The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. The GU60A & GU60A-65 Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. The GF50A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications. The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
Device Story
Digital X-ray imaging systems (stationary and mobile) capture radiographic images of human anatomy. X-rays pass through the patient to a detector, converting radiation into electrical signals. Signals are processed by the S-station (Operation Software) and saved as DICOM files. An Image Post-processing Engine (IPE) performs region-of-interest extraction, tonescale mapping, noise reduction, and texture restoration. The device is operated by qualified doctors or technicians in clinical settings. The IPE's advanced noise reduction algorithm allows for reduced radiation dose (up to 47.5% for adult abdomen; up to 45% for pediatric abdomen, 15.5% for pediatric chest, and 27% for pediatric skull) while maintaining diagnostic image quality. Output is sent to PACS for radiologist review. Dose reduction benefits patients by minimizing radiation exposure while maintaining diagnostic confidence.
Clinical Evidence
Evidence includes non-clinical phantom testing (TOR CDR phantom) and clinical evaluations. Phantom testing measured CNR, DCC, and MTF, showing improved CNR with the new IPE. Clinical validation for adult abdomen used anthropomorphic phantoms and clinical images scored by radiologists on a 5-point scale, confirming 47.5% dose reduction. Pediatric population evaluation used a noise simulation tool and clinical trial data; three pediatric radiologists assessed dose-simulated images to determine optimal diagnostic dose, confirming dose reductions of 45% (abdomen), 15.5% (chest), and 27% (skull).
Technological Characteristics
Stationary and mobile digital X-ray systems. High-frequency generators (up to 82kW). X-ray tubes with Rhenium-Tungsten targets. Indirect CsI or Gd2O2S detectors (140um pixel pitch). Connectivity via wired/wireless DICOM. Software includes S-station OS and IPE with advanced noise reduction algorithm. Standards: ES 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54, ISO 14971.
Indications for Use
Indicated for generating radiographic images of human anatomy, including adult abdominal and pediatric chest, abdomen, and skull exams. Not for mammography. For use by qualified/trained doctors or technicians.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo, which features a stylized human figure. The logo on the right is the FDA U.S. Food & Drug Administration logo, with the FDA acronym in a blue square and the full name of the administration written in blue text.
December 7, 2018
Samsung Electronics Co., LTD. Jaesang Noh Senior Professional, Regulatory Affairs 129, Samsung-Ro, Yeongtong-Gu Suwon-Si, 16677 Republic of Korea
Re: K182183
Trade/Device Name: GC70, GU60A, GU60A-65, GF50, GR50A ; GR40CW ; GM85 ; GC85A Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary X-Ray System Regulatory Class: Class II Product Code: KPR, MOB, IZL Dated: July 31, 2018 Received: August 13, 2018
Dear Jaesang Noh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind vou, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hol. 2. Nils
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name
GC85A
Indications for Use (Describe)
The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
> Prescription Use (Part 21 CFR 801 Subpart D) ] Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
PRAStaff@fda.hhs.gov
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name GC70
Indications for Use (Describe)
The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name GU60A, GU60A-65
Indications for Use (Describe)
The GU60A & GU60A-65 Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="padding-right: 5px;"> </span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="padding-right: 5px;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{5}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name GF50
Indications for Use (Describe)
The GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| <span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> </span> | <span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> </span> <span> </span> <span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> <span style="font-size: 10px;"> <span style="text-decoration: underline;"></span> </span> </span> <span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block;"> <div style="display:inline-block;"> <div style="display:inline-block;"> <img alt="checkbox" src="checkbox_checked.png" style="width:10px;height:10px;"/> </div> </div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div style="display:inline-block;"> <div style="display:inline-block;"> <div style="display:inline-block;"> <img alt="checkbox" src="checkbox.png" style="width:10px;height:10px;"/> </div> </div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{6}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name GF50A
Indications for Use (Describe)
The GF50A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K182183
Device Name GR40CW
Indications for Use (Describe)
The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="padding-right: 20px;"> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{8}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182183
Device Name GM85
Indications for Use (Describe)
The GM85 Digital X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div><span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the Samsung logo. The logo is in blue and the word "SAMSUNG" is written in white, bold letters. The logo is tilted slightly to the right.
510(k) Premarket Notification - Traditional
### Section 5: 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted accordance with requirements of 21 CFR 807.92
- 1. Date: July 31, 2018
#### 2. Submitter
- A. Company Name: SAMSUNG ELECTRONICS Co., Ltd.
