The Hyper GP (tisilfocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia. The Hyper GP (tisilfocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters. The Hyper GP (tisilfocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters. The Hyper GP (tisilfocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eves that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery. The Hyper GP (tisilfocon A) ORTHOKERATOLOGY contact lenses are indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule. Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
Device Story
Hyper GP is a lathe-cut, rigid gas permeable (RGP) contact lens made of siloxanyl fluoromethacrylate copolymer (tisilfocon A). Lenses are tinted for visibility and optionally include a benzophenone UV-blocking monomer or Tangible Hydra-PEG surface coating (a 30nm hydrophilic PEG-based polymer). Lenses are prescribed by eyecare practitioners for daily wear to correct refractive errors or manage irregular corneas; orthokeratology configurations provide temporary myopia reduction. Lenses are shipped dry or wet in polypropylene cases. Practitioners use the lenses to focus light on the retina; patients benefit from improved visual acuity or corneal management. The device is a passive optical refractive medium; no electronic or software components are involved.
Clinical Evidence
No new clinical data provided. Safety and effectiveness of tisilfocon A material previously established. Non-clinical bench testing included biocompatibility, toxicology, and physicochemical characterization (sessile drop contact angle, light transmission, oxygen permeability). Results demonstrate the device is non-toxic, non-irritating, and physically stable.
Indicated for daily wear for correction of myopia, hyperopia, presbyopia, and/or astigmatism (up to 10D) in aphakic/non-aphakic patients with non-diseased eyes. Also indicated for management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-graft, post-refractive surgery) and orthokeratology for temporary reduction of myopia up to 5.00D.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Rigid Gas Permeable Contact Lenses (tisilfocon A) (K103561)
OPTIMUM GP OK Daily Wear Contact Lens (roflufocon D & E) (K160859)
Optimum GP with HPT (roflufocon C, D, and E) Daily Wear Contact Lenses (K161100)
Submission Summary (Full Text)
{0}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
November 23, 2018
Contamac Ltd. Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln West Linn, OR 97068
Re: K182304
Trade/Device Name: Hyper GP (tisilfocon A) Daily Wear Contact Lens Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid gas permeable contact lens Regulatory Class: Class II Product Code: HQD, MUW Dated: October 24, 2018 Received: October 26, 2018
Dear Bret Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K182304
#### I. SUBMITTER
- October 24th, 2018 Date Prepared:
| Name: | Contamac Ltd. |
|-----------------|---------------------------------------------------------------------------------|
| Address: | Carlton House<br>Shire Hill<br>Saffron Walden<br>Essex CB11 3AU |
| Contact Person: | Robert McGregor<br>Managing Director |
| Phone number: | 01799 514800 |
| Consultant: | Bret Andre<br>EyeReg Consulting, Inc.<br>6119 Canter Ln.<br>West Linn, OR 97068 |
| Phone number: | (503) 372-5226 |
#### II. DEVICE
| Trade Name: | Hyper GP (tisilfocon A) Daily Wear Contact Lens |
|-------------------------|-----------------------------------------------------|
| Common<br>Name: | Daily wear rigid gas permeable contact lens |
| Classification<br>Name: | Rigid gas permeable contact lens. (21 CFR 886.5916) |
| Regulatory<br>Class: | Class II |
| Product Code: | HQD; MUW |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is substantially equivalent to the following predicate devices:
- . "Rigid Gas Permeable Contact Lenses (tisilfocon A)" By Menicon Co., Ltd. 510(k) number; K103561
- . "OPTIMUM GP OK Daily Wear Contact Lens (roflufocon D & E)" by Contamac Ltd. 510(k) number; K160859
- . "Optimum GP with HPT (roflufocon C. D. and E) Daily Wear Contact Lenses" by Contamac Ltd. 510(k) number; K161100
#### IV. DEVICE DESCRIPTION
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is manufactured from a machine latheable rigid gas permeable material composed of siloxanyl fluoromethacrylate copolymer that is tinted for visibility and available with or without an ultraviolet (UV) light absorber. The lenses may be plasma treated during the manufacturing process.
