← Product Code [GEI](/productcode/GEI) · K182365

# TempSure System (K182365)

_Cynosure, Inc. · GEI · Oct 24, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182365

## Device Facts

- **Applicant:** Cynosure, Inc.
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** Oct 24, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids. The 25mm, and 30mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation. The massage device is intended to provide a temporary reduction in the appearance of cellulite. The following surgical modes are applicable to the generator: Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed. Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty. Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps. Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis. Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage.

## Device Story

TempSure System is a radiofrequency (RF) generator for aesthetic and surgical applications; delivers RF energy to patient tissue to induce heating and cellular alteration. System includes generator, temperature-sensing handpieces, footswitch, neutral pad, and electrosurgical accessories. Used in clinical settings by physicians; energy output controlled via GUI and switches. Surgical modes include Cut, Coagulation, Blend, Fulgurate, and Bipolar. Temperature-sensing handpieces provide feedback for aesthetic treatments (wrinkles, pain relief, cellulite reduction). Output affects tissue via thermal effects; provider monitors device settings to achieve desired clinical outcome. Benefits include precise tissue heating and surgical hemostasis/cutting.

## Clinical Evidence

No clinical data submitted. Substantial equivalence supported by non-clinical bench testing, including ex-vivo tissue samples (liver, kidney, muscle) to compare thermally affected zones (TAZ) against predicate devices, and electrical safety/EMC testing per IEC 60601-1, 60601-1-2, and 60601-2-2.

## Technological Characteristics

RF generator; Monopolar and Bipolar modalities. Output: 4.0 MHz Sin-wave CW, Fully/Partially Rectified; 1.7 MHz for Bipolar. Max power 300W. Materials: Gold-Plated Brass, Polytheramide, Loctite, Delrin. Features temperature-sensing handpieces with <1s response time. Connectivity: Standalone. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Cynosure TempSure ([K171262](/device/K171262.md))
- Cynosure dba Ellman Surgitron 4.0 Dual RF S5 - Pelleve ([K123366](/device/K123366.md))
- ValleyLab ForceTriad ([K051644](/device/K051644.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. The FDA logo is on the right, with the letters "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

October 24, 2018

Cynosure, Inc Amy Tannenbaum Regulatory Affairs Specialist III 5 Carlisle Road Westford, Massachusetts 01886

Re: K182365

Trade/Device Name: TempSure System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 29, 2018 Received: August 30, 2018

Dear Amy Tannenbaum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely. Jennifer R. Stevenson For Binita S. Ashar, M.D., M.B.A., F.A.C.S.

Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K182365

Device Name TempSure System

#### Indications for Use (Describe)

The 10mm, 15mm, and 20mm handpieces are indicated for non-ablative treatment of mild to moderate facial wrinkles and rhytids.

The 25mm, and 30mm handpieces provide heating for the purpose of elevating tissue temperature for selected medical conditions such as temporary relief of pain, muscle spasms, and increase in local circulation.

The massage device is intended to provide a temporary reduction in the appearance of cellulite

The following surgical modes are applicable to the generator:

Coagulation/Hemostasis: General surgical procedures including urologic, thoracic, plastic, reconstructive, and gynecological procedures where electrosurgical coagulation of tissue is performed

Cutting: snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage. skin incisions, biopsy, cysts, abscesses. tumors. cosmetic repairs, development of skin flaps. skin tags and blepharoplasty.

Blended Cutting and Coagulation: submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment, turbinate shrinkage, skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma. nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and development of skin flaps.

Fulguration: basal cell carcinoma. papilloma. cyst destruction, tumors. verrucae, hemostasis.

Bipolar: pinpoint precise coagulation. pinpoint hemostasis in any, field (wet or dry). snoring, submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage control. epistaxis treatment and turbinate shrinkage

| Type of Use (Select one or both, as applicable)                                                     |                                                                                                    |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial, sans-serif;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# 510(k) Summary for Cynosure TempSure System

A summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.

