Portable Electro-Stimulation Therapy Device, model LGT-235 is indicated for the Symptomatic relief of chronic intractable pain, Post-traumatic pain and Post-surgical pain of the arms and legs.
Device Story
Portable Electro-Stimulation Therapy Device (LGT-235) is a handheld TENS and Microcurrent (MCR) stimulator for home use. System comprises a main unit, electrodes (knee brace or self-adhesive), and 'MStim Arth' mobile app. Device delivers electrical impulses to sensory nerves via skin electrodes to modify pain perception; does not cure underlying conditions. User operates device via app to select pre-programmed output modes and treatment time, or via main unit to adjust intensity and power. Output affects sensory nerves responsible for pain signal transmission. Benefits include symptomatic pain relief for chronic, post-traumatic, or post-surgical conditions in arms and legs.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included electrical safety, mechanical testing, software verification, and biocompatibility (ISO 10993-5/10). Compliance with IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-2-10 standards confirmed.
Technological Characteristics
Handheld TENS/MCR stimulator; 3.7V rechargeable lithium battery. Electrodes: OK cloth/SBR/dacron knee brace with silver/nylon conductive fabric or conductive hydrogel self-adhesive pads. Symmetrical biphasic waveform. Connectivity: Bluetooth/mobile app for control. Standards: ISO 10993, ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11.
Indications for Use
Indicated for adults experiencing chronic intractable pain, post-traumatic pain, or post-surgical pain in the arms and legs.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 26, 2018
Guangzhou Longest Science & Technology Co., Ltd. % You Yijie Manager Oimmig Medical Consulting Service Co., Ltd RM.1711, Building K, NO.101 Science Ave International Creative Valley Guangzhou, 510663 Cn
Re: K182108
Trade/Device Name: Portable Electro-Stimulation Therapy Device Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: GZJ Dated: March 29, 2019 Received: March 29, 2019
Dear You Yijie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Pamela D. Scott -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182108
Device Name Portable Electro-Stimulation Therapy Device
#### Indications for Use (Describe)
Portable Electro-Stimulation Therapy Device, model LGT-235 is indicated for the Symptomatic relief of chronic intractable pain, Post-traumatic pain and Post-surgical pain of the arms and legs.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
Date: April 24, 2019
#### 1. Submitter's Information
#### Establishment Registration Information
Name: Guangzhou Longest Science & Technology Co., Ltd. Address: 5&6f, Building B4, No.11, Kaiyuan Avenue Science City, Hi-Tech Industrial Zone Guangzhou, Guangdong CHINA
#### Contact Person of applicant
Name: Xiaobinq Luo Address: 5&6f, Building B4, No.11, Kaiyuan Avenue Science City, Hi-Tech Industrial Zone Guangzhou, Guangdong CHINA TEL: +86 020-66353999 FAX: +86 020-66353999 Email: qzlongest@126.com
#### Contact Person of the Submission:
Name: You Yijie Address: RM.1711, Building K, NO.101 Science Ave International Creative Valley Development Zone, Guangzhou China TEL: +86 020-8224 5821 FAX: +86 020-8224 5821 Email: Jet.you@gimmiq-med.com
#### 2. Device Information
Type of 510(K) submission: Traditional Trade Name: Portable Electro-Stimulation Therapy Device Model: LGT-235 Regulation name: Transcutaneous electrical nerve stimulator for pain relief. Review panel: Neurology Product code: GZJ Requlation Class: II Regulation Number: 21 CFR 882.5890
#### 3. Predicate Device Information
510(k) submitter/holder: Mettler Electronics Corp. 510(K) Number: K111482 Regulation Class: 2 Device: Sonicator Plus 920 Sonicator Plus 920, Model ME 920 Trade name:
{4}------------------------------------------------
### 4. Device description
The Portable Electro-Stimulation Therapy Device, model LGT-235 is a stimulator which sends gentle electrical current to underlying nerves via electrodes (Knee brace or self-adhesive electrode) applied on the skin with two operational modes and powered by rechargeable lithium battery; the device system is made up of mobile app MStim Arth, main unit and electrodes (Knee brace and self-adhesive electrode).