- Address: 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, 16677, B. Republic of Korea
#### 3. Primary Contact Persons
- A. Name: JAESANG NOH & CHULSIN KIM
- B. Title: Regulatory Affairs Senior Professional
- Phone Number: +82-2-2193-2444 & +82-2-2193-2437 ﻥ
- D. FAX Number: +82-31-200-6401 E-Mail: jaesang.noh@samsung.com & chulsin.kim@samsung.com
#### 4. Secondary Contact Person
- A. Name: Genci Omari
- B. Title: Manager, Regulatory Affairs
- C. Phone Number: 978-564-8602
- D. FAX Number: 978-560-0602 E-Mail: gomari@samsungneurologica.com
#### 5. Proposed Device
| | #1 | #2 | #3 | #4 | #5 | #6 | #7 |
|--------------------|-------------------------------------------|-------------------------------------------|-------------------------------------------|-------------------------------------------|--------------------------------|--------------------------------------------------------|-------------------------------------------|
| Trade<br>Name | GC70 | GU60A&<br>GU60A-<br>65 | GF50 | GF50A | GR40C<br>W | GM85 | GC85A |
| Device<br>Name | GC70 | GU60A&<br>GU60A-<br>65 | GF50 | GF50A | GR40C<br>W | GM85 | GC85A |
| Common<br>Name | Digital<br>Diagnos<br>tic X-ray<br>System | Digital<br>Diagnos<br>tic X-ray<br>System | Digital<br>Diagnos<br>tic X-ray<br>System | Digital<br>Diagnos<br>tic X-ray<br>System | Retrofit<br>Kit | Digital<br>Diagnos<br>tic<br>Mobile<br>X-ray<br>System | Digital<br>Diagnos<br>tic X-ray<br>System |
| Regulation<br>Name | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Mobile<br>X-ray<br>System | Stationar<br>y X-ray<br>System |
| Product<br>Code | KPR | KPR | KPR | KPR | MQB | IZL | KPR |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.
510(k) Premarket Notification - Traditional
| Regulation | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.172<br>0 | 21 CFR<br>892.168<br>0 |
|------------|------------------------|------------------------|------------------------|------------------------|------------------------|------------------------|------------------------|
|------------|------------------------|------------------------|------------------------|------------------------|------------------------|------------------------|------------------------|
#### 6. Predicate Devices
| | | Predicate Devices | | | | | | |
|----------------------------|--------------------------------|--------------------------------|--------------------------------|--------------------------------|------------------------------------|---------------------------|--------------------------------|--|
| | #1 | #2 | #3 | #4 | #5 | #6 | #7 | |
| Device<br>Name | GC70 | GU60A&<br>GU60A-<br>65 | GF50 | GF50A | GR40C<br>W | GM85 | GC85A | |
| Classificati<br>on Name | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Stationar<br>y X-ray<br>System | Solid<br>State X-<br>ray<br>Imager | Mobile<br>X-ray<br>System | Stationar<br>y X-ray<br>System | |
| Product<br>Code | KPR | KPR | KPR | KPR | MQB | IZL | KPR | |
| Regulation | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.168<br>0 | 21 CFR<br>892.172<br>0 | 21 CFR<br>892.168<br>0 | |
| 510(K)# | K180543 | K180543 | K180543 | K180543 | K181631 | K181626 | K181629 | |
| 510(K)<br>Decision<br>Date | May 24,<br>2018 | May 24,<br>2018 | May 24,<br>2018 | May 24,<br>2018 | July 20,<br>2018 | July 20,<br>2018 | July 20,<br>2018 | |
#### 7. Device Description
GC70, GU60&GU60A-65, GF50, GF50A, GR40CW, GM85 and GC85A are used to capture images by transmitting X-ray to a patient's body. The X-ray passing through a patient's body is sent to the detector and then converted into electrical signals. These signals go through the process of amplification and digital data conversion in the signal process on the S-station, which is the Operation Software (OS) of Samsung Digital Diagnostic X-ray System, and save in DICOM file, a standard for medical imaging. The captured images are tuned up by an Image Post-processing Engine (IPE) which is exclusively installed in S-station, SAMSUNG digital X-ray operation software, and sent to the Picture Archiving & Communication System (PACS) sever for reading images.
