Retrospective medical records; Clinical laboratory data; Patient treatment history
Retrospective analysis of serial serum samples from patients with multiple myeloma to evaluate the device's performance as an aid in monitoring disease response compared to clinical assessments based on IMWG/NCCN guidelines.
Retrospective study; Multiple Myeloma; Serial monitoring; Clinical performance
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical study of N Latex FLC assays in MM/LCMM monitoring; Retrospective observational study; Follow-up/Duration: 88 to 1338 days
98 subjects (81 IgG, IgA, IgM MM and 17 Kappa/Lambda LCMM subjects) with confirmed diagnosis; Sample Size: 98 subjects (391 serum samples)
Clinical response categories based on IMWG/NCCN guidelines
Agreement between N Latex FLC assay-based response categories and clinician-assigned disease response status
Indications for Use
In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle-enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings. The response category assignment of Complete Response for the monitoring of MM, is reliant upon the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments.
Device Story
In-vitro diagnostic reagents for quantitative determination of free light chains (FLC) in human serum and EDTA-plasma. Principle: particle-enhanced immunonephelometry; polystyrene particles coated with monoclonal antibodies agglutinate with FLC in samples; light scattering intensity proportional to protein concentration. Used on BN II and BN ProSpec systems in clinical laboratory settings. Output: quantitative FLC concentration (mg/L). Healthcare providers use results alongside clinical history, protein electrophoresis, immunofixation, bone marrow, imaging, and urine assessments to monitor multiple myeloma therapy response and aid in amyloidosis diagnosis. Benefits include standardized monitoring of disease progression or response to therapy.
Clinical Evidence
Retrospective clinical study of 98 subjects (104 MM, 21 LCMM) with 391 serum samples. Evaluated performance as an aid in monitoring MM disease status compared to IMWG clinical response criteria. Clinical sensitivity for detecting progressive disease was 26% (95% CI: 13.9%–40.3%) and clinical specificity was 97% (95% CI: 95.1%–98.8%). Comparison with predicate device showed 89% overall agreement in response category assignment.
Technological Characteristics
Particle-enhanced immunonephelometry. Reagents: polystyrene particles coated with mouse monoclonal anti-human FLC kappa or lambda antibodies. Energy source: light beam for nephelometric scattering. Dimensions/form factor: liquid reagents for automated BN II/BN ProSpec systems. Connectivity: standalone/integrated into clinical laboratory instrument workflow.
Indications for Use
Indicated for use as an aid in the diagnosis and monitoring of multiple myeloma (MM) and diagnosis of amyloidosis (AL) in patients, in conjunction with other laboratory and clinical findings. For prescription use only.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Predicate Devices
The Binding Site Freelite® Human Kappa Free Kit for use on the Siemens BN™ II (K031016)
The Binding Site Freelite® Human Lambda Free Kit for use on the Siemens BN™ II (K031016)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K182098
B. Purpose for Submission:
Modification of previously cleared devices – addition of intended use as aid in monitoring of multiple myeloma
C. Measurand:
Kappa (κ) Free Light Chain (FLC)
Lambda (λ) Free Light Chain (FLC)
D. Type of Test:
Nephelometry, quantitative
E. Applicant:
Siemens Healthcare Diagnostics Products GmbH
F. Proprietary and Established Names:
N Latex FLC kappa assay
N Latex FLC lambda assay
G. Regulatory Information:
1. Regulation section:
21 CFR § 866.5550 – Immunoglobulin (light chain specific) immunological test system
2. Classification:
Class II
3. Product code:
DFH, Kappa, antigen, antiserum, control
DEH, Lambda, antigen, antiserum, control
4. Panel:
Immunology (82)
{1}
H. Intended Use:
1. Intended uses:
N Latex FLC kappa and N Latex FLC lambda assays
In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle-enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings.
The response category assignment of Complete Response for the monitoring of MM, is reliant upon the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments.
2. Indications for use:
Same as Intended Use
3. Special conditions for use statements:
For prescription use only
Warning: The result of the FLC Kappa or FLC Lambda in a given specimen determined with assays and/or instrument platforms from different manufacturers can vary due to differences in assay methods and reagent specificity. The results reported by the laboratory to the physician must include the identity of the assay used. Values obtained with different assay methods cannot be used interchangeably. If, in the course of serially monitoring a patient, the assay method used for determining the FLC Kappa and FLC Lambda levels is changed, additional sequential testing should be carried out. Prior to changing assays, the laboratory MUST confirm baseline values for patients being serially monitored.
