Erran Dental Zirconia are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facing, and veneers.
Device Story
Erran Dental Zirconia consists of zirconia-based ceramic blocks and powder used by dental professionals or laboratories to fabricate all-ceramic dental restorations, including crowns, bridges, veneers, and inlays/onlays. The device is derived from yttria-stabilized tetragonal zirconia (Y-TZP) powder. Blocks are pre-formed; powder is used to directly shape teeth. The material is sintered to achieve final mechanical properties. The device serves as a restorative material to replace missing or damaged tooth structure, benefiting patients by providing durable, biocompatible prosthetic teeth. It is a single-use, non-sterile material processed in a dental laboratory setting.
Clinical Evidence
Bench testing only. Performance verified against ISO 6872-2015 (Dentistry: Ceramic Materials) and ISO 13356-2015 (Implants for Surgery-Ceramic Materials). Biocompatibility testing conducted per ISO 7405-2008 and ISO 10993-10, including cytotoxicity, mucosal irritation, and sensitization. All test results met acceptance criteria.
Technological Characteristics
Material: Yttria-stabilized tetragonal zirconia (ZrO2+Y2O3+HfO2+Al2O3) conforming to ISO 13356-2015. Form factor: Cuboid blocks, cylindrical blocks, and powder. Sintering temperature: 1350-1600°C. Post-sintering density: >6.02 g/cm³. Fracture toughness: >800 MPa. Elastic modulus: 210 GPa. Porosity: 0%.
Indications for Use
Indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facing, and veneers.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a stylized depiction of a human figure, while the FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are placed side-by-side, with the HHS logo on the left and the FDA logo on the right. The logos are commonly used to represent the two government agencies.
December 13, 2018
Hangzhou Erran Technology Co., Ltd. Tingkai Li Chief Technology Officer Room F3022, Floor 3, Building 1 (south) No. 368 Liuhe Road Hangzhou, 310053 Cn
Re: K182068
Trade/Device Name: Erran Dental Zirconia Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: July 28, 2018 Received: August 1, 2018
Dear Tingkai Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Mary S. Runner Digitally signed by Mary S.
> Runner -s3 -S3 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182068
Device Name Erran Dental Zirconia
Indications for Use (Describe)
Erran Dental Zirconia are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facing, and veneers.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo of Hangzhou Erran Technology Co., Ltd. The logo consists of a blue square with a green circle inside, and a horizontal green line across the middle of the circle. To the right of the logo is the company name, "Hangzhou Erran Technology Co.,Ltd" in black text.
#### 510(k) Summary K182068
1. Date
August 28, 2018
### 2. Submitter
Hangzhou Erran Technology Co.,Ltd. Room F3022, Floor 3, Building 1 (south)No. 368 Liuhe Road, Binjiang District Hangzhou, 310053 zhejiang, china
### 3. Official contact person
Tingkai Li Chief Technology Officer tingkai.li@errantech.com Tel: +86-15968174962 (China) 1-360 216 1286 (USA) Fax: +86-571-88055546
#### 4. Product information
Proprietary Name Erran Dental Zirconia
Common Name Dental Zirconia Ceramics
Classification Name (Number) Procelain powder for clinical use (872.6660)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo of Hangzhou Erran Technology Co., Ltd, which features a blue and green icon. Below the company name, it states "Proposed Regulatory Class" followed by "Class II". The text is in a clear, readable font, and the overall layout is simple and professional.
Product Code EIH
Predicate Devices Aidite Zirconia Dental Ceramics (K111291)
#### 5. Device Description
Normally, Erran Dental Zirconia are offered in blocks and powder and the size as shown below in Table 5-1, and are similar to the sizes and shapes of the predicate device.
There are two types of our device, one is block, another is powder. The main difference is that the blocks are made by powder, zirconia blocks is prepared and have a fixed shape, while zirconia powder is used to directly product the shape of the tooth.
