Cervella Cranial Electrotherapy Stimulator (CES) is indicated for treatment of insomnia, depression, or anxiety.
Device Story
Cervella is a Cranial Electrotherapy Stimulator (CES) delivering small pulses of electrical current to the brain via conductive electrodes placed bilaterally on the mastoid process. Electrodes are integrated into stereo over-ear noise-canceling headphones. The device is powered by an internal rechargeable Li-ion battery. Operation is controlled via a smartphone app communicating via Bluetooth LE; the app allows users to adjust stimulation intensity, frequency, and duration. The device includes a separate Bluetooth connection for audio/music playback. Treatment history is automatically stored in the app for healthcare provider review. The device is intended for patient use to manage insomnia, depression, or anxiety, potentially benefiting patients by providing a non-pharmacological treatment option with integrated audio features for comfort during sessions.
Clinical Evidence
Bench testing only. Conformance to ES60601-1:2005/(R)2012, IEC 60601-1-2 Edition 4.0, IEC 60601-2-10 Edition 2.1, and IEC 62133 Edition 2.0 was confirmed.
Technological Characteristics
Plastic enclosure; Li-ion battery (3.7V) power source; symmetrical biphasic square wave output; 0-500µA current intensity; 0.5Hz, 1.5Hz, 100Hz frequency selections; Bluetooth LE connectivity for app control; Bluetooth for audio. Electrodes placed on mastoid process. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, and IEC 62133.
Indications for Use
Indicated for treatment of insomnia, depression, or anxiety in patients.
Regulatory Classification
Identification
A cranial electrotherapy stimulator is a prescription device that applies electrical current that is not intended to induce a seizure to a patient's head to treat psychiatric conditions.
Special Controls
In combination with the general controls of the FD&C Act, the virtual reality behavioral therapy device for pain relief is subject to the following special controls:
*Classification.* (1) Class II (special controls) when intended to treat insomnia and/or anxiety. The special controls for this device are:(i) A detailed summary of the clinical testing pertinent to use of the device to demonstrate the effectiveness of the device to treat insomnia and/or anxiety.
(ii) Components of the device that come into human contact must be demonstrated to be biocompatible.
(iii) The device must be designed and tested for electrical safety and electromagnetic compatibility (EMC) in its intended use environment.
(iv) Appropriate software verification, validation, and hazard analysis must be performed.
(v) The technical parameters of the device, including waveform, output mode, pulse duration, frequency, train delivery, maximum charge, and energy, must be fully characterized and verified.
(vi) The labeling for the device must include the following:
(A) The intended use population and the intended use environment;
(B) A warning that patients should be monitored by their physician for signs of worsening;
(C) A warning that instructs patients on how to mitigate the risk of headaches, and what to do should a headache occur;
(D) A warning that instructs patients on how to mitigate the risk of dizziness, and what to do should dizziness occur;
(E) A detailed summary of the clinical testing, which includes the clinical outcomes associated with the use of the device, and a summary of adverse events and complications that occurred with the device;
(F) Instructions for use that address where to place the electrodes, what stimulation parameters to use, and duration and frequency of treatment sessions. This information must be based on the results of clinical studies for the device;
(G) A detailed summary of the device technical parameters, including waveform, output mode, pulse duration, frequency, train delivery, and maximum charge and energy; and
(H) Information on validated methods for reprocessing any reusable components between uses.
(vii) Cranial electrotherapy stimulator devices marketed prior to the effective date of this reclassification must have an amendment submitted to the previously cleared premarket notification (510(k)) demonstrating compliance with these special controls.
(2) Class III (premarket approval) when intended to treat depression.
