MONET Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate
K182151 · Ctl Medical Corporation · OVE · Oct 11, 2018 · Orthopedic
Device Facts
Record ID
K182151
Device Name
MONET Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate
Applicant
Ctl Medical Corporation
Product Code
OVE · Orthopedic
Decision Date
Oct 11, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.
Device Story
System comprises interbody fusion cage and supplementary fixation plate; plate and bone screws attach to cage via C-shaped arms and waist tracks. Implanted in cervical spine (C2-C7) to provide mechanical stability and facilitate intervertebral body fusion. Used by surgeons in clinical settings. Provides structural support to disc space; promotes bony fusion. Modular design allows for anatomical variation. Implantation requires device-specific and universal stainless steel instruments.
Clinical Evidence
Bench testing only. Performed static compression, static compression shear, static torsion, dynamic axial compression, and dynamic torsion per ASTM F2077. Additional testing included plate pull-off and screw push-out.
Technological Characteristics
Modular assembly: titanium alloy (Ti-6Al-4V ELI, ASTM F136) cage and fixation plate. Instruments: stainless steel (ASTM F899) and silicone. Mechanical stabilization via C-shaped attachment arms and waist tracks. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients with cervical degenerative disc disease (DDD) and radicular symptoms at one level (C2-C7). Requires six weeks of failed non-operative treatment. Used with autograft bone.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
MONET™ Anterior Cervical Interbody Fusion Cage System (K172788)
MATISSE™ Anterior Cervical Interbody Fusion Cage System (K121569)
MATISSE™ Anterior Cervical Interbody Fusion Cage System (K162682)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 11, 2018
CTL Medical Corporation % Mr. Barry E. Sands President and Founder ROMIS. Inc. 110 Haverhill Road, Suite 526 Amesbury, Massachusetts 01913
Re: K182151
Trade/Device Name: MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: August 1, 2018 Received: August 8, 2018
Dear Mr. Sands:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Melissa Hall -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182151
#### Device Name
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate
#### Indications for Use (Describe)
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.
Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## CTL Medical Corporation's MONET™ ACIF Cage with Supplementary Fixation Plate
#### Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
CTL Medical Corporation Andrea Wilcox 4550 Excel Parkway Site 300 Addison, TX 75001 Phone: 214-545-5820 Fax: 888-831-4892
Contact Person: Mr. Barry E Sands President RQMIS Inc. 110 Haverhill Rd., Suite 526 Amesbury, MA 01913 Phone: 978-358-7307 Facsimile: 978-447-0206 Email: barrysands@rqmis.com
Date Prepared: July 30, 2018
#### Name of Device and Name/Address of Sponsor
Trade Name: MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate CTL Medical Corporation 4550 Excel Parkway Site 300 Addison, TX 75001
#### Common or Usual Name
Intervertebral Body Fusion Device
#### Classification Name
21 CFR 888.3080 Intervertebral Body Fusion Device, Cervical
#### Product Codes: ODP, OVE
{4}------------------------------------------------
## Predicate Devices
## Primary:
K172788, MONET™ Anterior Cervical Interbody Fusion Cage System Additional:
K121569, MATISSE™ Anterior Cervical Interbody Fusion Cage System K162682, MATISSE™ Anterior Cervical Interbody Fusion Cage System
## Intended Use / Indications for Use
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is indicated for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine with accompanying radicular symptoms at one-disc level. DDD is defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly is used to facilitate intervertebral body fusion in the cervical spine at the C2 to C7 disc levels using autograft bone. Patients should have at least six (6) weeks of non-operative treatment prior to treatment with an intervertebral cage. If the MONET Anterior Cervical Interbody Fusion Cage System is used without the Supplementary Fixation Plate assembly, note that supplementary fixation is still required.
## Device Description
The subject device consists of a plate and bone anchoring screws which are made of titanium alloy (Ti-6Al-4V ELI) that conforms to ASTM F136. The anterior fixation plates can only be assembled with the predicate K172788 cage prior to final use. The final assembly constructs are implanted using a combination of device specific and universal instruments manufactured from stainless steel materials that conform to ASTM F899.
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate is considered an enhanced and modular line-extension to already cleared MONET™ Anterior Cervical Interbody Fusion Cage System (K172788).
The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate have the same risk classification, intended use, and generic designation, as the predicate system the MONET™ Anterior Cervical Interbody Fusion Cage System (K172788). The system is offered in various sizes and configurations to accommodate anatomical variation in the cervical vertebral levels and/or patient anatomy.
{5}------------------------------------------------
## Technological Characteristics
MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate assembly consists of two modular components: The first component is the previously cleared K172788, the MONET™ Anterior Cervical Interbody Fusion Cage and the second component is the complimentary reinforcement hardware.
MONET™ ACIF cage and the supplementary fixation plate are mechanically held together by the C-shaped "attachment arms" that mate with the "waist tracks". The Cshaped mating features provide additional stability to the MONET™ ACIF cage by increasing the footprint of the device within the disc space.
When fully assembled, the MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate provides mechanical support/stability to cervical spine segment to allow for bony fusion to occur.
## Material Composition
The subject supplementary fixation plate hardware (modular disc plates and bone anchoring screws) are made of the same implantable grade titanium alloy (Ti-6Al-4V, ASTM F136).
All subject instruments that are included as part of this submission and comes in contact with human tissue are manufactured from stainless steel materials. Instrument handles that will never touch human tissue are manufactured from stainless steel and silicone.
## Performance Data
The following test were performed per ASTM F2077:
- Static Compression .
- Static Compression Shear
- Static Torsion
- Dynamic Axial Compression ●
- Dynamic Torsion
In addition, the following tests were performed.
- Plate Pull-Off
- . Screw Push out
#### Substantial Equivalence
The MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate is as safe and effective as the predicate MONET™ and MATISSE™
{6}------------------------------------------------
Anterior Cervical Interbody Fusion Cage System, Titanium and PEEK (K172788, K162682 and K121569). The MONET™ Anterior Cervical Interbody Fusion Cage System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The reinforcement related technological differences between the MONET™ Anterior Cervical Interbody Fusion Cage System with Supplementary Fixation Plate and its predicate device do not raise different questions of safety or effectiveness. Thus, the MONET™ Anterior Cervical Interbody Fusion Cage System is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.