K182170 · Nuvasive Specialized Orthopedics, Inc. · HSB · Feb 5, 2019 · Orthopedic
Device Facts
Record ID
K182170
Device Name
PRECICE® BONE TRANSPORT™ System
Applicant
Nuvasive Specialized Orthopedics, Inc.
Product Code
HSB · Orthopedic
Decision Date
Feb 5, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PRECICE® BONE TRANSPORT™ System is indicated for limb-lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions or bone transport of long bones.
Device Story
The PRECICE® BONE TRANSPORT™ System is an intramedullary fixation device used for limb correction and bone transport. The system consists of an intramedullary nail, locking screws, end caps, surgical instruments, and an External Remote Controller (ERC). The nail contains an internal rare earth magnet, distraction rod, and planetary gearing. It is implanted into the intramedullary canal of long bones and secured with locking screws. The ERC is used non-invasively to adjust the distraction rod, allowing for gradual lengthening or compression of bone segments. The device is used in a clinical setting by a surgeon. The system benefits patients by providing a non-invasive method for bone transport and limb lengthening, reducing the need for repeated surgical interventions for adjustments.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included static and dynamic compression, bending strength, torsion, torque resistance, axial pullout, biocompatibility (ISO 10993), sterilization (ISO 11137), package distribution (ASTM D4169), distraction force, retention force, fluid ingress, and biomechanical axial compression.
Technological Characteristics
Intramedullary fixation rod with rare earth magnet, distraction rod, and planetary gearing. Materials conform to ASTM F1264 and ASTM F543. Sterilization via gamma irradiation (VDmax 25 kGy). Non-invasive adjustment via external electromagnetic controller. Slotted housing tube design for bone transport.
Indications for Use
Indicated for patients requiring limb-lengthening, open/closed fracture fixation, pseudoarthrosis, mal-unions, non-unions, or bone transport of long bones.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA) along with the Department of Health & Human Services logo. The FDA logo features the agency's acronym in a blue square, followed by the full name "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The Department of Health & Human Services logo is on the left side of the image and features a symbol with three stylized human profiles. The logos are placed side by side.
NuVasive Specialized Orthopedics, Inc. Regualtory Affairs Specialist 101 Enterprise. Suite 100
February 5, 2019
# Re: K182170
Aliso Viejo, California 92656
Vidhi Bhatt
Trade/Device Name: PRECICE® BONE TRANSPORT™ System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB, HWC Dated: September 21, 2018 Received: September 24, 2018
Dear Vidhi Bhatt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel S. Ramsey -S 2019.02.05 16:31:33 -05'00'
FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182170
Device Name
PRECICE® BONE TRANSPORT™ System
Indications for Use (Describe)
The PRECICE® BONE TRANSPORT™ System is indicated for limb-lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, non-unions or bone transport of long bones.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
| <div><span style="font-size:16px">✖</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized, abstract shape on the left, resembling a curved, open form in purple and silver. To the right of the shape, the word "NUVASIVE" is written in a sans-serif font, with a silver gradient effect. Below "NUVASIVE", the words "SPECIALIZED ORTHOPEDICS, INC." are written in a smaller, simpler font.
## PRECICE® Bone Transport System 510(k) Summary - K182170
In accordance with Title 21 of the Code of Federal Regulations. Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:
#### Submitted by: A.
Vidhi Bhatt Regulatory Affairs Specialist NuVasive Specialized Orthopedics, Inc. 101 Enterprise, Suite 100 Aliso Viejo, CA 92656 Telephone: (949) 563-0655
Date Prepared: February 5, 2019
#### B. Device Name
| Trade or Proprietary Name: | PRECICE® Bone Transport System |
|----------------------------|------------------------------------------------------------------------|
| Common or Usual Name: | Intramedullary Fixation Rod |
| Classification Name: | Rod, Fixation, Intramedullary and Accessories<br>Screw, Fixation, Bone |
| Device Class: | Class II |
|-----------------|-------------------|
| Classification: | 21 CFR § 888.3020 |
| Product Code: | HSB, HWC |
#### C. Predicate Devices
| Predicate Device | PRECICE® System (K173129) |
|------------------|----------------------------------|
| Reference Device | PRECICE® Stryde System (K180503) |
#### D. Device Description
The PRECICE® Bone Transport System includes the PRECICE® Bone Transport Nail, locking screws, end caps, surgical instruments, and external remote controller (ERC). The PRECICE® Bone Transport nails and end caps are supplied sterile by gamma radiation while the locking screws and instruments are supplied non-sterile and must be sterilized prior to use. The system is designed to achieve limb correction through gradual lengthening or compression of the intercalary bone segment and providing internal fixation for fractures of long bones. The PRECICE® Bone Transport intramedullary nail is implanted using locking screws, end caps, and reusable surgical instruments. The PRECICE® Bone Transport nail contains an enclosed rare earth magnet, distraction rod, and planetary gearing which allows the extension of the distraction rod to be adjusted non-invasively by the External Remote Controller (ERC). The PRECICE® Bone Transport nail is available in various diameters, lengths and screw hole configurations to accommodate a variety of patient anatomies and implantation methods. The locking screws are also available in a variety of diameters, lengths, and thread styles. The ERC is available in several compatible models.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic to the left of the company name. The text "NUVASIVE" is in a sans-serif font, and below it, in smaller letters, is "SPECIALIZED ORTHOPEDICS, INC."
