K182306 · Chongqing Sunkingdom Medical Instrument Co., Ltd. · HJO · Nov 22, 2018 · Ophthalmic
Device Facts
Record ID
K182306
Device Name
Sunkingdom Slit Lamp
Applicant
Chongqing Sunkingdom Medical Instrument Co., Ltd.
Product Code
HJO · Ophthalmic
Decision Date
Nov 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1850
Device Class
Class 2
Indications for Use
The AC-powered slit lamp bio-microscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affects the structural properties of the anterior eye segment.
Device Story
Sunkingdom Slit Lamp LS-1A/LS-1B are portable, AC-powered biomicroscopes; used by clinicians for anterior eye segment examination. Device projects thin, intense LED light beam into patient eye via control diaphragm; clinician views eye structures through binocular microscope. Components include optical body, movement mechanism, illumination system, and handle. Used in clinical settings to aid diagnosis of ocular disease or trauma. Provides magnified visualization of cornea, iris, lens, and anterior chamber; assists physician in identifying structural abnormalities. Benefits include portable, high-intensity visualization for diagnostic assessment.
Clinical Evidence
Bench testing only. Biocompatibility testing (ISO 10993-5, ISO 10993-10) confirmed no cytotoxicity, skin irritation, or sensitization. Electrical safety and EMC testing (ANSI AAMI IEC 60601-1-2, ANSI AAMI ES60601-1) confirmed compliance. Optical performance testing (ISO 10939) and light hazard protection testing (ISO 15004-2) confirmed safety, with a defined 83-second exposure limit at maximum intensity.
Technological Characteristics
Portable binocular slit-lamp biomicroscope; LED light source; battery-powered (7.4V/680mAh). Features include 10X/16X magnification, variable slit width (0.1-12mm), and horizontal slit angle adjustment (±60°). Materials include patient-contacting forehead rest and chinrest. Complies with ISO 10939, ISO 15004-2, and IEC 60601-1 standards. Sterilization not applicable; device is a reusable ophthalmic instrument.
Indications for Use
Indicated for eye examination of the anterior eye segment (cornea epithelium to posterior capsule) to aid in the diagnosis of diseases or trauma affecting the structural properties of the anterior eye segment in patients requiring such examination.
Regulatory Classification
Identification
An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Special Controls
*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2018
Chongqing Sunkingdom Medical Instrument Co Ltd Joy Tse Sales Director 1012, Block A of China Resource Center No.55 of XieJiaWan,JiuLongPo Chongqing, 400050 Cn
Re: K182306
Trade/Device Name: Sunkingdom Slit Lamp Regulation Number: 21 CFR 886.1850 Regulation Name: AC-Powered Slitlamp Biomicroscope Regulatory Class: Class II Product Code: HJO Dated: August 24, 2018 Received: August 24, 2018
Dear Joy Tse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours.
Bradley S. Cunningham -A
for Malvina B. Evdelman, M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K182306
Device Name Sunkingdom Slit Lamp LS-1A,LS-1B
#### Indications for Use (Describe)
The Chongqing Sunkingdom Slit Lamp LS-1A, LS-1B are AC --- powered slit lamp Biomicroscopes intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma which affects the structural properties of the anterior eye segment.
| Type of Use (Select one or both, as applicable) | <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
{3}------------------------------------------------
# 510(k) Summary
This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92
| Submitter(510K Owner): | Chongqing Sunkingdom Medical Instrument Co.,Ltd<br>35-2,YingTian GuangDian GongGu,CaiJiaGang<br>Industry Zone,BeiBei District,ChongQing,China<br>Contact Person: Joy Tse<br>Phone: +86 23 68643990<br>Fax: +86 23 68102793<br>E-Mail: joy.t@cqsunkingdom.com<br>Registration # 3010610157 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official correspondent: | Amado DiazJr<br>6095 N.W. 167 Street,Unit D-9,<br>Hialeah, FL 33015<br>Phone: 305-542-7556<br>Fax:(305)863-0122 |
| Dateof Preparation: | August 24,2018 |
| Proprietary Name: | Sunkingdom Slit LampLS-1A,LS-1B |
| Common or usual<br>name: | AC- Powered Slit-Lamp Biomicroscope |
| ClassificationName: | Biomicroscope, Slit-Lamp,AC-Powered<br>21 C.F.R. 886.1850<br>Class II |
| Productcode: | HJO |
{4}------------------------------------------------
# Predicate Devices:
| Predicate devices | Suzhou 66 Vision Tech YZ Slit Lamp YZ3 |
|-------------------|----------------------------------------|
| Manufacturer | Suzhou 66 Vision-Tech Co.,Ltd |
| K# | K131711 |
# Device Description:
Sunkingdom slit lamp LS-1A, LS-1B are an AC-powered slit lamp biomicroscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affects the structural properties of the anterior eye segment.
