K182252 · Cook Incorporated · FOZ · May 17, 2019 · General Hospital
Device Facts
Record ID
K182252
Device Name
Cook Unimpregnated Central Venous Catheter
Applicant
Cook Incorporated
Product Code
FOZ · General Hospital
Decision Date
May 17, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only The device is a short-term use catheter, intended for less than 30 days. The Cook dilator is used for dilating puncture sites or catheter tracts. The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations. The Cook Unimpregnated Central Venous Catheter with power injection is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only 7. Power injection of contrast media* *The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0, and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0, and 10.0 French catheters. The device is a short-term use catheter, intended for less than 30 days. The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter. The Cook dilator is used for dilating puncture sites or catheter tracts. The Cook Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
Device Story
Single-use, sterile, intravascular catheter; inserted via Seldinger technique; tip positioned at SVC-RA junction. Configurations: single, dual, triple, or five-lumen; materials include polyethylene, ethylene-vinyl acetate, polyurethane. Used in clinical settings for drug delivery, blood monitoring, and hyperalimentation. Power-injectable versions allow contrast media delivery at defined flow rates (up to 10 mL/sec) and pressure limits (up to 325 psi). Includes accessories like dilators, inner catheters, and wire guides. Healthcare providers use the device to establish vascular access; output is the delivery of fluids or monitoring of venous pressure. Benefits include reliable short-term venous access for critical care and diagnostic imaging.
Clinical Evidence
Bench testing only. Performance testing included tensile strength (BS EN ISO 10555-1), liquid/air leakage, kink radius (BS EN 13868), gravity flow rate, static burst, and hub pressure for power injection. Biocompatibility testing included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, implantation, and hemocompatibility per ISO 10993-1.
Technological Characteristics
Materials: Polyethylene, ethylene-vinyl acetate, polyurethane (catheter); nylon (inner catheter). Dimensions: 2.5-10 French, 5-40 cm length. Sensing/Actuation: Passive fluid delivery/pressure monitoring. Connectivity: None. Sterilization: Ethylene oxide (EtO). Standards: BS EN ISO 10555-1, BS EN 13868, ISO 594-1/2, ISO 11135-1.
Indications for Use
Indicated for adult and pediatric patients requiring short-term (<30 days) vascular access for drug infusion, CVP monitoring, hyperalimentation, blood sampling, and blood product delivery. Power-injectable models are indicated for contrast media injection at specified flow/pressure limits. Contraindicated for power injection in non-power-injectable models.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Cook Central Venous Catheters with or without Heparin (K081113)
Turbo-Ject Peripherally Inserted Central Venous Catheter (PICC) (K072625)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
May 17, 2019
Cook Incorporated Chelsea Woods Regulatory Affairs Specialist 750 Daniels Way P.Box 489 Bloomington, Indiana 47404
Re: K182252
Trade/Device Name: Cook Unimpregnated Central Venous Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular catheter Regulatory Class: Class II Product Code: FOZ, DRE Dated: April 15, 2019 Received: April 16, 2019
Dear Chelsea Woods:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Tina Kiang, Ph.D Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182252
Device Name
Cook Unimpregnated Central Venous Catheter
Indications for Use (Describe)
- The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:
- 1. Continuous or intermittent drug infusions
- 2. Central venous blood pressure monitoring (CVP)
- 3. Acute hyperalimentation
- 4. Blood sampling
- 5. Delivery of whole blood or blood products
- 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
The device is a short-term use catheter, intended for less than 30 days.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182252
Device Name
Cook Unimpregnated Central Venous Catheter
Indications for Use (Describe)
- The Cook Unimpregnated Central Venous Catheter with power injection is used for:
- 1. Continuous or intermittent drug infusions
- 2. Central venous blood pressure monitoring (CVP)
- 3. Acute hyperalimentation
- 4. Blood sampling
- 5. Delivery of whole blood or blood products
- 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
- 7. Power injection of contrast media*
*The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0. 8.0. 9.0. and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0, and 10.0 French catheters.
The device is a short-term use catheter, intended for less than 30 days.
The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Cook Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span></span> | <span></span> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Cook Medical logo. The logo is white text on a red background. The word "COOK" is in large, bold letters, and the word "MEDICAL" is in smaller letters below it.
