← Product Code [IYN](/productcode/IYN) · K182165

# Color Doppler Ultrasound System; Model: mEye1 and mEye2 (K182165)

_Shenzhen Well.D Medical Electronics Co., Ltd. · IYN · Dec 20, 2018 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182165

## Device Facts

- **Applicant:** Shenzhen Well.D Medical Electronics Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Dec 20, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular, trans-vaginal and trans-rectal applications in B, M, PWD and Color Doppler modes.

## Device Story

Color Doppler Ultrasound System (mEye1, mEye2) acquires ultrasound signals via convex or linear array probes; transforms signals into B, M, PWD, and Color Doppler imaging modes; displays images on integrated screen. Used in clinical settings by healthcare professionals for diagnostic imaging and fluid flow analysis. Output assists clinicians in visualizing internal structures and blood flow to support diagnostic decision-making. Benefits include non-invasive diagnostic visualization of various anatomical regions.

## Clinical Evidence

Bench testing only. No clinical data included.

## Technological Characteristics

Integrated color ultrasound system; convex (3.5MHz) and linear (7.5MHz) array probes; B, M, PWD, and Color Doppler modes. Power: AC 100-240V or DC 14.4V. Screen resolution: 1024x768. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO 10993-10, NEMA UD 2.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Apogee 1200 Digital Color Doppler Ultrasound Imaging System ([K113613](/device/K113613.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue.

December 20, 2018

Shenzhen Well.D Medical Electronics Co., LTD. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CHINA

Re: K182165

Trade/Device Name: Color Doppler Ultrasound System Model; mEye1 and mEye2 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 22, 2018 Received: December 3, 2018

Dear Diana Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Hse 2. Nils

for Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K182165

Device Name Color Doppler Ultrasound System Model: mEye1 and mEye2

#### Indications for Use (Describe)

Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular, transvaginal and trans-rectal applications in B, M, PWD and Color Doppler modes.

| Type of Use (Select one or both, as applicable)                          |                                                                         |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <span></span> <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Format

System: Color Doppler Ultrasound System

Transducer: 5C2AN convex array probe

| Clinical Application        |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only)   | Specific<br>(Track 1&3)            | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                  | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
| Fetal<br>Imaging &<br>Other | Fetal                              | N                 |   | N   |     | N                |                       | THI                 |
|                             | Abdominal                          | N                 |   | N   |     | N                |                       | THI                 |
|                             | Intra-operative(Specify)           |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative(Neuro)             |                   |   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
|                             | Pediatric                          | N                 |   | N   |     | N                |                       | THI                 |
|                             | Small Organ(Specify)               |                   |   |     |     |                  |                       |                     |
|                             | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph.(non-Card.)            |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) | N                 |   | N   |     | N                |                       | THI                 |
|                             | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Cardiac                     | Cardiac Adult                      |                   |   |     |     |                  |                       |                     |
|                             | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

{4}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Format

System: Color Doppler Ultrasound System

Transducer: 12L5AN linear array probe

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application        |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|-----------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
|                             | Specific<br>(Track 1&3)            | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| General<br>(Track 1 Only)   |                                    |                   |   |     |     |                  |                       |                     |
| Ophthalmic                  | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
| Fetal<br>Imaging &<br>Other | Fetal                              |                   |   |     |     |                  |                       |                     |
|                             | Abdominal                          |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative(Specify)           |                   |   |     |     |                  |                       |                     |
|                             | Intra-operative(Neuro)             |                   |   |     |     |                  |                       |                     |
|                             | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
|                             | Pediatric                          |                   |   |     |     |                  |                       |                     |
|                             | Small Organ(Specify)               | N                 |   | N   |     | N                |                       | THI                 |
|                             | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                             | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                             | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                             | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph.(non-Card.)            |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                     |
|                             | Musculo-skeletal<br>(Superficial)  | N                 |   | N   |     | N                |                       | THI                 |
|                             | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Cardiac                     | Cardiac Adult                      |                   |   |     |     |                  |                       |                     |
|                             | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                     |
|                             | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                             | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                       |                     |
|                             | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral<br>Vessel        | Peripheral vessel                  | N                 |   | N   |     | N                |                       | THI                 |
|                             | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

{5}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: K182165

- 1. Date of Preparation: 12/20/2018
- 2. Sponsor Identification

### SHENZHEN WELL.D MEDICAL ELECTRONICS CO., LTD.

Well.d Park, Qinglan 3 Rd, National Biopharmaceutical Industrial Base, Pingshan New Area, Shenzhen, 518118, China.

Establishment Registration Number: 3008350665

Contact Person: Ning Bai Position: Management Representative Tel: +86-755-36900033 Fax: +86-755-36900033 Email: weld-200@163.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Chengyu Wang (Alternative Contact Person)

## Mid-Link Consulting Co., Ltd

P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 001-3609253199 Email: info@mid-link.net

{6}------------------------------------------------

#### 4. Identification of Proposed Device

Trade Name: Color Doppler Ultrasound System Model: mEye1 and mEye2 Classification Name: Ultrasonic pulsed doppler imaging system Classification: II Product Code: IYN, IYO, and ITX Regulation Number: 21 CFR 892.1550, 21 CFR 892.1560 and 21 CFR 892.1570 Review Panel: Radiology

Indications for Use:

Diagnostic ultrasonic imaging for abdominal, pediatric, small organ, musculo-skeletal, cardiac, peripheral vascular, trans-vaginal and trans-rectal applications in B, M. PWD and Color Doppler modes.