(c) Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before March 19, 2020, for any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976, or that has, on or before March 19, 2020, been found to be substantially equivalent to any cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section, that was in commercial distribution before May 28, 1976. Any other cranial electrotherapy stimulator device with an intended use described in paragraph (b)(2) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right of it is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 7, 2019
Innovative Neurological Devices LLC Bart Waclawik President 13295 Illinois Street, Suite 312 Carmel, Indiana 46032
Re: K182311
Trade/Device Name: Cervella Regulation Number: 21 CFR 882.5800 Regulation Name: Cranial Electrotherapy Stimulator Regulatory Class: Class III Product Code: JXK Dated: February 1, 2019 Received: February 4, 2019
Dear Bart Waclawik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Pamela D. Scott Digitally signed by Pamela D.
Scott Scott Scott Scott -s Date: 2019.03.07 22:01:53 -5 -05'00 for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K183211
Device Name Cervella
Indications for Use (Describe)
Cervella Cranial Electrotherapy Stimulator (CES) is indicated for treatment of insomnia, depression, or anxiety.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR § 807.92(c), 510(k) summary is provided.
#### l. DATE PREPARED: March 5, 2019
### II. SUBMITTER
INNOVATIVE NEUROLOGICAL DEVICES LLC 13295 Illinois St, Suite 312 Carmel, IN 46032 (855) 413-3300 Phone: Contact Person: Bart Waclawik
### III. OFFICIAL CORRESPONDENCE/CONTACT PERSON
Bart Waclawik President | CEO INNOVATIVE NEUROLOGICAL DEVICES LLC Phone: (855) 413-3300 e-mail: regulatory@cervella.us
#### IV. DEVICE
| Brand Name of Device: | Cervella™ |
|-----------------------|----------------------------------------------------------------------------------------------|
| Common or Usual Name: | Cranial Electrotherapy Stimulator (CES) |
| Classification Name: | Stimulator, Cranial, Electrotherapy, a preamendment<br>Class III device per 21 CFR §882.5800 |
| Regulatory Class: | III |
| Product Code: | JXK |
## V. PREDICATE DEVICES
| Device Trade Name: | CES Ultra™ |
|--------------------|------------------------------|
| Device Company: | Neuro-Fitness, LLC |
| 510(k) Numbers: | K062284 |
| Device Trade Name: | Alpha-Stim CS |
| Device Company: | Electromedical Products Inc. |
| 510(k) Numbers: | K903014 |
{4}------------------------------------------------
## VI. DEVICE DESCRIPTION
Cervella Cranial Electrotherapy Stimulator (CES) is a device that delivers small pulses of electrical current through patient's brain. The stimulator is powered by an internal rechargeable battery which provides a low-level constant current to the cranium via a pair of conductive electrodes placed bilaterally on the mastoid process. The electrodes are incorporated into earpads of stereo over-ear noisecancelling headphones. The patient can use the audio (through a separate dedicated Bluetooth connection) and noise cancelling features of the stereo headphones during treatment (e.g. music listening). The stimulator is controlled via a software application (app) installed on the patient's smart device (e.g. smartphone). The app communicates with the device through a dedicated Bluetooth® LE connection that is independent of the optional Bluetooth connection for audio listening purposes. The patient uses the app to adjust the intensity level, frequency, and duration of the treatment. The app also allows the patient to automatically store the treatment history for review by the patient's healthcare provider.
#### INDICATIONS FOR USE VII.
Cervella is indicated for the treatment of insomnia, depression, or anxiety.
#### SUBSTANTIAL EQUIVALENCE VIII.
The Cervella CES is substantially equivalent with respect to indications for use, stimulation parameters (i.e. current levels, frequencies, pulse width and amplitude), and electrode placement to predicate devices CES Ultra cleared by K062284 and Alpha-Stim CS cleared by K903014.
## IX. PERFORMANCE DATA DEMONSTRATING SUBSTANTIAL EQUIVALENCE
The Cervella device has the same intended use and operating principles, with similar design features, and functional and performance characteristics as the previously-cleared devices. Cervella is designed to comply with relevant federal and international safety and performance standards.