# 510(k) SUMMARY
# K182252 Cook Unimpregnated Central Venous Catheter 21 CFR §807.92 Date Prepared: May 16, 2019
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|-------------------------------------------|
| Applicant: | Cook Incorporated |
| Applicant Address: | Cook Incorporated |
| | 750 Daniels Way |
| | Bloomington, IN 47404 |
| Primary Contact: | Chelsea Woods |
| Secondary Contact: | Ram Iyer, Ph.D. RAC |
| Email: | regsubmissions@cookmedical.com |
| Contact Phone Number: | (812) 335-3575 x104707 |
| Contact Fax Number: | (812) 332-0281 |
# Device Information:
| Trade Name: | Cook Unimpregnated Central Venous Catheter |
|-----------------------|-----------------------------------------------------------------------|
| Common Name: | Catheter, Intravascular, Therapeutic, Short-Term Less Than<br>30 Days |
| Classification Name: | Intravascular catheter |
| Regulation: | 21 CFR §880.5200 |
| Product Code: | FOZ; DRE |
| Device Class: | II |
| Classification Panel: | General Hospital |
# Predicate Device:
The Cook Unimpregnated Central Venous Catheter is substantially equivalent to the following devices:
- Cook Central Venous Catheters with or without Heparin (K081113, Cook -Incorporated) cleared on July 30, 2008
{5}------------------------------------------------
Cook Incorporated - Traditional 510(k) Cook Unimpregnated Central Venous Catheter May 16, 2019
#### Device Description:
The Cook Unimpregnated Central Venous Catheter is a single use sterile intravascular catheter designed to be inserted into a patient's vascular system. The CVC is inserted into the vasculature using the Seldinger technique. The tip of the catheter is then advanced until it is above the superior vena cava-right atrium (SVC-RA) junction. The CVC configurations include a single, dual, triple, or five-lumen shaft manufactured from polyethylene, ethylene-vinyl acetate, or polyurethane tubing. The CVC is manufactured with an outside diameter between 2.5 and 10 French and has a length between 5 and 40 centimeters dependent on the configuration. Single lumen CVCs are designed with a premolded winged hub on the proximal end. Dual, triple, and five-lumen CVCs are designed with a manifold assembly, which is comprised of a winged manifold connected to extension tubes. Each extension tube has a slide clamp and is manufactured with a proximal winged hub. Certain CVC configurations are power injectable. The CVC may be packaged as a convenience kit with various other components which may include a dilator, inner catheter, wire guide, access needle, injection caps, clamps, catheter securement device, connecting tube, and other convenience accessories. The subject device inner catheter, supplied with the 9 and 10 French subject device catheters, is used to assist advancement of the main catheter over the wire guide. The inner catheter, manufactured from nylon, is designed to lock with the hub of the polyurethane catheter. The subject device dilator, used to dilate the access site during the procedure, is designed with a radiopaque extruded polyethylene shaft and a pre-molded polyethylene proximal hub. It is manufactured with an outside diameter ranging from 3.5 to 12.0 French and has a length between 6 and 11 centimeters. For non-power injectable catheters, the 2.5 French, 3.0 French and 4.0 French catheters are recommended for patients from birth and older (or weighing at least 2.4 kilograms), the 5.0 French catheters are recommended for patients aged 2 years and older (or weighing at least 10 kilograms), the 6.0 French and 6.3 French catheters are recommended for patients aged 4 years and older (or weighing at least 13 kilograms), and the 7.5 French catheters are recommended for patients aged 10 years and older (or weighing at least 50 kilograms). For power injectable catheters, the 4.0 French catheters are recommended for patients from birth and older (or weighing at least 2.4 kilograms), the 5.0 French catheters are recommended for patients aged 2 years and older (or weighing at least 10 kilograms), the 7.0 French and 8.0 French catheters are recommended for patients aged 10 years and older (or weighing at least 50 kilograms),
{6}------------------------------------------------
and the 9.0 French and 10.0 French catheters are recommended for patients aged 21 years and older (adults only).
The Cook Unimpregnated Central Venous Catheter is a short-term device, sterilized by ethylene oxide, and intended for single use.
## Indications for Use (Non-Power Injectable):
The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:
- 1. Continuous or intermittent drug infusions
- 2. Central venous blood pressure monitoring (CVP)
- 3. Acute hyperalimentation
- 4. Blood sampling
- 5. Delivery of whole blood or blood products
- 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
The device is a short-term use catheter, intended for less than 30 days.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Cook unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations.
## Indications for Use (Power Injectable):
The Cook Unimpregnated Central Venous Catheter with power injection is used for:
- 1. Continuous or intermittent drug infusions
- 2. Central venous blood pressure monitoring (CVP)
- 3. Acute hyperalimentation
- 4. Blood sampling
- 5. Delivery of whole blood or blood products
- 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
- 7. Power injection of contrast media*
*The flow rate may not exceed 3 mL/sec 4.0Fr and 5.0Fr catheters and 10 mL/sec for 7.0, 8.0, 9.0, and 10.0Fr catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0, and 10.0 French catheters.
The device is a short-term use catheter, intended for less than 30 days.