In the Hyper GP (tisilfocon A) Daily Wear Contact Lens with UV Blocker, a Benzophenone UV blocking monomer is used to block >99% of UV radiation in the UVB range (280nm = 315nm) and >85% in the UVA range (316 - 380nm).
The Hyper GP (tisilfocon A) Daily Wear Contact Lens incorporates a visibility tint to make the lens more visible for handling. The tinted lenses contain one or more of the following color additives: D&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2 and D&C Red No. 17.
The Hyper GP (tisilfocon A) Daily Wear Contact Lens may be prescribed in a daily wear orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in nondiseased eyes.
The Hyper GP (tisilfocon A) Daily Wear Contact Lens may be optionally treated to incorporate Tangible™ Hydra-PEG-which is a thin polyethylene glycol (PEG)-based polymer that is covalently bonded to the surface of the contact lens and is designed to enhance the surface properties of the contact lens while retaining the mechanical properties of the underlying material. When treated with Tangible™ Hydra-PEG, the underlying material (tisilfocon A) is encapsulated in a thin layer of polymer that results in measurable improvement of wettability (sessile drop contact angle) compared to untreated lenses. The resulting layer is hydrophilic and approximately 30nm in thickness. The following table depicts the contact angle of the coated vs. uncoated lenses:
{5}------------------------------------------------
| | Hyper GP tisilfocon A | |
|------------------------------------------------------|-----------------------|-------------------------------|
| | Uncoated | Tangible™ Hydra-PEG<br>Coated |
| Average Sessile Drop Contact Angle<br>(degrees) n=30 | 106.2 | 37.34 |
| Standard Deviation | 5.31 | 5.30 |
The Hyper GP (tisilfocon A) Daily Wear Contact Lens may be packaged and shipped "dry" or "wet" in a polypropylene contact lens case. The primary container for shipping the Hyper GP (tisilfocon A) Daily Wear Contact Lens is the Poly Vial Contact Lens Case. When shipped "wet", the Hyper GP (tisilfocon A) Daily Wear Contact Lens may be packaged and shipped in the Unique pH contact lens care system by Menicon Co., Ltd. The active ingredients in Unique pH solution are Edetate Disodium 0.01% and Polyquaternium 10.0011%.
The physical properties of the Hyper GP (tisilfocon A) Daily Wear Contact Lens is as follows:
| | Hyper GP (tisilfocon A) | Hyper GP (tisilfocon A) with<br>TangibleTM Hydra-PEG |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| Refractive Index (dry) | 1.4378 | 1.4398 |
| Light Transmission (@380-780nm) | 91.694 | 93.427 |
| UVA Transmission (@316-380nm) | 14.241 | 16.105 |
| UVB Transmission (@280-315nm) | 0.011 | 0.020 |
| Oxygen Permeability<br>(Dk)<br>ISO/FATT Method | $180 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @<br>35°C) | $180 \times 10^{-11}$ (cm²/sec) (ml O₂/ml x mm Hg @<br>35°C) |
| Visitint lenses contain one or more of the<br>following color additives conforming to:<br>21 CFR Part 73 & 74, Subpart D | D&C Green No. 6, C.I. Solvent<br>Yellow No. 18, D&C Violet No. 2 and<br>D&C Red No. 17 | D&C Green No. 6, C.I. Solvent<br>Yellow No. 18, D&C Violet No. 2<br>and D&C Red No. 17 |
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is available in the Spherical, Toric, Multifocal/Bifocal, Irregular Cornea (Scleral) design configurations, within the following lens parameters:
| Parameter | Range | Tolerance |
|-------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve | 4.00mm to 12.00mm | ± 0.05 mm |
| Center Thickness | Varies | ± 0.02 mm |
| Chord Diameter | 7.00mm to 22.00mm | ± 0.10mm |
| Spherical Power | -30.00 D to +30.00 D<br>(in 0.25D steps) | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
| Cylindrical Power | Up to -10.00 D<br>(in 0.25 D steps) | ± 0.25 (0 to = 2D)<br ± 0.37 (2 to = 4D)<br ± 0.50 (over 4D) |
| Cylindrical Axis | 1° to 180° (in 1° steps) | ± 5° |
| Multifocal Power | +1.00 D to 4.00 D<br>(in 0.25 D steps) | ± 0.25D |
{6}------------------------------------------------
The Hyper GP (tisilfocon A) Daily Wear Contact Lens for daily wear orthokeratology is available in following lens parameters:
| Parameter | Range | Tolerance |
|-------------------|-----------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve (BC) | 4.0mm to 12.00mm | ± 0.05 mm |
| Center Thickness | 0.10mm to 0.70mm | ±0.02 mm |
| Diameter | 6.50mm to 11.50mm | ±0.10mm |
| Secondary Curves | 0.10mm to 2.00mm (flatter or steeper than BC) | ±0.10mm |
| Peripheral Curves | 0.10mm to 2.00mm (flatter or steeper than BC) | ±0.10mm |
| Spherical Power | -10.00 D to +3.00 D<br>(in 0.25D steps) | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
#### V. INDICATIONS FOR USE
The Hyper GP (tisilfocon A) SPHERICAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia.