#### Device Description

The Color Doppler Ultrasound Systems are integrated preprogrammed color ultrasound imaging system, capable of producing a resolution intended for clinical diagnostic imaging applications. Their basic function is to acquire ultrasound signal and display the images in the imaging operation modes of B, M, PWD and Color Doppler.

- 5. Identification of Predicate Device
Predicate Device 510(k) Number: K113613 Product Name: Apogee 1200 Digital Color Doppler Ultrasound Imaging System

#### 6. Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > IEC 60601-1-2005+CORR.1:2006+CORR.2:2007+A1:2012, Medical Electrical Equipment- Part 1: General requirements for basic safety and essential performance
- A IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility- Requirements and tests
- > IEC 60601-2-37: 2007 Medical electrical equipment - Part2 -37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- A ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

{7}------------------------------------------------

- ♪ ISO 10993-10:2010 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization
- > NEMA UD 2-2004 (R2009) Acoustic output measurement standard for diagnostic ultrasound equipment
- Clinical Test Conclusion 7.

No clinical study is included in this submission.

- Substantially Equivalent (SE) Comparison 8.

| ITEM                | Proposed Device                                                                                                                                                                                                       | Predicate Device<br>K113613                                                                                                                                                                                                            |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code        | IYN, IYO and ITX                                                                                                                                                                                                      | IYN, IYO and ITX                                                                                                                                                                                                                       |
| Class               | II                                                                                                                                                                                                                    | II                                                                                                                                                                                                                                     |
| Indications for Use | Diagnostic ultrasonic imaging for<br>abdominal, pediatric, small organ,<br>musculo-skeletal, cardiac, peripheral<br>vascular, trans-vaginal and trans-rectal<br>applications in B, M, PWD and Color<br>Doppler modes. | Diagnostic ultrasonic imaging for<br>abdominal, pediatric, small organ,<br>musculo-skeletal, cardiac, peripheral<br>vascular, trans-vaginal and trans-rectal<br>applications in B, M, PWD, CWD, Color<br>Doppler and 3D imaging modes. |
| Configuration       | Host and Probe                                                                                                                                                                                                        | Host and Probe                                                                                                                                                                                                                         |
| Sterile             | No                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                     |
| Single Use          | No                                                                                                                                                                                                                    | No                                                                                                                                                                                                                                     |
| Probe Type          | Convex array: 3.5MHz<br>Linear array: 7.5MHz                                                                                                                                                                          | Convex array: 2.0-6.1 MHz<br>Linear array: 4.0-15.0 MHz<br>Phase array: 2.0-4.4 MHz                                                                                                                                                    |
| Operation Mode      | B, M, PWD and Color Doppler modes                                                                                                                                                                                     | B, M, PWD, CWD, Color Doppler and 3D<br>imaging modes                                                                                                                                                                                  |
| Applications        | Abdominal, pediatric, small organ,<br>musculo-skeletal, cardiac, peripheral<br>vascular, trans-vaginal and trans-rectal                                                                                               | Abdominal, pediatric, small organ,<br>musculo-skeletal, cardiac, peripheral<br>vascular, trans-vaginal and trans-rectal                                                                                                                |
| Screen Resolution   | 1024 x 768                                                                                                                                                                                                            | 1024 x 768                                                                                                                                                                                                                             |
| Power Supply        | AC 100-240V, 50/60Hz<br>DC 14.4V                                                                                                                                                                                      | 100-240V, 50/60Hz                                                                                                                                                                                                                      |
| Electrical Safety   | Complied with IEC 60601-1                                                                                                                                                                                             | Complied with IEC 60601-1                                                                                                                                                                                                              |
| EMC                 | Complied with IEC 60601-1-2                                                                                                                                                                                           | Complied with IEC 60601-1-2                                                                                                                                                                                                            |
| Performance         | Complied with IEC 60601-2-37                                                                                                                                                                                          | Complied with IEC 60601-2-37                                                                                                                                                                                                           |

## Table 2 Substantially Equivalent Comparison

{8}------------------------------------------------

| Software Level   | Moderate level of concern  |
|------------------|----------------------------|
| Biocompatibility |                            |
| Cytotoxicity     | Complied with ISO 10993-5  |
| Skin Irritation  | Complied with ISO 10993-10 |
| Sensitization    | Complied with ISO 10993-10 |

The operation modes of the proposed device do not contain CWD and 3D imaging modes, which is different from that of predicate. There are only two types of probes in proposed device, which are less than that of predicate device. And it is not known whether the predicate device can be powered by battery or DC. However, the proposed device has been demonstrated that comply with requirements of electrical safety, EMC, performance and acoustic output. Therefore, the differences do not impact the safety and effectiveness.

#### Substantially Equivalent (SE) Conclusion 9.

Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K182165](https://fda-staging.innolitics.com/device/K182165)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