4. Special instrument requirements:
BN II (K943997) and BN ProSpec Systems (K001647)
I. Device Description:
The N Latex FLC kappa and N Latex FLC lambda assays are comprised of the following reagents in liquid form:
N latex FLC reagents: suspension of polysterene particles coated with mouse monoclonal antibodies to either human FLC Kappa or human FLC Lambda with preservative.
N FLC supplementary reagents: mouse immunoglobulin in buffered solution with preservatives.
{2}
N FLC standard SL and N FLC controls SL1 and SL2: human free light chains proteins, human serum albumin and protease inhibitors.
# J. Substantial Equivalence Information:
1. Predicate device names and 510(k) numbers:
The Binding Site Freelite® Human Kappa Free Kit and Freelite® Human Lambda Free Kit for use on the Siemens BN™ II - K031016
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device N Latex FLC kappa N Latex FLC lambda | Predicate Freelite® Human Kappa Free and Freelite® Human Lambda Free kits on the Siemens BN™II |
| Analyte | Kappa: Kappa FLC Lambda: Lambda FLC | Same |
| Measurement | Quantitative | Same |
| Detection Method | Nephelometric | Same |
| Calibrator | One level | Same |
| Reference material | Internal Reference preparation | Same |
| Units of measure | mg/L | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Indication for use | In-vitro diagnostic reagents for the quantitative determination of free light chains (FLC), type kappa or type lambda in human serum and EDTA-plasma by means of particle- enhanced immunonephelometry using the BN Systems. FLC measurements are used as an aid in the diagnosis and monitoring of multiple myeloma (MM) and as an aid in the diagnosis of amyloidosis (AL) in conjunction with other laboratory and clinical findings.The response category assignment of ‘Complete Response’ for the monitoring of MM, is reliant upon | This kit is intended for the quantitation of kappa free light chains or lambda free light chains in serum and urine on the Siemens BN™ II. Measurement of free light chains aids in the diagnosis and monitoring of multiple myeloma, lymphocytic neoplasms, Waldenstrom’s macroglobulinemia, AL amyloidosis, light chain deposition disease and connective tissue diseases such as systemic lupus erythematosus in conjunction with other laboratory and clinical findings. |
{3}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | the combination of clinical history and other tests including protein electrophoresis, immunofixation and bone marrow, imaging and urine assessments. | |
| Sample type | Serum and EDTA plasma | Serum and urine |
| Detection antibody | Kappa: Monoclonal mouse anti-human FLC kappa antibody coated onto polysterene particles
Lambda: Monoclonal mouse anti-human FLC lambda antibody coated onto polysterene particles | Kappa: Polyclonal sheep anti-human kappa antibody coated onto latex particles
Lambda: Polyclonal sheep anti-human lambda antibody coated onto latex particles |
| Analytical measuring ranges | Kappa: 3.4 – 110 mg/L
Lambda: 1.9 – 60 mg/L | Kappa: 5.9 – 190 mg/L
Lambda: 5.0 – 160 mg/L |
| Instrument system | Siemens BN II and BN ProSpec Systems | Siemens BN II |
| Reference interval | Kappa: 8.24 – 28.90 mg/L
Lambda: 9.10 – 32.60 mg/L
Ratio: 0.53 – 1.51 | Kappa: 3.30 to 19.40 mg/L
Lambda: 5.71 to 26.30 mg/L
Ratio: 0.26 – 1.65 |
## K. Standard/Guidance Document Referenced (if applicable):
## L. Test Principle:
The FLC test systems are based upon the principles of particle-enhanced immunonephelometry. Polystyrene particles coated with monoclonal antibodies to human free light chains (FLCs), type kappa or lambda, are agglutinated when mixed with samples containing FLCs. These aggregates scatter a beam of light passed through the sample. The intensity of the scattered light is proportional to the concentration of the respective protein in the sample. The result is evaluated by comparison with a standard of known concentration.