Cervella has undergone standard engineering bench testing to confirm conformance to design specifications as well as independent laboratory testing to insure conformance with applicable mandatory and voluntary medical device safety standards as follows:
{5}------------------------------------------------
- ES60601-1:2005/(R)2012 and A1:2012 ●
- IEC 60601-1-2 Edition 4.0 2014-02
- IEC 60601-2-10 Edition 2.1 2016-04
- IEC 62133 Edition 2.0 2012-1.
Conformance to mandatory and voluntary standards, combined with no changes in the Indications for Use and no change in the fundamental scientific technology demonstrates substantial equivalence to the predicate devices.
| Property | Cervella | CES Ultra | Alpha-Stim CS | Differences and<br>Comments |
|-------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Indications<br>for Use | Cervella is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | CES Ultra is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | Alpha-Stim CS is<br>indicated for the<br>treatment of<br>insomnia,<br>depression, or<br>anxiety. | None |
| Waveform | Symmetrical<br>Biphasic<br>Square Wave | Symmetrical Biphasic<br>Square Wave | Symmetrical<br>Biphasic Square<br>Wave | None |
| Current<br>Intensity<br>Range | 0µA – 500µA<br>adjustable in<br>50µA<br>increments | 0µA — 1500µA<br>continually adjustable | 0µA — 500µA<br>adjustable in 50μA<br>increments | Same range as<br>Alpha-Stim,<br>narrower range<br>than CES Ultra |
| Pulse Width<br>Range | 5ms - 1s | 2ms | 250ms - 1s | Pulse width varies<br>depending on<br>frequency<br>selection |
| Number of<br>electrodes | Two | Two | Two | None |
| Electrode<br>Placement | Head (Mastoid<br>process) | Head (Mastoid<br>process or Earlobes) | Head (Earlobes) | Same placement<br>as CES Ultra |
| Power<br>Source | Li-ion Battery<br>(3.7V) | Alkaline Battery (9V) | Alkaline Battery<br>(3V) | Cervella battery is<br>rechargeable |
# X. SUMMARY OF SIMILARITIES AND DIFFERENCES
{6}------------------------------------------------
# INNOVATIVE NEUROLOGICAL DEVICES LLC | Premarket Notification 510(k) Submission Under 21 CFR § 807.87 for Cervella
| Frequency<br>selections | 0.5Hz, 1.5Hz,<br>100Hz | 100Hz | 0.5Hz, 1.5Hz, 100Hz | Same frequency<br>selection as<br>Alpha-Stim CS |
|-------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|---------------------------------|-------------------------------------|-------------------------------------------------|
| Treatment<br>Range | 10 min – 60 min<br>in 10 min<br>intervals | 30 min, 60min, or<br>continuous | 10, 20, 60 minutes<br>or continuous | Same min-max<br>range as Alpha-<br>Stim |
| Unit<br>Controls | Via Bluetooth-<br>enabled smart<br>device featuring<br>Cervella app | Built into the device | Built into the device | Wireless<br>operation |
| Treatment<br>history<br>logging | Automatic | Not available | Not Available | New feature |
| Ability for<br>patient to<br>use audio<br>and noise<br>cancelling<br>feature of<br>headset<br>during<br>treatment | Optional via<br>dedicated<br>separate<br>Bluetooth<br>connection | Not available | Not Available | New feature |
| Dimensions<br>and weight | 7cmx7cmx2cm | 13.5cmx6.4cmx3.3cm | 9.8cmx6.3cmx2cm | Cervella is slightly<br>smaller |
| Enclosure | Plastic | Plastic | Plastic | None |
# XI. CONCLUSIONS
Cervella CES is substantially equivalent to the listed predicate device without raising any new issues of safety or effectiveness. The new device has the same intended use and operating principles, with similar features, and functional and performance characteristics as the predicates. The device was designed to comply with relevant federal and international safety and performance standards and has been tested to ensure conformity with all applicable medical device safety standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.