An AC-powered Slit lamp Bio-microscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.
Components: The optical body, movement mechanism, illumination system, portable handle, power supply.
# Intended Use/ Indication for use
The AC-powered slit lamp bio-microscope is intended for use in eye examination of the anterior eye segment, from the cornea epithelium to the posterior capsule. It is used to aid in the diagnosis of diseases or trauma, which affects the structural properties of the anterior eye segment.
# Substantial Equivalence
The Portable slitlamp LS-1A,LS-1B is substantially equivalent to the predicate device Suzhou 66VisionTech YZ Slit LampYZ3 as they have the exactly same indication for use, which is to examine the anterior eye segment for diagnostic purposes. Also Sunkingdom slit lamp use similar technology and perform similar functions to provide the physician with the necessary information to make a diagnosis, the slit lamp project a beam of light into the patient's eve through a control diaphragm.
The Slit Lamp LS-1A,LS-1B has the same intended use and indications for use, technological characteristics, and principles of operation as the previously cleared predicates.
{5}------------------------------------------------
# Comparation Table
| Descriptive<br>information | Proposed Device | Predicate Device | Description of differences<br>and discussion |
|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Chongqing<br>Sunkingdom<br>Medical Instrument<br>Co.,Ltd | 66 Vision Tech<br>Co.,Ltd. | |
| 510(k)Number | | K131711 | |
| Proprietary or<br>Model Name | LS-1A,LS-1B<br>Portable Slit Lamp<br>microscope | YZ3 Portable Slit<br>Lamp microscope | |
| Indications for<br>Use | The Sunkingdom<br>Portable Slit lamp is<br>an AC-powerslit<br>lamp biomicroscope<br>intended for use in<br>eye examination of<br>the anterior eye<br>segment, from the<br>cornea epithelium to<br>the posterior<br>capsule. It is used to<br>aid in the diagnosis<br>of diseases or<br>trauma which<br>affects the structural<br>properties of the<br>anterior eye<br>segment. | Same | |
| Flammability<br>of materials<br>near the light<br>source | None | None | |
| Descriprtive<br>information | Proposed Device | Predicate Device | Description of differences<br>and discussion |
| Maximum<br>temperature of<br>parts of the<br>device held by<br>the operator or<br>accessible to<br>the patient | Maximum<br>temperature of parts<br>of the device held<br>by the operator:<br>-Eyepiece:35 °<br>-Grip:35 °<br>-Slit width control<br>ring:35 °<br>Maximum<br>temperature of parts<br>of the device<br>accessible to the<br>patient:<br>Forehead rest 35 ° | Same | |
| Brightness<br>controls | Maximal<br>Illumination≥25000<br>0Lx | Maximal<br>Illumination≥30000Lx | The brightness controls are by a<br>variable continuous<br>potentiometer. It allow the user<br>to select a convenient<br>brightness to achieve optimum<br>results.<br>The maximum brightness is<br>significantly different.<br>The predicate device YZ3 has<br>a maximumbrightness<br>≥30000LUX, the LS-1A<br>and LS-1B has a maximum<br>brightness of 250000LUX.<br>For light hazard, test according<br>to ISO 15004-2 is proceed, LS-<br>1A and LS-1B are classified<br>into Group 2 intruments.<br>Exposure to light from this<br>instrument when operated at<br>maximum intensity will exceed<br>the safety guideline after 83<br>seconds. Manufacturer will<br>indicate this caution on label to<br>remind the practice.<br>LS-1A and LS-1B are conform<br>to the relevant electronic and<br>optical safety standard of ANSI<br>AAMI IEC60601-1-2,<br>AAMIES60601-1:2005, ISO<br>15004-2, and ISO10939.<br>It's approved to be safe, and no<br>issues of safety and<br>effectiveness. |
| Descriptive<br>information | Proposed Device | Predicate Device | Description of differences<br>and discussion |