## M. Performance Characteristics:
1. Analytical performance:
a. Precision/Reproducibility:
See K171742
b. Linearity/assay reportable range:
See K171742
{4}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
See K171742
d. Detection limit:
See K171742
e. Analytical specificity:
See K171742
f. Assay cut-off:
See expected values/reference range below.
2. Comparison studies:
a. Method comparison with predicate device:
See K171742
b. Matrix comparison:
See K171742
3. Clinical studies:
The performance of the N Latex FLC kappa assay and the N Latex FLC lambda assay as an aid in monitoring of disease status in multiple myeloma (MM) and light chain multiple myeloma (LCMM) subjects was determined from a retrospective clinical study. Changes in FLC Kappa and FLC Lambda levels in serial serum samples from subjects diagnosed with MM and LCMM during treatment were compared to clinical response categories based on International Myeloma Working Group (IMWG) as determined by the clinicians.
The patient inclusion and exclusion criteria were as follows:
Inclusion criteria:
- Confirmed diagnosis of MM or LCMM
- At least four blood collections in intervals of ≥ three weeks
- Informed consent available
- Clinical information of response level available
Exclusion criteria:
{5}
- Less than four blood collections
- No informed consent available
- Patients subsequently diagnosed with another disease
A total of 125 (104 MM and 21 LCMM) subjects were enrolled in the study. Among them, 27 subjects were excluded: 17 subjects had less than four sample draws; four MM subjects were re-diagnosed to Waldenstrom Macroglobulinemia; two subjects had no informed consent available; and four subjects had no available clinical information about response level. Therefore, a total 98 subjects (81 IgG, IgA and IgM MM and 17 Kappa and Lambda LCMM subjects) with a total of 391 serum samples were used in the study analysis.
Of the 98 subjects, 47 (47%) were male and 51 (52%) were female. The mean age was 65 years and the range was 35 to 92 years. The majority of subjects (n=70, 71%), were Caucasian, the remaining subjects included 18 (18%) African American, one (1%) Asian and nine (9%) Unknown.
A total of 391 draws were collected from 98 subjects. The mean number of draws per subject was five, ranging from four to seven draws. The duration of the monitoring days ranged from 88 days to 1338 days.
The information collected for each sample included the following: (a) general information including patient's demographic information and date of blood sample drawn; (b) clinical information including patient's diagnosis report/medical notes, treatment initiated, and diagnostic response criteria; c) laboratory data including serum electrophoresis results, serum/ urine immunofixation electrophoresis, serum FLC Kappa and FLC Lambda, CRAB (Calcium, Renal Failure, Anemia, Bone lesions) status, and plasma cell count.
## Results:
The performance of the N Latex FLC kappa and N Latex FLC lambda assays as aid in monitoring MM was evaluated by testing a total of 391 samples from 98 subjects consisting of 98 baseline values and 293 monitoring observations and comparing to clinical disease response status.
At each follow-up visit, the percentage change of the Involved FLC Kappa (iFLC Kappa) or Involved FLC Lambda (iFLC Lambda) was calculated by comparing the test result to the test result from the previous visit. The monitoring response criteria based on the FLC results were based on the FLC ratio, % relative change of difference between both FLC (rd_FLC), M protein level and IFE monoclonal protein detection.
The clinical assessment for each monitoring observation was provided by the attending physician who was responsible for the corresponding therapy decisions based on therapy response criteria defined according to International Myeloma Working Group (IMWG) 2011/National Comprehensive Cancer Network (NCCN) Clinical Practice Guidelines in Oncology on Multiple Myeloma. Clinical status of each monitoring event was
6
{6}
categorized into one of six response categories: Stringent Complete Response (sCR), Complete Response (CR), Very Good Partial Response (VGPR), Partial Response (PR), Stable Disease (SD), and Progressive Disease (PD).
Data were analyzed based on the comparison of disease category determined from N Latex assays to clinical assessment.