| 807.92(a)(1) Submitter Information                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant                                                                                                                                                                                                                                                                      | Cynosure, Inc                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Address                                                                                                                                                                                                                                                                        | 5 Carlisle Road, Westford MA, 01886                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Phone Number                                                                                                                                                                                                                                                                   | 781-993-2454                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Fax Number                                                                                                                                                                                                                                                                     | 978-256-6556                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Establishment Registration Number                                                                                                                                                                                                                                              | 1222993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Contact Person                                                                                                                                                                                                                                                                 | Amy Tannenbaum                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Preparation Date                                                                                                                                                                                                                                                               | August 29, 2018                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| 807.92(a)(2) Name of Device                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Trade or Proprietary Name                                                                                                                                                                                                                                                      | TempSure System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Common or Usual Name                                                                                                                                                                                                                                                           | Surgical RF Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name                                                                                                                                                                                                                                                            | Electrosurgical, Cutting / Coagulation / Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Classification Panel                                                                                                                                                                                                                                                           | General & Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regulation                                                                                                                                                                                                                                                                     | 21 CFR 878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Regulatory Class                                                                                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code(s)                                                                                                                                                                                                                                                                | GEI, PBX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| 807.92 (a)(3) Legally marketed device(s) to which equivalence is claimed                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate Devices                                                                                                                                                                                                                                                              | Cynosure TempSure (K171262)<br>Cynosure dba Ellman Surgitron 4.0 Dual RF S5 - Pelleve (K123366)<br>ValleyLab ForceTriad (K051644)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 807.92(a)(4) Device Description                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                | The Cynosure TempSure™ System is a radiofrequency generator with<br>a variety of applications both aesthetic and surgical procedures. The<br>intended action is achieved through application of radiofrequency<br>energy to the patient which results in minimization of heat dissipation<br>and cellular alteration. Output of energy is controlled via the guided<br>user interface (GUI) and the foot and/or hand-switch.<br><br>The TempSure generator has been upgraded to include additional<br>surgical modes – Cut, Bipolar, Fulgurate, and Blend, along with the<br>existing Coag. The power of the device for Surgical applications is a<br>maximum of 300W. There have been no changes to the Smart<br>Handpiece treatment with temperature sensitive handpieces. |
|                                                                                                                                                                                                                                                                                | TempSure system is used with the same existing Ellman                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                | electrosurgical accessories.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                | The TempSure™ System includes:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                                                                                                                                                                                                                                | TempSure™ Generator<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                                                                                                                                                                                                                                                | Temperature Sensing Handpieces<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                                                                                                                                                                                                                                | IEC Power Cord<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                                                                                                                                                                                                                                                | Footswitch<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                | Disposable/Reusable Neutral Pad<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                                                                                                                                                                                                                                                                | Surgical Fingerswitch/Foot Controlled Handpieces<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                | Monopolar Cables<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                                                                                                                                                                                                                                | Disposable/Reusable Electrodes, Forceps<br>•                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 807.92(a)(5) Intended Use of the Device                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                | The TempSure System has the following indications for use:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                | The 10mm, 15mm, and 20mm handpieces are indicated for non-<br>ablative treatment of mild to moderate facial wrinkles and rhytids.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                | The 25mm, and 30mm handpieces provide heating for the purpose of<br>elevating tissue temperature for selected medical conditions such as<br>temporary relief of pain, muscle spasms, and increase in local<br>circulation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                                                                                                                                                                                                                                                                | The massage device is intended to provide a temporary reduction in<br>the appearance of cellulite                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                                                                                                                                                                                                                | The following surgical modes are applicable to the generator:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                                                                                | <b>Coagulation/Hemostasis:</b> General surgical procedures including<br>urologic, thoracic, plastic, reconstructive, and gynecological<br>procedures where electrosurgical coagulation of tissue is performed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                                                                                                                                                                                                                                                                                | <b>Cutting:</b> snoring. submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage<br>control. epistaxis treatment, turbinate shrinkage. skin incisions,<br>biopsy, cysts, abscesses. tumors. cosmetic repairs, development of'<br>skin flaps. skin tags and blepharoplasty.                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                                                                                                                                                                                                                                | <b>Blended Cutting and Coagulation:</b> snoring. submucosal palatal<br>shrinkage, traditional uvulopalatoplasty (RAUP). myringotomy with<br>effective hemorrhage control. epistaxis treatment, turbinate shrinkage,<br>skin tags, papilloma keloids. keratosis, verrucae, basal cell carcinoma.                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| nevi. fistulas. epithelidma. cosmetic repairs. cysts. abscesses. and<br>development of skin flaps.                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| <b>Fulguration:</b> basal cell carcinoma. papilloma. cyst destruction,<br>tumors. verrucae, hemostasis.                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| <b>Bipolar:</b> pinpoint precise coagulation. pinpoint hemostasis in any,<br>field (wet or dry). snoring. submucosal palatal shrinkage, traditional<br>uvulopalatoplasty (RAUP). myringotomy with effective hemorrhage<br>control. epistaxis treatment and turbinate shrinkage |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

{6}------------------------------------------------

### 807.92(b)(1) Non-clinical tests submitted

The following non-clinical tests have been included in this 510k submission in determination of substantial equivalence between the test device and the referenced predicates. These tests have been conducted in accordance with the FDA Guidance "Premarket Notification 510k Submissions for electrosurgical Devices for General Surgery" - Section XI. Performance Data. E - System Testing, Thermal Effects on Tissue. See Section 18 - Performance Testing, Bench. A brief summary is provided below.