The IPE operates, from the input image, the roles of a region-of-interest extraction, tonescale mapping, noise reduction and texture restoration. The IPE employing an advanced noise reduction algorithm (hereinafter "new IPE") is shown that the image quality of PA radiograph for average adult chest, exposed at the condition of 50% lower dose at Entrance Skin Exposure (ESE) in comparison with the condition of the conventional noise reduction algorithm (hereinafter "old IPE"), is substantially equivalent. It was cleared with K172229 that using the new IPE is able to reduce dose of 50% for chest PA of average adult in the proposed device, GC85A. The scope of the claim, 50% dose reduction by using the new IPE, was expanded at K180543, from GC85A to GC70, GU60&GU60A-65, GF50, GF50A, GR40CW and GM85. In K180543, it was evaluated that the IPE had no relation with imaging chain and the images acquired by a various imaging chain were substantially equivalent to GC85A (K172229).
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the Samsung logo. The logo is in blue and features the word "SAMSUNG" in bold, white letters. The word is slightly tilted to the right and is enclosed in an elliptical shape.
510(k) Premarket Notification - Traditional
This submission is purposed to get 510(k) clearance for expanding the scope of the dose reduction claims, the chest PA of average adult to abdominal radiograph of adult, cleared with K172229 and K180543, and chest, abdomen and skull of pediatric populations.
The principles of operation and technological characteristics of the proposed devices and the predicate devices are the same. The proposed device is identical to the predicate devices since there are the same devices including the new IPE; however, Samsung would like to expand dose reduction claims to the adult's abdominal radiograph, and the chest, abdomen and skull of pediatric populations for marketing purposes.
The non-clinical and clinical image evaluations show that the proposed devices with the new IPE employing and advanced noise reduction algorithm are able to reduce x-ray dose up to 47.5% for abdominal radiographs of average adult and up to 45% for pediatric abdomen, 15.5% for pediatric chest, and up to 27% for pediatric skull exams, compared with radiation doses needed to obtain similar image quality using the old IPE on the same x-ray systems.
Note: The dose reduction claim is based on non-clinical (phantom) testing and on clinical testing conducted at one medical site with a specific x-ray imaging system and specific technique factors. In practice, the values of dose reduction may vary accordingly. The clinical site is responsible for determining whether the particular radiographic imaging needs are not impacted by such x-ray dose reduction. The pediatric subgroup, neonate under 1 month old, was not studied.
#### 8. Intended Use
The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The GU60A & GU60A-65 Digital X-ray Imaging Systems are intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
GF50 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The GF50A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
The GR40CW Digital X-ray Imaging System is intended for use in general projection radiographic applications wherever conventional screen-film systems or CR systems may be used. This device is not intended for mammographic applications.
The GM85 Digital Mobile X-ray imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image features the word "SAMSUNG" in bold, white letters against a blue, rounded background. The font is sans-serif and the letters are evenly spaced. The blue background is a rounded shape, resembling an oval or elongated circle, and it provides a contrasting backdrop that makes the white text stand out.
510(k) Premarket Notification - Traditional
device is not intended for mammographic applications.
The GC85A Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications.
- 9. Summary of Technological characteristic of the proposed devices compared with the predicate devices
The proposed devices, GC70, GU60&GU60A-65, GF50, GF50A, GR40CW, GM85 and GC85A, are identical to the predicate devices recently cleared with K180543 (GC70, GU60&GU60A-65, GF50 and GF50A), K181631 (GR40CW), K181626 (GM85) and K181629 (GC85A), including the new IPE with employing an advanced noise reduction algorithm.
| Specification | Proposed Device #1 | Predicate Device #1 (PD #1) | Discussion |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name | GC70 | GC70 | |
| Manufacturer | SAMSUNG ELECTRONICS Co., Ltd. | SAMSUNG ELECTRONICS Co., Ltd. | |
| 510(k) Number | None | K180543 | |
| Intended Use | The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | The GC70 Digital X-ray Imaging System is intended for use in generating radiographic images of human anatomy by a qualified/trained doctor or technician. This device is not intended for mammographic applications. | Same as PD#1 |
| High Voltage Generator | | | |
| Type | High Frequency | High Frequency | Same as PD#1 |
| Max. Power | 82kW | 82kW | Same as PD#1 |
| Output Range | Tube Voltage<br>40-150kV | Tube Voltage<br>40-150kV | Same as PD#1 |
| | Tube Current<br>10-1000mA | Tube Current<br>10-1000mA | Same as PD#1 |
| | Exposure Time<br>1msec-10sec | Exposure Time<br>1msec-10sec | Same as PD#1 |
| AEC (Automatic Exposure Control) | Yes | Yes | Same as PD#1 |
#### A. Compared with Predicate Devices
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the Samsung logo. The word "SAMSUNG" is written in white, bold letters. The background is a blue, rounded rectangle.
