K182118 · Mavrik Dental Systems, Ltd. · EIE · May 3, 2019 · Dental
Device Facts
Record ID
K182118
Device Name
armor LC
Applicant
Mavrik Dental Systems, Ltd.
Product Code
EIE · Dental
Decision Date
May 3, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6300
Device Class
Class 1
Indications for Use
armor LC™ is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues.
Device Story
armor LC™ is a light-cured, resin-based protective barrier; used by dental professionals to seal and protect gingival tissue during clinical procedures. Supplied in 1 mL syringe; applied via applicator tip to gingival margins. Material flows to form seal at base of teeth; rapidly light-cured in situ to create flexible barrier. After procedure, barrier is removed in one piece using standard dental instruments. Benefits include protection of soft tissue from dental materials; single-use, non-sterile.
Clinical Evidence
Bench testing only. Performance evaluated via appearance, handling, removal, curing temperature, ambient light sensitivity, viscosity, depth of cure, and flexural strength. Biocompatibility testing performed per ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 7405, including cytotoxicity, irritation, and skin sensitization assessments on both uncured and cured materials.
Technological Characteristics
Resin-based barrier composed of urethane dimethacrylate (UDMA), cetyl alcohol, light-curing chemistry, and flavor. Viscous foam form factor. Light-cured polymerization (<20s setting time). Non-sterile, single-use. Biocompatible per ISO 10993 and ISO 7405 standards.
Indications for Use
Indicated for use as a protective barrier for dental materials or tissues during professional dental procedures. Prescription use only.
Regulatory Classification
Identification
A rubber dam and accessories is a device composed of a thin sheet of latex with a hole in the center intended to isolate a tooth from fluids in the mouth during dental procedures, such as filling a cavity preparation. The device is stretched around a tooth by inserting a tooth through a hole in the center. The device includes the rubber dam, rubber dam clamp, rubber dam frame, and forceps for a rubber dam clamp. This classification does not include devices intended for use in preventing transmission of sexually transmitted diseases through oral sex; those devices are classified as condoms in § 884.5300 of this chapter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2019
Mavrik Dental Systems Ltd % Sherwyn Davidson QA/RA Director 25 Ha Haroshet Street 6th Fl., Raanana, 4365613 Israel
Re: K182118
Trade/Device Name: armor LCTM Regulation Number: 21 CFR 872.6300 Regulation Name: Rubber Dam And Accessories Regulatory Class: Class I Product Code: EIE Dated: April 3, 2019 Received: April 3, 2019
Dear Sheila Hemeon-Heyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Malvina Eydelman, M.D. Director OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182118
Device Name armor LCTM
#### Indications for Use (Describe)
armor LCTM is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues.
Type of Use (*Select one or both, as applicable*)
| <div> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
### Mavrik Dental Systems Ltd.'s armor LCTM Device
| Applicant's name: | Mavrik Dental Systems Ltd. | | | armor LCTM device<br>(K182118) | Primary Predicate<br>Triumph Resin Dam<br>(K151945) | Reference Predicate:<br>OpalDam<br>(K971284) | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Contact Person: | Sherwyn Davidson<br>QA/RA Director<br>25 Ha Haroshet Street 6th Fl.,<br>Raanana, 4365613, Israel<br>Tel. +972-9-9740100<br>Fax +972-9-9740101<br>Sherwyn.davidson@mavrikdental.com | | Device Class | Class I | Class I | Class I | |
| Date Prepared: | April 30, 2019 | | | Classification<br>Panel | Dental | Dental | Dental |
| Name of Device: | armor LC TM | | | Product Code | EIE | EIE | EIG |
| Common or Usual Name: | Rubber dam and accessories | | Regulation<br>Description | Rubber dam and<br>accessories | Rubber dam and<br>accessories | Dental hand instrument | |
| Classification: | Product Code: | EIE | Regulation<br>number | 21 C.F.R. §872.6300 | 21 C.F.R. §872.6300 | 21 CFR 872.4565 | |
| | Regulation No: | 21 C.F.R. §872.6300 | Indications for<br>Use | armor LC™ is a resin-<br>based, light-cured<br>protective barrier material<br>recommended to protect<br>dental materials or<br>tissues. | Triumph Resin Dam is a<br>resin-based, light-cured<br>protective barrier material<br>recommended to protect<br>dental materials or<br>tissues. | OpalDam resin is used to<br>protect soft tissue<br>adjacent to the teeth<br>during in-office bleaching.<br>OpalDam is used with<br>Opalescence Xtra (35%<br>hydrogen peroxide)<br>"power" bleaching agent. | |
| | Class: | Class: I | Device Type | Light cured rubber dam | Light cured rubber dam | Light-cured rubber dam | |
