The Elekta Medical Linear Accelerator is indicated to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g. lesions, arterio-venous malformations, malignant and benign tumors), whilst sparing surrounding normal tissue and critical organs from excess radiation for treatment that includes but is not limited to, malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.
Device Story
Medical linear accelerator system for image-guided radiation therapy; delivers ionizing radiation to defined target volumes (lesions, tumors, malformations) while sparing surrounding healthy tissue. Used in clinical settings by licensed practitioners (radiation oncologists/therapists). System includes beam shaping components, imaging, and patient positioning accessories. Device transforms treatment plans into precise radiation delivery. Output assists clinicians in targeting malignancies; enables SRS and SBRT procedures. Benefits include localized tumor control with reduced collateral damage to critical organs. Subject of this notification is an update to the Integrity control software to improve error detection and unify the codebase across hardware platforms.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included module, integration, and system-level testing per FDA Quality System Regulation (21 CFR 820), ISO 13485, ISO 14971, and IEC 62304. Conformance to consensus standards (IEC 60601-1, IEC 60601-2-1, IEC 60601-1-6, IEC 62366-1) confirmed basic safety and essential performance.
Technological Characteristics
Medical linear accelerator with beam shaping and imaging. Software: Integrity control software. Standards: ISO 13485, ISO 14971, IEC 62304, IEC 60601-1, IEC 60601-2-1, IEC 60601-1-6, IEC 62366-1. No changes to hardware technological characteristics.
Indications for Use
Indicated for patients requiring radiation therapy for malignant and benign tumors, including brain tumors, brain metastases, spine lesions (SRS), head and neck squamous cell carcinoma, lung, breast, pancreatic, hepatic malignancies (SBRT), prostate cancer, and bone metastases.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. Food & Drug Administration' in blue text.
September 5, 2018
ELEKTA LIMITED % Mr. Andrew Hedges Principal Engineer - RA and Compliance Linac House, Fleming Way Crawley, West Sussex RH10 9RR UNITED KINGDOM
Re: K182138
Trade/Device Name: Precise Treatment System™; Synergy® platform; Synergy®; Infinity TM; VersaHD TM Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: Class II Product Code: IYE Dated: July 31, 2018 Received: August 7, 2018
Dear Mr. Hedges:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm_identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Bollyer
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
K182138
#### Device Name
Precise Treatment System, Synergy Platform, Synergy, Infinity and Versa HD
#### Indications for Use (Describe)
The Elekta Medical Linear Accelerator is indicated to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g. lesions, arterio-venous malignant and benign tumors), whilst sparing surrounding normal tissue and critical organs from excess radiation for treatment that includes but is not limited to, malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
The following information follows the format of 21 CFR 807.92
# Date of preparation of summary: 31th July 2018
Submitted by:
| | Elekta Limited<br>Linac House, Fleming Way, Crawley, West Sussex<br>RH10 9RR, United Kingdom<br>Telephone: +44 (0)1293 654201 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | Fax: +44 (0)1293 654321 |
| Contact name: | Andrew Hedges |
| Trade name: | Precise Treatment System™, Elekta Synergy® Platform,<br>Elekta Synergy®, Elekta Infinity ™ and Versa HD™ |
| Common Name: | Medical Linear Accelerator System |
| Classification Name: | Medical Charged-Particle Radiation Therapy System<br>Accelerator, Linear, Medical, 21CFR 892.5050 |
| Product Code: | 90 IYE |
| Predicate Device: | Elekta Limited, K123808 Agility, K051932 Elekta Synergy |
# Product Description:
The Elekta Medical Linear Accelerator system is an image guided Radiation Therapy device to assist a licensed practitioner in the delivery of ionizing radiation to a defined target volume. The system consists of components of the accelerator, such as, beam shaping, with imaging and accessories for patient positioning and set-up to deliver therapeutic treatments. The Elekta Medical Linear Accelerator System is currently available in the following model variants – Precise Treatment System, Elekta Synergy Platform, Elekta Infinity and Versa HD.
# Intended Use / Indication For Use Statement:
The Elekta Medical Linear Accelerator is indicated to assist a licensed practitioner in the delivery of radiation to defined target volumes (e.g. lesions, arterio-venous malformations, malignant and benign tumors), whilst sparing surrounding normal tissue and critical organs from excess radiation for treatment that includes but is not limited to, malignant and benign brain tumors, brain metastases, spine lesions treated using SRS, squamous cell carcinoma of the head and neck, lung, breast, pancreatic, hepatic malignancies treated using SBRT, prostate, and bone metastases.
# Summary of Technological Characteristics:
This premarket notification is not related to any change in the technological characteristics of the medical linear accelerator system and these are unchanged from those of the previously cleared medical device. Elekta has introduced changes to the control software, Integrity™, primarily to provide tighter error detection and to merge the codebase to provide a single release that supports the current hardware platform. There are no novel forms of technology introduced in this premarket notification.
# Substantial Equivalence
The functionality of the Elekta Medical Linear Accelerator system is substantially equivalent to that of its predicate device Agility K123808 and its associated linear accelerator, devices in safety and effectiveness. The intended use of the system, principles of operation, technological characteristics are substantially equivalent.
{4}------------------------------------------------
# Summary of performance testing (non clinical)
Testing in the form of module, integration and system level verification was conducted in accordance with FDA Quality System Regulation (21 CFR §820), ISO 13485 Quality Management System standard, ISO 14971 Risk Management Standard, IEC 62304 Software life-cycle processes, and the other FDA recognised consensus standards which includes but is not limited to IEC 60601-1, IEC 60601-2-1, IEC 60601-1-6, IEC 62366-1. Software verification testing was conducted and documented in accordance with FDA's "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" for devices that pose a major level of concern (Class C per IEC 62304). Basic safety and essential performance have been satisfied through conformance with device-specific recognised consensus standards, as well as applicable general and collateral safety and essential performance standards for medical devices.
Validation of the integrated system under clinically representative conditions has been performed by competent and professionally qualified personnel. Results from verification and validation testing demonstrate that conformance to applicable technical design specifications have been met and safety & effectiveness has been achieved.
#### Conclusion:
The results of verification, validation and safety standard testing demonstrate that the Elekta Medical Linear Accelerator system is substantially equivalent to their predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.