Erran Dental Zirconia are derived from Zirconia powder that has been processed into their final net shapes. Erran Dental Zirconia have are then being further fabricated in to all-ceramic restoration such as crowns, bridge, veneers, inlay/onlay. The zirconia powder is composed of ZrO2+Y2O3+HfO2+Al2O3 with its composition conforms to ISO13356-2015, Implants for Surgery-Ceramic Materials Based in Yttria-Stabilized Tetragonal Zirconia(Y-TZP). The performance of the dental zirconia conforms to ISO 6872-2015, Dentistry: Ceramic Materials.
| | Model (Device size, unit: mm) |
|--------------------|---------------------------------------------------------------------|
| Cuboid blocks | (Length - width - height, unit: mm) 20-14-15; 20-19-15;<br>40-14-15 |
| Cylindrical blocks | (Diameter - Thickness, unit: mm) D15-16; D18-14; D98-14 |
| Powder | (Weight, unit: g) 60 |
Table 5-1 The normal model of Erran Dental Zirconia
#### 6. Indications for Use
Erran Dental Zirconia are intended for use for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facing, and veneers.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Hangzhou Erran Technology Co., Ltd. The logo features a stylized letter "E" formed by a blue rectangle and a green semi-circle. The text "Hangzhou Erran Technology Co.,Ltd" is written in black, with the company name following the logo. The text is aligned to the right of the logo.
#### 7. Summary of Device Testing
Bench testing was performed per ISO 6872-2015 and interal procedures to ensure that the Erran Dental Zirconia met its specification. All tests were verified to meet acceptance criteria. Biocompatibility testing was performed to verify the equivalent safety of the materials that are used.
The following table 5-2 shows test requirement and result of our device. Tests were conducted following the recommended procedures outline in the FDA recognized consensus standard of ISO 6872-2015, ISO7405-2008 and ISO 10993-1-2009, and results meet all relevant requirements in the test standards.
| Item | The Requirements | Result | Conclusion |
|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Chemical<br>Composition(%) | Meet the ISO 13356-2015 | ZrO2+HfO2 ≥99<br>+Y2O3<br>Y2O3 4.5~6.6<br>HfO2 ≤5<br>Al2O3 ≤0.5<br>SiO2 ≤0.3<br>Fe2O3 ≤0.2 | Meet the<br>ISO<br>13356-2015 |
| Appearance | There is no spot cracks and<br>foreign matter on the surface<br>of Erran Dental Zirconia. | There is no spot cracks and<br>foreign matter on the<br>surface of Erran Dental<br>Zirconia. | Conform |
| Size(mm) | The size deviation of<br>zirconia blocks is within<br>1.0mm.<br>The weight deviation of<br>zirconia powder is within<br>10g. | Meet the requirement. | Conform |
| Density (g/cm³) | Not less than 2.6, the<br>deviation is less than 0.05<br>compared with marked<br>value(3.16). | Meet the requirement. | Conform |
| Sintering<br>density<br>(g/cm³) | Not less than 6.02 | 6.025 | Conform |
| Fracture<br>Toughness (post<br>sintering)<br>(MPa) | At least 8 samples of the 10<br>samples have a fracture<br>toughness greater than 800 | The samples met the<br>criteria. | Conform |
| Chemical<br>solubility<br>(µg/cm²) | <100 | 12.42 | Conform |
| Shrinkage | At least 8 samples of the 10<br>samples have a Shrinkage<br>which should between<br>19~27 | Average is 19.61, all<br>samples are between 19 - 27<br>range. | Conform |
| Mixing<br>and<br>condensation<br>properties(only<br>for<br>zirconia<br>powder) | After the zirconia powder is<br>compacted, it should not be<br>cracked or broken, and after<br>forming compaction, it may<br>not be agglomerated. | The samples meet the<br>requirements. | Conform |
| The coefficient<br>of thermal<br>expansion (only<br>for<br>zirconia<br>powder) | $6.5 × 10^{-6} K^{-1}~11.5 × 10^{-6}$<br>$K^{-1}$ , the deviation is less<br>than $0.5 × 10^{-6} K^{-1}$ | $10.5 × 10^{-6} K^{-1}$ | Conform |
| Radioactivity | The activity concentration of<br>U-238 is not more than 1.0 | <0.005 | Conform |
| Cytotoxicity<br>Irritation Oral<br>Mucosa<br>Irritation<br>Sensitization<br>Subacute<br>Subchronic<br>Toxicity<br>Genotoxicity | Biocompatibility testing<br>including cytotoxicity,<br>mucosal irritation,<br>sensitization conducted in<br>accordance with ISO<br>7405-2008 and ISO<br>10993-10. | We conducted comparative<br>performance testing for<br>biocompatibility testing<br>including cytotoxicity,<br>mucosal irritation,<br>sensitization and etc.<br>conducted in accordance<br>with ISO 7405-2008 and<br>ISO 10993-10, and meet the<br>requirements. | Conform |
#### Table 5-2 Test summary table
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a blue square containing a stylized letter 'E'. The 'E' is formed by a green semi-circle overlapping a white shape, creating the impression of the letter. The logo is simple and modern, using a combination of geometric shapes and contrasting colors.
