The Solero Microwave Tissue Ablation (MTA) System is indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not indicated for cardiac use.
Device Story
Solero MTA Cart is an optional transport and storage accessory for the Solero Microwave Tissue Ablation (MTA) System. The cart features a shelf for the generator, an integrated IV pole for chilled saline, a cable management system, and a tray for manuals. It sits on four locking casters. The cart is used in clinical settings (e.g., OR) by physicians or clinical staff to maneuver and position the generator at an appropriate working height. It does not contact the patient. The cart facilitates system organization and transport, ensuring the generator and accessories are securely held and accessible during procedures. It does not perform ablation itself; the ablation is performed by the Solero MTA System generator and applicators.
Clinical Evidence
Bench testing only. Testing included IEC 60601-1 (basic safety and essential performance), packaging testing, and functional testing (including electromagnetic compatibility and coexistence in a simulated use environment). No clinical data was required or provided.
Technological Characteristics
Hardware accessory; non-sterile; no patient contact. Features four rotating locking casters (tested for 5,000 cycles). Includes integrated IV pole and cable management system. Designed for transport and storage of the Solero MTA Generator. No electronic components; no shelf-life. Complies with IEC 60601-1.
Indications for Use
Indicated for ablation of soft tissue during open procedures. Not for cardiac use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Solero Microwave Tissue Ablation (MTA) System (K162449)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 28, 2019
AngioDynamics, Inc. Kasey Newcomb Regulatory Affairs Specialist II 26 Forest St. Marlborough, Massachusetts 01752
Re: K182250
Trade/Device Name: Solero MTA Cart, Solero MTA System, Solero microwave Tissue Ablation Applicator (14cm), , Solero microwave Tissue Ablation Applicator (19cm), Solero microwave Tissue Ablation Applicator (29cm) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: January 31, 2019 Received: February 1, 2019
Dear Kasey Newcomb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K182250
Device Name Solero Microwave Tissue Ablation (MTA) System and Accessories
Indications for Use (Describe)
The Solero Microwave Tissue Ablation (MTA) System is indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not indicated for cardiac use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
PSC Polylishing Sections (101)-443-6749 EF
{3}------------------------------------------------
# SOLERO MTA SYSTEM CART ACCESSORY 510(K) SUMMARY
#### SPONSOR A.
AngioDynamics, Inc. 26 Forest St. Marlborough, MA 01752 USA
#### B. CONTACT
Kasey E. Newcomb Specialist II, Global Regulatory Affairs Tel: 508.658.7813 Email: knewcomb@angiodynamics.com
#### DEVICE NAME ﺯ
| Trade Name: | Solero Microwave Tissue Ablation (MTA) System and Accessories |
|-----------------------|-----------------------------------------------------------------------------------------------|
| Common/Usual Name: | Microwave Tissue Ablation System and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>(21 CFR § 878.4400, Class II, Pro-Code NEY) |
| Classification Panel: | General Surgery |
#### D. PREDICATE DEVICE
| 510(k) Number: | K162449 |
|-----------------------|-----------------------------------------------------------------------------------------------|
| Trade Name: | Solero Microwave Tissue Ablation (MTA) System |
| Common/Usual Name: | Microwave Tissue Ablation (MTA) System |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>(21 CFR § 878.4400, Class II, Pro-Code NEY) |
| Classification Panel: | General Surgery |
#### E. REFERENCE DEVICE
| 510(k) Number: | K133821 |
|-----------------------|-----------------------------------------------------------------------------------------------|
| Trade Name: | Emprint Ablation System |
| Common/Usual Name: | Microwave Ablation System and Accessories |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>(21 CFR § 878.4400, Class II, Pro-Code NEY) |
| Classification Panel: | General Surgery |
#### F. DEVICE DESCRIPTION
The device description of the Solero MTA System from the previous cleared K162449, has not been modified with the exception of the addition of the Solero MTA Cart as an optional accessory to the Solero MTA System.
The Solero MTA Cart is an optional accessory, that allows for transport, storage, and maneuverability of the system pre- and post-use. The Solero MTA Cart provides a shelf that is designed to securely hold the Solero MTA Generator in place. Once on the cart, the Solero MTA Generator is at an appropriate working height. The Solero MTA Cart has an incorporated IV pole for holding the chilled saline source used for maintaining the Solero Applicator at an appropriate temperature. There is a cable management system built into the Solero MTA Cart that allows for the mains power cable and footswitch to be stored for transporting the device or storage. The Solero MTA System can be transported using the handle on the cart. There is a tray on the stand of the cart intended to hold
{4}------------------------------------------------
Directions for Use and Operators Manual. The Solero MTA Cart sits on four rotating casters that can all be locked into place during use, storage, or as appropriate during transit.
#### G. INDICATION FOR USE
The Solero Microwave Tissue Ablation (MTA) System and Accessories are indicated for the ablation of soft tissue during open procedures. The Solero MTA System is not intended for cardiac use.
#### H. STERILIZATION/CLEANING/SHELF LIFE
The Solero MTA Cart is a hardware device with no patient contact and is provided non-sterile, therefore sterilization validation was not required or performed.
The Solero MTA Cart Operators Manual states that the cart should be cleaned by wiping down all accessible surfaces with a clean cloth that is slightly dampened with 70% isopropyl alcohol cleaning solution.
The Solero MTA Cart is not manufactured with any electronic or materials that are expected to expire or degrade over time. The Solero MTA Cart's casters have successfully completed 5,000 full "on" and "off" cycles without failure of the locking holding power, wear or permanent deformation that would adversely affect the locking performance of the casters. There is no assigned shelf-life.
#### I. BIOCOMPATIBILITY
The Solero MTA Cart is a hardware device with no patient contact and provided non-sterile. Biocompatibility testing was not performed.
#### J. SUMMARY OF SIMILARITIES AND DIFFERENCES IN TECHNOLOGY CHARACTERISTICS
Reference device, Emprint Ablation System Cart, cleared via K133821, was used to support safety and effectiveness of the subject device. Both the subject device and specified reference device include the following technological characteristics:
- The carts worksurface heights are appropriate for performing microwave tissue ablation . procedures
- . No rough surfaces, sharp corners and/or edges
- Four (4) rotating locking casters ●
- Frame/shelf that accommodates a specific microwave tissue ablation generator. The ● microwave tissue ablation generators are not permanently installed on the cart.
- Includes tray/shelf for accessories storage
- Cable management system (hooks) that allow power cords, connection leads, cables, ● footswitch, etc, to be properly stored for either transporting or storage of microwave tissue ablation system
- . Cart movement is manageable by a single user
Technological characteristic that are different between the subject and specified reference device are as follows:
- Solero MTA Cart incorporates an IV pole for holding the chilled saline source used for . maintaining the Solero Applicator at an appropriate temperature. The specified reference device utilizes a separate IV pole that is positioned next to the cart.
- All isolation electronics are contained within the Solero MTA Generator. The specified . reference cart includes an isolation transformer. The Solero MTA Generator contains appropriate electrical isolation; therefore, a separate isolation transformer is not required.
#### K. PERFORMANCE DATA
The following performance testing was used to substantiate safety and effectiveness:
- . IEC 60601-1: General Requirements for Basic Safety and Essential Performance
- . Packaging Testing
- . Functional Testing including (e.g. electromagnetic compatibility and coexistence in a simulated use environment)
{5}------------------------------------------------
#### L. CONCLUSIONS
The results of comparison of technology similarities and the non-clinical performance testing demonstrates that the subject device (cart accessory), when used as an accessory for the predicate device (Solero MTA System), is substantially equivalent to the predicate device when used without it.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.