The Mobile App MStim Arth provides access to treatment controls for Portable Electro-Stimulation Therapy Device from a compatible mobile device for selecting a pre-programmed output mode and treatment time.
The main unit provides access to adjust the intensity up or down, put ON/OFF the main unit.
The two modes that Portable Electro-Stimulation Therapy Device, model LGT-235 employs are transcutaneous electrical stimulation (TENS) and Microcurrent (MCR).
TENS and MCR are both specifically targets the sensory nerves, which are responsible for sending pain signals to the brain. TENS and MCR use tiny electrical impulses sent through the skin to nerves to modify the pain perception. TENS and MCR do not cure any physiological problem; it only helps control the pain, this activates the underlying sensory nerves. Knee brace is placed on the leg or arm close to the area of pain.
### 5. Principle of operation:
The Portable Electro-Stimulation Therapy Device, model LGT-235 sends gentle electrical current to underlying nerves and muscle group via electrodes (Knee brace or self-adhesive electrode) applied on the skin and powered by rechargeable lithium battery. When used in TENS or MCR mode, it specifically targets the sensory nerves, which are responsible for sending pain signals to the brain; and it uses tiny electrical impulses sent through the skin to nerves to modify pain perception and finally helps control the pain.
#### 6. Indications for Use
Portable Electro-Stimulation Therapy Device, model LGT-235 is indicated for the Symptomatic relief of chronic intractable pain, Post-traumatic pain and Post-surgical pain of the arms and legs..
#### 7. Summary of technological characteristics of device compared to the predicate devices (K111482)
| Characteristic | Subject device | Predicate device | Discussion of difference | |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Present application<br>(Portable<br>Electro-Stimulation<br>Therapy Device, model | (K111482, Sonicator<br>Plus 920, Model ME<br>920 (TENS and<br>Microurrent mode)) | | |
| | LGT-235) | | | |
| Classification | 21 CFR 882.5890 | 21 CFR 890.5890 | Same | |
| Product Code | GZJ | GZJ | Same | |
| FDA Class | 2 | 2 | Same | |
| Intended Use | Symptomatic relief of<br>chronic intractable pain<br>2. Post-traumatic pain<br>3. Post-surgical pain | Symptomatic relief of<br>chronic intractable pain<br>2. Post-traumatic pain<br>3. Post-surgical pain | Same | |
| Principle of operation | sends gentle electrical<br>current to muscle group<br>via electrodes applied<br>on the skin | sends gentle electrical<br>current to muscle group<br>via electrodes applied<br>on the skin | Same | |
| target population | adults | adults | Same | |
| anatomical site | Arm and legs | Arm and legs | Same | |
| Material of Patient<br>contact components | <b>Electrodes:</b><br>Knee brace: OK cloth +<br>SBR + dacron<br>Conductive fabric:<br>Silver wire + nylon<br>Self-adhesive electrode:<br>conductive hydrogel | Electrode: conductive<br>hydrogel | Different<br>Knee brace was<br>demonstrated<br>biocompability safety by<br>passing ISO 10993-5 and<br>ISO 10993-10 tests. The<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. | |
| where used | home | home | Same | |
| Design | Handheld | Desk type | Different<br>This difference existed will<br>not affect the safety and<br>effectiveness of the<br>proposed device | |
| Power Source | Rechargeable lithium<br>battery 3.7V | AC line | Different<br>The proposed device was<br>demonstrated electrical<br>safety by passing ANSI<br>AAMI ES60601-1. The<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. | |
