← Product Code [GCJ](/productcode/GCJ) · K182160

# 1688 Camera Control Unit ; 1688 AIM 4K Camera Head, C-Mount ; 1688 AIM 4K Camera Head with Integrated Coupler ; AIM 4K Coupler, 20mm, C-Mount ; L11 LED Light Source with AIM (K182160)

_Stryker · GCJ · Nov 5, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K182160

## Device Facts

- **Applicant:** Stryker
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Nov 5, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

1688 4K Camera System with Advance Imaging Modality: The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectorny, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists. L11 LED Light Source with Advanced Imaging Modality: The L11 AIM Light Source and SafeLightTM Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The L11 AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the L1 AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. The L11 AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

## Device Story

AIM System provides real-time 4K visible white light and near-infrared (NIR) illumination/imaging for endoscopic surgery. Components: 1688 Camera Control Unit, 1688 AIM 4K Camera Head, L11 LED Light Source, SafeLight Cable, laparoscopes, IRIS Ureteral Kit, and ICG imaging agent. Operation: Light projected via scope/coupler onto CMOS sensors; continuous image data stream processed into video for display. Light source uses RGB LEDs and NIR laser. Surgeons use system in OR to visualize anatomy, assess tissue perfusion/blood flow, identify biliary ducts via fluorescence, and transilluminate ureters. Output assists surgical navigation and decision-making during minimally invasive procedures.

## Clinical Evidence

No clinical studies required. Evidence consists of bench testing (electrical safety, EMC, laser safety, sterilization per ISO 14937), software validation (IEC 62304), usability testing (IEC 62366-1), and animal performance testing comparing the device to legally marketed predicates.

## Technological Characteristics

System includes 4K CMOS camera, RGB LED/NIR laser light source, fiberoptic cables, and laparoscopes. Connectivity: Digital video output. Standards: IEC 60601-1 (electrical safety), IEC 60601-2-18 (endoscopic equipment), IEC 60601-1-2 (EMC), IEC 60825-1 (laser safety), ISO 14937 (sterilization).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Stryker Infrared Fluorescence Imaging (IRF) System ([K173866](/device/K173866.md), [K151243](/device/K151243.md), [K142310](/device/K142310.md), [K132785](/device/K132785.md))
- PINPOINT Endoscopic Fluorescence Imaging System ([K150956](/device/K150956.md), [K161792](/device/K161792.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 5, 2018

Stryker April Malmborg Director, Regulatory Affairs 5900 Optical Court San Jose, California 95138

Re: K182160

Trade/Device Name: 1688 Camera Control Unit ; 1688 AIM 4K Camera Head, C-Mount ; 1688 AIM 4K Camera Head with Integrated Coupler ; AIM 4K Coupler, 20mm, C-Mount ; L11 LED Light Source with AIM

Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: GCJ, OWN, FCS, FCW Dated: August 7, 2018 Received: August 9, 2018

Dear April Malmborg:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Jennifer R. Stevenson -53

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182160

#### Device Name

1688 Camera Control Unit ; 1688 AIM 4K Camera Head, C-Mount ; 1688 AIM 4K Camera Head with Integrated Coupler ; AIM 4K Coupler, 20mm, C-Mount ; L11 LED Light Source with AIM

#### Indications for Use (Describe)

1688 4K Camera System with Advance Imaging Modality:

The 1688 Video Camera is indicated for use in general laparoscopy, ear endoscopy, ear endoscopy, and plastic surgery whenever a laparoscope/ endoscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectorny, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

#### L11 LED Light Source with Advanced Imaging Modality:

The L11 AIM Light Source and SafeLightTM Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The L11 AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the L1 AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

The L11 AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

| Type of Use (Select one or both, as applicable)                                         |                                                                                        |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <div style="text-align: left;">    Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div style="text-align: left;">    Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of 21 C.F.R Part 807.92(c).

### Submitter:

| Applicant:      | Stryker Endoscopy<br>5900 Optical Court<br>San Jose, CA 95138                                                                                    |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | April Malmborg, RAC<br>Director, Regulatory Affairs<br>Phone: (408) 754-2473<br>Facsimile: (408) 754-2598<br>Email: april.malmborg(a)stryker.com |
| Date Prepared:  | October 29, 2018                                                                                                                                 |

### Subject Device:

The subject device is the AIM (Advanced Imaging Modality) System, and specifically the following system components:

| Name of Device:         | 1688 4K Camera System with Advance Imaging Modality            |
|-------------------------|----------------------------------------------------------------|
| Common or<br>Usual Name | 3-chip Video Camera                                            |
| Classification<br>Name: | Laparoscope, General and Plastic Surgery (21 C.F.R. §876.1500) |
| Regulatory Class:       | II                                                             |
| Product Code:           | GCJ                                                            |
| 510(k) Review<br>Panel: | General & Plastic Surgery                                      |

| Name of Device:         | L11 LED Light Source with Advanced Imaging Modality                                                                                                                      |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or<br>Usual Name | Light Source, Illuminator                                                                                                                                                |
| Classification<br>Name: | Confocal Optical Imaging1 (21 C.F.R. §876.1500)<br>Fiberoptic light ureteral catheter2 (21 C.F.R. §876.4020)<br>Light Source, Fiberoptic, Routine3 (21 C.F.R. §876.4020) |
| Regulatory Class:       | II                                                                                                                                                                       |
| Product Code:           | OWN1<br>FSC2<br>FCW3                                                                                                                                                     |
| 510(k) Review<br>Panel: | General & Plastic Surgery1<br>Gastroenterology/ Urology2,3                                                                                                               |

