K182142 · Jiangsu Yoai Technology Co., Ltd. · HEB · Aug 28, 2019 · Obstetrics/Gynecology
Device Facts
Record ID
K182142
Device Name
FEMME Applicator Tampon
Applicant
Jiangsu Yoai Technology Co., Ltd.
Product Code
HEB · Obstetrics/Gynecology
Decision Date
Aug 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5470
Device Class
Class 2
Indications for Use
The FEMME Applicator Tampon is an unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the tampon correctly into the vagina.
Device Story
FEMME Applicator Tampon is an unscented, menstrual tampon designed for intravaginal absorption of menstrual or vaginal discharge. Device consists of an absorbent viscose pledget, overwrap, removal string, and a three-part plastic applicator (barrel, grip, plunger). Applicator features a rounded, mushroom-shaped tip to facilitate insertion. Available in Regular, Super, and Super plus absorbency levels. Used by women for personal hygiene; OTC availability. Device functions by physical absorption of fluids; applicator serves as a mechanical aid for correct placement. Benefits include ease of use and effective fluid management. No electronic or software components.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-5, -10, -11), microbiological safety (bioburden, effect on vaginal microflora, S. aureus growth, and TSST-1 production), chemical residue analysis (PCDD/Fs per EPA 1613B, pesticides per EN 15662), and physical performance (Syngyna absorbency per 21 CFR 801.430, removal string strength per AS/NZS 2869, fiber shedding, and integrity).
Technological Characteristics
Materials: Viscose pledget, polyester/polyethylene overwrap, cotton/polyester removal string, LDPE/thermoplastic elastomer applicator. Design: Compact, mushroom-shaped tip, straight grooves. Absorbency: 6.0-15.0g (Regular to Super plus). Standards: ISO 10993 (biocompatibility), ISO 11737-1 (bioburden), 21 CFR 801.430 (absorbency), AS/NZS 2869 (string strength). Non-sterile. No software or energy source.
Indications for Use
Indicated for women for personal hygiene, specifically for the intravaginal absorption of menstrual or other vaginal discharge.
Regulatory Classification
Identification
An unscented menstrual tampon is a device that is a plug made of cellulosic or synthetic material that is inserted into the vagina and used to absorb menstrual or other vaginal discharge. This generic type of device does not include menstrual tampons treated with scent (i.e., fragrance materials) or those with added antimicrobial agents or other drugs.
Predicate Devices
Opal by Femtex and Private Label Unscented Compact Applicator Tampon (K151470)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
August 28, 2019
Jiangsu YOAI Technology Co., Ltd. c/o Field Fu Official Correspondent Shenzhen Joyantech Consulting Co., Ltd. Room 1122, No. 55 Shizhou Middle Road, Nanshan District Shenzhen, GD755 Guangdong CHINA
Re: K182142
Trade/Device Name: FEMME Applicator Tampon Regulation Number: 21 CFR 884.5470 Regulation Name: Unscented Menstrual Tampon Regulatory Class: II Product Code: HEB Dated: July 10, 2019 Received: July 29, 2019
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sharon Andrews Assistant Division Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K182142
Device Name FEMME Applicator Tampon
Indications for Use (Describe)
The FEMME Applicator Tampon is an unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the tampon correctly into the vagina.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 1. Submission Sponsor
-
| Applicant Name | Jiangsu Yoai Technology Co., Ltd. |
|----------------|------------------------------------------------------------------------------------|
| Address | Building No.3, No.33 Jing You Road, Jiangning<br>District, Nanjing, Jiangsu, China |
| Phone No. | 86-025-69620808 |
| Contact Person | Ms. Huan Wang |
| Email | huan.wang@yoai.com |
| Date Prepared | August 22, 2019 |
### 2. Submission correspondent
| Name | Shenzhen Joyantech Consulting Co., Ltd |
|----------------|----------------------------------------------------------------------------------------------------|
| Address | Room 1122, No.55 Shizhou Middle Road, Nanshan<br>District,<br>Shenzhen,<br>Guangdong,<br>P.R.China |
| Image: [logo] | |
| Post Code | 518000 |
| Phone No. | 86-755-86069197 |
| Contact Person | Mr. Field Fu; Ms. Jessie You; Ms. Elly Xu |
| Email | Jessie@cefda.com; elly@cefda.com |
## 3. Device Identification
| Trade name | FEMME Applicator Tampon |
|---------------------|-------------------------------|
| Common name | Unscented Menstrual Tampon |
| Specifications | Regular, Super and Super plus |
| Classification | II |
| Classification name | Tampon, Menstrual, Unscented |
| Product code | HEB |
| Regulation Number | 21 CFR 884.5470 |
{4}------------------------------------------------
Subject device: FEMME Applicator Tampon
### 4. Predicate Device Information
Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Light Absorbency; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Regular & Super; Opal by Femtex and Private Label Unscented Compact Applicator Tampon, Super Plus; Opal by Femtex and Private Label Unscented (K151470). The predicate device has not been subject to a design-related recall.
