HardyDisk AST Eravacycline 20µg (ERV20)

K182357 · Hardy Diagnostics · JTN · Sep 26, 2018 · Microbiology

Device Facts

Record IDK182357
Device NameHardyDisk AST Eravacycline 20µg (ERV20)
ApplicantHardy Diagnostics
Product CodeJTN · Microbiology
Decision DateSep 26, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1620
Device ClassClass 2

Indications for Use

HardyDisk™ AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococus spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae. Use of HardyDisk AST Eravacycline 2015 (ERV20) for in vitro agar diffusion susceptibility testing is indicated when there is need to determine the susceptibility of bacteria among the Enterobacteriaceae to Eravacycline. HardyDisk AST Eravacycline at concentration 20μg can be used to determine the zone diameter (mm) of Eravacycline against the following bacteria among the Enterobacteriaceae. Eravacycline has been shown to be active both clinically and in vitro against the following organisms: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Eravacycline has been shown to be active in vitro against most of the following bacteria, but their clinical significance is unknown: Citrobacter koseri Klebsiella (Enterobacter) aerogenes.

Device Story

HardyDisk AST Eravacycline 20μg consists of 6-mm filter paper disks impregnated with 20μg of Eravacycline; used in clinical microbiology laboratories by trained personnel. Principle of operation is Kirby-Bauer agar diffusion; disks are placed on inoculated Mueller Hinton agar plates. After 16-18 hours incubation at 35±2°C, user measures zone of inhibition diameter (mm). Results are compared to established interpretive criteria to classify bacteria as Susceptible. Provides clinicians with susceptibility data to inform antibiotic treatment decisions for Enterobacteriaceae infections.

Clinical Evidence

No clinical data. Substantial equivalence supported by analytical data, including reproducibility, quality control, and disk-to-MIC correlation studies evaluated by CDER to establish breakpoints and QC ranges.

Technological Characteristics

Antimicrobial susceptibility test disk containing 20μg of Eravacycline. The device is a paper disk designed for use in standardized Kirby-Bauer agar diffusion procedures. It is a manual, non-automated, non-software-based diagnostic tool.

Indications for Use

Indicated for in vitro susceptibility testing of Enterobacteriaceae to Eravacycline using agar diffusion (Kirby-Bauer) methods. Applicable to Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca, Klebsiella pneumoniae, Citrobacter koseri, and Klebsiella (Enterobacter) aerogenes.