| Method of line<br>current isolation | N/A (3.7V rechargeable<br>lithium battery operated<br>device) | Double isolation | Different<br>This difference existed will<br>not affect the safety and<br>effectiveness of the<br>proposed device. | |
| Patient<br>Leakage | Normal<br>condition | < 1µA | Meet requirement of<br>ANSI/AAMI ES60601-1 | Similar<br>The proposed device was |
| Current | Single fault<br>condition | < 1μΑ | >50μΑ | demonstrated electrical<br>safety by meet the<br>requirement of chapter<br>8.7.4.7 of ANSI AAMI<br>ES60601-1. The<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. |
| Number<br>of Output<br>Channels | Number | One | two | Different<br>This will not affect the<br>safety and effectiveness<br>of the proposed device |
| | Synchronous<br>s or<br>Alternating? | Synchronous | Synchronous | Same |
| Number of Output<br>Modes | | two (TENS + MCR) | two (TENS + MCR<br>modes) | Same |
| Regulated Current or<br>Regulated Voltage? | | Current | Current | Same |
| | Software/Firmware/Micr<br>oprocessor Control? | Yes | Yes | Same |
| Automatic Overload<br>Trip? | | Yes | Yes | Same |
| Automatic No-Load<br>Trip? | | Yes | Yes | Same |
| Automatic Shut Off? | | "On/Off" button | "On/Off" button | Same |
| | User Override Control? | No, On/Off | No, On/Off, Hold or Stop | Same |
| Indicator<br>Display | On/Off<br>Status? | Yes | Yes | Same |
| | Low Battery? | Yes | N/A | Different<br>This difference existed will<br>not affect the safety and<br>effectiveness of the<br>proposed device |
| | Voltage/<br>Current<br>Level? | Yes (on app) | Yes | Same<br>System validation<br>testing scenarios<br>covering mitigation of<br>wireless risks in<br>accordance with<br>RED were added to our<br>full system testing<br>protocol to ensure safe<br>and effective use. |
| | Timer Range (minutes) | 1~60 minutes | 0~60 minutes | Same |
| ANSI AAMI ES60601-1 | Yes | Yes | Same | |
| IEC 60601-1-2 | Yes | Yes | Same | |
| IEC 60601-2-10 | Yes | Yes | Same | |
| Weight | main unit: 120g<br>Charging case: 150g | 11(lbs.) | Different<br>The Weight will not affect<br>the safety and<br>effectiveness of the<br>proposed device | |
| Dimensions<br>(W x H x D) | 59mm (W) × 59mm (L) ×<br>22mm (H) (mm) | 4.9 x 13.6 x10.5(in.) | Different<br>The dimensions will not<br>affect the safety and<br>effectiveness of the<br>proposed device | |
| Operating<br>condition | Temperature: 5 to 40°C;<br>Rel. humidity: ≤80%;<br>Atmosphere Pressure:<br>86.0 to 106.0kPa. | Temperature of use from<br>10°C to 40°C<br>Max. relative humidity<br>from 30% to 75%<br>Atmospheric pressure<br>from 700 hPa to1060 hPa | Similar<br>The operating condition of<br>subject device has passed<br>the safety test, and the<br>Instructions for Use<br>provides the operating<br>condition, so the<br>difference between the<br>operating conditions of<br>subject device and<br>predicate device will not<br>affect the safety and<br>effectiveness of subject<br>device | |
#### Basic Device Characteristics - Comparison with Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
# Output Specifications – Comparison with Predicate Device
| Characteristic | Subject device<br>Present application<br>(Portable<br>Electro-Stimulation<br>Therapy Device, model<br>LGT-235) | Predicate device<br>(K111482, Sonicator<br>Plus 920, Model ME<br>920 (TENS and<br>Microurrent mode)) | Discussion of difference | |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Waveform | Symmetrical biphasic | Symmetrical Biphasic | Same | |