| Slit Width | 0.1mm,0.2mm,0.8m<br>m, 1mm, 5mm,<br>12mm | 0 to 12mm<br>continuously<br>adjustable | LS-1A & LS-1B from SK<br>have 6 step of the slit<br>width, the maximum slit<br>width is 12mm which is<br>same like predict device of<br>66 vision, all other step<br>which is in the width area<br>of 66 vision.<br>The differences are not<br>significant, There are no<br>new issues of safety and<br>effectiveness. |
| Slit Length | 1mm,5mm,8mm,12<br>mm | 0.2mm,1mm,2mm,12<br>mm | The differences are not<br>significant. As the<br>illumination field is defined<br>by a circular aperture, the<br>slit length cannot be larger<br>than the slit width. The<br>maximum slit length of SK<br>MED is also the same as<br>predict device which is<br>12mm.<br>There are no new issues of<br>safety and effectiveness. |
| Radial<br>movement of<br>the slit light<br>illumination<br>relative to the<br>microscope<br>axis | Horizontal±60 ° | Horizontal ±30 ° | The predicate device has<br>horizontal±30 °Whereas,th<br>e LS-1A,LS-1B slit lamp<br>has horizontal±60.<br>When examing the conea,<br>the proposed device have<br>larger angle to ensure the<br>clear view of conea<br>formed.<br>SK MED's parameter is<br>better than the predicate<br>device.<br>There are no new issues of<br>safety and effectiveness. |
| Stereo angle | 13 ° | Same | The differences are not<br>significant<br>There are no new issues of<br>safety and effectiveness. |
| Descriprtive<br>information | Proposed Device | Predicate Device | Description of differences<br>and discussion |
| Light sources | LED | LED | LED is a safety light<br>source, it already filter the<br>IR and UV light which is<br>harmful to eye.<br><br>Image: Spectral Power Distribution Characteristics<br><br>Both SK MED and<br>predicate device are using<br>LED light source.<br><br>The differences are not<br>significant , There are no<br>new issues of safety and<br>effectiveness. |
| Pupil-distance | 49-75mm | 50 to 75mm | The differences are not<br>significant, There are no<br>new issues of safety and<br>effectiveness. |
| Eye piece | 12.5X | 10X,16X | The differences are not<br>significant, There are no<br>new issues of safety and<br>effectiveness. |
| Objective | 1X | Same | There are no new issues of<br>safety and effectiveness. |
| Total<br>magnifications | LS-1A:10X<br>LS-1B:10X,16X | 10X,16X(Optional) | The differences are not<br>significant.There are no<br>new issues of safety and<br>effectiveness. |
| Filter | Neutral density,<br>red-free, cobalt blue | Cobalt blue, Red-free,<br>Color Temperature<br>Compensation | The differences are not<br>significant. There are no<br>new issues of safety and<br>effectiveness. |
| Descriprtive<br>information | Proposed Device | Predicate Device | Description of differences<br>and discussion |
| Optical<br>Radiation<br>Emissions | Conform to ISO<br>15004-2 | Conform to ISO<br>15004-2 | The predict device and<br>LS-1A & LS-1B are all<br>conform toISO15004-2.<br>The differences are not<br>significant.<br>There are no new issues of<br>safety and effectiveness. |
| Working<br>distance | 100mm(10X),<br>80mm(16X) | 60mm | The differences are not<br>significant. There are no<br>new issues of safety and<br>effectiveness. |
| Power | 7.4V/680mAh,AA<br>Battery | 7.4V 2200mA Li<br>Battery, Rechargeable | The differences are not<br>significant.<br>Predicate device has 7.4V<br>2200mA Li Battery,<br>whereas the slit lamp<br>LS-1A,LS-1B has<br>7.4V/680mAh,AA Battery.<br>As all slit lamps conform to<br>the relevant standard ISO<br>10939. There are no new<br>issues of safety and<br>effectiveness. |
| Working time | 4 hours | 2.5 hours | The proposed device<br>working longer than<br>predicate device which is a<br>better performance. The<br>differences are not<br>significant.<br>There are no new issues of<br>safety and effectiveness. |
| Net Weight | 890g(with battery) | 900g | There are no new issues of<br>safety and effectiveness. |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Model | | LS-1B | LS-1A |