Regulatory Classification

Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: K182357 B. Purpose for Submission: To obtain a substantial equivalence determination for Eravacycline Antimicrobial Susceptibility Test Disk C. Measurand: Eravacycline 20μg D. Type of Test: Antimicrobial Susceptibility Test Disks E. Applicant: Hardy Diagnostics F. Proprietary and Established Names: HardyDisk AST Eravacycline 20μg (ERV20) G. Regulatory Information: 1. Regulation section: 21 CFR 866.1620 Antimicrobial Susceptibility Test Disc 2. Classification: Class II 3. Product code: JTN 4. Panel: 83, Microbiology 1 {1} # H. Intended Use: # 1. Intended use(s): HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae. # 2. Indication(s) for use: Use of HardyDisk AST Eravacycline 20μg (ERV20) for in vitro agar diffusion susceptibility testing is indicated when there is need to determine the susceptibility of bacteria among the Enterobacteriaceae to Eravacycline. HardyDisk AST Eravacycline at concentration 20μg can be used to determine the zone diameter (mm) of Eravacycline against the following bacteria among the Enterobacteriaceae. Eravacycline has been shown to be active both clinically and in vitro against the following organisms: Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Among the Enterobacteriaceae, Eravacycline has been shown to be active in vitro against most of the following bacteria, but their clinical significance is unknown: Citrobacter koseri Klebsiella (Enterobacter) aerogenes HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae. # 3. Special conditions for use statement(s): For prescription use only 2 {2} 4. Special instrument requirements: Not applicable. ### I. Device Description: HardyDisk AST Disks utilize 6-mm diameter white filter paper disks. The disks are prepared by impregnating absorbent paper with a known concentration of 20 \( \mu \) g Eravacycline. The disks are marked with the code ERV20, on both sides. HardyDisk AST Disks are supplied in plastic cartridges containing 50 disks each. They are also packaged as one cartridge per vial with desiccant or five cartridges per vial with desiccant. ### J. Substantial Equivalence Information: 1. Predicate device name(s): HardyDisk Tigecycline 15μg 2. Predicate 510(k) number(s): K062245 3. Comparison with predicate: Table 1: Comparison with the Predicate | Similarities | | | | --- | --- | --- | | Item | DeviceK182357HardyDiskEravacycline 20μg | PredicateK062245HardyDiskTigecycline 15μg | | Test Method | Antimicrobial Susceptibility Testing using paper disks impregnated with an antimicrobial agent | Same | | Methodology | Kirby-Bauer Disk Diffusion Susceptibility Test Protocol requires the user to determine categorical interpretations (S/I/R) using the measured zone diameters. | Same | | Inoculum | Prepared from pure isolated colonies to match the turbidity equivalent of a 0.5 McFarland in Tryptic Soy Broth. | Same | | Inoculation Method | Dip a sterile swab into the prepared inoculum and streak an appropriate agar plate's surface three times. Add | Same | 3 {3} | Similarities | | | | --- | --- | --- | | Item | Device K182357 HardyDisk Eravacycline 20μg | Predicate K062245 HardyDisk Tigecycline 15μg | | | the disks impregnated with the antimicrobial agent to the surface of the plate. Incubate the agar plate agar side up in a 35 ± 2°C incubator for 16-18 hours. | | | Reading Method | The user will interpret the zone diameters according established interpretive criteria for the drug. | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Product Name | HardyDisk Eravacycline 20μg (ERV20) | HardyDisk Tigecycline | | Antimicrobial Agent | Eravacycline | Tigecycline | | Concentration | 20μg | 15μg | ### K. Standard/Guidance Document Referenced (if applicable): CLSI M100 28th Edition, Performance Standards for Antimicrobial Susceptibility Testing ### L. Test Principle: The HardyDisk AST Disk is based on the agar diffusion (Kirby-Bauer) methodology. It utilizes dried filter paper disks impregnated with a known concentration of an antimicrobial agent that are placed onto the test medium surface. Mueller Hinton agar is recommended for agar diffusion testing of non-fastidious organisms and Mueller Hinton with 5% Sheep Blood is recommended for Streptococcus spp. Three to five similar colonies are transferred to 4-5 mL of a suitable broth medium. The broth is incubated at 35°C for 2-6 hours to develop a turbidity that exceeds or is equivalent to a 0.5 McFarland standard. Alternatively, a direct broth or saline suspension of colonies may be prepared from an overnight culture. The final inoculum density should be equivalent to a 0.5 McFarland turbidity standard. The inoculum density may also be standardized photometrically. Within 15 minutes of inoculum preparation, the Mueller Hinton agar plate is streaked with an inoculated swab to obtain an even inoculation of organism. Disks are aseptically placed onto the agar surface with a disk dispenser and the disks are pressed down with a sterile needle or forceps to make contact with the agar surface. Agar plates are incubated in an ambient air incubator at 35±2°C for 16 - 18 hours. Fastidious organisms are tested using appropriate media incubated in an atmosphere enriched with 5% CO₂, as recommended in the CLSI M02 approved standard document. 4 {4} After incubation the agar medium is examined for a zone of inhibition around the disks. The zones of inhibition are measured to the nearest millimeter and compared to recognized zone size ranges for the antimicrobial agent being tested. ### M. Performance Characteristics (if/when applicable): Descriptive characteristics were sufficient for the HardyDisk Eravacycline 20 \( \mu \) g (ERV20) disk based on extensive data from several microbiology disk studies evaluated by CDER which were used to generate the breakpoints and quality control (QC) expected ranges used for this subject device. The disk data used to support this submission included data from testing Enterobacteriaceae within the spectrum of activity of Eravacycline. Data was obtained from reproducibility, quality control and disk to MIC correlation studies and was generated in accordance with the CDER Clinical/Antimicrobial guidance, Microbiology Data for Systemic Antibacterial Drugs- Development, Analysis, and Presentation to ensure precise, accurate, and reproducible results. For this review, the interpretative criteria are applied broadly to include the Enterobacteriaceae family according to the FDA STIC website. Testing has been expanded to include members of the family and is not limited only to the indicated species. The following statements are added as footnotes to the Eravacycline 20 \( \mu \) g interpretative criteria table for Enterobacteriaceae in the HardyDisk AST package insert: The safety and efficacy of Eravacycline in treating clinical infections due to organisms other than Citrobacter freundii, Enterobacter cloacae, Escherichia coli, Klebsiella oxytoca and Klebsiella pneumoniae, may or may not have been established in adequate and well-controlled clinical trials. The clinical significance of susceptibility information in such instances is unknown. The current absence of resistant isolates precludes defining any results other than “Susceptible”. Isolates yielding results other than “Susceptible” should be submitted to a reference laboratory for further testing. Activity of eravacycline was demonstrated in vitro against Enterobacteriaceae in the presence of certain beta-lactamases, including extended spectrum beta-lactamases, and AmpC. However, some beta-lactamase-producing isolates may confer resistance to eravacycline via other resistance mechanisms. #### 1. Analytical performance: a. Precision/Reproducibility: Not applicable b. Linearity/assay reportable range: 5 {5} Not applicable c. *Traceability, Stability, Expected values (controls, calibrators, or methods):* Not applicable d. *Detection limit:* Not applicable e. *Analytical specificity:* Not applicable f. *Assay cut-off:* Not applicable # 2. Comparison studies: a. *Method comparison with predicate device:* Not applicable b. *Matrix comparison:* Not applicable # 3. Clinical studies: a. *Clinical Sensitivity:* Not applicable b. *Clinical specificity:* Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Not applicable # 4. Clinical cut-off: Not applicable 6 {6} ### 5. Expected values/Reference range: The Eravacycline interpretative criteria for disk diffusion is shown in Table 2 below. Table 2: Interpretative Criteria for Eravacycline Disk Diffusion \( ^{a} \) (Zone diameter in mm) | Organism(s) | Interpretative Criteria | | --- | --- | | Enterobacteriaceae | Susceptible | | | ≥15 | \( ^{a} \) FDA-recognized Antimicrobial Susceptibility Test Interpretive Criteria Website https://www.fda.gov/Drugs/DevelopmentApprovalProcess/DevelopmentResources/ucm575163.htm. The QC strains and expected ranges are the same as recommended in CLSI M100 28 \( ^{th} \) edition. ## N. Proposed Labeling: The labeling supports the finding of substantial equivalence for this device. ## O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 7
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...