← Product Code [NQQ](/productcode/NQQ) · K182261

# NvisionVLE Imaging System, NvisionVLE Optical Probe (K182261)

_Ninepoint Medical, Inc. · NQQ · Oct 16, 2018 · Radiology · SESU_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182261

## Device Facts

- **Applicant:** Ninepoint Medical, Inc.
- **Product Code:** [NQQ](/productcode/NQQ.md)
- **Decision Date:** Oct 16, 2018
- **Decision:** SESU
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The NvisionVLE Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross sectional, real-time depth visualization. The Nvision VLE Imaging System is intended to provide an image of the tissue microstructure. The safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.

## Device Story

NvisionVLE Imaging System is a high-resolution volumetric imaging system using swept-source optical coherence tomography (SS-OCT), also known as Optical Frequency Domain Imaging (OFDI). The system captures tissue microstructure data via a scanning optical probe inserted through an endoscope working channel. The console processes reflected light signals to reconstruct real-time, cross-sectional imagery. Components include a mobile console with touchscreen interfaces, proprietary software, a single-use sterile optical probe, an inflation system, and a probe lock. Used in clinical settings by physicians; the system provides visual feedback to assist in evaluating tissue microstructure. The device does not provide diagnostic analysis for specific diseases. The current enhancement introduces balloon-less, low-profile optical probes in various focal lengths to improve procedural utility.

## Clinical Evidence

No clinical data. Bench testing only. Performance testing confirmed proper image acquisition, lateral resolution, back reflection, transmission rates, and mechanical integrity (tensile testing) of the low-profile probes. Compliance with voluntary standards including ISO 10993-7, ISO 11135:2014, IEC 60601-1, IEC 60601-1-2, ISO 62304, IEC 60601-2-18, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-6, and IEC 62366-1 was demonstrated.

## Technological Characteristics

SS-OCT/OFDI imaging system. Probe assembly: fiber optic probe in flexible stainless steel torque shaft; distal optics in stainless steel hypotube. Guide sheath: PET balloon/nylon shaft or balloon-less low-profile configuration. Connectivity: standard fiber optic/catheter connectors to console. Sterilization: Ethylene oxide (ANSI/AAMI/ISO 10993-7). Software: ISO 62304 compliant. Energy: Laser (IEC 60825-1).

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- NvisionVLE Imaging System ([K143678](/device/K143678.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure. To the right of the symbol is a blue square containing the acronym "FDA" in white letters. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.

October 16, 2018

NinePoint Medical, Inc. Eman Namati President & CEO 12 Oak Park Drive Bedford, Massachusetts 01730

Re: K182261

Trade/Device Name: NvisionVLE Imaging System, NvisionVLE Optical Probe Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: NOO Dated: August 20, 2018 Received: August 21, 2018

Dear Eman Namati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at

https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpm/pmn.cfm_identifies_combination_product submissions. The general controls provisions of the Act include requirements for annual registration, isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou. however, that device labeling must be truthful and not misleading.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warning section of the device's labeling:

{1}------------------------------------------------

- The Nyision VLE Imaging System is intended to provide an image of the tissue microstructure. The 1. safety and effectiveness of this device for diagnostic analysis (i.e. differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.
Furthermore, the indication for use "The Nvision System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross sectional, realtime depth visualization" must be prominently displayed in all abeling, including pouch box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to the market. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification if the limitation statement described above is added to your labeling.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website

{2}------------------------------------------------

(http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FDA

William H. Maisel -S

William H. Maisel, MD, MPH Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

| 510(k) Number (if known): | This application K182261                                                                                                                                                                                                                                 |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:              | <i>NvisionVLE Imaging System</i>                                                                                                                                                                                                                         |
| Indications for Use:      | The <i>NvisionVLE Imaging System</i> is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross-sectional, real-time depth visualization. |

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-the -Counter Use

(Part 21 CFR 801 Subpart D)

(21CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{4}------------------------------------------------

- 1. Basic Information-Submitter:

| 510(k) Owner:     | NinePoint Medical Inc.                                                                                                                                   |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:          | 12 Oak Park Drive<br>Bedford, MA 01730<br>(617) 250-7190 (main number)<br>(617) 250-7199 (fax)                                                           |
| Official Contact: | Eman Namati, PhD<br>President and CEO<br>NinePoint Medical, Inc.<br>(617) 250-7190 (main number)<br>(617) 250-7199 (fax)<br>enamati@ninepointmedical.com |

Date Summary Prepared: 20 August 2018

- 2. Device Name:
Trade Name: NvisionVLE® Imaging System Common Name: Optical Coherence Tomography Imaging System Classification Name: Ultrasonic pulsed echo imaging system Requlation Number: 21 CFR 892.1560 Product Code: NQQ Classification: Class II

- 3. Predicate Devices:
NvisionVLE® Imaging System, K143678

- 4. Device Description:
The NinePoint Medical NvisionVLE® Imaging System is a high-resolution volumetric imaging system based on optical coherence tomography (OCT). In an analogous fashion to ultrasound imagery, OCT images are formed from the time delay and magnitude of the signal reflected from the tissue of interest. The NvisionVLE® Imaging System employs an advanced form of OCT known as swept-source OCT (SS-OCT), or Optical Frequency Domain Imaging (OFDI),

{5}------------------------------------------------

in combination with a scanning optical probe to acquire high-resolution, crosssectional, real-time imagery of tissue called Volumetric Laser Endomicroscopy (VLE).

