K182070 · BioHorizons Implant Systems, Inc. · DZE · Oct 10, 2018 · Dental
Device Facts
Record ID
K182070
Device Name
BioHorizons Tapered IM Implants
Applicant
BioHorizons Implant Systems, Inc.
Product Code
DZE · Dental
Decision Date
Oct 10, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K182070 · Oct 10, 2018
BioHorizons Tapered IM Implants
BioHorizons Implant Systems, Inc.
Published clinical literature (Nevins et al. 2008, Ricci et al. 2000)
The sponsor relies on published clinical literature to support the safety and effectiveness of the Laser-Lok surface technology, specifically regarding connective tissue attachment and bone response.
Literature review; Human histologic evidence; Bone response
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Nevins et al. Human Histologic Evidence of a Connective Tissue Attachment to a Dental Implant; Human histologic study
Human subjects
Not applicable for this study
Connective tissue attachment to dental implant
Ricci et al. Bone Response to Laser Microtextured Surfaces; Clinical/Experimental study
—
Not applicable for this study
Bone response to laser microtextured surfaces
Indications for Use
BioHorizons Tapered IM Implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The implants may be restored immediately (1) with a temporary prosthesis that is not in functional occlusion or (2) when splinted together for multiple tooth replacement or when stabilized with an overdenture supported by multiple implants.
Device Story
BioHorizons Tapered IM Implants are machined titanium alloy (ASTM F136) screw-form endosseous dental implants. Features include Resorbable Blast Texture (RBT) surface treatment and Laser-Lok micro-machined grooves on the collar to inhibit epithelial downgrowth and promote connective tissue attachment. Used by dental surgeons in the mandible or maxilla for tooth replacement or dental retention. Implants are provided sterile via gamma irradiation. Clinicians use the device as an artificial root structure; the output is the physical implant itself, which provides a foundation for prosthetic restoration. Clinical benefit includes restoration of chewing function and dental aesthetics.
Clinical Evidence
Bench testing only. Comparative dimensional assessment confirmed structural equivalence to predicates. Dynamic mechanical fatigue testing (ISO 14801) performed on worst-case implant-abutment assemblies, surviving 5 million cycles. Sterilization validated per ANSI/AAMI/ISO 11137-1. Endotoxin testing performed per ANSI/AAMI ST72, meeting USP <161> limits. No clinical data presented.
Indicated for use in the mandible or maxilla as an artificial root structure for single tooth replacement, fixed bridgework, or dental retention. May be restored immediately with a non-functional temporary prosthesis, or when splinted for multiple tooth replacement, or stabilized with an overdenture supported by multiple implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
BioHorizons Tapered Internal Plus Implants (K121787)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 10, 2018
BioHorizons Implant Systems, Inc. Michael Davis Director, Regulatory Affairs 2300 Riverchase Center Birmingham, Alabama 35244
Re: K182070
Trade/Device Name: BioHorizons Tapered IM Implants Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: September 7, 2018 Received: September 11, 2018
Dear Michael Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Indications for Use | Food and Drug Administration | | 510(K) Number (if known)<br>Device Mame | | | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|--|-------------------------------------------------------------------------------------------------------------------|--|--|--|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| See PRA Statement below | isquration Date: 06/30/2020 | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | BioHorizons Tapered IM Implants | |
| | | | | | | | . Mar starting and oversioner supportunes superformands.<br>tooth replacement or for fixed bridgework and dental retention.<br>Indications for Use (Describe) | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| The implants may be restored immediately ( ) with a | | | and and surver the first in as a stlixsur on in in as a referent and and and the more the more of a | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | ro memore of the register of the redware redw ( ) 10 noisules and right in tour and the reserved vir of moders of | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | Type of Use (Select one or poth, as applicable) | |
| | | | | | | | | |
| (D) Presscription Use (Part 21 CFR 801 Subpart D) | | | | | | | | |
| I Over-The-Counter Use (2) CFR 801 Suppart C) | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| OD | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| see single and the sense and to strement sed the sell general see and | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| A would and and and and of the on and of the submit of the only | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| itime to respent and mistuism burses, gather sunses, gather states needs needs meets and<br>and pribulari in a sund en a super of beteritise si noitemoni in noisello sing not not amit nebud edit | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| of this information collection, including suggestions for reducing this purder, to: | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Delivery of Health and Human Services | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Food and Drug Administration | | | | | | | | |
| to rest to the collection of internations send comments would be the survey of the | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Office of Chief Information Officer | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| Paperwork Reduction Act (PRA) State | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| PRASStaff@gda.hhs.gov | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
୍ର ପ
,以
ﺒﻨﺎﻧﺔ ﺍﻟ
。
{3}------------------------------------------------
## 510(k) Summarv 21 CFR 807.92
### Submitter's Name & Address
| Manufacturer: | BioHorizons Implant Systems, Inc.<br>2300 Riverchase Center<br>Birmingham, AL 35244<br>Phone (205) 967-7880<br>Fax (205) 870-0304 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Official contact: | Michael Davis, Director, Regulatory Affairs |
| Date prepared: | October 10, 2018 |
# Name of the Device
| Trade Name: | BioHorizons Tapered IM Implants |
|------------------------|---------------------------------|
| Common or Usual Name: | Screw-type dental implant |
| Classification Name: | Endosseous dental implant |
| Classification Number: | Class II (21 CFR 872.3640) |
### Predicate Devices
Primary Predicate Device:
- 1. K121787, BioHorizons Tapered Internal Plus Implants, September 5, 2012.
