K182133 · Shenzhen AOJ Medical Technology Co., Ltd. · FLL · Nov 20, 2018 · General Hospital
Device Facts
Record ID
K182133
Device Name
Infrared Thermometer
Applicant
Shenzhen AOJ Medical Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Nov 20, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared thermometers (AOJ-20A and AOJ-20B) take human body temperature via the eardrum or forehead. They apply to all age groups except for babies under three months. Both devices apply to both professional use and home use.
Device Story
Handheld infrared thermometer; models AOJ-20A and AOJ-20B; measures human body temperature via eardrum or forehead. Device uses thermopile sensor to detect infrared thermal energy emitted from skin or eardrum; converts energy into temperature reading in Celsius or Fahrenheit. Powered by two 1.5V AAA alkaline batteries; results displayed on LCD. Used in clinical or home settings by professionals or patients. Device includes memory for 20 measurements. Provides non-invasive temperature monitoring; aids healthcare providers in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical study conducted per ASTM E1965-98 using a randomized, single-blind, homologous control, pairing design. Study included at least 90 subjects (1/3 children, 1/3 adults). Primary endpoint: clinical accuracy compared to reference device. Acceptance criteria: accuracy ±0.2°C within 32°C–42.9°C range, 99% confidence. All testing passed pre-specified criteria.
Indicated for human body temperature measurement via eardrum or forehead in all age groups except infants under three months. Suitable for professional and home use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 20, 2018
Shenzhen AOJ Medical Technology Co., Ltd. Qihuan Zhao General Manager Room 202, HaoGu Industry Park, 2037 Guanguang Road, Guangming District, Shenzhen Guangdong, 518105 China
Re: K182133
Trade/Device Name: Infrared Thermometer, models AOJ-20A and AOJ-20B Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: October 12, 2018 Received: October 22, 2018
Dear Oihuan Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Geeta K. Pamidimukkala -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182133
Device Name
Infrared thermometer, models AOJ-20A and AOJ-20B
Indications for Use (Describe)
The Infrared thermometers (AOJ-20A and AOJ-20B) take human body temperature via the eardrum or forehead. They apply to all age groups except for babies under three months. Both devices apply to both professional use and home use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# K182133
# Prepared in accordance with the requirements of 21 CFR Part 807.92
| 1. Submitter: | Shenzhen AOJ Medical Technology Co., Ltd. |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Room 202, HaoGu industry park, 2037 Guanguang road, Guangming district,<br>518105, Shenzhen, P.R. China |
| | Tel.: +86 18603031299 |
| Contact Person: | Qihuan Zhao |
| Prepare date: | 2018-11-15 |
| 2. Device name<br>and classification: | Device Name: Infrared Thermometer<br>Models: AOJ-20A, AOJ-20B |
| | Regulation Number: 21 CFR 880.2910 |
| | Regulation Name: Clinical Electronic Thermometer |
| | Product code: FLL |
| | Regulatory Class: Class II |
| 3. Predicate<br>Device(s): | Shenzhen Brav Electronic Technologies Co., Ltd., EFT-165 Infrared Thermometer<br>/K171214 |
| 4. Device<br>Description: | The infrared thermometers (AOJ-20A and AOJ-20B) are handheld instruments,<br>which can measure human body's temperature either via the eardrum or the<br>forehead for clinical or home use. The intended patient population is all age<br>groups except for babies under three months. |
| | The thermometers are powered by AAA 1.5V×2 alkaline batteries. The results<br>can be displayed on LCD. |
| | A thermopile sensor is employed to detect or monitor the infrared thermal energy<br>emitted from the eardrum or the surface of the skin of the forehead, which is<br>converted into temperature measurement with the unit of °C or °F. |
| 5. Indications for<br>Use: | The Infrared thermometers (AOJ-20Aand AOJ-20B) take human body temperature<br>via the eardrum or forehead. They apply to all age groups except for babies under<br>three months. Both devices apply to both professional use and home use. |
{4}------------------------------------------------
# 6.Predicate Device Comparison
| ITEM | Proposed Device<br>AOJ-20A&AOJ-20B | Predicate Device<br>EFT-165/K171214 | Comparison<br>Result |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Manufacture | Shenzhen AOJ Medical<br>Technology Co., Ltd. | Brav Electronic Technologies<br>Co., Ltd. | --- |
| Indications for<br>Use | The Infrared thermometers<br>(AOJ-20Aand AOJ-20B) take<br>human body temperature via the<br>eardrum or forehead. They apply<br>to all age groups except for<br>babies under three months. Both<br>devices apply to both<br>professional use and home use. | The infrared thermometer is<br>intended for the measurement<br>and monitoring of human body<br>temperature by doctor or<br>customers in the hospital or<br>home. | Different 1 |
| Operational Specifications | | | |
| Operational<br>Principle | Infrared radiation detection | Infrared radiation detection | Same |
| Measuring Mode | Forehead and ear | Forehead and ear | Same |
| Measurement<br>Range | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | 32.0°C~42.9°C<br>(89.6°F~109.2°F) | Same |
| Measurement<br>Distance | 0 cm | 0 cm | Same |
| Accuracy | ±0.2°C(0.4°F) | ±0.2°C(0.4°F) | Same |
| Memory Data<br>Limit | memorize 20 measurements<br>automatically | memorize 20 measurements<br>automatically | Same |
| Product<br>configuration | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | It is mainly composed with<br>infrared sensor, signal receiving<br>processor, buttons, buzzer, LCD<br>display, battery and etc. | Same |
