesis Systems
The following Digital Breast Tomosynthesis systems have been tested and are compatible with PowerLook Tomo Detection V2 software:
- . Hologic Selenia Dimensions
- . GE Pristina
The PowerLook Tomo Detection V2 software is designed as a stand-alone executable operating within the larger software framework provided by PowerLook AMP1.
#### Intended Use / "Indications for Use"
PowerLook® Tomo Detection V2 software is a computer-assisted detection and diagnosis (CAD) software device intended to be used concurrently by interpreting physicians while reading digital breast tomosynthesis
<sup>1</sup> iCAD PowerLook AMP is a Class I medical device exempt per 21 CFR § 892.2010 (Medical image storage device) and 21 CFR § 892.2020 (Medical image communications device). The PowerLook AMP is a device that provides electronic storage functions for medical images and also provides electronic transfer of medical image data between medical devices
{6}------------------------------------------------
(DBT) exams from compatible DBT systems. The system detects soft tissue densities (masses, architectural distortions and asymmetries) and calcifications in the 3D DBT slices. The detections and Certainty of Finding and Case Scores assist interpreting physicians in identifying soft tissue densities and calcifications that may be confirmed or dismissed by the interpreting physician.
#### Comparison with Predicate Device
The Summary of Substantial Equivalence Table below details the similarities and differences between the subject device. PowerLook Tomo Detection V2. and the predicate device, OsteoDetect. classified under De Novo request DEN180005.
PowerLook Tomo Detection V2 and the predicate device have the same indication for use. Both devices are intended to aid in the detection, localization, and characterization of disease specific findings on acquired medical images, per 21 CFR 892.2090.
From a technology standpoint, PowerLook Tomo Detection V2 is the same as the predicate per 21 CFR 892.2090; both devices are radiological computer assisted detection and diagnostic software, intended to aid in the detection, and characterization of disease specific findings on acquired medical images. The outputs of both devices serve as a concurrent read. The output is used to inform the clinical user (who themselves make the primary diagnostic and patient management decisions) and will not replace the clinical expertise and judgment of the clinical user. Where any differences may occur between the subject device and the predicate, special controls established for Radiological Computer Assisted Detection and Diagnosis Software are in place to further mitigate any risks in these differences.
#### Summary of Indications for Use
The subject device and the predicate device are both classification name and number: Radiological Computer Assisted Detection and Diagnosis Software 21 CFR 892.2090. Both devices are intended to aid in the detection, and characterization of disease specific findings on acquired medical images.
The subject device and the predicate may differ in their detection of disease specific findings however, special controls are in place to mitigate any risk for this difference.
## Summary of Technological Characteristic
The technological characteristics of PowerLook Tomo Detection V2 are the same as the predicate OsteoDetect classified under De Novo request DEN180005. Per 21 CFR 892.2090, both devices are radiological computer assisted detection and diagnostic software intended to aid in the detection, localization, and characterization of disease specific findings on acquired medical
{7}------------------------------------------------
images. The outputs of both devices serve as a concurrent read. The output is used to inform the clinical user (who themselves make the primary diagnostic and patient management decisions) and will not replace the clinical expertise and judgment of the clinical user.
#### General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis which is used to identify and mitigate potential hazards. Any potential hazards are controlled via software development, verification and validation testing. In addition, general controls of the FD&C Act, and special controls established for Radiological Computer Assisted Detection and Diagnosis Software are in place to further mitigate any safety and or effectiveness risks.
#### Assessment of Non-Clinical Performance Data
PowerLook Tomo Detection V2 has been verified and validated according to iCAD's design control processes. All supporting documentation has been included in this 510(k) Premarket Notification. Verification activity included unit, integration, and system level testing. Validation testing included standalone testing and performing a pivotal reader study to compare the clinical performance of radiologists using CAD detections and Certainty of Finding and Case Scores from the PLTD V2 CAD system with DBT images to that of radiologists using DBT without CAD
#### . Standalone Performance on a Screening Population Dataset
## Hologic DBT
A standalone study, which evaluated the performance of PLTD without a radiologist, was conducted with a sample of 655 Hologic DBT cases, including 235 cancer cases with 242 malignant lesions. These 655 cases were used in the standalone study with PLTD V2. A stratified bootstrap procedure was used to estimate over a screening patient population. The bootstrap procedure limits the number of cases in a particular category when computing performance measures.
