The UPnRIDE™ stand-up power wheelchair is intended for medical purposes to provide mobility to persons restricted to a sitting position. The product changes position from Sitting to Standing and Standing to Sitting but also any position in between. The product provides indoor and outdoor mobility.
Device Story
UPnRIDE is a stand-up power wheelchair for disabled/elderly individuals. Input: user joystick/button commands via R-NET controller; IMU sensor data (pitch/roll of ground and user). Operation: Main controller processes sensor inputs and user commands to drive motors and linear actuators; proprietary balancing system maintains user in vertical position relative to earth on skewed surfaces; COG maintained at center. Output: motorized mobility in sitting, standing, or intermediate positions. Used in indoor/outdoor urban environments; operated by the user. Benefits: enables mobility for those restricted to sitting; provides standing capability; enhances safety on uneven terrain via active balancing.
Clinical Evidence
No clinical testing was required for this submission. Performance was validated through bench testing and a usability study.
Technological Characteristics
Steel frame; two front motorized wheels; 21.6VDC/29.7Ah Li-Ion batteries. Features IMU for pitch/roll sensing, five actuator drivers, and R-NET controller. Balancing system uses two linear actuators to maintain verticality. Complies with ISO 7176 series (1-6, 8-11, 13-16, 21, 25) and ANSI/RESNA WC-1:2009.
Indications for Use
Indicated for individuals who need a power wheelchair and cannot stand up on their own, requiring mobility in sitting or standing positions.
Regulatory Classification
Identification
A standup wheelchair is a device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position. The device incorporates an external manually controlled mechanical system that is intended to raise a paraplegic to an upright position by means of an elevating seat.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 12, 2019
UPnRIDE Robotics Ltd. % Yoram Levy Qsite General Manager Osite 31 Haavoda St. Binyamina, 30500 Israel
### Re: K182257
Trade/Device Name: UPnRIDE Regulation Number: 21 CFR 890.3900 Regulation Name: Standup Wheelchair Regulatory Class: Class II Product Code: IPL Dated: July 15, 2019 Received: July 22, 2019
### Dear Yoram Levy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vivek Pinto, Ph.D. Director (Acting) DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182257
Device Name UPnRIDE
Indications for Use (Describe)
The UPnRIDE™ stand-up power wheelchair is intended for medical purposes to provide mobility to persons restricted to a sitting position. The product changes position from Sitting to Standing and Standing to Sitting but also any position in between. The product provides indoor and outdoor mobility.
Type of Use (Select one or both, as applicable)
| <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for UpnRIDE Robotics LTD. The logo has the word "UP" in gray, followed by a blue square with a lowercase "n" in white. The word "RIDE" is in purple, and below it is the text "ROBOTICS LTD." in a smaller font, also in purple.
# 510(K) SUMMARY
# UPnRIDE
# 510(k) Number K182257
| Applicant's Name: | UPnRIDE Robotics Ltd.<br>Kochav Yokneam Bld.<br>Yokneam Ilit 2069204<br>Israel.<br>TEL: (972)4-611-0061<br>FAX: (972)4-611-3313 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Yoram Levy, Qsite<br>31 Haavoda St.<br>Binyamina, Israel 3054431<br>Tel (972)4-638-8837; Fax (972)4-638-0510<br>Yoram@qsitemed.com |
| Trade Name: | UPnRIDE™ |
| Summary<br>Preparation Date: | August 8, 2018 |
| Classification: | Classification name: Standup wheelchair<br>Product Code: IPL<br>Regulation No: 21 CFR 890.3900<br>Class: II<br>Panel: Physical Medicine |
| Device Description: | The UPnRIDE™ device enables disabled and elderly<br>individuals to be mobile in a sitting or standing position<br>indoors and outdoors.<br><br>The device design is based on the following principles:<br>1. Using motorized jointed braces and harnessing to safely<br>support the user while sitting, shifting to standing, standing<br>and returning to sitting |
- 2. Center of gravity (COG) is maintained substantially at the center of the device in all positions, thus maximizing stability and safety.
- 3. Balancing the platform which supports the user so that the user always remains vertical even on skewed surfaces.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for "UPnRIDE ROBOTICS LTD." The "UP" is in a dark gray font, and the "n" is in a blue rounded square. The "RIDE" is in a purple font, and "ROBOTICS LTD." is in a smaller font below "RIDE".
- 4. Implementing an operation and safety algorithms to minimize risks and hazards to the user, even beyond the provisions of standard wheelchairs.
The product is comprised of the following modules:
- Driving module: a motorized driving chassis that 1. includes two front motorized wheels, two high capacity Li-Ion batteries 21.6VDC / 29.7Ah and a steel frame which is installed as a chassis for the device's components.
- Inertial Measurement Unit ("IMU") for measuring 2. ground slopes (pitch and roll) and user's tilt angles (pitch and roll).
- Standing and sitting module: a motor-based lifting 3. mechanism for shifting between standing and sitting positions.
- 4. Balancing module: a platform that balances the user while sitting or standing, relative to earth.
- 5. User interface module: a unit through which the user controls the UPnRIDE™. This module has an "R-NET" controller that contains a Joystick, commanding buttons and a display for viewing indications and status and a switching box for controlling the various modes of operation.
- Actuator drivers: five units that control the lifting 6. motors.
- 7. Main Controller: a unit which reads all sensors and user inputs, performs the balancing and safety algorithms, and controls all the chairs function.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for UpnRIDE Robotics LTD. The logo has the word "UP" in gray, a blue square with a lowercase "n" in white, and the word "RIDE" in purple. Underneath the word "RIDE" is the text "ROBOTICS LTD." in a smaller font.
