VITROS Chemistry Products CRBM Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy. VITROS Chemistry Products CREA Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Product CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5.1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. VITROS Chemistry Products TBIL Slides: Rx Only. For in vitro diagnostic use only. VITROS Chemistry Products TBIL Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hematological and metabolic disorders, including hepatitis and gall bladder block. VITROS XT 7600 Integrated System: Rx Only. For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest.
Device Story
VITROS XT 7600 is a fully automated, computer-controlled clinical chemistry/immunodiagnostic analyzer. It processes biological fluids (serum, plasma, urine, CSF) using MicroSlide, MicroTip, or MicroWell reagents. Input samples are programmed, metered, and processed via four detection systems: reflection densitometry (MicroSlides), transmission spectrophotometry (MicroTip), chemiluminescence (MicroWell), and electrometer (ISE). The system performs auto-dilution, repeat/reflex testing, and sample quality monitoring (hemolysis, icterus, turbidity). Results are reported to the operator via a touch-screen interface and can be integrated with LIS/LAS. Predictive alerts are sent to service groups via eConnectivity. The device aids clinicians in monitoring drug therapy, diagnosing renal/liver disease, and managing metabolic disorders, providing rapid, high-throughput diagnostic data.
Clinical Evidence
Bench testing only. Precision studies (CLSI EP05-A3) performed over 20 days (n=80) showed acceptable CVs. Linearity (CLSI EP06-A) and detection limits (CLSI EP17-A2) were verified. Method comparison studies against the VITROS 5600 predicate (n=116-125 samples) demonstrated high correlation (r=1.00) and slopes near 1.00 for all analytes. Interference testing (CLSI EP07-A2) identified specific drug/metabolite limitations.
Technological Characteristics
Fully automated clinical chemistry analyzer. Detection systems: reflection densitometry (MicroSlides), transmission spectrophotometry (MicroTip), chemiluminescence (MicroWell), and electrometer (ISE). Throughput: 845 tests/hour. Connectivity: eConnectivity for remote monitoring. User interface: 17-inch touch screen. Software-controlled sample/reagent metering and dilution (1:400).
Indications for Use
Indicated for quantitative measurement of carbamazepine, creatinine, and total bilirubin in serum, plasma, or urine to monitor therapy, diagnose renal disease, or assess liver/metabolic disorders. For prescription use only in clinical laboratory settings.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
VITROS Chemistry Products CRBM Slides (k160495)
VITROS Chemistry Products CREA Slides (k063591)
VITROS Chemistry Products TBIL Slides (k840880)
VITROS 5600 Integrated System (k081543)
Submission Summary (Full Text)
{0}
# **510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE**
# **A. 510(k) Number:**
k182063
# **B. Purpose for Submission:**
Adding previously cleared assays to a new instrument platform
# **C. Measurand:**
Carbamazepine, Creatinine, and Total Bilirubin
# **D. Type of Test:**
Carbamazepine assay: quantitative competitive immunoassay
Creatinine assay: quantitative colorimetric assay
Total Bilirubin assay: quantitative colorimetric assay
# **E. Applicant:**
Ortho-Clinical Diagnostics, Inc.
# **F. Proprietary and Established Names:**
VITROS Chemistry Products CRBM Slides
VITROS Chemistry Products CREA Slides
VITROS Chemistry Products TBIL Slides
VITROS XT 7600 Integrated System
# **G. Regulatory Information:**
| Analyte | Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- | --- |
| Creatinine | JFY | II | 21 CFR 862.1225 | Chemistry (75) |
| Total Bilirubin | CIG | II | 21 CFR 862.1110 | Chemistry (75) |
| Carbamazepine | KLT | II | 21 CFR 862.3645 | Toxicology (91) |
| Discrete photometric chemistry analyzer for clinical use | JJE | I | 21 CFR 862.2160 | Chemistry (75) |
1
{1}
# **H. Intended Use:**
# 1. Intended use(s):
See Indication(s) for use below.
# 2. Indication(s) for use:
Carbamazepine: Rx Only. For *in vitro* diagnostic use only. VITROS Chemistry Products CRBM Slides quantitatively measure carbamazepine (CRBM) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements obtained are used in monitoring levels of carbamazepine to help ensure appropriate therapy.
Creatinine: Rx Only. For *in vitro* diagnostic use only. VITROS Chemistry Product CREA Slides quantitatively measure creatinine (CREA) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Total Bilirubin: Rx Only. For *in vitro* diagnostic use only. VITROS Chemistry Products TBIL Slides quantitatively measure total bilirubin (TBIL) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Measurements of the levels of bilirubin are used in the diagnosis and treatment of liver, hemolytic hematological and metabolic disorders, including hepatitis and gall bladder block.
VITROS XT 7600 Integrated System: Rx Only. For *in vitro* diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest.
# 3. Special conditions for use statement(s):
For prescription use only.
# 4. Special instrument requirements:
VITROS XT 7600 Integrated System
2
{2}
# I. Device Description:
# VITROS Chemistry Products CRBM Slides
The VITROS CRBM Slide is a multilayered, analytical element coated on a polyester support. The reactive ingredients per cm² are immobilized mouse monoclonal anti-carbamazepine antibody 0.02 mg; carbamazepine-horseradish peroxidase conjugate 1.6 ng; and 2-(3,5- dimethoxy-4-hydroxyphenyl)-4,5-bis(4-dimethylaminophenyl) imidazole (leuco dye) 0.02 mg.
# VITROS Chemistry Products CREA Slides
The VITROS CREA Slide is a multilayered, analytical element coated on a polyester support. The reactive ingredients per cm² are creatinine amidohydrolase (Flavobacterium sp.) 0.20 U; creatine amidinohydrolase (Alcaligenes sp.) 3.6 U; sarcosine oxidase (Bacillus sp.) 0.55 U; peroxidase (horseradish root) 1.6 U and 2- (3,5-dimethoxy-4-hydroxyphenyl)-4,5-bis(4- dimethylaminophenyl) imidazole (leuco dye) 32 μg.