The purpose of the standalone study was to assess the standalone performance of PLTD V2 on a screening population.
Results from the standalone study showed that Case-Level Sensitivity, Lesion-Level Sensitivity, FP Rate in Non-Cancer Cases, and Specificity met design specifications. The detailed results are in the User Manual.
{8}------------------------------------------------
#### GE DBT
A standalone study, which evaluated the performance of PLTD V2, was conducted with a sample of 610 GE DBT cases, including 204 cancer cases with 221 malignant lesions. These 610 cases were used in the standalone study PLTD V2. A stratified bootstrap procedure was used to estimate performance over a screening patient population. The bootstrap procedure limits the number of cases in a particular category when computing performance measures.
The purpose of the standalone study was to assess the standalone performance of PLTD V2 on a screening population.
Results from the standalone study showed that Case-Level Sensitivity, Lesion-Level Sensitivity, FP Rate in Non-Cancer Cases, and Specificity met design specifications. The detailed results are in the User manual.
#### Comparison of Performance Between GE DBT and Hologic DBT (Control Group)
A comparison was made between the standalone performance of PowerLook Tomo Detection 2.0 on Hologic DBT images and PowerLook Tomo Detection 2.0 on GE DBT images. For this comparison, the performance on Hologic is considered the control group and performance on GE is the test group. The test is to determine if the difference between the control group and the test group is within the margin of non-inferiority for Sensitivity and AUC, and FPPI.
Standalone testing was performed for the control group and the test group individually. Key performance measures were then compared by subtracting the test group performance from the control group performance. This was done in a way to produce not just an estimate of the mean difference but also a distribution of the expected differences. In order to show statistical significance, the two-sided 95% confidence interval boundaries must be within the margin of non-inferiority.
Three measures were used to compare the performance of PowerLook Tomo Detection 2.0 with GE DBT images to PowerLook Tomo Detection 2.0 with Hologic images. Each of the three measures produced differences that were within the margin of non-inferiority. Therefore, in the areas of Sensitivity, FPPI, and AUC, PowerLook Tomo Detection 2.0 with the GE DBT system is not inferior to PowerLook Tomo Detection 2.0 with the Hologic DBT system.
{9}------------------------------------------------
#### Assessment of Clinical Performance Data
#### . Pivotal Reader Study
A pivotal reader study, which was a retrospective, fully-crossed, multi-reader, multi-case (MRMC) study of iCAD's PowerLook® Tomo Detection (PLTD) V2 computer-assisted detection (CAD) system, was conducted with 24 tomosynthesis radiologist readers and an enriched sample of 260 Hologic digital breast tomosynthesis (DBT) cases, including 65 cancer cases with 66 malignant lesions. The purpose of the pivotal study was to compare clinical performance of radiologists using CAD detections and Certainty of Finding and Case Scores from the PLTD V2 CAD system with DBT images to that of radiologists using DBT without CAD. The results of the pivotal study will support a regulatory submission for the PLTD V2 CAD system.
The objectives of this pivotal reader study were the following:
A. Co-primary objectives. The co-primary objectives were to determine:
- 1. Whether radiologist performance when using CAD with DBT images is non-inferior to radiologist performance when using DBT images without CAD and
- 2. Whether radiologist reading time when using CAD with DBT images is superior to (shorter than) radiologist reading time when using DBT images without CAD.
Radiologist performance was assessed by measuring case-level area under the receiver operating characteristic (ROC) curve (AUC) for the detection of malignant lesions, where malignant lesion localization was required for a reader to correctly detect cancer in a case.