|-----------------------------------------------------------------------------------------------|-------------------------|--------------------------------------------|---------------|
| Micros<br>c ope<br>Type | Type | Binocular Convergence Biomicroscope | |
| | Eyepiece | 13° | |
| | Magnification | 10X, 16X | 10X |
| | Work distance | 16X:80mm,10X:100m<br>m | 10X:100m<br>m |
| | Pupil distance | 49mm~75mm | |
| | Diopter | -7D~+7D | |
| | Slit width | 0.1mm,0.2mm,0.8mm,1mm,5mm,12mm | |
| | Slit length | 1mm,5mm,8mm,12m<br>m | |
| Illumin<br>a tion<br>System | Slit angle | Horizontal ±60° | |
| | Filters | Neutral density,Redfree(green),Cobalt blue | |
| | Light source | LED | |
| | Brightness | ≥250000Lux | |
| | | | |
| Power<br>Suppl<br>y | LED<br>wor<br>k voltage | DC7.4V,680mAh | |
| | Battery | 7.4V/680mAh,AA Battery | |
| | | | |
| Slit lamp LS-1A,LS-1B are equivalent to each other, the devices all have the same indications | | | |
| for use and the devices are equivalent to the predicate devices. | | | |
# Difference between each model
# Performance Data
The following perfomance data were provided in support of the substantial quivalencedetermination.
# Biocompatibilitytesting
The biocompatibility evaluation for the Sunkingdom Slit Lamp LS-1A,LS-1B wasconductedinaccordancewiththe21CFR58GoodLaboratoryPracticefor Nonclinical Laboratory Studies,ISO 10993-10:2010 "Biological Evaluation of Medical Devices-Part 10:Tests For Irritation And Skin Sensitization",Jul,26,2016,and Standard ISO 10993-5:2009 "Biological Evaluation of Medical Devices-Part 5:Tests For In Vitro Cytotoxicity", Dec,23,2016,as recognized by FDA.The portable slit lamp of testing included the followingtests:
- . VitroCytotoxicity
- . SkinIrritation
- SkinSensitization
The test article Chinrest pat and Forhead band extract did not show potential toxicitytoL-929cells.TheextractofappliedsampleChinrestpatandForehead banddidnotinduceskinirritationinrabbitskin.ChinrestpatandForheadband extract showed no significant evidence of causing skin sensitization in the guineapig.
{11}------------------------------------------------
The material of forhead rest bracket of Portable Slit Lamp LS-1A/LS-1B which is in contact with patient is the same material as the material of test report.
## Electrical safety and electromagnetic compatibility(EMC)
Electrical safety and EMC testing were conducted on the Sunkingdom Slit Lamp LS-1A, LS-1B. The system complies with the "ANSI AAMI IEC60601-1-2:2007/(R)2012,Jun,27,2016, standard for EMC and the ANSI AAMIES60601-1:2005/(R)2012AndA1:2012,Jul,09,2014,standardforsafety.
## Performance testing
The performance testing was conducted on the Sunkingdom Slit Lamp LS-1A.LS-1B. consisting of the optical body, movement mechanism, illumination system and portable handle. The system complies with Standard ISO 10939 "Ophthalmic instruments-Slit-Lamp microscopes",Dec,19,2017, in accordance with Slit lamp guidance - VI References.
# Light Hazard Protection
The light hazard protection for Ophthalmic Instruments was conducted on the Sunkingdom Slit Lamp LS-1A, LS-1B. The system complies with the ISO 15004-2:2007 Ophthalmic instruments -- Fundamental requirements and test methods --Part 2: Light hazard protection .
It is classified into Group 2 instruments according to the requirement of ISO 15004-2, The light emitted from this instrument is potentially hazardous. The longer the duration of exposure, the greater the risk of ocular damage. Exposure to light from this instrument when operated at maximum intensity will exceed the safety guideline after 83 seconds.
# Conclusions
In accordance to 21 CFR 807.92(d) and based on the technical characteristics and the results of the performance tests we conclude that Portable slit lamp LS-1A,LS-1B are substantially equivalent and as safe and effective as the predicate device 66 Vision Tech Co.,Ltd YZ3 Portable Slit Lamp.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.