| APR<br>(Anatomically<br>Programmed<br>Radiography) | | | Yes | | | Yes | | | Same as PD#1 |
|----------------------------------------------------|-----------------------------------------|------------------------------------|------------------------------------|------------------------------------|------------------------------------|------------------------------------|------------------------------------|--------------|--------------|
| X-ray tube | | | | | | | | | |
| | Name | E7869X | E7252X | E7869X | E7252X | Same as PD#1 | | | |
| | kV range | 40-150kV | 40-150kV | 40-150kV | 40-150kV | Same as PD#1 | | | |
| Max.<br>Power | Large focus | 100kW | 75kW | 100kW | 75kW | Same as PD#1 | | | |
| | Small focus | 40kW | 27kW | 40kW | 27kW | Same as PD#1 | | | |
| Max.<br>mA | Large focus | 1000mA | 1000mA | 1000mA | 1000mA | Same as PD#1 | | | |
| | Small focus | 500mA | 400mA | 500mA | 400mA | Same as PD#1 | | | |
| | Anode target<br>material | Rhenium-<br>Tungsten | Rhenium-<br>Tungsten | Rhenium-<br>Tungsten | Rhenium-<br>Tungsten | Same as PD#1 | | | |
| Collimator | | | | | | | | | |
| | Name | SDR-OGCL83U | SDR-OGCL83U | Same as PD#1 | | | | | |
| | Overall<br>Size(mm) | H212 X W306<br>X D179 | H212 X W306<br>X D179 | Same as PD#1 | | | | | |
| | Beam Limiting<br>Blade<br>Moving Method | | Motorized<br>/Manual | Motorized<br>/Manual | Same as PD#1 | | | | |
| | Manual<br>Operation<br>Method | | Volume | Volume | Same as PD#1 | | | | |
| Collimator<br>Rotation | | ±45 | ±45 | Same as PD#1 | | | | | |
| | Beam Light<br>Source | LED | LED | Same as PD#1 | | | | | |
| | Light Field<br>Indicator Timer | O | O | Same as PD#1 | | | | | |
| | | O | O | Same as PD#1 | | | | | |
| | Side Lamp | Laser Module | Laser Module | Same as PD#1 | | | | | |
| | Field Size / SID<br>Display | Color LCD | Color LCD | Same as PD#1 | | | | | |
| Detector | | | | | | | | | |
| | Name | S433<br>5-W | S434<br>3-W | S302<br>5-W | S433<br>5-W | S434<br>3-W | S302<br>5-W | Same as PD#1 | |
| | | Csl | Csl | Csl | Csl | Csl | Csl | Same as PD#1 | |
| | Detector Type | Indirect | Indirect | Indirect | Indirect | Indirect | Indirect | Same as PD#1 | |
| | Detector Area | 14"X1<br>7"<br>(345 mmX<br>425m m) | 17"X1<br>7"<br>(425 mmX<br>425m m) | 10"X1<br>2"<br>(245 mmX<br>295m m) | 14"X1<br>7"<br>(345 mmX<br>425m m) | 17"X1<br>7"<br>(425 mmX<br>425m m) | 10"X1<br>2"<br>(245 mmX<br>295m m) | Same as PD#1 | |
| | Number of<br>pixels | 2466<br>X3040 | 3036<br>X3040 | 1750<br>X2108 | 2466<br>X3040 | 3036<br>X3040 | 1750<br>X2108 | Same as PD#1 | |
{14}------------------------------------------------
Image /page/14/Picture/1 description: The image shows the SAMSUNG logo in white text on a blue background. The logo is stylized with a slightly tilted, oval shape. The text is bold and sans-serif. The overall impression is clean and modern.