| Temperature unit<br>and conversion | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Dual temperature units “°C” and<br>“°F” optional, and the two units<br>can convert by the conversion<br>key automatically | Same |
| Applicable<br>Standards | IEC 60601-1, IEC 60601-1-2,<br>IEC 60601-1-11, and ASTM<br>E1965-98 | AAMI ANSI ES60601-1, IEC<br>60601-1-2, IEC 60601-1-11, and<br>ASTM E1965-98 | Same |
| Display | 0.1°C/°F, LCD | 0.1°C/°F, LCD | Same |
| Operating<br>Environment | Temperature: 10°C~40°C<br>(50°F-104°F)<br>Humidity: 15%-95% RH,<br>non-condensing<br>Atmospheric pressure: 70–106<br>kPa | 15°C ~ 35°C (59°F to 95 °F)<br><85% moisture condensation | Different 2 |
| Storage<br>Environment | Ambient Temperature: -20°C to<br>55°C (-4°F-131°F)<br>Relative Humidity: 0-95% RH,<br>non-condensing<br>Atmospheric pressure: 50kPa to<br>106kPa | -20°C~55°C(-4°F~131°F)≤90%<br>moisture condensation | Same |
Table 1 Comparison between main predicate EFT-165 and the subject device
{5}------------------------------------------------
| Physical Specifications | | | |
|------------------------------------|-----------------------------------------------------------|-----------------------------------------------------------|-------------|
| Weight | 90g (battery included) | 72 g (without battery) | Different 3 |
| Dimensions | 146mm *52 mm *40 mm | 152mm * 44 mm *30mm | |
| Biological Specifications | | | |
| Patient<br>Contacting<br>Materials | PC+ABS | PC+ABS | Same |
| Patient<br>Contacting | Surface-contacting,<br>Less than 24 h | Surface-contacting,<br>Less than 24 h | Same |
| Biocompatibility<br>Standard | ISO 10993-5,<br>ISO 10993-10 | ISO 10993-5,<br>ISO 10993-10 | Same |
| Biocompatibility<br>Testing Items | In vitro Cytotoxicity<br>Skin Sensitization<br>Irritation | In vitro Cytotoxicity<br>Skin Sensitization<br>Irritation | Same |
#### Justification for the differences:
- 1) Different Indications for Use
As indicated in the comparison table, the subject device and the predicate device have similar indications for use, the subject device includes the patient population and areas to take temperature. The intended use is the same.
- 2) Different Operation&Storage&Transport Environments
Minor difference to operation environments (including Temperature and Relative Humidity) for the subject device, but the system has demonstrated to be as safe and effective as the predicate device since the safety testing was conducted under the suggested environment; Moreover, environment testing data shows the device can work as intended under the suggested conditions. So those changes do not raise any safety and effectiveness concerns.
- 3) Different Physical Specifications
The subject and the predicate are of different size but proximity. Moreover, such engineering design has been verified during development by compliance with the applicable international standards, so such minor different will not raise any new problems.
As seen in the comparison tables, the subject and predicate devices have almost the same design features and performance specifications. Moreover, as demonstrated in the non-clinical and clinical testing, AOJ-20A&AOJ-20B Infrared thermometer system can perform as intended.
## 7.Performance Testing
## 7.1 Non-Clinical Data
The following safety standards are conducted on the subject device support of the substantial equivalence determination, including performance testing and bench testing.
## Performance Testing Information:
- (1) ISO 10993-5 Biological evaluation for medical device Part 5 Test for In Vitro Cytotoxicity
- (2) ISO 10993-10 Biological evaluation for medical device Part 10 Test for irritation and skin sensitization
- (3) AAMI TIR12:2004 Designing, testing, and labeling reusable medical devices for reprocessing in health care facilities: A guide for medical device manufacturers, Approved 23 December 2004.
{6}------------------------------------------------
## Bench Testing Information:
- (1) ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009((R)2012 and A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
- (2) IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- (3) IEC 60601-1-11: 2010 MEDICAL ELECTRICAL EQUIPMENT -Part 1-11: General requirements for basic safety and essential performance –Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- (4) IEC 60601-1-2: 2014 Medical electrical equipment =Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
- (5) ISO 80601-2-56: 2009 Medical electrical equipment Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement for performance effectiveness.
- (6) Software verification and validation testing per FDA's Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued on May 11, 2005.
# 7.2 Clinical Data
Clinical testing is conducted per ASTM E 1965-98 (Reapproved 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consists of a minimum of 90 subjects, of which 1/3 are children and the rest 1/3 are adults (NOTE: Infants---newborn to one year; Children--- greater than one to five years; Adults--- greater than five years old.). with no single measurement error exceeding the allowable limit, provides a 99% confidence. A reference device is introduced as required by the standard, and the acceptance criteria is that clinical accuracy of the subject device is at least the same as that of the reference, and the accuracy is ±0.2℃ during the claimed measurement range 32℃ - 42.9℃.
# 8.Conclusion
Verification and validation testing was conducted on the subject device and all testing passed pre-specified criteria. This premarket notification submission demonstrates that the AOJ-20A & AOJ-20B Infrared Thermometer is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.