B. Secondary objectives. The secondary objectives of the pivotal reader study included the following for radiologists when using CAD with DBT compared to using DBT without CAD:
- 1. Superiority of case-level AUC
- 2. Non-inferiority (with non-inferiority margin delta = 0.05) of sensitivity at the case level
- 3. Superiority of sensitivity at the case level
- 4. Non-inferiority (with non-inferiority margin delta = 0.05) of sensitivity at the lesion level
- 5. Superiority of sensitivity at the lesion level
- 6. Non-inferiority (with non-inferiority margin delta = 0.05) of specificity (case-level)
- 7. Non-inferiority (with non-inferiority margin delta = 0.05) of recall rate in non-cancers (case-level)
{10}------------------------------------------------
The pivotal study results showed that both co-primary endpoints were met. Specifically, the pivotal study showed that:
- 1. Radiologist performance using CAD with DBT was non-inferior to, and statistically significantly superior to, radiologist performance using DBT without CAD. Radiologists had superior per-subject average area under the receiver operating characteristic (ROC) curve (AUC) with CAD, 0.852, versus without CAD, 0.795. The average difference in AUC was 0.057 (95% CI: 0.028, 0.087; non-inferiority p < 0.01 for non-inferiority margin delta = 0.05, and p < 0.01 for test of difference).
- 2. Radiologist reading time when using CAD with DBT is superior to (shorter than) radiologist reading time when using DBT without CAD. Reading time improved 52.7% with CAD (95% CI: 41.8%, 61.5%; p < 0.01). *
* Interpreting physicians reading times may vary based on the specific functionality of the viewing application used for interpretation.
All pre-specified secondary endpoints also were met. In addition to superiority of case-level AUC, the pivotal study showed that:
- Radiologists had superior sensitivity at the case level with CAD. Average sensitivity . increased by 0.080 (95% CI: 0.026, 0.134; non-inferiority p < 0.01 for non-inferiority margin delta = 0.05, and p < 0.01 for test of difference). Average case-level sensitivity was 0.770 without CAD and 0.850 with CAD.
- At the lesion level, radiologists also had superior sensitivity with CAD. Average per-. lesion sensitivity across readers increased by 0.084 (95% CI: 0.029, 0.139; noninferiority p < 0.01 for non-inferiority margin delta = 0.05, and p < 0.01 for test of difference), from 0.769 without CAD to 0.853 with CAD.
- Radiologists had non-inferior specificity with CAD. Specificity was 0.627 without CAD and 0.696 with CAD, for an average increase of 0.069 (95% CI: 0.030, 0.108: non-inferiority p < 0.01 for non-inferiority margin delta = 0.05).
- Finally, radiologists had non-inferior recall rate in non-cancer cases with CAD. In non-● cancer cases, lower recall rates are better than higher recall rates. Average recall rate in non-cancer cases was 0.380 without CAD and 0.309 with CAD, for an average reduction of 0.072 (95% CI: 0.031, 0.112; non-inferiority p < 0.01 for non-inferiority margin delta = 0.05).
In this study the following were observed:
- . Average sensitivity increased by 0.120 (SE=0.040) in the subgroup of 15 cancer cases with only calcifications.
- . Average sensitivity increased by 0.068 (SE=0.031) in the subgroup of 50 cancer cases with at least one soft tissue density or mixed lesion.
{11}------------------------------------------------
- Average specificity decreased by 0.027 (SE=0.038) in the subgroup of 24 benign and . recalled (non-cancer) cases with only calcifications.
- Average specificity increased by 0.079 (SE=0.028) in the subgroup of 62 benign and ● recalled (non-cancer) cases with at least one soft tissue density or mixed lesion.
- Average specificity increased by 0.084 (SE=0.021) in the subgroup of 109 non-cancer . cases with no lesions.