| Pixel Pitch(um) | 140 | 140 | 140 | 140 | 140 | 140 | Same as PD#1 |
|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-------------------------|-------------------------|-------------------------|-------------------------|-------------------------|---------------|
| High Contrast<br>Limiting<br>Resolution<br>(LP/mm) | 3.57 | 3.57 | 3.57 | 3.57 | 3.57 | 3.57 | Same as PD#1 |
| Communication | Wired /<br>Wirel<br>ess | Wired /<br>Wirel<br>ess | Wired /<br>Wirel<br>ess | Wired /<br>Wirel<br>ess | Wired /<br>Wirel<br>ess | Wired /<br>Wirel<br>ess | Same as PD#1 |
| Image Post-processing Engine | | | | | | | |
| Noise reduction<br>algorithm | Image Post-<br>processing Engine<br>with an advanced<br>noise reduction<br>algorithm | | | | | | Same as PD#1 |
| Claim of dose<br>reduction | 50% dose reduction<br>for chest PA of<br>average adult | | | | | | Same as PD #1 |
| | Up to 47.5% dose<br>reduction for<br>abdominal<br>radiographs of adult | | | | | | Different(1) |
| | Up to 45% dose<br>reduction for pediatric<br>abdomen, 15.5% for<br>pediatric chest, and<br>up to 27% for pediatric<br>skull exams | | | | | | Different(2) |
| Specification | Proposed Device #2 | | Predicate Device #2<br>(PD #2) | | Discussion | | |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------|-------------------|---------------|
| | GU60A | GU60A-65 | GU60A | GU60A-65 | | | |
| Device Name | GU60A & GU60A-65 | | GU60A & GU60A-65 | | | | |
| Manufacturer | SAMSUNG<br>ELECTRONICS Co.,<br>Ltd. | | SAMSUNG<br>ELECTRONICS Co.,<br>Ltd. | | | | |
| 510(k) Number | None | | K180543 | | | | |
| Intended Use | The GU60A &<br>GU60A-65 Digital X-<br>ray Imaging Systems<br>are intended for use in<br>generating<br>radiographic images<br>of human anatomy by<br>a qualified/trained<br>doctor or technician.<br>This device is not<br>intended for<br>mammographic<br>applications. | | The GU60A &<br>GU60A-65 Digital X-<br>ray Imaging Systems<br>are intended for use in<br>generating<br>radiographic images<br>of human anatomy by<br>a qualified/trained<br>doctor or technician.<br>This device is not<br>intended for<br>mammographic<br>applications. | | Same as PD#2 | | |
| High Voltage Generator<br>Type | High Frequency | | High Frequency | | Same as PD#2 | | |
| | Max. Power | 50kW/68kW | 65kW | 50kW/68kW | 65kW | Same as PD#2 | |
| | Tube<br>Voltage | 40 to 150kVp | 40 to 150kVp | 40 to 150kVp | 40 to 150kVp | Same as PD#2 | |
| Output<br>RANGE | Tube<br>Current | 10 - 630mA/<br>10 - 800mA | 10 - 800mA | 10 - 630mA/<br>10 - 800mA | 10 - 800mA | Same as PD#2 | |
| | Exposure<br>Time | 1msec-<br>6.3sec/<br>1msec-<br>10sec | 1msec-<br>6.3sec | 1msec-<br>6.3sec/<br>1msec-<br>10sec | 1msec-<br>6.3sec | Same as PD#2 | |
| X-ray tube | | | | | | | |
| | Name | E7252X | | E7252X | | Same as PD#2 | |
| | kV range | 40-150kV | | 40-150kV | | Same as PD#2 | |
| Max.<br>Power | Large<br>focus | 75kW | | 75kW | | Same as PD#2 | |
| r | Small<br>focus | 27kW | | 27kW | | Same as PD#2 | |
| Max.<br>mA | Large<br>focus | 1000mA | | 1000mA | | Same as PD#2 | |
| | Small<br>focus | 400mA | | 400mA | | Same as PD#2 | |
| Anode target<br>material | | Rhenium-Tungsten | | Rhenium-Tungsten | | Same as PD#2 | |
| Collimator | | | | | | | |
| | Name | SDR-OGCL82U | | SDR-OGCL82U | | Same as PD#2 | |
| | Overall<br>Size(mm) | H212 X W306<br>X D179 | | H212 X W306<br>X D179 | | Same as PD#2 | |
| | Beam Limiting<br>Blade<br>Moving Method | Motorized<br>/Manual | | Motorized<br>/Manual | | Same as PD#2 | |
| | Manual<br>Operation<br>Method | Volume | | Volume | | Same as PD#2 | |
| | Collimator<br>Rotation | ±45 | | ±45 | | Same as PD#2 | |
| | Beam Light<br>Source | LED | | LED | | Same as PD#2 | |
| | Light Field<br>Indicator Timer | O | | O | | Same as PD#2 | |
| | Side Lamp | Laser Module | | Laser Module | | Same as PD#2 | |
| | Field Size / SID<br>Display | Color LCD | | Color LCD | | Same as PD#2 | |
| Detector | | | | | | | |
| | Name | S433<br>5-W | S434<br>3-W | S302<br>5-W | S433<br>5-W | Same as PD#2 | |
| | | Csl | Csl | Csl | Csl | Same as PD#2 | |
| | 14"X | 17"X | 10"X1 | 14"X | 17"X1 | 10"X1 | |
| | 17" | 17" | 2" | 17" | 7" | 2" | |
| | (345 | (425 | (245…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.