The Hyper GP (tisilfocon A) TORIC Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with nondiseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10.00 diopters.
The Hyper GP (tisilfocon A) MULTIFOCAL/BIFOCAL Rigid Gas Permeable (RGP) Contact Lens is indicated for daily wear for the correction of refractive error in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The Hyper GP (tisilfocon A) IRREGULAR CORNEA Daily Wear Contact Lens may be prescribed in otherwise non-diseased eves that require a rigid gas permeable lens for the management of irregular corneal conditions such as; keratoconus, pellucid marginal degeneration or following penetrating keratoplasty or refractive (e.g. LASIK) surgery.
The Hyper GP (tisilfocon A) ORTHOKERATOLOGY contact lenses are indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system.
{7}------------------------------------------------
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE VI. DEVICE
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is substantially equivalent to the Rigid Gas Permeable Contact Lenses (tisilfocon A) (cleared under K103561) in terms of the following:
- Intended use - daily wear contact lenses
- 트 Indications for use
- Actions
- 트 USAN - contact lens material composition
- I Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- I Production method - lathe cut
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is substantially equivalent to the OPTIMUM GP OK Daily Wear Contact Lens (roflufocon D & E) (cleared under K160859) in terms of the following:
- I Intended use - daily wear contact lenses
- 트 Indications for use - daily wear orthokeratology
- I Contact lens design - orthokeratology
- 트 Actions
- I Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- 트 Production method - lathe cut
The Hyper GP (tisilfocon A) Daily Wear Contact Lens is substantially equivalent to the Optimum GP with HPT (roflufocon C, D, and E) Daily Wear Contact Lenses (cleared under K161100) in terms of the following:
- I Intended use – daily wear contact lenses
- 트 May be surface coated with Tangible Hydra-PEG
- 트 Actions
- 트 Classification - Lenses, Rigid Gas Permeable, Daily Wear (21 CFR 886.5916)
- 트 FDA material group - group # 3 fluoro silicone acrylate
- I Production method - lathe cut
{8}------------------------------------------------
The following matrix illustrates the production method, lens function and material characteristics of the Hyper GP (tisilfocon A) Daily Wear Contact Lens, as well as the predicate devices.