| Maximum Output<br>Voltage (± 20%) | TENS:<br>50V @ 500 Ω<br>115 V @ 2 kΩ<br>125 V @ 10 kΩ<br>MCR:<br>0.45 V@ 500 Ω<br>1.8 V@ 2 kΩ<br>7.4 V @ 10 kΩ | TENS:<br>50 V @ 500 Ω<br>115 V @ 2 kΩ<br>125 V @ 10 kΩ<br>MCR:<br>0.37 V @ 500 Ω<br>1.5 V @ 2 kΩ<br>7.4 V @ 10 kΩ | Similar<br>The error existed in MCR<br>mode between proposed<br>device and predicate<br>device are acceptable.<br>The difference does not<br>raise the issue of<br>product's safety and<br>effectiveness. | |
| Maximum Output | TENS: | TENS: | | |
| Current (± 20%) | | 100 mA@ 500 Ω<br>58 mA@ 2 kΩ<br>13 mA@ 10 kΩ<br>MCR:<br>0.92 mA@ 500 Ω<br>0.90 mA@ 2 kΩ<br>0.74 mA@ 10 kΩ | 100 mA@ 500 Ω<br>58 mA@ 2 kΩ<br>13 mA@ 10 kΩ<br>MCR:<br>0.74 mA@ 500 Ω<br>0.75 mA@ 2 kΩ<br>0.74 mA@ 10 kΩ | The error existed in MCR<br>mode between proposed<br>device and predicate<br>device are acceptable.<br>The difference does not<br>raise the issue of<br>product's safety and<br>effectiveness. |
| Pulse Width | | TENS: 50 to 500 μs | TENS: 100 ~ 600us | Similar |
| | | MCR: 4.16 ms ~ 0.5s | MCR: 1.25ms ~ 1.67s | The predicate device is<br>safe and effective in pulse<br>width range 100 ~ 600us<br>for TENS mode and<br>1.25ms ~ 1.67s for MCR<br>mode, since the pulse<br>width range of proposed<br>device are in the range of<br>predicate device, the<br>proposed device is also<br>safe and effective with its<br>pulse width, therefore, the<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. |
| Frequency | | TENS: 1 ~ 120 Hz | TENS: 0.5 ~ 250 Hz | Similar |
| | | MCR: 1 ~ 120 Hz | MCR: 0.3 ~ 400 Hz | The frequency range of<br>predicate device is<br>overriding the range of<br>proposed device, since<br>the predicate device is<br>safe and effective,<br>therefore the proposed<br>device is also safe and<br>effective, the difference<br>does not raise the issue of<br>product's safety and<br>effectiveness. |
| Maximum intensity | | 100mA | 100 mA | Same |
| For<br>multiphasic | Symmetrical<br>phases? | Yes | Yes | Same |
| waveforms<br>only | Phase<br>Duration<br>(include<br>units)<br>(state<br>range, if<br>applicable<br>) (both<br>phases, if<br>asymmetric<br>al) | TENS: 50 to 500 µs<br>MCR: 4.16 ms ~ 0.5s | TENS: 100 ~ 600us<br>MCR: 1.25ms ~ 1.67s | Similar |
| | | | | The predicate device is<br>safe and effective in pulse<br>width range 100 ~ 600us<br>for TENS mode and<br>1.25ms ~ 1.67s for MCR<br>mode, since the pulse<br>width range of proposed<br>device are in the range of<br>predicate device, the<br>proposed device is also<br>safe and effective with its<br>pulse width, therefore,<br>difference existed does<br>not raise the issue of<br>product's safety and<br>effectiveness. |
| Stimulating surface area<br>of electrode (minimum) | | 2500mm² (50mm,<br>square) | 2015.80mm² (50.8mm<br>diameter, round) | Similar<br>The smaller of the surface<br>area the higher risk is, the<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. |
| Net Charge (µC/pulse) | | TENS:0µC @ 500Ω<br>MCR: 0µC@ 500Ω | 0 µC @ 500Ω | Same. |
| Maximum Phase<br>Charge (µC) | | TENS: 48 (µC) @ 500Ω<br>MCR: 34 (µC) @ 500Ω | TENS: 60 (µC) @ 500Ω<br>MCR: 75 (µC) @ 500Ω | Similar<br>The Maximum Phase<br>Charge of predicate<br>device is overriding the<br>proposed device, since<br>the predicate device is<br>safe and effective,<br>therefore the proposed<br>device is also safe and<br>effective, difference<br>existed does not raise the<br>issue of product's safety<br>and effectiveness. |
| Maximum Current<br>Density,<br>(mA / cm², r.m.s.) | | TENS: 0.81mA/cm2 @<br>500Ω<br>MCR: 0.036mA/cm2@<br>500Ω | TENS: 1.97 (mA/cm2)<br>@ 500Ω<br>MCR: 0.026mA/cm2 @<br>500Ω | Similar<br>For TENS mode, the<br>maximum current density<br>of proposed device is<br>overriding the proposed<br>device which means the<br>proposed device is safe<br>and effective. As for mode<br>MCR, the maximum<br>current density of<br>proposed device is little<br>bigger than predicate<br>device, however the<br>proposed device was<br>demonstrated safety by |
