ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card
K182123 · Advin Biotech, Inc. · DJG · Mar 29, 2019 · Clinical Toxicology
Device Facts
Record ID
K182123
Device Name
ATTEST Drug Screen Cup, ATTEST Drug Screen Dip Card
Applicant
Advin Biotech, Inc.
Product Code
DJG · Clinical Toxicology
Decision Date
Mar 29, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3650
Device Class
Class 2
Indications for Use
The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, d/l-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows: [Table of analytes/cutoffs]. The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board adulteration/specimen validity tests (SVT) in the dip card format or in the cup format. Only one cutoff concentration per analyte will be included per device. The drug screen tests are intended for prescription use only. The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results. The tests are not intended to differentiate between illicit and prescription use of benzodiazepines, barbiturates, buprenorphine, oxycodone, propoxyphene and tricyclic antidepressants. There are no uniformly recognized cutoff levels for these drugs in urine.
Device Story
Rapid lateral flow immunoassay for qualitative detection of drugs of abuse in human urine; utilizes competitive binding principle. Device consists of test strips with drug-protein conjugates on nitrocellulose membrane and anti-drug antibody-colloidal gold conjugates on a pad. Urine sample rehydrates gold conjugate; if target drug is below cutoff, gold conjugate binds to immobilized drug-protein conjugate, forming visible line (negative). If drug exceeds cutoff, antibody sites are saturated, preventing line formation (preliminary positive). Used in professional clinical settings; operated by healthcare professionals. Provides preliminary results requiring confirmation via GC/MS or LC/MS. Benefits include rapid, point-of-care screening for multiple drug analytes (up to 16) and optional specimen validity testing (SVT) to detect tampering.
Clinical Evidence
Bench testing only. Precision/reproducibility evaluated at three sites using 180 replicates per format over 10 days. Analytical specificity and cross-reactivity tested against structurally similar compounds and endogenous substances. Method comparison performed against LC/MS quantitative methods using clinical urine specimens. No clinical studies performed.
Technological Characteristics
Lateral flow immunochromatographic assay; competitive binding. Formats: Cup and Dip Card. Qualitative visual readout. Internal control line (C-line) for migration/volume verification. No external controls provided. Stable for 28 months. No electronic components or software algorithms.
Indications for Use
Indicated for qualitative detection of drugs of abuse and their metabolites in human urine for professional, prescription-only use. Target population includes individuals undergoing drug screening. Contraindications: Not for differentiating between illicit and prescription use of specific drugs (benzodiazepines, barbiturates, buprenorphine, oxycodone, propoxyphene, tricyclic antidepressants).
Regulatory Classification
Identification
An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.
Special Controls
*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
Advin Multi-Drug Screen Test Cup (k122809)
Advin Multi-Drug Screen Test Dipcard (k122809)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k182123
B. Purpose for Submission:
Modification of a previously cleared device (k122809) to add Amphetamine (1000 ng/mL cutoff), Methamphetamine (1000 ng/mL cutoff), Cocaine (300 ng/mL cutoff), and 6-Acetylmorphine devices (10 ng/mL cutoff), and to change the formulation of the previously cleared Morphine device.
C. Measurands:
Amphetamine, Methamphetamine, Cocaine, Morphine, and 6-Acetylmorphine
D. Type of Test:
Qualitative, lateral flow immunoassay
E. Applicant:
Advin Biotech, Inc.