## 5. Device Description
The FEMME Applicator Tampon is an unscented, menstrual tampon. The tampon is comprised of an absorbent viscose pledget, an overwrap, a removal string, and an applicator. The applicator contains 3 parts: a barrel, grip, and plunger. The tampon was designed as a compact tampon with a mushroom-shaped tip. The pledget is of the traditional cylindrical bullet-like shape and is covered with a thin layer of overwrap. The applicator has a standard rounded and smooth tip to ease insertion. Each tampon is individually wrapped and packaged in multi-unit container for retail sale. There are 3 specifications of tampon with different absorbencies: Regular, Super and Super plus. Each absorbency specification has a different color removal string and applicator.
## 6. Indications for Use Statement
The FEMME Applicator Tampon is an unscented tampon for women's personal hygiene with respect to intra vaginal absorption of menstrual or other vaginal discharge. The plastic applicator is for easing the placement of the tampon correctly into the vagina.
#### 7. Substantial Equivalence Discussion
| | Subject Device | Predicate Device |
|---------------|----------------------------|----------------------------------------------------------------------------|
| | FEMME Applicator<br>Tampon | Opal by Femtex and Private<br>Label Unscented Compact<br>Applicator Tampon |
| 510(k) number | K182142 | K151470 |
{5}------------------------------------------------
## Subject device: FEMME Applicator Tampon
| Regulation | | | | |
|---------------------|----------------------------|------------------------|----------------------------------|---------------|
| Product Code | HEB | | HEB | |
| Indications for use | The FEMME Applicator | | Unscented tampon for | |
| | | Tampon is an unscented | women's personal hygiene with | |
| | tampon for women's | | respect to intra vaginal | |
| | personal hygiene with | | absorption of menstrual or | |
| | respect to intra vaginal | | other vaginal discharge. The | |
| | absorption of menstrual or | | plastic applicator is for easing | |
| | other vaginal discharge. | | the placement of the tampon | |
| | The plastic applicator is | | correctly into the vagina. | |
| | for easing the placement | | | |
| | of the tampon correctly | | | |
| | into the vagina. | | | |
| Components | The FEMME Applicator | | The Unscented Compact | |
| | Tampon is comprised of | | Applicator Tampon is | |
| | an absorbent pledget and | | comprised of an absorbent | |
| | an applicator. | | pledget and an applicator. | |
| Sterile? | No | | No | |
| Design | Compact tampon with | | Compact tampon with | |
| | mushroom-shaped tip and | | mushroom-shaped tip and | |
| | straight grooves. | | straight grooves. | |
| Syngyna Absorbency, | N/A | | Light | < 6.0 |
| grams | Regular | 6.0 - 9.0 | Regular | 6.0 - 9.0 |
| | Super | 9.0 - 12.0 | Super | 9.0 - 12.0 |
| | Super plus<br>12.0 - 15.0 | | Super plus | 12.0 - 15.0 |
