K182224 · Lina Medical Aps · FYD · Feb 7, 2019 · General, Plastic Surgery
Device Facts
Record ID
K182224
Device Name
SafeAir Smoke Evacuator compact
Applicant
Lina Medical Aps
Product Code
FYD · General, Plastic Surgery
Decision Date
Feb 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5070
Device Class
Class 2
Indications for Use
SafeAir Smoke Evacuator compact is designed to remove and filter smoke generated during electrosurgical and laser procedures.
Device Story
Device is a vacuum-pump-based smoke evacuation system for hospital/clinic use; removes surgical smoke plume via diathermy pencil or tubing set connected to the unit. Operates via manual front-panel controls or footswitch; synchronizes with electrosurgical units. Features replaceable ULPA filter with active carbon. Healthcare providers use the device to clear the surgical field of smoke/aerosols during procedures; benefits include improved visibility and reduced exposure to surgical smoke. Device output is the suction of smoke through the filter system.
Clinical Evidence
Bench testing only. Performance verified through simulated use testing comparing flow rates and under-pressure parameters using standard accessories (smoke evacuation pencils/tubing). Electrical safety and electromagnetic compatibility verified per AAMI/ANSI ES60601-1 and IEC 60601-1-2. Software validated per IEC 62304.
Technological Characteristics
Vacuum pump system; ULPA filter (99.999% efficiency at 0.1 microns) with active carbon. Electrical power source. Manual and footswitch activation. Compliant with IEC 60601-1 (electrical/mechanical safety) and IEC 60601-1-2 (EMC). Non-patient contact device. Software validated per IEC 62304.
Indications for Use
Indicated for the removal and filtration of surgical smoke generated during electrosurgical and laser procedures in a hospital or clinic environment.
Regulatory Classification
Identification
Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 7, 2019
LiNA Medical APS % Kevin MacDonald Regulatory Consultant Kevin F. MacDonald 4297 D Street Sacramento, California 95819
Re: K182224
Trade/Device Name: SafeAir Smoke Evacuator compact Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus for a Surgical Operating Room Regulatory Class: Class II Product Code: FYD Dated: January 8, 2019 Received: January 9, 2019
Dear Kevin MacDonald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray Iii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182224
Device Name SafeAir Smoke Evacuator compact
Indications for Use (Describe)
SafeAir Smoke Evacuator compact is designed to remove and filter smoke generated during electrosurgical and laser procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary: K182224
| 510(k) Owner: | LiNA Medical ApS<br>Formervangen 5<br>DK-2600 Glostrup, Denmark<br>Telephone: +48 61 222 21 43 |
|---------------------------|------------------------------------------------------------------------------------------------|
| Contact Person: | Kevin MacDonald,<br>Regulatory Consultant<br>Email: kma@lina-medical.com<br>Tel. 415-609-9875 |
| Date Summary<br>Prepared: | 04 Feb 2019 |
| Trade Name of<br>Device: | SafeAir Smoke Evacuator compact |
| Common Name : | Air-handling apparatus for a surgical operating room |
| Classification: | Class II |
| Product Code: | FYD |
| Panel: | General and Plastic Surgery Devices |
| Regulation Number: | 21CFR 878.5070 |
| Predicate Device: | Visiclear Smoke Evacuation System - K131402 |
#### Device Description
SafeAir Smoke Evacuator compact is a device with a vacuum pump that aspirates the surgical smoke produced during cauterization.
The main purpose of smoke evacuation systems is to remove surgical smoke plume as close as possible to the surgical site. The component of the system ensuring that this is achieved is a diathermy pencil with integrated smoke evacuation function or a click-on tubing set connected to a surgical smoke evacuator.
SafeAir Smoke Evacuator compact is designed for use in the hospital/clinic environment. It is built around a vacuum pump with a replaceable filter at the entry point. A specially designed pump control is centrally placed in the machine. The SafeAir Smoke Evacuator compact is designed with a pump with different activation functions; manual activation on the front panel, footswitch activation through diathermy pencil with smoke evacuation tubing. Synchronization is possible with most of the electrosurgical units.
| Table 1: Models number – main device | |
|--------------------------------------|---------------------------------|
| Reference number | Description |
| SFR-compact-US | SafeAir Smoke Evacuator compact |
{4}------------------------------------------------
| Table 2: Accessories | |
|----------------------|-----------------------------------------------|
| Reference Number | Description |
| SFR-FSW-US | SafeAir footswitch |
| SFR-FIL-C-US | SafeAir ULPA filter with active carbon |
| SFR-CON-US | SafeAir tubing for bottle connection |
| SFR-TUB-7-US | SafeAir 7m Tubing Set for external exhaust |
| SFR-GC-US | SafeAir Grounding cable |
| SFR-compact-KIT-1-US | SafeAir 1 Year Service KIT for SFR-compact-US |
| SFR-compact-KIT-3-US | SafeAir 3 Year Service KIT for SFR-compact-US |
#### Indications for Use
SafeAir Smoke Evacuator compact is designed to remove and filter smoke generated during electrosurgical and laser procedures.
#### Technological Characteristic Comparison
Table 3 compares the SafeAir Smoke Evacuator compact to the predicate devices with respect to indications for use, principles of operation, technological characteristics, and performance testing.