F. Proprietary and Established Names:
ATTEST Drug Screen Cup
ATTEST Drug Screen Dip Card
G. Regulatory Information:
| Regulation names | Product Code | Class | Regulation | Panel |
| --- | --- | --- | --- | --- |
| Opiate Test System | DJG | II | 862.3650 | Toxicology (91) |
| Amphetamine Test system | DKZ | II | 862.3100 | Toxicology (91) |
| Cocaine and metabolites Test System | DIO | II | 862.3250 | Toxicology (91) |
| Methamphetamine Test System | DJC | II | 862.3610 | Toxicology (91) |
{1}
H. Intended Use:
1. Intended use(s):
Refer to Indications for Use
2. Indication(s) for use:
The ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card are rapid lateral flow immunoassays for the qualitative detection of 6-Acetylmorphine, d-Amphetamine, Benzoylecgonine, Buprenorphine, EDDP, d/l-Methadone, d-Methamphetamine, d/l-Methylenedioxymethamphetamine, Morphine, Nortriptyline, Oxazepam, Oxycodone, Phencyclidine, d-Propoxyphene, Secobarbital and THC-COOH in human urine. The test cut-off concentrations and the compounds the tests are calibrated to are as follows:
| Analyte | Calibrator | Cutoff (ng/mL) |
| --- | --- | --- |
| 6-Acetylmorphine | 6-monoacetylmorphine | 10 |
| Amphetamine | d-Amphetamine | 500 |
| Amphetamine | d-Amphetamine | 1,000 |
| Barbiturates | Secobarbital/Pentobarbital | 300 |
| Benzodiazepines | Oxazepam | 300 |
| Buprenorphine | Buprenorphine | 10 |
| EDDP | 2-ethylidene-1,5-dimethyl-3-3-diphenylpyrrolidine | 300 |
| Cocaine | Benzoylecgonine | 150 |
| Cocaine | Benzoylecgonine | 300 |
| Ecstasy | d,l-Methylenedioxymethamphetamine | 500 |
| Methamphetamine | d-Methamphetamine | 500 |
| Methamphetamine | d-Methamphetamine | 1,000 |
| Marijuana | 11-nor-Δ⁹-THC-9-COOH | 50 |
| Methadone | d/l-Methadone | 300 |
| Opiates | Morphine | 300 |
| Opiates | Morphine | 2,000 |
| Oxycodone | Oxycodone | 100 |
| Phencyclidine | Phencyclidine | 25 |
| Propoxyphene | Propoxyphene | 300 |
| Tricyclic Antidepressants | Nortriptyline | 1,000 |
The single or multi-test panels can consist of the above listed analytes in any combination, up to a maximum of 16 analytes, with and without on-board
{2}
adulteration/specimen validity tests (SVT) in the dip card format and in the cup format. The drug screen tests are intended for prescription use only.
The tests provide only a preliminary result. A more specific alternative chemical method should be used in order to obtain a confirmed presumptive positive result if the donor doesn't admit use or anytime required by testing procedures. Gas Chromatography / Mass Spectrometry (GC/MS), Liquid Chromatography / Mass Spectrometry (LC/MS) and their tandem mass-spectrometer versions are the preferred confirmatory methods. Careful consideration and judgment should be applied to any drugs of abuse screen test result, particularly when evaluating preliminary positive results.
3. Special conditions for use statement(s):
This assay is intended for prescription use
4. Special instrument requirements:
Not applicable.
I. Device Description:
The ATTEST Drug Screen Cup and ATTEST Drug Screen Dip Card devices are immunochromatographic assays that use a lateral flow system for the qualitative detection of Amphetamine, Methamphetamine, Barbiturates, Benzodiazepine, Cannabinoids, Cocaine, Methadone, Tricyclic antidepressants, Buprenorphine, EDDP, MDMA, Morphine, Oxycodone, Phencyclidine, Propoxyphene, and 6-Acetylmorphine. The ATTEST Drug Screen Cup device consists of a cup device and a package insert. The ATTEST Drug Screen Dip Card device consists of a Dip Card device, a package insert and a urine cup for sample collection.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Advin Multi-Drug Screen Test Cup, Advin Multi-Drug Screen Test Dipcard
2. Predicate 510(k) number(s):
k122809
{3}
4
3. Comparison with predicate:
Predicate Similarities and Differences Table for All Assays
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device (k122809) |
| Intended Use | Same | Qualitative detection of drugs of abuse and/or their metabolites in urine |
| Test Principle | Same | Lateral flow immunochromatographic assay based on competitive binding |
| Matrix | Same | Urine |
| Analyte cutoffs (ng/mL) | Same except for the following analytes: 6-AM - 10 Amphetamine - 1,000 Cocaine - 300 Methamphetamine - 1,000 Opiates - 300 | Amphetamine - 500 Barbiturates - 300 Benzodiazepine - 300 Buprenorphine - 10 Cocaine - 150 EDDP - 300 Ecstasy - 500 Methamphetamine - 500 Methadone - 300 Morphine - 300 Opiates - 2,000 Oxycodone - 100 Phencyclidine - 25 Propoxyphene - 300 Tricyclics - 1,000 Marijuana - 50 |
| Differences | | |
| Formats | Cup and Dip Card only | Cup, Dip Card, Cassette |
K. Standard/Guidance Document Referenced (if applicable):
None referenced
L. Test Principle:
Each device employs lateral flow immunochromatographic technology and is based on the principle of competitive binding. Drugs, if present in concentrations below the cutoff level, will not saturate the binding sites of the antibody coated particles on the drug specific test strips. The goat-anti-rabbit IgG antibody-coated particles will then be captured by immobilized drug-specific conjugate. If the level of drug in the urine specimen is below the cutoff concentration, the T line appears as a visible burgundy line. If the level of drug in the urine specimen is above the cutoff, no T line develops.