| Pledget length (mm) | N/A | | Light | 40.0 - 45.0 |
| | Regular | 42.0 - 52.0 | Regular | 40.0 - 45.0 |
| | Super | 47.0 - 52.0 | Super | 43.0 - 46.0 |
| | Super plus | 47.0 - 52.0 | Super plus | 43.0 - 46.0 |
| Pledget diameter | | N/A | Light | 8.7 – 9.7 |
| | Regular | 10.6 – 12.6 | Regular | 8.7 – 9.7 |
| | Super | 12.0 - 14.0 | Super | 11.0 - 13.0 |
| | Super plus | 14.0 - 16.0 | Super plus | 13.0 - 15.0 |
| | N/A | | Light | 66.0 — 68.0 x |
| | | | | 10.5 - 11.5 |
| Applicator length x | Regular | 78.5 — 79.5 x | Regular | 66.0 — 68.0 x |
| diameter (mm, Inner | | 6.4 - 6.5 | | 10.5 — 11.5 |
{6}------------------------------------------------
#### Subject device: FEMME Applicator Tampon
| tube-plunger) | | Super | $78.5 - 79.5 x$<br>$6.4 - 6.5$ | Super | $64.0 - 66.0 x$<br>$13.5 - 14.5$ |
|-----------------------------------------------------|-----------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | | Super plus | $78.5 - 79.5 x$<br>$6.4 - 6.5$ | Super plus | $65.5 - 68.5 x$<br>$15.5 - 16.5$ |
| Applicator length x<br>diameter (mm, Outer<br>tube) | | N/A | | Light | $58.0 - 60.0 x$<br>$13.0 - 14.0$ |
| | | Regular | $77.0 - 78.0 x$<br>$14.5 - 15.0$ | Regular | $58.0 - 60.0 x$<br>$13.0 - 14.0$ |
| | | Super | $77.0 - 78.0 x$<br>$14.5 - 15.0$ | Super | $59.0 - 61.0 x$<br>$15.0 - 16.0$ |
| | | Super plus | $77.5 - 78.5 x$<br>$16.5 - 17.0$ | Super plus | $60.0 - 62.0 x$<br>$17.0 - 18.0$ |
| Removal String | | Cotton, Polyester<br>Violet, green, and blue<br>colorants with the<br>following characteristics:<br>• 21 CFR 178.3297<br>compliant | | 100% Hydrophobic Cotton | |
| Pledget | | Viscose | | Viscose | |
| Overwrap | | Polyester, Polyethylene | | Polyethylene/Polyethylene<br>terephthalate | |
| Wrapper (Primary<br>Packaging) | | Polyethylene | | Polypropylene | |
| Applicator | Plastic<br>Tube | Low Density Polyethylene<br>(LDPE), Thermoplastic<br>elastomer | | Low Density Polyethylene | |
| | Colorant | Purple, green, and blue<br>colorants with the<br>following characteristics: | | Green, Pink, Blue, and Pearl<br>White colorants with the<br>following characteristics:<br>• Meets Toxics Substance<br>Control Act Inventory (TSCA)<br>• 21 CFR 178.3297 and/or EU<br>Resolution (89)1 compliant<br>• Extractables testing<br>confirmed that no harmful<br>compounds are<br>bioavailable | |
{7}------------------------------------------------
Subject device: FEMME Applicator Tampon
| | | 178 |
|--|--------------------------|-----------------------------------------------------------------------|
| | Mold<br>Release<br>Agent | N/A<br>Food grade mold release agent—conforms with 21 CFR 177 – 178 |
| | Antistatic<br>Agent | N/A<br>Food grade antistatic agent—<br>conforms with 21 CFR 177 – 178 |
### 8. Non-Clinical Performance Data
The evaluation is in accordance with FDA guidance: Menstrual Tampons and Pads: Information for Premarket Notification Submissions (510(k)s).
