K182129 · Siemens Medical Solutions USA, Inc. · LNH · Oct 12, 2018 · Radiology
Device Facts
Record ID
K182129
Device Name
MAGNETOM Sola
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Oct 12, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Sola is a 1.5T MRI system utilizing XJ gradient hardware; inputs include RF signals from patient anatomy via specialized coils; system transforms signals into cross-sectional images/spectra using syngo MR XA11A software platform; operated by healthcare professionals in clinical settings; features include BioMatrix sensors/tuners, Dot Engines for workflow automation, and advanced pulse sequences (e.g., Compressed Sensing, CAIPIRINHA, SMS); output displayed on monitors for physician interpretation to assist in diagnosis; supports interventional procedures with MR-compatible equipment; benefits include improved image quality, workflow efficiency, and patient-specific adjustments.
Clinical Evidence
Bench testing only. Clinical study of 40 individuals conducted solely to determine nerve stimulation thresholds for gradient system output (PNS compliance per IEC 60601-2-33). No diagnostic clinical trials performed; sample clinical images provided to support new coil performance.
Technological Characteristics
1.5T MRI system; XJ gradient system; 32 RF channels. Conforms to IEC 60601-1, IEC 60601-2-33, ISO 14971, IEC 62304, NEMA MS 4, MS 9, and DICOM standards. Software platform syngo MR XA11A. Features BioMatrix technology (sensors, tuners, interfaces).
Indications for Use
Indicated for patients requiring diagnostic MR imaging of the head, body, or extremities; used for cross-sectional imaging, spectroscopy, and interventional procedures; compatible with MR-safe devices; no specific age or gender restrictions.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
October 12, 2018
Siemens Medical Solutions USA, Inc. Cordell Fields Regulatory Specialist 65 Valley Stream Parkway 65-1A MALVERN, PA 19355
Re: K182129
Trade/Device Name: MAGNETOM Sola Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH Dated: October 2, 2018 Received: October 3, 2018
Dear Cordell Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Hsl 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use Section 4 Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K182129 Device Name
MAGNETOM Sola
Indications for Use (Describe)
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremittes. Other physical parameters derived from the images and or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
| Type of Use (Select one or both, as applicable) | <table><tr><td><span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span></td><td><span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span></td></tr></table> | <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
| FORM FDA 3881 (7/17) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|
#### Siemens 510(k) Traditional Premarket Notification
August 3, 2018
Siemens MR System MAGNETOM Sola (1.5T) with XJ Gradient System
{3}------------------------------------------------
# Section 5 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act 1990 and 21 CFR § 807.92.
### I. General Information
| Establishment | Siemens Medical Solutions USA, Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA |
|----------------------|----------------------------------------------------------------------------------------------------------|
| Registration Number: | 2240869 |
| Date Prepared | August 3, 2018 |
Siemens Healthcare GmbH Manufacturer Henkestrasse 127 Erlangen, Bayern, Germany 91052 Registration Number: 3002808157
- Contact Person Cordell L. Fields, Esq. Regulatory Affairs Technical Specialist Siemens Medical Solutions USA, Inc. Cordell.Fields@siemens-healthineers.com
- Device Name MAGNETOM Sola MAGNETOM Sola Trade Name
- Classification Name: Magnetic Resonance Diagnostic Device (MRDD) Classification Panel: Radiology 21 CFR § 892.1000 CFR Code: Classification: Class II Product Code: Primary: LNH, Secondary: LNI, MOS
{4}------------------------------------------------
### II. Safety and Effectiveness Information Supporting Substantial Equivalence
#### Intended Use
The indications for use for the subject device are the same as the predicate device and are as follows:
Your MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
#### Device Description
MAGNETOM Sola with XJ gradient system is similar to the predicate device MAGNETOM Aera with syngo MR E11C (K153343) except for some new and modified software and hardware. A high level summary of the hardware and software compared to the MAGNETOM Aera with syngo MR E11C device is included below:
#### Hardware
#### New Hardware
- Main components such as: -
- magnet
- gradient coil -
- patient table (fixed and mobile) -
- New coils: -
- Body Coil -
- BM Head/Neck 20 -
- -BM Spine 48
- BM Spine 32 -
- UltraFlex Large 18 -
- UltraFlex Small 18 -
{5}------------------------------------------------
- Shoulder Shape 16 -
- TX/RX Knee 18 -
- BM Spine 24 -
- -BM Body 12
- Peripheral Angio 16 -
- Head/Neck 16 MR Coil 1.5T (Same as with predicate device, -MAGNETOM Aera)
- Other components such as:
- helium compressor -
- cover includes new touch displays and redesigned control island -
- computers and monitor -
- Respiratory Sensors (respiratory sensor loops) -
- camera
- Multi-Channel Interface -
#### Modified Hardware
- Main components such as gradient power amplifier and patient communication unit as well as 32 independent RF channels
- Other components such as RF power amplifier, magnet supervision, electronic cabinet and measurement control components
#### Software
#### New Features and Applications
- । SliceAdjust (framework for pulse sequence types which allows applying adjustment settings dynamically to measured sub-volumes during the measurement.)