{7}------------------------------------------------
The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Cook Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
## Comparison to Predicate Device:
The Cook Unimpregnated Central Venous Catheter is substantially equivalent to the predicate device, Cook Central Venous Catheters with or without Heparin (K081113, Cook Incorporated), cleared on July 30, 2008. The subject device is identical to the predicate device in intended use, principle of operation, and fundamental technologies. The differences between the subject device and the predicate device (K081113), including pediatric indication, materials, lumen number, and dimensions do not raise any different questions of safety and/or effectiveness. The substantial equivalence comparison of the subject device to the predicate devices is provided in the table below. To further support the safety and effectiveness of the subject device dilators, Dilator Sets cleared under K183036 are provided as a reference device.
{8}------------------------------------------------
| | PREDICATE DEVICE | SUBJECT DEVICE | |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| | Cook Central Venous Catheters with or<br>without Heparin<br>(K081113) | Cook Unimpregnated Central Venous<br>Catheter | |
| Regulation | 21 CFR §880.5200 | IDENTICAL TO PREDICATE | |
| Product Code | FOZ<br>Catheter, Intravascular, Therapeutic, Short-<br>Term Less Than 30 Days | IDENTICAL TO PREDICATE | |
| Classification | II | IDENTICAL TO PREDICATE | |
| Indications for<br>Use | The Cook Central Venous Catheter is used<br>for:<br>- Continuous or intermittent drug<br>infusions<br>- Central venous blood pressure<br>monitoring (CVP)<br>- Acute hyperalimentation<br>- Blood sampling<br>- Delivery of whole blood or blood<br>products<br>- Power injection of contrast media*<br>The device is a short-term use catheter.<br>*The flow rate of the Cook Central Venous<br>Catheters may not exceed 10 mL/sec. | The Cook Unimpregnated Central Venous<br>Catheter (Non-Power Injectable) is used for:<br>1. Continuous or intermittent drug infusions<br>2. Central venous blood pressure monitoring<br>(CVP)<br>3. Acute hyperalimentation<br>4. Blood sampling<br>5. Delivery of whole blood or blood<br>products<br>6. Simultaneous, separate infusion of drugs<br>for multi-lumen catheters only<br><br>The device is a short-term use catheter,<br>intended for less than 30 days.<br><br>The Cook dilator is used for dilating puncture<br>sites or catheter tracts.<br><br>The Cook unimpregnated Central Venous<br>Catheter (Non-Power Injectable) is intended<br>for adult and pediatric populations | |
| | | The Cook Unimpregnated Central Venous | |
| | | Catheter with power injection is used for: | |
| | | 1. Continuous or intermittent drug infusions | |
| | | 2. Central venous blood pressure monitoring | |
| | | (CVP) | |
| | | 3. Acute hyperalimentation | |
| | | 4. Blood sampling | |
| | | 5. Delivery of whole blood or blood | |
| | | products | |
| | | 6. Simultaneous, separate infusion of drugs<br>for multi-lumen catheters only | |
| | | 7. Power injection of contrast media* | |
| | | *The flow rate may not exceed 3 mL/sec for | |
| | | 4.0 and 5.0 French catheters and 10 mL/sec | |
| | | for 7.0, 8.0, 9.0, and 10.0 French catheters | |
| | | Verify prior to use that the maximum safety | |
| | | cut-off pressure limit is set at or below 250 psi | |
| | | for 4.0 and 5.0 French catheters and 325 psi | |
| | | for 7.0, 8.0, 9.0, and 10.0 French catheters. | |
| | | The device is a short-term use catheter, | |
| | | intended for less than 30 days. | |
| | | The 9.0 and 10.0 French catheters include an | |
| | | inner catheter to facilitate insertion of the | |
| | | main catheter. | |
| | | The Cook dilator is used for dilating puncture | |
| | | sites or catheter tracts. | |
| | | The Cook Unimpregnated Central Venous | |
| | | Catheter with power injection is intended for | |
| | | adult and pediatric populations. | |
| | | Catheter Size | Recommended Minimum<br>Age / Weight |
| Patient Population | Not specified | 2.5Fr | Birth / < 2.4 kg |
| | | 3.0Fr | Birth / 2.4 kg |
| | | 4.0Fr | |
| | | 5.0Fr | 2 yrs / 10 kg |
| | | 6.3Fr | 4 yrs / 13 kg |
| | | 7.0Fr | |
| | | 7.5Fr | 10 yrs / 50 kg |
| | | 8.0Fr | |
| | | 9.0Fr | 21 yrs (adult only) |
| | | 10.0Fr | |
| Device is for One-<br>Time Use | Yes | IDENTICAL TO PREDICATE | |
| Catheter<br>Placement Method | Percutaneous via Seldinger technique | IDENTICAL TO PREDICATE | |
| Catheter Tip<br>Target Anatomy | SVC-RA junction | IDENTICAL TO PREDICATE | |
| Catheter Tip<br>Location<br>Confirmation<br>Method | | | |
| Catheter Distal<br>End Configuration | Straight and tapered | IDENTICAL TO PREDICATE | |
| Catheter Ink<br>Markings | Yes | IDENTICAL TO PREDICATE…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.