#### 8.1 Biocompatibility testing
Biocompatibility assessments for the FEMME Applicator Tampon were conducted in accordance with FDA guidance document Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". Biocompatibility testing was selected for a surface contacting device having "limited contact" (i.e., < 24 hrs) with intact skin and mucous membranes. The battery of testing included the following tests:
- A. FEMME Applicator Tampon - applicator
- In Vitro Cytotoxicity (ISO 10993-5: 2009) 1)
- 2) Intracutaneous study (ISO 10993-10: 2010)
- 3) Guinea pig maximization sensitization study (ISO 10993-10: 2010)
- FEMME Applicator Tampon pledget B.
- 1) Cytotoxicity study (ISO 10993-5: 2009)
- 2) Vaginal irritation study (ISO 10993-10: 2010)
- 3) Guinea pig maximization sensitization study (ISO 10993-10: 2010)
- Systemic toxicity study (ISO 10993-11: 2017) 4)
#### 8.2 Performance testing - Bench
The performance tests were conducted to demonstrate the effectiveness of device:
- A. Microbiological testing:
- 1) Bioburden (ISO 11737-1: 2006): the bioburden of the subject device was 5.6 CFU/tampon:
{8}------------------------------------------------
Subject device: FEMME Applicator Tampon
- 2) The effect of tampon on growth of normal vaginal microflora (USP 71): the subject device does not alter the growth of normal vaginal microflora;
- 3) The effect of tampon on growth of Staphylococcus aureus (USP 71): The subject device does not enhance the growth of Staphylococcus aureus (S. aureus) compared to an S. aureus control;
- 4) The effect of tampon on production of S. aureus TSST-1: the subject device did not increase production of Toxic Shock Syndrome Toxin-1 (TSST-1) compared to an S. aureus control.
- B. Chemical residue testing:
- Polychlorinated dibenzo-p-dioxins and polychlorinated dibenzofurans 1) residues (PCDD/Fs) testing was performed on the subject device per EPA 1613B:1997: PCDD/Fs concentrations were N.D. (not detectable).
- 2) Pesticide and CI residue testing was performed on the subject device per EN 15662: 2008: Pesticides and Cl residue concentrations were N.D. (not detectable).
- C. Physical performance testing:
- Absorbency testing for the FEMME Tampon Applicator was performed per 1) the 21 CFR 801.430 Syngyna absorbency protocol. The absorbency ranges for all specifications met the requirements of 21 CFR 801.430.
- 2) Removal string strength for the subject device was performed per AS/NZS 2869: 1998. Each tampon tested for every absorbency level met the Standard's minimum attachment strength requirement of 22.4 Newtons (N). Each absorbency data set met an average value of ≥ 28 N, which is the standard's requirement for each production lot.
- Fiber shedding testing was evaluated using internal protocols, since no 3) recognized standard is available. The subject device does not exhibit the potential to shed more than the predicate fiber.
- 4) Tampon integrity was evaluated using internal protocols, since no recognized standard is available. The subject device meets the visual and observational requirements set for an acceptable subject device.
#### 9. Conclusion
The FEMME Applicator Tampon characteristics and comparisons detailed above demonstrate that the subject device has the same fundamental design, Indications for Use, technology, and performance specifications as the predicate device. Non-Clinical Performance Testing demonstrates that the new device meets the absorbency
{9}------------------------------------------------
#### Subject device: FEMME Applicator Tampon
requirements defined by 21 CFR 801.430, which confirms the device's effectiveness for the Indications for Use. Biocompatibility testing, microbiology testing, and physical performance testing demonstrate that the new device has a safe biocompatibility profile, does not increase the potential to promote TSS relative to the predicate device, and has a tampon integrity profile that is equivalent to the predicate device. The data support the conclusion that the FEMME Applicator Tampon by Jiangsu Yoai Technology Co., Ltd. is substantially equivalent to the predicate device and is safe and effective for its Indications for Use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.