#### Conclusion:
Based upon the information presented in this submission, it is concluded that PowerLook® Tomo Detection V2 is substantially equivalent to the named predicate device and that the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device
{12}------------------------------------------------
# Summary of Substantial Equivalence:
| Features and<br>Characteristics | Subject Device | Predicate Device | Discussion of Differences and<br>Comments |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | iCAD Inc.<br>PowerLook® Tomo Detection V2<br>Software | Imagen Inc.<br>OsteoDetect<br>DEN180005 | |
| Regulation<br>Number/Name | 21 CFR 892.2090 / Radiological Computer Assisted<br>Detection and Diagnosis Software. | Same | NA |
| Regulation<br>Description | A radiological computer assisted detection and diagnostic<br>software is an image processing device intended to aid in<br>the detection, localization, and characterization of fracture,<br>lesions, or other disease specific findings on acquired<br>medical images (e.g. radiography, MR, CT). The device<br>detects, identifies and characterizes findings based on<br>features or information extracted from images, and<br>provides information about the presence, location, and<br>characteristics of the findings to the user. The analysis is<br>intended to inform the primary diagnostic and patient<br>management decisions that are made by the clinical user.<br>The device is not intended as a replacement for a complete<br>clinician's review or their clinical judgment that takes into<br>account other relevant information from the image or<br>patient history. | Same | NA |
| Intended Use | PowerLook® Tomo Detection V2 software is a computer-<br>assisted detection and diagnosis (CAD) software device<br>intended to be used concurrently by interpreting physicians<br>while reading digital breast tomosynthesis (DBT) exams<br>from compatible DBT systems. The system detects soft<br>tissue densities (masses, architectural distortions and<br>asymmetries) and calcifications in the 3D DBT slices. The<br>detections and Certainty of Finding and Case Scores assist<br>interpreting physicians in identifying soft tissue densities<br>and calcifications that may be confirmed or dismissed by<br>the interpreting physician. | OsteoDetect analyzes wrist radiographs<br>using machine learning techniques to identify<br>and highlight distal radius fractures during<br>the review of posterior-anterior (PA) and<br>lateral (LAT) radiographs of adult wrists. | Both devices have the same<br>intended use per 21 CFR<br>892.2090. |
| Type of CAD<br>Software | Radiological computer assisted detection and<br>diagnostic software | Same | NA |
| Mode of Action | Image processing device intended to aid in the<br>detection, localization, and characterization of soft<br>tissue densities (masses, architectural distortions<br>and asymmetries) and calcifications in the 3D<br>DBT slices. | Image processing device intended to aid in<br>identifying and highlighting distal radius<br>fractures during the review of posterior-<br>anterior (PA) and lateral (LAT) radiographs<br>of adult wrists. | Both devices are intended to aid in the<br>detection, localization, and characterization<br>of disease specific findings on acquired<br>medical images per 21 CFR 892.2090.<br>PLTD detects a different disease specific<br>finding. However, special controls are in<br>place to mitigate any risk for this difference. |
| Clinical Output | To inform the primary diagnostic and patient<br>management decisions that are made by the<br>clinical user. | Same | The outputs of both devices serve as a<br>concurrent read. The output is used to<br>inform the clinical user (who themselves<br>make the primary diagnostic and patient<br>management decisions) and will not replace<br>the clinical expertise and judgment of the<br>clinical user. |
| Patient<br>Population | Symptomatic and asymptomatic women<br>undergoing mammography. | Adult men and women undergoing<br>radiographs of adult wrists. | PLTD is intended for a different patient<br>population however special controls are<br>in place to mitigate any risk for this<br>difference. |
| End Users | Interpreting Physicians Radiologists | Clinicians | Same |
| Image Source<br>Modalities | Digital breast tomosynthesis slices | 2D X-ray images | The Image Source Modalities are different.<br>However, special controls are in place to<br>mitigate any risk for this difference. |
| Output Device | Softcopy Workstation | Softcopy Workstation, PACS, RIS | NA |
| Deployment | Stand-alone computer | Same | NA |
| Software Level<br>of Concern | Moderate | Moderate | NA |
{13}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.