| | Classification Panel: | Dental | Composition | Urethane dimethacrylate,<br>cetyl alcohol, light curing<br>chemistry, flavor | Urethane acrylates, light-<br>curing chemistry, silica,<br>titanium dioxide | Diurethane<br>dimethacrylate, cetyl<br>alcohol, light-curing<br>chemistry, titanium<br>dioxide | |
| Predicate Device: | Primary predicate: Triumph Resin Dam (K151945)<br>Reference predicate: OpalDam (K971284) | | Light Cure<br>Capabilities: | Yes | Yes | Yes | |
| Light Cure<br>Setting Time: | <20 seconds | <20 seconds | <20 seconds | | | | |
| Flexibility to<br>allow removal<br>in one piece,<br>when applied<br>around a dental<br>tooth model | Yes | Yes | Yes | | | | |
| Biocompatible | Yes | Yes | Yes | | | | |
| Sterilization | Non-sterile<br>Not intended to be<br>sterilized by end-user | Non-sterile<br>Not intended to be<br>sterilized by end-user | Non-sterile<br>Not intended to be<br>sterilized by end-user | | | | |
#### Intended Use / Indications for Use
armor LC™ is a resin-based, light-cured protective barrier material recommended to protect dental materials or tissues.
#### Device Description
armor LC™ is a light-cured resin dam designed to seal around and protect the gingival tissue during professional dental procedures. It consists of a viscous urethane dimethacrylate (UDMA)based resin foam supplied in a black 1 mL syringe. armor LC™ is only for prescription use and is a single-use, non-sterile, product. Using an applicator tip, the uncured product is applied over the gingival margin where it flows to form a seal at the base of the teeth. It is then rapidly light cured in situ to form a flexible material barrier. After the dental procedure it is readily removable with simple dental instruments.
{4}------------------------------------------------
# Substantial Equivalence
The following table provides a comparison of the armor LC™ to the primary and reference predicate devices:
{5}------------------------------------------------
## Substantial Equivalence Discussion:
armor LC™ device has the same indications for use statement as the primary predicate device, Triumph Resin Dam, which was found to be substantially equivalent to the reference predicate device, OpalDam. The only technological difference between armor LC™ and the two named predicate devices is minor differences in their chemistry. All three products include acrylates as the base monomer, light curing chemistry for polymerization, and fillers to achieve the desired mechanical properties. While the exact chemical compositions of the predicate devices is unknown, the main difference appears to be that armor LC™ does not include titanium dioxide, which is added to some resins as a filler or pigment. armor LC™ also includes a rum flavor ingredient, which has no impact on the device properties. Side-by-side bench testing comparing armor LC™ to the predicate devices demonstrated substantial equivalence of the handling and technical characteristics.
## Non-Clinical Performance Data
## Bench Performance Testing
Mavrik Dental Systems conducted several performance tests to demonstrate that armor LCTM device complies with performance standards and that it functions as intended. Bench performance tests included:
- . Appearance and handling
- Removal in one piece ●
- Curing temperature ●
- Sensitivity to ambient light ●
- Viscosity
- Depth of cure
- . Flexural strength
In all instances, armor LC™ device functioned as intended and met the pre-defined acceptance criteria.
## Biocompatibility
armor LC™ is a surface contact device in limited contact with mucosal membranes. The testing required by FDA and ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and testing Within A Risk Management Process" are cytotoxicity, irritation, and skin sensitization. Testing was conducted on the final, finished armor LCTM device in accordance with the international recognized consensus standards listed below and was found to be biocompatible.
- ISO 10993-1 Fourth edition 2009-10-15, Biological evaluation of medical devices .
- ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices -● Part 5: Tests for in vitro cytotoxicity
- o Uncured material - Direct contact method with exposure of 5 minutes. The material must be cured within 1 minute of application in clinical use.
{6}------------------------------------------------
- o Cured material ISO elution method with exposure of 48 hours
- · ISO 10993-10 Third Edition 2010-08-01, Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- Uncured material 4 hours
- o Cured material 7 days
- · ISO 7405 Second edition 2008-12-15, Dentistry Evaluation of biocompatibility of medical devices used in dentistry [Including: Amendment 1 (2013)]
## Conclusion
The armor LCTM device is substantially equivalent to the predicate device in design, composition, principle of operation, biocompatibility, performance, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.