The control line (C line) serves as an internal quality control. The control line should always
{4}
appear as a burgundy-colored band regardless of the presence of the drug, if enough sample volume has been added to the test and if the sample has correctly migrated up the test strip. Testing is based on the principle of a competitive immunochemical reaction between a chemically labeled drug (drug-protein conjugate) and the drug or drug metabolites which may be present in the urine sample competing for the limited antibody binding sites.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision study of the ATTEST Drug Screen Cup and Dip card were evaluated for Amphetamine (1000 ng/mL cutoff, AMP 1000), Methamphetamine (1000 ng/mL cutoff, MET 1000), Cocaine (300 ng/mL cutoff, COC 300), Opiates (300 ng/mL cutoff, OPI 300) and 6-Acetylmorphine devices (10 ng/mL cutoff, 6-AM 10) at three external testing sites. Samples tested were drug free urine spiked with certified reference standards to the target concentrations. The test solutions contain the following drug concentrations: negative, +/-75% cutoff, +/-50% cutoff, +/-25% cutoff, and +100% cutoff. All test solutions were confirmed by a chemically specific quantitative method to verify the drug concentrations.
Three operators performed the study using three lots of ATTEST Cup Drug Screen and three lots of ATTEST Card Drug Screen. Each sample was tested in 20 replicates for each device format at each of three sites. A total of 180 specimens for each device format were tested over 10 non-consecutive days. The results are shown in the tables below. Performance data for Amphetamine (500 ng/mL cutoff), Barbiturates, Benzodiazepines, Buprenorphine, EDDP, Cocaine (150 ng/mL cutoff), Ecstasy, Methamphetamine (500 ng/mL cutoff), Marijuana, Methadone, Opiates (2000 ng/mL cutoff), Oxycodone, Phencyclidine, Propoxyphene, and Tricyclic Antidepressants was provided in k122809.
Precision Study Result of ATTEST Cup:
| Drug Test | Results | -100% Cutoff | -75% Cutoff | -50% Cutoff | -25% Cutoff | Cutoff | +25% Cutoff | +50% Cutoff | +75% Cutoff | +100% Cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 6-AM 10 | Neg | 60 | 60 | 60 | 58 | 32 | 0 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 2 | 28 | 60 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 97% | --- | 100% | 100% | 100% | 100% |
| AMP 1000 | Neg | 60 | 60 | 60 | 58 | 5 | 1 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 2 | 55 | 59 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 97% | --- | 98% | 100% | 100% | 100% |
{5}
6
| COC 300 | Neg | 60 | 60 | 60 | 59 | 22 | 1 | 0 | 0 | 0 |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Pos | 0 | 0 | 0 | 1 | 38 | 59 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 98% | --- | 98% | 100% | 100% | 100% |
| MET 1000 | Neg | 60 | 60 | 60 | 58 | 26 | 3 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 2 | 34 | 57 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 97% | --- | 95% | 100% | 100% | 100% |
| OPI 300 | Neg | 60 | 60 | 60 | 56 | 10 | 4 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 4 | 50 | 56 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 93% | --- | 93% | 100% | 100% | 100% |
Precision Study Result of ATTEST Dip Card:
| Drug Test | Results | -100% Cutoff | -75% Cutoff | -50% Cutoff | -25% Cutoff | Cutoff | +25% Cutoff | +50% Cutoff | +75% Cutoff | +100% Cutoff |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| 6-AM 10 | Neg | 60 | 60 | 60 | 59 | 23 | 1 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 1 | 37 | 59 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 98% | --- | 98% | 100% | 100% | 100% |