1When used for assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging

2When used to transilluminate the ureter during open or laparoscopic surgical procedures

3When used to provide standard endoscopic visible light to support real-time endoscopic visible imaging.

{4}------------------------------------------------

# Predicate Device(s):

| Stryker® Infrared Fluorescence Imaging (IRF) System | K173866, K151243,<br>K142310, K132785 |
|-----------------------------------------------------|---------------------------------------|
| PINPOINT Endoscopic Fluorescence Imaging System     | K150956, K161792                      |

NOTE: The predicate devices have not been subject to a design-related recall.

# Device Description:

The AIM (Advanced Imaging Modality) System is an endoscopic real-time 4K visible white light and near-infrared light illumination and imaging system. Near-infrared illumination is used for both fluorescence imaging using indocyanine green (ICG) and transillumination of the ureters during minimally invasive and open surgical procedures, respectively. The AIM (Advanced Imaging Modality) System includes the following components: (1) A Camera System for processing near-infrared and visible light images; (2) A Light Source and SafeLight Cable for emitting light within the visible light as well as near-infrared light spectrum; (3) A Laparoscope for visible light and near-infrared light illumination and imaging; (4) The IRIS Ureteral Kit for transillumination of the ureters; and, (5) Imaging Agent Kits containing ICG used for fluorescence imaging.

# Indications for Use:

# 1688 4K Camera System with Advance Imaging Modality:

The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion. Anterior cruciate ligament reconstruction. Knee arthroscopy, Small joint arthroscopy, Decompression fixation, Wedge resection, Lung biopsy, Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass, Coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists, cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

# L11 LED Light Source with Advanced Imaging Modality:

The L11 AIM Light Source and SafeLight™ Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The L11 AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging.

{5}------------------------------------------------

Fluorescence imaging of biliary ducts with the L11 AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.

The L11 AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

| Item                      | Subject Device                                                                                                                                                                                   | Predicate Devices                                                                                                                                                                                                                                                              |                                                                                                        |                        |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|------------------------|
|                           | AIM (Advanced Imaging Modality) System                                                                                                                                                           | Stryker Infrared Fluorescence (IRF) Imaging System                                                                                                                                                                                                                             | PINPOINT Endoscopic Fluorescence Imaging System                                                        |                        |
| Manufacturer              | Stryker                                                                                                                                                                                          | Same as subject device                                                                                                                                                                                                                                                         | Novadaq Technologies (now a part of Stryker)                                                           |                        |
| Submission Reference      | Current Submission                                                                                                                                                                               | K173866, K151243, K142310, K132785                                                                                                                                                                                                                                             | K150956                                                                                                |                        |
| Intended Use              | Endoscopic visible and near-infrared light illumination and imaging during surgical endoscopic procedures                                                                                        | Same as subject device                                                                                                                                                                                                                                                         | Same as subject device                                                                                 |                        |
| Indications for Use       | NOTE 1                                                                                                                                                                                           | Same as subject device                                                                                                                                                                                                                                                         | NOTE 2 (same as L11 LED Light Source)                                                                  |                        |
| Imaging Modes             | White Light<br>Near-infrared – fluorescence<br>Near-infrared - transillumination                                                                                                                 | Same as subject device                                                                                                                                                                                                                                                         | White Light<br>Near -infrared - fluorescence                                                           |                        |
| Safety Standards          | IEC 60601-1<br>IEC 60601-2-18<br>IEC 60601-1-2<br>IEC 60825-1                                                                                                                                    | Same as subject device                                                                                                                                                                                                                                                         | Same as subject device                                                                                 |                        |
| System Components         | Camera System<br>Light Source and SafeLight Cable<br>Laparoscopes<br>IRIS Ureteral Kits<br>Imaging Agent Kits                                                                                    | Same as subject device                                                                                                                                                                                                                                                         | Video Processor/ Illuminator<br>Camera Head<br>Light Guide Cable<br>Laparoscopes<br>Imaging Agent Kits |                        |
| Camera System             | Principles of Operation                                                                                                                                                                          | Via an optical scope and coupler, light is projected onto one or more complementary metal oxide semiconductor image sensors which acquire a continuous stream of image data. The image data is processed to provide a video stream that is then sent to a display for viewing. | Same as subject device                                                                                 | Same as subject device |
|                           | Image Sensor                                                                                                                                                                                     | CMOS image sensor                                                                                                                                                                                                                                                              | Same as subject device                                                                                 | Same as subject device |
|                           | Image Processing/ Video Output                                                                                                                                                                   | Digital                                                                                                                                                                                                                                                                        | Same as subject device                                                                                 | Same as subject device |
|                           | Resolution                                                                                                                                                                                       | 4K (up to3840 x 2160)                                                                                                                                                                                                                                                          | Up to 1920 x 1080                                                                                      | Up to 1920 x 1080      |
| Light Source              | Principles of Operation                                                                                                                                                                          | An electronic driver controls Red/Green/Blue LEDs & a near-infrared                                                                                                                                                                                                            | Same as subject device                                                                                 | Same as subject device |
| Item                      | Subject Device                                                                                                                                                                                   | Predicate Devices                                                                                                                                                                                                                                                              |                                                                                                        |                        |
|                           | AIM (Advanced Imaging Modality)<br>System                                                                                                                                                        | Stryker Infrared<br>Fluorescence (IRF)<br>Imaging System                                                                                                                                                                                                                       | PINPOINT Endoscopic<br>Fluorescence Imaging<br>System                                                  |                        |
|                           | dichroic mirrors and projected onto an<br>output light collimator. A fiber output<br>bundle can be inserted into the light<br>source to couple light to the distal end<br>and into an endoscope. |                                                                                                                                                                                                                                                                                |                                                                                                        |                        |
| Light<br>Source/<br>Laser | RGB LEDs<br>Infrared Laser                                                                                                                                                                       | Same as subject<br>device                                                                                                                                                                                                                                                      | Same as subject device                                                                                 |                        |