Results of N Latex FLC kappa and N Latex FLC lambda assays in combination with other laboratory results were used to assign a response category and compared to the clinical assessment according to the table below:
| Response Category | Response Criteria based on N Latex FLC, SPE, IFE Results | Clinical Assessment based on IMWG |
| --- | --- | --- |
| Good Response | FLC ratio normal, M protein not detectable and IFE negative* | sCR |
| | | CR |
| Moderate Response | ≥ 50 to ≥ 90% reduction of rd_dFLC** and ≥ 50 to ≥ 90% reduction of M protein | VGPR |
| | | PR |
| Stable Disease | 50% reduction to 25% increase of rd_dFLC and 50% reduction to 25% increase of M protein | SD |
| Progressive Disease | ≥ 25% increase of rd_dFLC and ≥ 25% increase of M protein AND increase of d_dFLC*** ≥ 100 mg/L and increase of M protein ≥ 0.5 g/L | PD |
*Applies only to MM population
**rd_dFLC = (dFLC t2 - dFLC t1) / dFLC t1
\*\*\*d_dFLC = dFLC t2 - dFLC t1 (with 'dFLC = iFLC - ni FLC')
# Definitions:
- t1 = time point 1; t2 = time point 2
- rd = relative difference
- iFLC = involved FLC; niFLC = not involved FLC
- Relative change of involved FLC: rd_iFLC = (iFLC t2 - iFLC t1)/iFLC t1
- Relative change of difference between both FLC: dFLC = iFLC - niFLC
The performance of N Latex FLC Kappa and FLC Lambda in combination with other laboratory tests compared to the clinical assessment is summarized in the following table:
| Response Category | Clinical Assessment | | | | |
| --- | --- | --- | --- | --- | --- |
| | Good Response | Moderate Response | Stable Disease | Progressive Disease | Total |
| Good Response | 21 | 3 | 6 | 1 | 31 |
{7}
8
| Moderate Response | 10 | 104 | 38 | 16 | 168 |
| --- | --- | --- | --- | --- | --- |
| Stable Disease | 3 | 24 | 126 | 18 | 171 |
| Progressive Disease | 0 | 0 | 9 | 12 | 21 |
| Total | 34 | 131 | 179 | 47 | 391 |
| Agreement | 62% | 79% | 70% | 26% | 67% |
## Discordance results:
The response criteria established by clinicians are often based on multiple factors besides FLC results, therefore FLC results alone might lead to a different classification.
The clinical disease response categories were condensed into two clinical status categories: “Progression” and “No-progression”. Subjects with ‘Progression’ consist of those monitoring events defined as ‘Progressive Disease’. Subjects with ‘No-Progression’ consist those monitoring events defined as ‘Good Response, Moderate Response, and Stable Disease Response’ categories. The performance N-Latex FLC results compared to the clinical status are summarized as follows:
| | Clinical Status | | | |
| --- | --- | --- | --- | --- |
| | | Progression | No Progression | Total |
| Change in N Latex FLC | Positive* | 12 | 9 | 21 |
| | Negative** | 35 | 335 | 370 |
| | Total | 47 | 344 | 391 |
| Clinical Sensitivity: 26% (12/47) (95% CI: 13.9% – 40.3%)
Clinical Specificity: 97% (335/344) (95% CI: 95.1%– 98.8%) | | | | |
*Positive: ≥25% rd_dFLC and d_dFLC ≥100 mg/L
**Negative: <25% rd_dFLC and d_dFLC FLC <100 mg/L
b. Other clinical supportive data (when a. and b. are not applicable):
The performance of N Latex FLC kappa and FLC lambda assays were also compared to the predicate device based on the disease response categories described above.
In this retrospective clinical study, all samples were tested with the N Latex FLC kappa and FLC lambda assays and with the predicate device. The performance of N Latex FLC kappa and N Latex FLC lambda assays compared to the predicate device based on the same disease response category is summarized as follows:
| Response based on N Latex FLC results | Response Based on Freelite Results | | | | |
| --- | --- | --- | --- | --- | --- |
| | Good Response | Moderate Response | Stable Disease | Progressive Disease | Total |
| Good Response | 27 | 2 | 0 | 1 | 31 |
| Moderate Response | 0 | 66 | 15 | 0 | 81 |
| Stable Disease | 0 | 13 | 238 | 7 | 258 |
| Progressive Disease | 0 | 0 | 5 | 16 | 21 |
| Total | 27 | 81 | 259 | 24 | 391 |
| Agreement | 100% | 81% | 92% | 67% | 89% |
{8}
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
See K171742
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.