VORTRAN APM-Plus

K182292 · Vortran Medical Technology 1, Inc. · CAP · Sep 18, 2019 · Anesthesiology

Device Facts

Record IDK182292
Device NameVORTRAN APM-Plus
ApplicantVortran Medical Technology 1, Inc.
Product CodeCAP · Anesthesiology
Decision DateSep 18, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.2600
Device ClassClass 2
AttributesPediatric

Indications for Use

The VORTRAN® APM-Plus is used to measure airway pressure for patients utilizing positive pressure devices where monitoring is desired.

Device Story

VORTRAN APM-Plus is a portable, battery-operated airway pressure monitor for patients ≥ 10 kg. It connects to positive-pressure devices (resuscitators, CPAP) via pneumatic tubing and a solid-state pressure transducer. The device operates in two user-selectable modes: PAP Mode (displays real-time pressure) and Resuscitator Mode (displays PIP, PEEP, respiratory rate, I-time, E-time, and I:E ratio). A microprocessor-controlled algorithm continuously monitors pressure data against user-defined alarm limits. If limits are exceeded, the device triggers visual (flashing LED, LCD) and audible alarms. Used by healthcare professionals in hospitals and sub-acute settings, the device provides real-time monitoring and backup safety, alerting clinicians to abnormal pressure conditions or non-cycling states, thereby supporting clinical decision-making and patient safety during respiratory support.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), alarm performance (IEC 60601-1-8), software validation (IEC 62304), risk management (ISO 14971), and accelerated shelf-life testing.

Technological Characteristics

ABS plastic enclosure; solid-state pressure transducer; 2x AA battery power. Dimensions: 110x64x33 mm; Weight: 150g. Connectivity: pneumatic tubing with luer/tee adapters. Software: microprocessor-based, moderate level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 62304, ISO 14971.

Indications for Use

Indicated for adult or pediatric patients ≥ 10 kg utilizing positive pressure devices (e.g., resuscitation bags, automatic resuscitators, CPAP). Used for monitoring pressure and/or rate, providing alarms when conditions exceed user-selected limits. For professional use only in hospitals and sub-acute institutions.