# Reference Predicate Devices:
- 1. K103691, BioHorizons Abutments for Zimmer, November 3, 2011.
- 2. K151621, BioHorizons CAD/CAM Abutments, January 13, 2016.
- 3. K071638. BioHorizons Tapered Internal Implant System, October 10, 2017.
- 4. K071161, Endosseous Dental Implant System, November 16, 2007.
### Device Description
The BioHorizons Tapered IM Implants are machined titanium, screw-form endosseous dental implants supplied in 7.0mm and 8.0mm diameters. Both implants include a 5.7mm prosthetic The implants are provided in 7.5mm, 9.0mm and 10.5mm lengths across both platform. diameters. Implant material is titanium alloy as specified in ASTM F136 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy (UNS R56401) for Surqical Implant Applications.
The devices are further processed by roughening the threaded surface with Resorbable Blast Texture (RBT) media (hydroxylapatite) and by micro-machining grooves, known as Laser-Loke microchannels, on the implant collar. The product is packaged using materials known in the industry to be appropriate for medical device packaging and is provided with a minimum sterility assurance level of 10°, validated in compliance with ANSI/AAMI/ISO 11137-1 Sterilization of healthcare products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
### Indications for Use
BioHorizons Tapered IM Implants are intended for use in the mandible or maxilla as an artificial root structure for single tooth replacement or for fixed bridgework and dental retention. The implants may be restored immediately (1) with a temporary prosthesis that is not in functional occlusion or (2) when splinted together for multiple tooth replacement or when stabilized with an overdenture supported by multiple implants.
{4}------------------------------------------------
# Technological Characteristics
The fundamental scientific technology of the BioHorizons Tapered IM endosseous dental implant devices subject to this 510(k) is substantially equivalent to the primary predicate device. The threaded portion of the implants is RBT-blasted, and Laser-Lok microchannels are applied to the implant collar.
Laser-Lok is a surface feature in which patterns of micro-machined grooves are applied to the collar of a dental implant, providing a roughened surface to establish a physical, connective tissue attachment. This tissue connection:
- 1) is functionally oriented,
- 2) inhibits epithelial cell downgrowth, and
- 3) enables crestal bone adjacent to the implant to attach and be retained
The preceding claims are supported in the published literature by Nevins et al. Human Histologic Evidence of a Connective Tissue Attachment to a Dental Implant, International Journal of Periodontics & Restorative Dentistry 2008; 28:111-121 and Ricci et al. Bone Response to Laser Microtextured Surfaces, Bone Engineering 2000:382-392.
All materials, suppliers, processing, packaging and sterilization methods remain the same as for the primary predicate device, BioHorizons Tapered Internal Plus Implants (K121787). The Laser-Lok feature is substantially equivalent to that cleared for the BioHorizons Tapered Internal Plus Implants. The BioHorizons Tapered IM Implants are substantially equivalent to the features of the predicate implant devices because of the similarities in design, materials and intended use. Refer to Table 1, Summary Table of Substantial Equivalence, immediately following.
### Summary of Testing
A comparative dimensional assessment was performed comparing the worst-case subject Tapered IM implant d…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.