| AMP 1000 | Neg | 60 | 60 | 60 | 60 | 18 | 2 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 0 | 42 | 58 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 100% | --- | 97% | 100% | 100% | 100% |
| COC 300 | Neg | 60 | 60 | 60 | 57 | 18 | 1 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 3 | 42 | 59 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 95% | --- | 98% | 100% | 100% | 100% |
| MET 1000 | Neg | 60 | 60 | 60 | 59 | 31 | 2 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 1 | 29 | 58 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 98% | --- | 97% | 100% | 100% | 100% |
| OPI 300 | Neg | 60 | 60 | 60 | 59 | 21 | 1 | 0 | 0 | 0 |
| | Pos | 0 | 0 | 0 | 1 | 39 | 59 | 60 | 60 | 60 |
| | Total | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 | 60 |
| | Agreement | 100% | 100% | 100% | 98% | --- | 98% | 100% | 100% | 100% |
b. Linearity/assay reportable range:
{6}
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
This device has internal process controls. A colored line appearing in the control region confirms sufficient sample volume and adequate membrane wicking. Users are informed that the test is invalid if a line fails to appear in the control region. There are no external controls supplied with the device.
Device stability has been evaluated through accelerated and real-time studies. Protocols and acceptance criteria were described and found to be acceptable. The manufacturer claims that the devices are stable for 28 months.
d. Detection limit:
Not applicable.
e. Analytical specificity:
Analytical specificity for both the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card was determined through the testing of contrived solutions made by spiking certified standards of chemically-related or structurally-similar compounds into drug free urine. The relative cross-reactivity represents the minimum concentration necessary to yield a result similar to the cutoff level of the respective assay. Identical results, provided in the table below, were obtained with the two devices. Performance data for Amphetamine (500 ng/mL cutoff), Barbiturates, Benzodiazepines, Buprenorphine, EDDP, Cocaine (150 ng/mL cutoff), Ecstasy, Methamphetamine (500 ng/mL cutoff), Marijuana, Methadone, Opiates (2000 ng/mL cutoff), Oxycodone, Phencyclidine, Propoxyphene, and Tricyclic Antidepressants was provided in k122809.
| Assay/Cutoff | Compound | Concentration (ng/mL) | Relative cross-reactivity (%) |
| --- | --- | --- | --- |
| 6-AM 10 | 6-acetylmorphine | 10 | 100 |
| | Diacetylmorphine (heroin) | 300 | 3.3 |
| | Morphine | >100,000 | <0.1 |
| | Codeine | >100,000 | <0.1 |
| | Oxycodone | >100,000 | <0.1 |
| | Oxymorphone | >100,000 | <0.1 |
| AMP 1000 | d-Amphetamine | 1,000 | 100 |
| | l-Amphetamine | 100,000 | 1 |
| | MDA | 15,000 | 6.7 |
| | Phentermine | 100,000 | 1.0 |
| COC 300 | Benzoylecgonine | 300 | 100 |
| | Cocaethylene | 100,000 | 0.3 |
| | Cocaine | 10,000 | 3 |
{7}
The potential interference from compounds chemically dissimilar to target drugs, and known endogenous agents was also determined for both the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card. Testing of compounds and endogenous agents was performed by spiking the substances noted in the table below into pooled urine containing target drugs at near-cutoff concentrations. Identical results, provided in the table below, were obtained with the two devices. Unless otherwise indicated, substances were tested for potential interference at a concentration of 100 µg/mL. Performance data for Amphetamine (500 ng/mL cutoff), Barbiturates, Benzodiazepines, Buprenorphine, EDDP, Cocaine (150 ng/mL cutoff), Ecstasy, Methamphetamine (500 ng/mL cutoff), Marijuana, Methadone, Opiates (2000 ng/mL cutoff), Oxycodone, Phencyclidine, Propoxyphene, and Tricyclic Antidepressants was provided in k122809.