- Whole-Body Dot Engine (supports multi-region examinations with consistent settings for spatial resolution, image contrast, and breath-hold capacity)
- Compressed Sensing GRASP-VIBE (conduct dynamic contrast-enhanced abdominal exams in free breathing)
- CoilShim (reduces patient induced strongly localized B0 inhomogeneities by generating the respective anatomy-specific B0 field with local shim coil elements integrated in the local Rx coil BM Head/Neck 20
{6}------------------------------------------------
- GOKnee3D (examination which comprises the AutoAlign knee localizer and two SPACE with CAIPIRINHA sequences to support fast high-resolution 3D exams of the knee)
- SPACE with CAIPIRINHA (3D SPACE pulse sequence type now offers the iPAT mode CAIPIRINHA)
- Respiratory Sensor Support (support for respiratory triggered measurements is provided in several SE-, GRE- and EPI-based pulse sequence types)
- MRSim / Synthetic CT (provides MR pulse sequences for the creation of Synthetic CT images based on the MR image input)
- Cardiac Dot Flow Add-In (extension of Cardiac Dot Engine to support blood flow measurements)
- PCASL mode (extension of ASL pulse sequence types by a new blood labeling mode)
- SMS in TSE (Simultaneous Multi Slice (SMS) support for TSE)
#### New Software / Platform
- Software platform (new software version syngo MR XA11A based on the Numaris/X software which combines the software of the predicate device and the reference device with modifications and extensions as well as changed structure and design of the user interface and changed post processing integration.)
- General workflow (scanning tab-card / left-hand side (LHS) and viewing and processing tab-card / right-hand side (RHS) concept with single / dual monitor workflow where patient handling and scanning is separated from post-processing and data handling. The scanning tab-card / "LHS" contains scan-related elements such as the Scheduler and scan UI and displays the inline processed images. At the viewing and processing tab-card / on the "RHS" additional results can be qenerated in the form of basic and advanced post-processing including data handling and result distribution.)
- User role concept (five user roles with different rights)
- Distribution step (allows the selection of data sets for archiving and DICOM transfer within MR View&GO)
- 4D Viewing (allows the visualization of 4D data such as different phases, b-। values or echoes of one DICOM series within MR View&GO and allows temporal (phase navigation in 4D data sets) and spatial scrolling.)
- Launcher Step (starts an advanced application as a viewing and processing tabcard / on the RHS)
- Multiframe DICOM format (now available in addition with improvements in performance, application support, navigation and data interchange.)
{7}------------------------------------------------
- Touch positioning (Select&GO 2.0) (to position the exam region in the isocenter by using the touch displays integrated in the system cover. Select&GO 2.0 has been extended to additional body area positions if dedicated coils are plugged in)
#### Modified Features and Applications
- Compressed Sensing Cardiac Cine (An option to enable Compressed Sensing Cardiac Cine is integrated into the BEAT pulse sequence type. The total acceleration factor is a user-interface parameter that controls the degree of kspace under-sampling. Compressed Sensing Cardiac Cine can be performed in single-shot and in multi-shot mode.)
- RetroGating (Compressed Sensing Cardiac Cine acquisitions which split the data acquisition over multiple heartbeats can now be configured to perform complete sampling of the cardiac cycle without prior definition of an acquisition window. Combination with arrhythmia rejection is possible.)
- Dixon fat/water separation (improvement for a more robust assignment of local fat and water regions to the respective image)
- Pre-Scan-Normalize (Improvements with the goal to correct MRI images for local coil sensitivity variations and to generate homogenous MRI images. Three different matrix sizes are available now.)