Regulatory Classification

Identification

An airway pressure monitor is a device used to measure the pressure in a patient's upper airway. The device may include a pressure gauge and an alarm.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 18, 2019 VORTRAN Medical Technology 1, Inc. James Lee Executive VP & COO 21 Goldenland Court, Suite 100 Sacramento, California 95834 Re: K182292 Trade/Device Name: VORTRAN APM-Plus Regulation Number: 21 CFR 868.2600 Regulation Name: Airway Pressure Monitor Regulatory Class: Class II Product Code: CAP Dated: August 30, 2019 Received: September 16, 2019 Dear James Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, James Lee Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K182292 Device Name VORTRAN® APM-Plus ### Indications for Use (Describe) The VORTRAN® APM-Plus is a multiple-patient-use device intended to monitor a positive-pressure device that connects to an adult or pediatric patient ≥ 10 kg. The device alarms when the conditions (pressure and/or rate, as applicable) are outside of the user selected high and low alarm limits. In PAP Mode, it displays only real-time breasure; in Resuscitator Mode, it displays ventilation parameters in addition to the pressure information. It may be used with positive pressure devices which do not include pressure measurement capabilities, e.g. resuscitation bags, automatic resuscitators, and CPAP, or as an independent backup pressure monitor for devices with pressure measurement capability. The device is designed to be portable and can be used in stationary or portable situations in hospitals and sub-acute institutions. For professional use only. | Type of Use (Select <i>one</i> or <i>both</i> , as applicable) | | |----------------------------------------------------------------|--| |----------------------------------------------------------------|--| X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### I. SUBMITTER VORTRAN® Medical Technology 1, Inc. 21 Goldenland Court, Suite 100 Sacramento, CA 95834 USA Phone: (800) 434-4034 Fax: (916) 648-9751 Contact Person: James Lee Contact Title: Executive Vice President & COO Date Prepared: August 17, 2018 #### II. DEVICE Name of Device: VORTRAN® APM-Plus Common or Usual Name: Airway Pressure Monitor Classification Name: Monitor, Airway Pressure (Includes Gauge And /or Alarm) Regulation Number: 21 CFR 868.2600 Regulatory Class: II Product Code: CAP Classification Advisory Committee: Anesthesiology #### PREDICATE DEVICE III. | K992101 | Criterion 40 Airway Pressure Monitor (Primary) | |---------|------------------------------------------------| | K103639 | VORTRAN® Airway Pressure Monitor (Reference) | #### DEVICE DESCRIPTION/PRINCIPLE OF OPERATION IV. The VORTRAN® APM-Plus is a battery (2 x AA) operated, microprocessor controlled, portable, self-contained device designed for monitoring a positive airway-pressure device such as a resuscitator or CPAP. The VORTRAN® APM-Plus is housed in a plastic enclosure. It connects to the positive-pressure device under monitor via pneumatic pressure tubing and receives the pressure signals using its solid-state pressure transducer. The pressure signals are used to calculate the related characteristics which are displayed on the LCD screen. The values are used to trigger alarms that alert the user of any abnormal pressure conditions. The software algorithm constantly monitors the data and if/when the pressure {4}------------------------------------------------ characteristics are outside of the user-set limits, an alarm will be activated. The alarms are both visual and audible. For ease of use, the VORTRAN® APM-Pus software provides two modes of operations: Resuscitator Mode and PAP Mode. When the device is turned on, the user can select one of the two modes. In each mode, the pressure data shown is suitable for the device under monitoring: - . Monitoring/Alarms, Resuscitator Mode: The VORTRAN® APM-Plus will display PIP, PEEP, Respiratory Rate, Inspiratory and Expiratory Time, and I:E Ratio. When a connected pressure-cycled device stops cycling for a user-set time or the pressure falls under the user-set limit, the VORTRAN® APM-Plus will activate the NON-CYCLING ALARM or LOW PEEP ALARM, respectively. In addition, when respiratory rate or the pressure exceeds their user-set limits, the HIGH RATE or HIGH PIP alarm will activate, respectively. - . Monitoring/Alarms, PAP Mode: The VORTRAN® APM-Plus will display real-time pressure. When a connected constant-pressure device falls outside of the user-set range, the HIGH PRESSURE or LOW PRESSURE ALARM will activate. The VORTRAN® APM-Plus has a flashing red LED and an audible sound, while simultaneously displaying the alarm condition on the LCD. The accessories were all cleared with the reference device's 510(k) submission K103639 (VORTRAN® Airway Pressure Monitor) and include: - Pressure tubing for connecting the VORTRAN® APM-Plus to a breathing circuit . - A hydrophobic filter for protecting the device from any condensate that may pass ● through the breathing circuit - . A luer connector for connecting the pressure tubing to the hydrophobic filter - . A tee adapter for connecting the pressure tubing in-line with a breathing circuit that does not have its own outlet #### V. INTENDED USE The VORTRAN® APM-Plus is used to measure airway pressure for patients utilizing positive pressure devices where monitoring is desired. {5}------------------------------------------------ #### VI. INDICATIONS FOR USE The VORTRAN® APM-Plus is a multiple-patient-use device intended to monitor a positive-pressure device that connects to an adult or pediatric