| | | | IEC 60601-2-10. The<br>difference does not raise<br>the issue of product's<br>safety and effectiveness. | |
| Maximum Power<br>Density (W/cm²) | TENS: 8.26 mW/cm²<br>@ 500Ω<br><br>MCR: 0.016 W/cm²<br>@ 500Ω | TENS: 39 mW/cm²<br>@ 500Ω<br><br>MCR: 0.007 mW/cm²<br>@ 500Ω | <b>Similar</b><br>For TENS mode, the<br>maximum power density<br>of proposed device is<br>overriding the proposed<br>device which means the<br>proposed device is safe<br>and effective. As for mode<br>MCR, the maximum<br>power density of proposed<br>device is bigger than<br>predicate device but far<br>less than 0.25 W/cm²<br>required by Guidance<br>Document for Powered<br>Muscle Stimulator (June<br>9, 1999), the difference<br>does not raise the issue of<br>product's safety and<br>effectiveness. | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
#### 8. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Non-clinical verification testing of the Portable Electro-Stimulation Therapy Device, model: LGT-235 included electrical, mechanical, and software tests to show the device meets its design specifications. Validation and performance testing validates that the device meets its user needs. Verification and validation test results established that the device meets its intended use, that it is as safe, as effective, and performs as well as the predicate devices, and that no new issues of safety and effectiveness were raised. The Portable Electro-Stimulation Therapy Device, model: LGT-235 was designed, verified, and validated according to the company's Design Control process and has been subjected to extensive safety and performance testing as shown in the test results provided in this submission. Verification and Validation testing data demonstrate that the device meets all of its specifications including compliance with the following standards:
- 1. ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicityISO 10993-10:2010 Biological evaluation of medical devices. Tests for irritation and skin sensitization
- 2. ANSI AAMI ES60601-1:2005/(R) 2012 and A1:2012 Medical electrical equipment. General requirements for basic safety and essential performance
- 3. IEC 60601-1-2:2014 Medical electrical equipment part 1-2: general requirements for safety - collateral standard: electromagnetic compatibility - requirements and tests
- 4. IEC 60601-2-10:2012 Medical Electrical equipment Part 2-10: Particular requirements for the safety of nerve and muscle stimulators
- 5. IEC 60601-1-11:2015 Medical electrical equipment Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
{11}------------------------------------------------
#### 9. Conclusions
The electrical safety, EMC, biocompatibility, software verification and validation, basic unit characteristics, and output specifications information provided is sufficient to demonstrate substantial equivalence to the predicate device. As the Portable Electro-Stimulation Therapy Device, model: LGT-235 is nearly identical to the predicate device, differences in their characteristics do not raise any raise new questions regarding safety and effectiveness with identical indications for use and essentially identical technological characteristics, the Portable Electro-Stimulation Therapy Device, model: LGT-235 is substantially equivalent to the predicate device Sonicator Plus 920, Model ME 920 (TENS and MCR mode).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.