# Hangzhou Erran Technology Co.,Ltd
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo of Hangzhou Erran Technology Co., Ltd. The logo features a stylized letter 'E' formed by blue and green shapes. The company name is written in a clear, sans-serif font to the right of the logo.
## 8. Comparison to Predicate Device
The following table 5-3 shows similarities and differences of use, design, and the materials between our device and the predicate devices.
| Description | Erran Dental Zirconia<br>K182068 | Predicate Device<br>K111291 | Comparison |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Indication for<br>use | Erran Dental Zirconia<br>Blocks are indicated for the<br>production of artificial teeth<br>in fixed or removable<br>dentures, or for jacket<br>crowns, facing, and veneers.<br>All Blocks are processed<br>thought dental laboratories<br>or by dental professionals. | Aidite Zirconia Dental<br>Ceramics are indicated for<br>the production of artificial<br>teeth in fixed or removable<br>dentures, or for jacket<br>crowns, facing, and<br>veneers. All Blanks are<br>processed thought dental<br>laboratories or by dental<br>professionals. | Same |
| Single Use | Yes | Yes | Same |
| Color | None | None, and Pre-shaded (for<br>Pre-shaded series) | Different, Does<br>not affect the<br>use of the<br>product |
| Type<br>Specification | Blocks, powder | Blocks, disc, and rod | Different, Does<br>not affect the<br>use of the<br>product |
| Materials | Zirconia(ZrO2+Y2O3+HfO2≥<br>99.0%) | Zirconia(ZrO2+Y2O3+HfO2<br>+Al2O3≥99.0%) | Different, Does<br>not affect the<br>use of the<br>product |
| Crystal<br>Morphology | Tetragonal | Tetragonal | Same |
| Porosity | 0% | 0% | Same |
| Sintering<br>temperature | 1350-1600°C | 1480°C | Different |
| Density (pre<br>sintering) | $3.10g/cm^3$ | $3.10g/cm^3$ | Same |
| Density (post<br>sintering) | $>6.02g/cm^3$ | $6.05g/cm^3$ | Different, Does<br>not affect the<br>use of the |
| | | | | Table 5-3 Comparison to Predicate Device(Zirconia Dental Ceramic) |
|--|--|--|--|-------------------------------------------------------------------|
|--|--|--|--|-------------------------------------------------------------------|
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo with a blue square in the background. A green circle is in the foreground, partially overlapping the blue square. A horizontal blue line runs through the center of the green circle and the blue square.
# Hangzhou Erran Technology Co.,Ltd
| | | | product |
|--------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|------------------------------------------------------------|
| Fracture<br>Toughness (pre<br>sintering) | 55Mpa | 55Mpa | Same |
| Bending<br>strength | > 800MPa; | > 800MPa; | Same |
| Shrinkage | 19%-27% | 22.00% | Different, Does<br>not affect the<br>use of the<br>product |
| Elastic modulus<br>(post sintering) | 210Gpa | 210Gpa | Same |
| Porosity | 0% | 0% | Same |
| Chemical<br>solubility | <100µg/cm² | <100µg/cm² | Same |
| Radioactive | The activity concentration of<br>U-238is not more than 1.0. | The activity concentration<br>of U-238is not more than<br>1.0. | Same |
| Cytotoxicity<br>Irritation Oral<br>Mucosa<br>Irritation<br>Sensitization<br>Subacute<br>Subchronic<br>Toxicity<br>Genotoxicity | We conducted comparative<br>performance testing for<br>biocompatibility testing<br>including cytotoxicity,<br>mucosal irritation,<br>sensitization and etc.<br>conducted in accordance<br>with ISO 7405-2008 and ISO<br>10993-10, and meet the<br>requirements. | No biocompatibility testing<br>provided in the submission | Different |
Erran Dental Zirconia meets the requirement per ISO6872-2015 and ISO7405-2015. Its performance meets the requirements of its predefined acceptance criteria and intended use, and results are also comparable to the predicate device as evidenced in comparative performance testing.
### 9. Substantial Equivalence Conclusion
Bases upon technology, indications for use, and non-clinical performance testing, Erran Dental Zirconia Blocks is substantially equivalent to Aidite Zirconia Dental Ceramics K111291.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.