The purpose of this submission is to implement design changes to the predicate PRECICE® System (K173129) and reference device PRECICE® Stryde System (K180503) to create a product line extension which includes the PRECICE® Bone Transport nail and compatible locking screws and end caps. The subject nail performance meets predicate device performance specifications. The design changes of subject nail inform revised contraindications, warnings. precautions and implantation procedure on the label to address new risks compared to predicate device.
These modifications to the PRECICE® System do not change the indications for use or intended use of the device, nor do they change the fundamental scientific technology of the device.
#### E. Indication for Use
The PRECICE® Bone Transport System is indicated for limb lengthening, open and closed fracture fixation, pseudoarthrosis, mal-unions, or bone transport of long bones.
#### F. Comparison of Technological Characteristics with the Predicate Device
The subject PRECICE® Bone Transport System is substantially equivalent to the predicate PRECICE® System (K173129) and reference device PRECICE® Stryde System (K180503) previously cleared by the FDA for commercial distribution in the United States.
Substantial equivalence between subject, predicate and reference device is established based on identical indications for use, material composition, technological characteristics, principles of operation and fundamental scientific principles.
Substantial equivalence is also based on device dimension, specifically, the nail diameters, locking screws, device configurations and end caps which are the same as the reference device. The predicate, reference and subject devices are inserted into the intramedullary canal of the long bones and secured with locking screws. The predicate, reference and subject devices are adjusted non-invasively by the External Remote Controller (ERC). The design modifications for the subject device, PRECICE® Bone Transport are as follows:
- . The subject nail is a slotted intramedullary (IM) nail. The housing tube of this IM nail has longitudinal transport slot.
- . The stroke length/transport length is independent of rod length since the distraction rod excursion is accommodated within the housing tube making the subject device a fixed length device.
- The distraction rod spans both the defect site and osteotomy site.
- Additional size of locking screw available with the subject device .
- Assembly of the gear train and distraction rod is reversed to incorporate the transport slot ● although the components remain the same as the predicate device
The labeling changes includes added contraindication and warning based on the design modifications of the subject device.
There are no changes being made to the ERCs or PRECICE® surgical instruments as a result of this submission.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Nuvasive Specialized Orthopedics, Inc. The logo features a stylized purple and gray graphic on the left, followed by the word "NUVASIVE" in gray, sans-serif font. Below "NUVASIVE" is the text "SPECIALIZED ORTHOPEDICS, INC." in a smaller, sans-serif font, also in gray.
Conclusions can be drawn from these comparisons that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate device.
#### G. Performance Data
Nonclinical performance verification testing was performed to demonstrate that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate PRECICE® device. The following testing shows that the subject device conforms to the following standards, practices and guidance:
| Testing Description | Applicable Standard |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Static Compression<br>Bending Strength | ASTM F1264 - Standard Specification<br>and Test Methods for Intramedullary<br>Fixation Devices |
| Dynamic Compression<br>Bending strength | |
| Torsion | |
| Torque Resistance | ASTM F543 - Standard Specification and<br>Test Methods for Metallic Medical Bone<br>Screws |
| Axial Pullout | |
| Biocompatibility | ISO 10993- Biological evaluation of<br>medical devices |
| Sterilization | Gamma Irradiation Sterilization Testing:<br>VDmax 25 kGy method per IS/EN/ISO<br>11137-1:2015 and IS/EN/ISO 11137-<br>2:2015 |
| Package Distribution | ASTM D4169-16 Standard Practice for<br>Performance Testing of Shipping<br>Containers and Systems |
In addition, the device also underwent the following performance testing to show that all requirements have been met:
- >> Distraction Force Testing
- Retention Force Testing A
- A Fluid Ingress Testing
- Biomechanical Axial Compression Testing A
The results demonstrate that the subject PRECICE® Bone Transport System is substantially equivalent to the predicate device.
#### H. Conclusions
The subject PRECICE® Bone Transport System has been shown to be substantially equivalent to the legally marketed predicate device for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.