- iPAT / TSE Reference Scan (Improvements in the VIBE and in the HASTE pulse sequence types to improve the image quality. Changes in the VIBE pulse sequence type concern the k-space reordering, the external GRE reference scan and the minimal slice oversampling. Changes in the HASTE, TSE, FAST TSE and TSE DIXON pulse sequence types includes the possibility to use a reference scan "TSE/Separate" for GRAPPA acquisition and reconstruction)
- HeartFreeze (extended to support multiple repeats and averaging of the same slice acquired at the same phase of the cardiac cycle in combination with nonselective inversion recovery imaging)
- Care Bolus in Angio Dot Engine (workflow support for bolus administration (bolus detection))
- MRCP in SPACE (improvement of image quality for MR Cholanqiopancreatography (MRCP) acquisitions based on the SPACE pulse sequence type)
- MR Elastography (Replacement of existing masking by a masking that is performed on the pre-scan images used also within the pre-scan normalize (PSN) functionality. Optimization of pulse sequence type timing. Changes in MEG time period (no longer fixed to the wavelength of the MEG and also implementation of a reduced MEG period))
{8}------------------------------------------------
#### Modified (general) Software / Platform
- Scheduler (two components (Patient Registration and Scheduler) consolidated.)
- Table Positioning Mode (Automated movement of table provided on examination level so that the scan is performed in the magnet isocenter. Additionally the "LocalRange" positioning mode can be used for smaller regions like e.g. the heart or the brain.)
- Spectroscopy Add-in (enables planning on non-distortion corrected images for spectroscopy.)
- MR View&GO (For quality assurance, image viewing, basic post-processing, printing and result distribution. The functionality is provided in workflow steps which provide guidance, allow independent work and does not require any reloading of data. Addition of Mosaic View (view mode to scroll through dimensions instead of space) and 4D Movie Toolbar (movie toolbar is usable to navigate through the 4th dimension))
- Prior Handling (for the display of available priors)
- System Start and Shutdown (improvements with a System Start Timer and shutdown unattended by the user)
- Dot Cockpit (additional features for handling of scan pulse sequences and offline Dot Cockpit)
#### Other Modifications and / or Minor Changes
- SAR Assistant (two additional options for selection)
- Noise masking (to remove the noise floor in outer regions)
- MAGNETOM RT Pro Edition marketing bundle (extension of the bundle)
- MAGNETOM Sola Cardiovascular Edition marketing bundle (bundle of components for cardiovascular MR imaging with extension with additional components)
- Siemens "BioMatrix" (BioMatrix consists of three core technologies, Sensors, Tuners and Interfaces, all of which address different aspects of patient biovariability with extension with additional components)
- Siemens "Healthineers" (new Siemens brand)
- Improved Adjustments (frequency adjustment optimized for more reliable water peak detection, and FastView adjustments extended to be available for all kind of data selectable in the user interface)
{9}------------------------------------------------
- Cooling Cabinet (increased cooling capacity and improvements)
- BEAT_IRTTT (extends BEAT_IRT sequence with a multi-slice functionality and introduces some parameters. "TTT" stands for the extension of BEAT IRT to three dimensions)
- Vacuum pump (automatic switching mechanism)
- teamplay Protocols Interface (interface to support external pulse sequences management systems)
- Unilateral Hip (added in Large Joint Dot Engine) (user workflow optimized, since information/settings from the patient registration are taken)
- GRE RefScan (external GRE RefScan has been extended to multiple pulse sequence types)
- Asymmetric saturation pulses (support for regional saturation with an asymmetric shape has been added for BOLD imaging)
- CP Mode modification ("RF Transmit Mode" is provided as part of the patient registration based on IEC 60601-2-33)
- MR Breast Biopsy (support of breast coils has been changed)
#### Technological Characteristics
MAGNETOM Sola with XJ gradient system has different technological characteristics than the predicate device MAGNETOM Aera (K153343; cleared April 15, 2016). New and modified software with respect to the predicate device systems and the different technological characteristics are described.
The subject device is substantially equivalent to the predicate device with regard to the operational environment, programming language, operating system and performance.
MAGNETOM Sola with XJ gradient system conforms to the standard for software medical devices (IEC 62304:2006) and IEC as well as NEMA standards.
This submission includes new and modified hardware including new coils in comparison to the predicate device.