### Comparison of Technological Characteristics with the Predicate Device:

{6}------------------------------------------------

NOTE 1: The 1688 Video Camera is indicated for use in general laparoscopy, nasopharyngoscopy, ear endoscopy, sinuscopy, and plastic surgery whenever a laparoscope/ arthroscope is indicated for use. A few examples of the more common endoscope surgeries are Laparoscopic cholecystectomy, Laparoscopic hernia repair, Laparoscopic appendectomy, Laparoscopic pelvic lymph node detection, Laparoscopically assisted hysterectomy, Laparoscopic and thorascopic anterior spinal fusion, Anterior cruciate ligament reconstruction, Knee arthroscopy, Small ioint arthroscopy. Decompression fixation. Wedge resection. Lung biopsy. Pleural biopsy, Dorsal sympathectomy, Pleurodesis, Internal mammary artery dissection for coronary artery bypass grafting where endoscopic visualization is indicated and Examination of the evacuated cardiac chamber during performance of valve replacement. The users of the 1688 Video Camera are general surgeons, gynecologists. cardiac surgeons, thoracic surgeons, plastic surgeons, orthopedic surgeons, ENT surgeons and urologists.

The L11 AIM Light Source and SafeLight™ Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The L11 AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visual light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the L11 AIM Light Source and SafeLight Cable is intended for use with standard-of-care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. The L1 AIM Light Source is also intended to transilluminate the ureter during open or laparoscopic surgical procedures.

NOTE 2: The PINPOINT Endoscopic Fluorescence Imaging System is indicated for use to provide real time endoscopic visible and near-infrared fluorescence imaging. The PINPOINT System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct or common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the PINPOINT System is intended for use with standard of care white light, and when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

# Performance Data:

| Test              | Method                                                                        | Result |
|-------------------|-------------------------------------------------------------------------------|--------|
| Electrical Safety | ANSI IEC 60601-1:2005 + A1:2012;<br>IEC 60601-2-18:2009<br>IEC 60601-1-6:2013 | PASS   |
| EMC Testing       | IEC 60601-1-2:2010                                                            | PASS   |
| Laser Safety      | IEC 60825-1:2014                                                              | PASS   |
| Sterilization     | ISO 14937:2009                                                                | PASS   |

Testing was completed in accordance with the following:

{7}------------------------------------------------

| Test                                  | Method                                                                                                      | Result |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------|--------|
| Software Validation &<br>Verification | IEC 62304:2006                                                                                              | PASS   |
| Usability                             | IEC 62366-1:2015                                                                                            | PASS   |
| Performance - Bench                   | In accordance with device performance<br>specifications                                                     | PASS   |
| Performance - Animal                  | In accordance with user needs;<br>Comparative testing to currently legally marketed<br>device in compliance | PASS   |

NOTE: The AIM (Advanced Imaging Modality) System does not require clinical studies to support the determination of substantial equivalence.

### Conclusions:

The AIM (Advanced Imaging Modality) System is substantially equivalent in design, intended use, principles of operation, technological characteristics and safety features to the predicate devices. There are no new issues of safety and/or effectiveness introduced by the 1688 4K Camera System or L11 LED Light Source when used as instructed.

---

**Source:** [https://fda-staging.innolitics.com/device/K182160](https://fda-staging.innolitics.com/device/K182160)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).