The following substances demonstrated no interference to the assays encompassed in this submission.
| Acetaminophen | Acetone | Acetylsalicylic acid |
| --- | --- | --- |
| Albumin | Ampicillin | Ascorbic acid |
| Aspartame | Atropine | Benzocaine |
| Bilirubin | Caffeine | Chloroquine |
| Chlorpheniramine | Creatine | Dexbrompheniramine |
| Dextromethorphan | Dimethylaminoantipyrine | Diphenhydramine |
| Dimenhydrinate | Dopamine | Isoproterenol |
| Ephedrine | Erythromycin | Ethanol |
| Furosemide | Gabapentin | Glucose |
{8}
| Guaiacol glyceryl ether | Hemoglobin | Ibuprofen |
| --- | --- | --- |
| Isoproterenol | Ketamine | Lidocaine |
| Methylephedrine | Naproxen | Niacinamide |
| Nicotine | Norephedrine | Oxalic acid |
| Pantoprazole | Penicillin-G | Pheniramine |
| Phenothiazine | 1-Phenylephrine | B-Phenylethylamine |
| Pregablin | Procaine | Quinidine |
| Ranitidine | Riboflavin | Sertraline |
| Sodium chloride | Sulindac | Theophylline |
| Tyramine | | |
Interference from pH and specific gravity variation was also evaluated for both the ATTEST Drug Screen Cup and the ATTEST Drug Screen Dip Card using pooled urine specimens containing target drugs at near-cutoff concentrations. The results demonstrated that pH levels of 4 to 9 and specific gravity levels of 1.003 to 1.030 do not affect the results of the assays for either of the devices. Performance data for Amphetamine (500 ng/mL cutoff), Barbiturates, Benzodiazepines, Buprenorphine, EDDP, Cocaine (150 ng/mL cutoff), Ecstasy, Methamphetamine (500 ng/mL cutoff), Marijuana, Methadone, Opiates (2000 ng/mL cutoff), Oxycodone, Phencyclidine, Propoxyphene, and Tricyclic Antidepressants was provided in k122809.
f. Assay cut-off:
For characterization of how the device performs analytically around the claimed cutoff concentration, please refer to section M.1.a., above.
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was conducted for the ATTEST Drug Screen Cup and ATTEST Drug Screen Dip card, using clinical urine specimens quantitated for the target drugs of abuse by chemically specific quantitative methods (e.g. LC-MS). Identical results, provided in the table below, were obtained with the two devices. Performance data for Amphetamine (500 ng/mL cutoff), Barbiturates, Benzodiazepines, Buprenorphine, EDDP, Cocaine (150 ng/mL cutoff), Ecstasy, Methamphetamine (500 ng/mL cutoff), Marijuana, Methadone, Opiates (2000 ng/mL cutoff), Oxycodone, Phencyclidine, Propoxyphene, and Tricyclic Antidepressants was provided in k122809.
{9}
10
| Device | Device Result | Concentration determined by chemically specific quantitative method (e.g. LC/MS) | | | |
| --- | --- | --- | --- | --- | --- |
| | | ≤50% cutoff concentration | Near Cutoff Negative (-50% to the cutoff concentration) | Near Cutoff Positive (cutoff to 150% of the cutoff) | High Positive (≥150% of the cutoff) |
| 6-AM/10 | Negative | 40 | 5 | 0 | 0 |
| | Positive | 0 | 0 | 5 | 35 |
| AMP/1000 | Negative | 40 | 6 | 0 | 0 |
| | Positive | 0 | 0 | 6 | 40 |
| COC/300 | Negative | 40 | 5 | 0 | 0 |
| | Positive | 0 | 0 | 5 | 35 |
| MET/1000 | Negative | 40 | 6 | 0 | 0 |
| | Positive | 0 | 0 | 5 | 40 |
| OPI/300 | Negative | 40 | 3 | 0 | 0 |
| | Positive | 0 | 1 | 4 | 46 |
Summary of Discordant Results:
| Drug / Cutoff (ng/ml) | ATTEST Cup Result | GC/MS result | |
| --- | --- | --- | --- |
| | | Drug Concentration (ng/ml) | Analyte |
| OPI/300 | Positive | 289 | Morphine |
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
{10}
Not applicable.
5. Expected values/Reference range:
Not applicable.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.