{5}------------------------------------------------
| Table 3 - Technological Characteristic Comparison Table | | | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Variable | SafeAir Smoke Evacuator compact<br>(Subject Device) | Visiclear Smoke Evacuation System<br>(Predicate - K131402) | Comparison |
| 510(k) number | K182224 | K131402 | |
| Product Code | FYD | Same | Same |
| Regulatory Class | II | Same | Same |
| Regulation Number | 21 CFR 878.5070 | Same | Same |
| Intended Use | SafeAir Smoke Evacuator compact is<br>designed to remove and filtrate smoke<br>generated during electrosurgical and<br>laser procedures in the vicinity of its<br>source. | The VISICLEAR Smoke Evacuation System is<br>designed to remove and filter smoke,<br>aerosols produced during electrosurgical<br>and laser procedures. | Same |
| Direct Body Contact<br>Materials | Not intended to come in contact with<br>patient | Same | Same |
| Energy used | Electrical current | Same | Same |
| Operational Settings | On/Off Switch, Suction level buttons for<br>Motor Control | Same | Same |
| Intended marketed<br>accessories | Electrosurgical pencils, tubing, hoses &<br>adaptors - sterile and non sterile | Same | Same |
| Filtration | The ULPA filter is 99.999% with<br>efficiency at 0.1 microns particle size | The ULPA-grade filter is 99.999% efficiency<br>at 0.1 to 0.2 micron particle size. | Same |
| Filter Life | 5 hours | 35 hours | Use life of the filter is established with performance<br>testing and is consistent with intended claims. |
| Table 3 - Technological Characteristic Comparison Table | | | |
| Variable | SafeAir Smoke Evacuator compact<br>(Subject Device) | Visiclear Smoke Evacuation System<br>(Predicate - K131402) | Comparison |
| Filter life indicator | Indicator light | Shut down | When the indicator light is on, the user is instructed<br>to change the filter. |
| Electrical safety | Tested and compliant with IEC 60601-1<br>and IEC 60601-1-2 | Same | Same |
| Mechanical safety | Tested and compliant with IEC 60601-1 | Same | Same |
| Chemical safety | neutral pH, non-patient contact | Same | Same |
| Mechanical safety | Tested and compliant with IEC 60601-1 | Same | Same |
| Chemical safety | neutral pH, non-patient contact | Same | Same |
| Thermal safety | Operation of the device does not result<br>in harmful temperatures tested and<br>compliant per IEC 60601-1 | Same | Same |
| Radiation safety | non-radioactive | Same | Same |
| Maximum flow rate<br>(claimed) | 110 l/min | 839 l/min | Flow rate for the subject device is demonstrated to<br>effectively entrain surgical smoke |
| Maximum flow rate<br>(suction pencil)* | 97.1 l/min | 93.2 l/min | Similar |
| Maximum flow rate<br>(suction tube)* | 66.2 l/min | 62.2 l/min | Similar |
| Minimum flow rate<br>(suction pencil)* | 41.2 l/min | 43.1 l/min | Similar |
| Minimum flow rate<br>(suction tube)* | 26.0 l/min | 25.9 l/min | Similar |
{6}------------------------------------------------
*All values obtained from simulated use testing
{7}------------------------------------------------
The SafeAir Smoke Evacuator compact shares the same indications for use, device operation, overall technical and functional capabilities as the predicate devices. In addition, the SafeAir Smoke Evacuator compact is similar in design and function to the Visiclear Smoke Evacuation System (K131402) in terms of mode of operation and use.
The technological differences between the SafeAir Smoke Evacuator Compact and predicate device are limited to maximum flow and under-pressure values defined by the manufacturers. These technological differences do not have an impact on the actual clinical use parameters. As confirmed through the design verification testing, both the predicate and subject devices performed as intended, meeting the performance specification independent of the technological differences.
In the design verification testing, accessories representative of components that would be used in a commercial setting were used to assess the clinically relevant flow rate and under-pressure parameters. The accessories included Smoke Evacuation pencils or Smoke Tubing, commonly used in electrosurgical or laser procedures. Both the predicate and subject devices were tested utilizing the same method and accessories and as reported in Table 3, the test results were comparable between the predicate and subject devices.
The SafeAir Smoke Evacuator compact shares the same indications for use, device operation, overall technical and functional capabilities as the predicate devices. In addition, the SafeAir Smoke Evacuator compact is similar in design and function to the Visiclear Smoke Evacuation System (K131402) in terms of mode of operation and use.
### Sterilization and Biocompatibility
Based on the intended use, both the predicate and subject devices are not intended to come in direct contact with the patient and therefore, do not require sterilization or to be tested for biocompatibility.
### Nonclinical Performance Data
The SafeAir Smoke Evacuator compact meets all the requirements for overall design, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The SafeAir Smoke Evacuator compact was subjected to safety performance testing in accordance with. AAMI / ANSI ES60601-1.
The SafeAir Smoke Evacuator compact passed all applicable testing in accordance with internal requirements, national standards, and international standards shown below:
- Electrical safety testing per AAMI / ANSI ES60601-1, IEC 60601-1-2,
- Software Validation and Verification per IEC 62304
### Conclusion
Based on the intended use, technological characteristics and non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.