| | Hyper GP<br>(tisilfocon A)<br>New Device | Menicon<br>(tisilfocon A)<br>Predicate Device | Optimum GP with<br>HPT<br>(roflufocon C, D, E)<br>Predicate Device | OPTIMUM GP OK<br>(Orthokeratology)<br>(roflufocon D & E)<br>Predicate Device |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Indicated for daily wear for the<br>correction of visual acuity in<br>aphakic and not aphakic<br>persons with non-diseased<br>eyes with myopia or hyperopia<br>and/or presbyopia. The lens<br>may also be prescribed for<br>management of irregular<br>corneal conditions such as<br>keratoconus and post graft<br>fitting. Indicated for daily<br>wear in an orthokeratology<br>fitting program for the<br>temporary reduction of myopia<br>of up to 5.00 diopters in non-<br>diseased eyes. To maintain the<br>orthokeratology effect of<br>myopia reduction, lens wear<br>must be continued on a<br>prescribed wearing schedule. | Indicated for daily wear for<br>the correction of<br>refractive error (myopia,<br>hyperopia, presbyopia<br>and/or astigmatism) in<br>aphakic and<br>non-aphakic persons with<br>non-diseased eyes. The<br>lenses may be prescribed<br>for daily wear in otherwise<br>non-diseased eyes that<br>require a rigid contact lens<br>for the management of<br>irregular corneal<br>conditions such as<br>keratoconus, pellucid<br>marginal degeneration, or<br>following penetrating<br>keratoplasty or<br>refractive (e.g., LASIK)<br>surgery. | Indicated for daily wear for<br>the correction of visual<br>acuity in aphakic and not<br>aphakic persons with non-<br>diseased eyes with myopia<br>or hyperopia and/or<br>presbyopia. The lens may<br>also be prescribed for<br>management of irregular<br>corneal conditions such as<br>keratoconus and post graft<br>fitting. | Indicated for daily wear in<br>an orthokeratology fitting<br>program for the temporary<br>reduction of myopia of up<br>to 5.00 diopters in non-<br>diseased eyes. To maintain<br>the orthokeratology effect<br>of myopia reduction, lens<br>wear must be continued on<br>a prescribed wearing<br>schedule. |
| Functionality | The contact lenses act as a<br>refractive medium that focus<br>light rays from near and distant<br>objects on the retina | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina |
| Indications | Daily Wear | Daily Wear | Daily Wear | Daily Wear |
| Production<br>Method | Lathe-Cut, custom<br>manufactured | Lathe-Cut, custom<br>manufactured | Lathe-Cut, custom<br>manufactured | Lathe-Cut, custom<br>manufactured |
| USAN name | tisilfocon A | tisilfocon A | roflufocon C, D, E | roflufocon D, E |
| Water Content<br>(%) | <1% | <1% | <1% | <1% |
| Specific Gravity | 1.20 | 1.20 | Roflufocon C: 1.178<br>Roflufocon D: 1.166<br>Roflufocon E: 1.155 | Roflufocon D: 1.166<br>Roflufocon E: 1.155 |
| Oxygen<br>Permeability<br>(cm²/sec) (ml O¸/ml x mm Hg<br>@ 35°C) | 180 x10-11 | 160 x10-11 | Roflufocon C: 65x10-11<br>Roflufocon D: 100x10-11<br>Roflufocon E: 125x10-11 | Roflufocon D: 100x10-11<br>Roflufocon E: 125x10-11 |
| UV Blocker | Yes, Optional | Yes | Yes. Optional | Yes, Optional |
| Plasma<br>Treatment | Yes, Optional | Yes | Yes, Optional | Yes. Optional |
| Hydra-PEG<br>Surface Coating | Yes, Optional | No | Yes, Optional | No |
### Substantial Equivalence Matrix
{9}------------------------------------------------
#### VII. PERFORMANCE DATA
### ~ Non-Clinical Studies ~
A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the Hyper GP (tisilfocon A) Daily Wear Contact Lens. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols. Test results of the nonclinical testing on the Hyper GP (tisilfocon A) Daily Wear Contact Lens demonstrate that:
- The finished lenses (uncoated and coated with Tangible™ Hydra-PEG) are not toxic and ● not irritating,
- The physicochemical, mechanical, and optical properties of the Tangible™ Hydra-PEG ● coated and uncoated lenses are substantially equivalent to the predicate device.
- Bioburden levels are below the acceptance criteria (<100 cfu/lens) initially and following 30 days of storage in Unique pH at ambient temperatures,
- Physical parameters are stable following 30 days of storage in Menicon Unique pH at ambient temperatures, and
- The Tangible "10 HydraPEG coated tisilfocon A lenses are physically compatible with currently marketed care solutions.
### ~ Clinical Studies ~
Clinical performance data to demonstrate the safety and effectiveness of contact lenses manufactured from tisilfocon A has been previously addressed.
### VIII. CONCLUSIONS
### Substantial Equivalence
Information presented in this Premarket Notification establishes that Hyper GP (tisilfocon A) Daily Wear Contact Lenses are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.
### Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.