#### Thermal Effects on Tissue

Testing was performed on ex-vivo tissue samples to compare the thermally affected zone (TAZ) of the TempSure and predicate devices. Measurements were taken in triplicate on three tissue samples (liver, kidney, and muscle), and three power settings. For each tissue and power setting combination, the TempSure and the predicate device(s) were compared. This test shows that the TempSure is substantially equivalent to the ForceTriad and Surgitron devices.

### Electromagnetic Compatibility and Electrical Safety

Electrical safety testing for the Cynosure TempSure System was also completed to prove the safe use of the device. These test reports are provided in accordance with FDA Guidance "Premarket Notification 510k Submissions for Electrosurgical Devices for General Surgery" - Section XII -Electrical Safety and Electromagnetic Compatibility". The following test reports are available in Section 17 - Electromagnetic Compatibility and Electrical Safety.

- IEC 60601-1, Medical Electrical Equipment Part 1: General requirements for basic safety . and essential performances
- IEC 60601-1-2, Medical Electrical Equipment Part 1 -2: General requirements for basic ● safety and essential performance - Collateral standard: Electromagnetic disturbances requirements and tests
- IEC 60601-2-2, Medical electrical equipment Part 2 -2: Particular requirements for the basic ● safety and essential performance of high frequency surgical equipment and high frequency surgical accessories

### 807.92(b)(2) Clinical tests submitted – N/A - No clinical tests submitted

### 807.92(b)(3) Conclusions drawn from clinical and non-clinical tests submitted

{7}------------------------------------------------

The nonclinical tests demonstrate that the TempSure™ system is safe and effective and performs as well as the legally marketed predicate devices. The Thermal Effects on Tissue test showed that the thermally affected zone created by the TempSure™ device was similar to that created by the predicate devices, Surgitron and ForceTriad, in ex-vivo tissue testing. Despite their difference in maximum output power, the devices created a similar TAZ and therefore the devices can be considered substantially equivalent. In addition, the Electromagnetic Compatibility and Electrical Safety testing shows that the device is safe to use and meets required standards. The non-clinical testing shows that the device meets design specifications as well as performance requirements.