The device consists of the following main components and accessories: (i) a mobile NvisionVLE Console with an integrated computer and two touchscreen interfaces; (ii) proprietary NvisionVLE Software used to acquire, process, and visualize VLE images; (iii) a single-use, sterile NvisionVLE Optical Probe that is inserted through the working channel of an endoscope: (iv) a single-use, sterile NvisionVLE Inflation System that is used to inflate the balloon as required, to facilitate placement; and (v) a Probe Lock Accessory to prevent longitudinal motion of the Probe within the endoscope.

- 5. Indications for Use Statement:
The NvisionVLE™ Imaging System is indicated for use as an imaging tool in the evaluation of human tissue microstructure, including esophageal tissue microstructure, by providing two-dimensional, cross sectional, real-time depth visualization.

Labeling limitation: The Nvision VLE Imaging System is intended to provide an image of the tissue microstructure. The safety and effectiveness of this device for diagnostic analysis (i.e.differentiating normal versus specific abnormalities) in any tissue microstructure or specific disease has not been evaluated.

- 6. Technological Characteristics:
The subiect of this Special 510(k) submission is a device enhancement to provide balloon-less Low-profile Optical Probes in multiple configurations which include 20mm, 14mm and Short Working Distance (SWD) focal length optics. There are no new materials introduced with this change.

The NvisionVLE™ Optical Probe is made up of an optical probe subassembly and a guide sheath. The optical probe subassembly is a fiber optic probe assembly secured inside a flexible, stainless steel torque shaft. The distal optics are housed in a stainless steel hypotube which is attached to the torque shaft. The proximal end of the optical fiber and torque shaft terminate in a standard fiber optic connector and catheter connector which interfaces with the system console. The optical probe subassembly transmits and focuses the optical light and detects the reflected light for image reconstruction.

The guide sheath is a coaxially-designed balloon sheath. The sheath is composed of a PET balloon and a nylon shaft, or is provided without the balloon for the Low-profile Optical Probe models. The inner lumen of the 

{6}------------------------------------------------

sheath is sealed, enclosing the optical probe subassembly. The guide sheath is positioned within the organ structure of interest and allows the probe to rotate in a helical pattern while positioned in the inner lumen allowing for image reconstruction of the targeted tissue.

The technological characteristics of this device are unchanged from the predicate device cleared under K143678. The balloon-less version of the probe does not affect the intended use or alter the fundamental scientific technology of the device.

- 7. Non-Clinical Performance data:
Functional testing of the Low-profile Optical Probes included:

- Confirmation of the proper image on the Console
- Acceptable probe lateral resolution, back reflection, and transmission . rate
- Mechanical integrity of the probe/sheath tip demonstrated through . tensile testing
- . Functional testing to ensure the probe loads and withdraws properly

The NvisionVLE Imaging System has also been tested against and complies with the following voluntary standards:

| Consensus<br>Standard       | Description                                                                                                                                                                                    |
|-----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI/ISO<br>10993-7    | Biological evaluation of medical devices: Part 7:<br>Ethylene oxide sterilization residuals                                                                                                    |
|                             | ANSI/AAMI/ISO<br>10993-1                                                                                                                                                                       |
| ANSI/AAMI/ISO<br>11135:2014 |                                                                                                                                                                                                |
|                             | IEC 60601-1                                                                                                                                                                                    |
| IEC 60601-1-2               | Medical Electrical Equipment, Part 1-2, General<br>Requirement for Basic Safety and Essential<br>Performance - Collateral Standard<br>Electromagnetic Compatibility -Requirements and<br>Tests |
| ISO 62304                   | Software Development Life Cycle                                                                                                                                                                |

{7}------------------------------------------------

| IEC 60601-2-18 | Medical Electrical Equipment – Part 2-18:<br>Particular requirements for basic safety and<br>essential performance of endoscopic equipment                                                   |
|----------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-2-22 | Medical Electrical Equipment – Part 2-22:<br>Particular requirements for the basic safety and<br>essential performance of surgical, cosmetic,<br>therapeutic and diagnostic laser equipment. |
| IEC 60825-1    | Safety of Laser Products – Part 1: Equipment<br>Classification and requirements                                                                                                              |
| IEC 60601-1-6  | Medical electrical equipment Part 1-6<br>General requirements for basic safety and<br>essential performance - Collateral Standard:<br>Usability                                              |
| IEC 62366-1    | Medical devices -- Part 1: Application of usability<br>engineering to medical devices                                                                                                        |

- 8. Substantial Equivalence:
The NvisionVLE Imaging System has the same intended use and same indication, principles of operation, and technological characteristics as the predicate device. The minor differences with the Low-profile Optical Probe do not raise any new or different questions of safety or effectiveness.

Nonclinical performance testing demonstrates that the NvisionVLE Imaging System is as safe and effective as the predicate device and is therefore considered substantially equivalent.

---

**Source:** [https://fda-staging.innolitics.com/device/K182261](https://fda-staging.innolitics.com/device/K182261)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