#### Nonclinical Tests
The following performance testing was conducted on the subject device
- Sample clinical images were taken in particular for the coils
- Software verification and validation testing was completed in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
{10}------------------------------------------------
- Performance tests were completed in accordance with the FDA guidance document, "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices" dated November 18, 2016
- Verification & Validation was completed for hardware modifications
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
#### Clinical Tests
A clinical study of 40 individuals was conducted to determine the nerve stimulation thresholds used to limit the gradient system output. The observed parameters were used to set the PNS (Peripheral Nerve Stimulation) threshold level which is required in IEC 60601-2-33.
No additional clinical tests were conducted to support the subject device and the substantial equivalence argument; however, sample clinical images were provided to support the new coils of the subject device per the MR guidance document.
#### Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis in compliance with ISO 14971:2007 to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized in hardware and software development, testing and product labeling. To minimize risks. Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the operators are healthcare professionals familiar with and responsible for the acquisition and post processing of magnetic resonance images.
MAGNETOM Sola with XJ gradient system conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regard to performance and safety as recommended by the respective MR FDA Guidance Document as stated in the following table.
{11}------------------------------------------------
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and date | Standards<br>Development<br>Organization |
|-----------------------|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| 19-4 | General | Medical electrical equipment -<br>part 1: general requirements for<br>basic safety and essential<br>performance | ES60601-<br>1:2005/(R) 2012<br>and A1:2012 | AAMI / ANSI |
| 19-8 | General | Medical electrical equipment -<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance - Collateral<br>Standard: Electromagnetic<br>disturbances - Requirements<br>and tests | 60601-1-2 Edition<br>4.0:2014-02 | IEC |
| 12-295 | Radiology | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic safety<br>and essential performance of<br>magnetic resonance equipment<br>for medical diagnosis | 60601-2-33 Ed.<br>3.2:2015 | IEC |
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>edition 2007-10 | ISO |
| 5-96 | General | Medical devices – Application of<br>usability engineering to medical<br>devices | 62366 Edition 1.0<br>2015 | AAMI<br>ANSI<br>IEC |
| 13-32 | Software | Medical device software -<br>Software life cycle processes | 62304:2006 | AAMI<br>ANSI<br>IEC |
| 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices | MS 4-2010 | NEMA |
| 12-288 | Radiology | Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images<br>(MRI) | MS 9-2008 | NEMA |
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 - 3.20<br>(2016) | NEMA |
| 2-156 | Biocompati<br>bility | Biological evaluation of medical<br>devices - part 1: evaluation and<br>testing within a risk management<br>process. (Biocompatibility) | 10993-1:2009/(R)<br>2013 | AAMI<br>ANSI<br>ISO |
{12}------------------------------------------------
#### Substantial Equivalence
MAGNETOM Sola with XJ gradient system includes most of the features of the predicate device and additional new and modified hardware and software as noted above.
| Predicate device | FDA Clearance<br>Number and Date | Product code | Manufacturer |
|-----------------------------------------------|-----------------------------------------|----------------|------------------------------------------------|
| MAGNETOM Aera with syngo<br>MR E11C | K153343,<br>cleared April 15,<br>2016 | LNH<br>LNI,MOS | Siemens AG /<br>Siemens Healthcare<br>GmbH |
| Reference Devices | FDA Clearance<br>Number and Date | Product code | Manufacturer |
| MAGNETOM Vida with software<br>syngo MR XA10A | K170396,<br>cleared June 14,<br>2017 | LNH<br>LNI,MOS | Siemens Healthcare<br>GmbH |
| MAGNETOM Amira with syngo<br>MR E11S | K173600<br>cleared December<br>19, 2017 | LNH<br>LNI,MOS | Siemens Shenzhen<br>Magnetic Resonance<br>Ltd. |
#### Conclusion as to Substantial Equivalence
MAGNETOM Sola with XJ gradient system has the same intended use and different technological characteristics than the predicate device system, MAGNETOM Aera with syngo MR E11C, with respect to the magnetic resonance features and functionalities. While there are some technical features that vary with respect to the predicate device MR System, the conclusions from all verification and validation data suggest that the features with different technological characteristics from the predicate device bear an equivalent safety and performance profile as that of the predicate device and the reference devices.
MAGNETOM Sola does not introduce any new issues of safety or effectiveness. Therefore, Siemens is of the opinion that MAGNETOM Sola with XJ gradient system is substantially equivalent to the legally marketed predicate device MAGNETOM Aera with software syngo MR E11C (K153343, cleared on April 15, 2016).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.