# Indications for Use:
The UPnRIDETM stand-up power wheelchair is intended for medical purposes to provide mobility to persons restricted to a sitting position. The product changes people's position from Sitting to Standing and Standing to Sitting but also any position in between. The product provides indoor and outdoor mobility.
### Predicate Devices:
Substantial equivalence to the following predicate and reference devices is
| | Device Name | 510k No | Date of Clearance |
|-----------|--------------------|---------|-------------------|
| Predicate | Levo comfort II | K051387 | June 10, 2005 |
| Reference | Quickie® Q700-UP M | K172384 | January 16, 2018 |
claimed:
## Comparison with Predicate Devices
## Intended use and indications for use
The UPnRIDE™ stand-up power wheelchair is a product which changes people's position from sitting to standing and standing to sitting but also any position in between, same as the cleared Levo comfort II (K051387). Both devices are intended to allow indoor and outdoor mobility.
The proposed UPnRIDE™ and the two predicate devices: the Levo comfort II and the Q700-UP M (K172384) wheelchairs are intended for any individual who needs a power wheelchair and cannot stand up on their own.
The UPnRIDE™ can be used on uneven ground, same as with the Levo comfort II and the Q700-UP M.
The UPnRIDE™ is classified as a standup wheelchair, under the regulation code IPL (regulation number 890.3900), same as the Levo comfort II (K051387) and the Q700-UP M (K172384). The devices are Class II devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for a company called "UPn RIDE ROBOTICS LTD." The word "UP" is in gray, and the letter "n" is inside a blue rounded square. The word "RIDE" is in purple, and the words "ROBOTICS LTD." are in a smaller font below the word "RIDE."
### Technical Comparison
Both the UPnRIDE™, the Levo Comfort II and the Q700-UP M devices contain two main modules. The first one is a driving module that includes a steel frame, two drive motors and batteries as a source of energy. The second is a sitting-standing module, which enables the user adjust the chair according to the user's demand. The components of the sitting-standing adjustment module of the UPnRIDE™ device are similar to the cleared reference device- Q700-UP M powered wheelchair and includes several motors, that allow the user to transfer from a sitting to a standing position and vice versa, to amend the positions of the leg support, back support and chair's horizontal and vertical locations while sitting and standing, like the Q700-UP M.
Both UPnRIDE™ and the predicate devices include a controller which contains a joystick and command buttons designed by PG Drives Technology. In addition, both UPnRIDETM device and Q700-UP M contain the same braking system which is based on PG Drives Technology.
UPnRIDE™ includes additional module - Balancing module. UPnRIDE™ proprietary balancing system enables the user to be mobile on almost any urban environment with increased safety. Using two linear actuators, the system balances the platform which supports the user so that the user always remains vertical even on skewed surfaces. The balancing operation was validated by multitude performance tests and a usability study to ensure its efficiency and to prevent any safety concerns.
Both UPnRIDE™ and the predicate devices have similar characteristics such as user weighing up, maximum curb height and distance range.
#### Performance Standards
UPnRIDETM complies with the following voluntary standards:
- 1. ISO 7176-1:1999 Wheelchairs Part 1: Determination of static stability.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "UPnRIDE ROBOTICS LTD." The words "UP" and "RIDE" are in large, bold fonts, with "UP" in black and "RIDE" in purple. Between the two words is a blue square with rounded corners, containing a white letter "n". Below the word "RIDE" is the text "ROBOTICS LTD." in a smaller font.
- 2. ISO 7176-2:2001 Wheelchairs Part 2: Determination of dynamic stability of electric wheelchairs.
- 3. ISO 7176-3:2012 Wheelchairs Part 3: Determination of effectiveness of brakes
- 4. ISO 7176-4:2008 Wheelchairs Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range.
- ર. ISO 7176-5:2008 Determination of dimensions, mass and maneuvering space.
- 6. ISO 7176-6:2001 Wheelchairs Part 6: Determination of maximum speed, acceleration and deceleration of electric wheelchairs.
- 7. ISO 7176-8:2014 Wheelchairs Part 8: Requirements and test methods for static, impact and fatigue strengths- clauses 8-9 (static, impact strength).
- 8. ISO 7176-9:2009 Wheelchairs Part 9: Climatic tests for electric wheelchairs
- 9. ISO 7176-10:2008 Wheelchairs Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs
- 10. ISO 7176-11:2012 -Wheelchairs Part 11: Test dummies
- 11. ISO 7176-13:2008 Determination of coefficient of friction of test surfaces
- 12. ISO 7176-14 :2008 Wheelchairs Part 14: Power and control systems for electrically powered wheelchairs and scooters -Requirements and test methods
- 13. ISO 7176-15 -: 1996 Wheelchairs Part 15: Requirements for information disclosure, documentation and labelling.
- 14. ISO 7176-16 :2012 Resistance to ignition of upholstered parts - Requirements and test methods.
- 15. ISO 7176-21-:2009 Wheelchairs Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers.
- 16. ISO 7176-25:2013 Requirements and test methods for batteries and battery chargers intended for use with electrically powered wheelchairs
- 17. ANSI/RESNA WC-1: 2009, Section 20 -American National Standard for Wheelchairs -Volume 1: Requirements and Test Methods for Wheelchairs
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for a company called "UPnRIDE ROBOTICS LTD." The logo has the word "UP" in gray, followed by a blue square with a lowercase "n" inside. The word "RIDE" is in purple, and the words "ROBOTICS LTD." are in a smaller font below the word "RIDE". The logo is simple and modern, and it uses a limited color palette.
# Summary of Clinical Performance Data:
No Clinical Testing is required for this submission
### Substantial equivalence conclusion
The performance tests and the clinical study that were conducted show that the UPnRIDE™ is substantially equivalent to the listed predicate devices without raising any new safety and effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.