{8}------------------------------------------------

### Cynosure TempSure K182365

| Characteristic                   | Cynosure TempSure™<br>(KPending)                                                                                                                                                                                                                                                                                                                           | Cynosure TempSure<br>(K171262)                                                                                                                                                                                                                     | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve<br>S5) (K123366)                                                                                                                                                                                                                                                                                                                                                 | Valley Labs Force<br>Triad (K051644)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use              | The 10mm, 15mm, and<br>20mm Smart handpieces<br>are indicated for non-<br>ablative treatment of mild<br>to moderate facial<br>wrinkles and rhytids                                                                                                                                                                                                         | The 10mm, 15mm, and<br>20mm handpieces are<br>indicated for non-ablative<br>treatment of mild to<br>moderate facial wrinkles<br>and rhytids                                                                                                        | <b>Cutting:</b> Snoring,<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effect hemorrhage<br>control, epistaxis<br>treatment, turbinate<br>shrinkage, skin<br>incisions biopsy, cysts,<br>abscesses, tumors,<br>cosmetic repairs,<br>development of skin<br>flaps, skin tags, and<br>blepharoplasty                                                     | General (including<br>urologic, thoracic,<br>plastic and<br>reconstructive,<br>arthroscopic),<br>laparoscopic, and<br>gynecological<br>procedures where<br>electrosurgical cutting<br>and coagulation of<br>tissue, and sealing<br>(fusion) of vessels and<br>tissue bundles is<br>performed, including<br>such procedures as<br>bowel resections,<br>hysterectomies (both<br>vaginal and<br>abdominal),<br>laparoscopic,<br>cholecystectomies,<br>laparoscopically<br>assisted vaginal<br>hysterectomies, gall<br>bladder procedures,<br>Nissen fundoplication,<br>adhesiolysis,<br>oophorectomy, etc. |
|                                  | The 25mm, and 30mm<br>Smart handpieces provide<br>heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such<br>as temporary relief of<br>pain, muscle spasms, and<br>increase in local<br>circulation.                                                                                                         | The 18mm, 25mm, and<br>30mm handpieces provide<br>heating for the purpose of<br>elevating tissue<br>temperature for selected<br>medical conditions such<br>as temporary relief of<br>pain, muscle spasms, and<br>increase in local<br>circulation. | <b>Blended Cutting and<br/>Coagulation:</b> Snoring,<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective<br>hemorrhage control,<br>epistaxis treatment,<br>turbinate shrinkage,<br>skin tags, papilloma<br>keloids, keratosis,<br>verrucae, basal cell<br>carcinoma, nevi,<br>fistulas, epithelidma,<br>cosmetic repairs, cysts,<br>abscesses, and | Vessels (arteries,<br>veins, lymph) 7mm<br>and smaller in<br>diameter, and bundles<br>as large as will fit in                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                  | The massage device is<br>intended to provide a<br>temporary reduction in the<br>appearance of cellulite                                                                                                                                                                                                                                                    | The massage device is<br>intended to provide a<br>temporary reduction in the<br>appearance of cellulite                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | The following surgical<br>modes are applicable to<br>the generator:                                                                                                                                                                                                                                                                                        | <b>Coagulation/Hemostasis:</b><br>Using the surgical<br>handpieces and<br>accessories, general<br>surgical procedures<br>including urologic,<br>thoracic, plastic,<br>reconstructive, and<br>gynecological procedures<br>where electrosurgical     |                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | <b>Coagulation/Hemostasis:</b><br>General surgical<br>procedures including<br>urologic, thoracic, plastic,<br>reconstructive, and<br>gynecological procedures                                                                                                                                                                                              |                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Characteristic                   | Cynosure TempSure™<br>(KPending)                                                                                                                                                                                                                                                                                                                           | Cynosure TempSure<br>(K171262)                                                                                                                                                                                                                     | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve S5) (K123366)                                                                                                                                                                                                                                                                                                                                                    | Valley Labs Force<br>Triad (K051644)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                  | where electrosurgical<br>coagulation of tissue is<br>performed                                                                                                                                                                                                                                                                                             | coagulation of tissue is<br>performed                                                                                                                                                                                                              | development of skin<br>tags.                                                                                                                                                                                                                                                                                                                                                                                 | the jaws of the devices<br>can be sealed with<br>Ligasure vessel sealing<br>(tissue fusion) output.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                  | <b>Cutting:</b> snoring.<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective hemorrhage<br>control. epistaxis<br>treatment, turbinate<br>shrinkage. skin incisions,<br>biopsy, cysts, abscesses.<br>tumors. cosmetic repairs,<br>development of skin<br>flaps. skin tags and<br>blepharoplasty.  |                                                                                                                                                                                                                                                    | <b>Hemostasis:</b> Control of<br>bleeding, epilation,<br>telangiectasia.                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  | <b>Blended Cutting and<br/>Coagulation:</b> snoring.<br>submucosal palatal<br>shrinkage, traditional<br>uvulopalatoplasty<br>(RAUP). myringotomy<br>with effective hemorrhage<br>control. epistaxis<br>treatment, turbinate<br>shrinkage, skin tags,<br>papilloma keloids.<br>keratosis, verrucae, basal<br>cell carcinoma. nevi.<br>fistulas. epithelidma |                                                                                                                                                                                                                                                    | <b>Fulguration:</b> Basal<br>cell carcinoma,<br>papilloma, cyst<br>destruction, tumors,<br>verrucae, hemostasis.                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  |                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    | <b>Bipolar:</b> Pinpoint<br>precise coagulation,<br>pinpoint hemostasis in<br>any field (wet or dry),<br>snoring, submucosal<br>palatal shrinkage,<br>traditional<br>uvulopalatoplasty<br>(RAUP), myringotomy<br>with effective<br>hemorrhaging control,<br>epistaxis treatment and<br>turbinate shrinkage.                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                  |                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                    | Wrinkles: Non-ablative<br>treatment of mild to<br>moderate facial<br>wrinkles and rhytids.                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Characteristic                   | Cynosure TempSure™<br>(KPending)                                                                                                                                                                                                                                                                                                                           | Cynosure TempSure<br>(K171262)                                                                                                                                                                                                                     | Ellman Surgitron 4.0<br>Dual RF S5 (Pelleve<br>S5) (K123366)                                                                                                                                                                                                                                                                                                                                                 | Valley Labs Force<br>Triad (K051644)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                  | cosmetic repairs. cysts.<br>abscesses. and<br>development of skin flaps.                                                                                                                                                                                                                                                                                   |…

---

**Source:** [https://fda-staging.innolitics.com/device/K182365](https://fda-staging.innolitics.com/device/K182365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).