patient ≥ 10 kg. The device alarms when the conditions (pressure and/or rate, as applicable) are outside of the user selected high and low alarm limits. In PAP Mode, it displays only real-time breathing circuit pressure; in Resuscitator Mode, it displays ventilation parameters in addition to the pressure information. It may be used with positive pressure devices which do not include pressure measurement capabilities, e.g. resuscitation bags, automatic resuscitators, and CPAP, or as an independent backup pressure monitor for devices with pressure measurement capability. The device is designed to be portable and can be used in stationary or portable situations in hospitals and sub-acute institutions. For professional use only. ### DIFFERENCES IN INDICATIONS FOR USE TO THE VII. PRIMARY PREDICATE DEVICE The VORTRAN® APM-Plus and its primary predicate device have the following differences in their Indications for Use: - . The Criterion 40 Airway Pressure Monitor displays real-time pressure as well as peak pressure. The VORTRAN® APM-Plus, when in PAP Mode, only displays the real-time pressure. - In Resuscitator Mode, the VORTRAN® APM-Plus displays Peak Inspiratory Pressure (PIP), Positive End-Expiratory Pressure (PEEP), Breath Rate, Inspiratory Time (I-Time), Expiratory Time (E-Time), and the ratio of Inspiratory Time to Expiratory Time (I:E Ratio), while the Criterion 40 only displays real-time pressure and peak pressure. These technological differences do not raise different questions of safety and effectiveness, as evidence by the substantial equivalence comparison to the reference predicate device, the VORTRAN® Airway Pressure Monitor, which has all the aforementioned parameters. {6}------------------------------------------------ ## VIII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The following tables provide a comparison between device characteristics: | General Characteristics: | | | | |----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------| | | NEW DEVICE | PRIMARY | REFERENCE | | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway<br>Pressure Monitor | VORTRAN® Airway<br>Pressure Monitor | | Patient Population | Adult or pediatric patients (10 kg and above) utilizing positive pressure devices | Any patient utilizing positive pressure devices<br>and the clinician desires to have pressure monitoring. | Adults 40 kg and above<br>(Model 3900) and<br>Pediatrics 10 kg and above<br>(Model 3800) | | Patient<br>Interface/Method of<br>Deployment | Connected via pneumatic tubing (of the connection kit) to a tee fitting or Luer outlet within the breathing circuit of a resuscitator, CPAP, or other positive pressure device. | Airway adapter placed in the circuit or connection to a face mask | Connects to any breathing circuit that is connected to a patient. | | Environment of Care | Hospitals and sub-acute institutions;<br>designed to be used in stationary or portable situations | Hospitals, sub-acute institutions, home care settings; stationary and intra-institution transport only. | To be used by properly trained personnel in an environment in which the clinician desires to have pressure monitoring. | | Sterility | Non-sterile (device and accessories) | Non-sterile | Non-sterile | | Type of Use | Prescription Use (21CFR801 Subpart D) | Not specified | Prescription Use<br>(21CFR801 Subpart D) | General Characteristics: Physical Characteristics: | Physical Characteristics. | | | | |---------------------------|----------------------------------------------------|-------------------------------------------------------|-----------------------------------------------------| | | NEW DEVICE | PRIMARY | REFERENCE | | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway Pressure Monitor | VORTRAN® Airway Pressure<br>Monitor | | Size | 110 mm x 64 mm 33 mm (4.3 in<br>x 2.5 in x 1.3 in) | 198 mm x 94 mm x 188 mm (7.9 in x<br>3.7 in x 7.4 in) | 81 mm x 56 mm x 48 mm (3.2 in x<br>2.2 in x 1.9 in) | | Weight | 150 g (with 2 AA batteries) | 680 g | 125 g (with 9V battery) | | Enclosure<br>Construction | ABS Plastic | Unspecified Plastic | ABS Plastic | {7}------------------------------------------------ Technological/Principles of Operation Characteristics: | | NEW DEVICE | PRIMARY | REFERENCE | |-------------------------|--------------------------------------|----------------------------------------------------|---------------------------------------| | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway<br>Pressure Monitor | VORTRAN® Airway<br>Pressure Monitor | | Method of<br>Connection | Female Luer connection | Female Luer<br>connection | Female Luer<br>connection | | Alarms | YES | YES | YES | | Measurement<br>Device | Solid state pressure sensor | Solid state pressure<br>sensor | Solid state pressure<br>sensor | | Output | Digital readout of pressures | Digital readout of<br>pressures | Digital readout of<br>pressures | | Control Method | Internal microprocessor | Internal<br>microprocessor | Internal<br>microprocessor | | Power Type | DC input (2 x AA alkaline batteries) | AC Mains input (and<br>DC input backup<br>battery) | DC input (1 x 9V<br>alkaline battery) | Performance Specifications: | | NEW DEVICE | PRIMARY | REFERENCE | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------| | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway<br>Pressure Monitor | VORTRAN® Airway<br>Pressure Monitor | | Connected Devices | Positive pressure devices (including<br>pressure-cycled devices) | Positive pressure<br>devices (e.g. bubble<br>CPAP, resuscitators) | Pressure-cycled devices<br>(e.g. resuscitators) | | Displayed<br>Measurements | Resuscitator Mode: PIP, PEEP,<br>Respiratory Rate, Inspiratory Time,<br>Expiratory Time, Inspiratory: Expiratory<br>Ratio<br>PAP Mode: Real-time Pressure | Real-time Pressure,<br>Peak Pressure | PIP, PEEP, Respiratory<br>Rate, Inspiratory Time,<br>Expiratory Time,<br>Inspiratory: Expiratory<br>Ratio | {8}------------------------------------------------ | Performance Specifications, continued: | | | | |----------------------------------------|---------------------------------------------------------|---------------------------------------------------------------------|-------------------------------------| | | NEW DEVICE | PRIMARY | REFERENCE | | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway<br>Pressure Monitor | VORTRAN® Airway Pressure<br>Monitor | | Pressure Display<br>Range | 0 to 55 cm-H2O | 0 to 99.5 cm-H2O | 0 to 50 cm-H2O | | Pressure Display<br>Accuracy | ± 1 cm-H2O | ± (1 cm-H2O + 3% of<br>reading) rounded up to<br>nearest 0.5 cm-H2O | ± 10% of reading | | Pressure Display<br>Resolution | Resuscitator Mode: 1 cm-<br>H2O<br>PAP Mode: 0.1 cm-H2O | 0.5 cm-H2O | 1 cm-H2O | | Respiratory Rate<br>Display Range | 0 to 99 BPM | N/A | 0 to 55 BPM | | Respiratory Display<br>Rate Accuracy | ± 10% | N/A | ± 10% | | Respiratory Rate<br>Display Resolution | 1 Breath per Minute | N/A | 1 Breath per Minute | | I-Time & E-Time<br>Display Range | 0 to 9.9 seconds | N/A | 0 to 9.9 seconds | | I-Time & E-Time<br>Display Accuracy | ± 10% | N/A | ± 10% | | I-Time & E-Time<br>Display Resolution | 0.1 second | N/A | 0.1 second | | I:E Ratio Display<br>Range | 1:0.0 to 1:9.9 | N/A | 1:0.0 to 1:9.9 | | I:E Ratio Display<br>Accuracy | ± 10% | N/A | ± 10% | ### Performance Specifications, continued: {9}------------------------------------------------ | | NEW DEVICE | PRIMARY | REFERENCE | |------------------------------------------------|--------------------------------------------|---------------------------------------------|------------------------------------------------------------------------------------------------| | Device Name | VORTRAN® APM-Plus | Criterion 40 Airway<br>Pressure Monitor | VORTRAN® Airway Pressure<br>Monitor | | I:E Ratio Display<br>Resolution | 0.1 | N/A | 0.1 | | Alarm Delay | Yes, Non-Cycling Alarm<br>Only: 1-20 s | Yes: 1-20 s | Yes, Non-Cycling Alarm Only<br>- 15 s for Adult Model 3900, 8<br>s for Pediatric Model 3800 | | High Pressure<br>Alarm Range and<br>Resolution | 1 to 50 cm-H2O,<br>Resolution of 1 cm-H2O | 5 to 99 cm-H2O,<br>Resolution of 1 cm-H2O | Fixed at 50 cm-H2O for Adult<br>Model 3900,<br>Fixed at 35 cm-H2O for<br>Pediatric Model 3800 | | Low Pressure<br>Alarm Range and<br>Resolution | 1 to 15 cm-H2O,<br>Resolution of 1 cm-H2O | 1 to 20 cm-H2O,<br>Resolution of 1 cm-H2O | N/A | | High Rate Alarm<br>Range and<br>Resolution | 12 to 99 BPM, Resolution<br>of 1 BPM | N/A | Fixed at 50 BPM for Adult<br>Model 3900,<br>Fixed at 55 BPM for Pediatric<br>Model | | Non-Cycling Alarm<br>Range and<br>Resolution | 1 to 20 seconds,<br>Resolution of 1 second | N/A | Fixed at 15 seconds for Adult<br>Model 3900,<br>Fixed at 8 seconds for Pediatric<br>Model 3800 | | Low Battery Alarm | YES | YES (when using internal<br>backup battery) | YES | | Battery Life | 164 Hours (Worst-Case) | Unknown | 27 Hours (Worst-Case) | | Calibration | None | Unknown | None | ### Performance Specifications, continued: {10}------------------------------------------------ #### IX. PERFORMANCE DATA The following performance data were provided in support of the substantial equivalence determination: #### Biocompatibility Testing . The patient contacting materials, considered to be in the patient's gas flow pathway, include the accessory kit (pressure tubing, hydrophobic filter, luer connector, and tee adapter) and solid-state pressure transducer. The materials in said components of the VORTRAN® APM-Plus, in their final finished form are identical to the accessory kit and solid-state pressure transducer of the cleared reference device in K103639, the VORTRAN® Airway Pressure Monitor, in formulation, processing, and sterilization, and geometry and no other chemicals have been added (e.g., plasticizers, fillers, additives, cleaning agents, mold release agents). #### . Electrical Safety and Electromagnetic Compatibility (EMC) The VORTRAN® APM-Plus was evaluated as medical electrical equipment, under the classification of "Internally Powered Device" and was found to be compliant with the following IEC standards (and applicable IEC collateral standards, as required by ISO 80601-2-55): - Electrical safety testing per IEC 60601-1 o - Electromagnetic compatibility testing per IEC 60601-1-2 O - Usability Study as per IEC 60601-1-6 O - O Alarms designed as per IEC 60601-1-8 - Software Development Process as per IEC 62304 о #### Software Verification and Validation Testing ● The embedded software was developed in accordance with FDA guidelines for a MODERATE Level of Concern. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software lifecycle process was assessed to be in accordance with IEC 62304. #### . Risk Management Risk Management and FMEA was assessed to (and found to be in compliance with) ISO 14971. - . Shelf Life An accelerated shelf life test was performed and met the acceptance criteria for a simulated age of 1 year. {11}------------------------------------------------ - . Clinical Studies None #### X. CONCLUSION The VORTRAN® APM-Plus's capabilities have been tested and compared to the predicate devices and the results support the substantial equivalence of the VORTRAN® APM-Plus.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...