# VITROS Chemistry Products TBIL Slides
The VITROS TBIL Slide is a multilayered, analytical element coated on a polyester support. The reactive ingredients per cm² are dyphylline 0.5 mg and 4-(N-carboxymethylaminosulfonyl) benzene diazonium hexafluorophosphate 57 μg.
# VITROS XT 7600 Integrated system:
The VITROS XT 7600 Integrated System is a fully automated, computer controlled, clinical chemistry and immunodiagnostic analyzer intended for the in vitro determination of a variety of analytes in biological fluids such as serum, plasma, urine and cerebral spinal fluid. The System operates in conjunction with reagents, calibrators and controls designed for use with the system in the MicroSlide format.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
VITROS CRBM Slides on the VITROS 5600 Integrated System
VITROS CREA Slides on the VITROS 5600 Integrated System
VITROS TBIL Slides on the VITROS 5600 Integrated System
VITROS 5600 Integrated System
2. Predicate 510(k) number(s):
k160495
k063591
k081543
k840880
3
{3}
# 3. Comparison with predicate:
Carbamazepine:
| Similarities and Differences | | |
| --- | --- | --- |
| Device Characteristic | Candidate Device VITROS CRBM Slides k182063 | Predicate Device VITROS CRBM Slides k160495 |
| Intended Use | For the quantitative measurement of carbamazepine (CRBM) concentration in serum and | Same |
| Measuring range | 3.0–20.0 (μg/mL) | Same |
| Basic principle | Multiple-point Immuno-rate | Same |
| Wavelength | 670 nm, 540 nm is also used for wash detection | Same |
| Sample type | Serum and plasma | Same |
| Sample volume | 11 μL | Same |
| Instrumentation | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ **XT 7600** Integrated System | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System |
Creatinine:
| Similarities and Differences | | |
| --- | --- | --- |
| Device Characteristic | Candidate Device VITROS CREA Slides k182063 | Predicate Device VITROS CREA Slides k063591 |
| Intended Use | For the quantitative measurement of creatinine (CREA) concentration in serum, plasma, and urine. | Same |
| Measuring range | Serum/Plasma 0.15–14.0 (mg/dL), Urine 3.2–346.5 (mg/dL) | Same |
| Basic principle | Two-point rate | Same |
| Wavelength | 670 nm | Same |
| Sample type | Serum, plasma, urine | Same |
| Sample volume | 6 μL | Same |
| Instrumentation | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ **XT 7600** Integrated System | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System |
4
{4}
Total Bilirubin:
| Similarities and Differences | | |
| --- | --- | --- |
| Device Characteristic | Candidate Device VITROS TBIL Slides k182063 | Predicate Device VITROS TBIL Slides k840880 |
| Intended Use | For the quantitative measurement of total bilirubin (TBIL) concentration in serum and plasma. | Same |
| Measuring range | 0.1–27.0 (mg/dL) | Same |
| Basic principle | Dual wavelength endpoint | Same |
| Wavelength | measured at 2 wavelengths, 460 and 540nm | Same |
| Sample type | Serum and plasma | Same |
| Sample volume | 10 μL | Same |
| Instrumentation | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600/ **XT 7600** Integrated System | VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System |
VITROS XT 7600 Integrated System:
| Similarities and Differences | | |
| --- | --- | --- |
| Device Characteristic | Candidate Device VITROS XT 7600 Integrated System k182063 | Predicate Device VITROS 5600 System k081543 |
| Intended use | For use in the measurement of a variety of analytes of clinical interest. | Same |
| Operating principle | Sample programming, sampling processing, result calculation, result reporting | Same |
| Modes of operation | Continuous, Random, STAT | Same |
| Throughput | 845 tests per hour | Same |
5
{5}
| Similarities and Differences | | |
| --- | --- | --- |
| Device Characteristic | Candidate Device VITROS XT 7600 Integrated System k182063 | Predicate Device VITROS 5600 System k081543 |
| Fundamental scientific technology | The analyzer uses four main detection systems: 1. Reflection densitometry for colorimetric and Immunorate VITROS MicroSlides. 2. Transmission spectrophotometry for VITROS MicroTip assays. 3. Enhanced chemiluminescent detection for VITROS MicroWell assays. 4. Electrometer for VITROS MicroSlide ion-selective electrode (ISE) assays. | Same |
| User interface | Touch screen (17-inch monitor), keyboard, ADD | Same |
| Sample and reagent volume verification | Verification to ensure sufficient quantity of sample and reagent to run requested assays | Same |
| On-Board Dilution Range | Dilution factor of 1: 400 | Same |
| Predictive alerts through eConnectivity | The predictive alerts are logged and are electronically sent to the equipment service group real time for monitoring. | Same |
| e-Connectivity | Yes | Same |
| The following subsystems of the VITROS 5600 Integrated System are modified | REFL – Reflectometer SLIN – Slide Incubator SLSU – Slide Supply SAIN – Sample Integrity SRME – Sample and Reagent Metering SWCT - System Control and Sample Processing Software SWIN – Software Infrastructure SWUI – Graphical User Interface Software ADDI – Assay Data Disk. | All modifications pertain solely to the MicroSlide processing center. There are no changes being made to the MicroTip and MicroWell processing centers. |
6
{6}
# **K. Standard/Guidance Document Referenced (if applicable):**
CLSI EP05-A3, Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline -Third Edition
CLSI EP06-A, Evaluation of Linearity of Quantitative Measurement Procedures, Approved Guideline; 1st Edition
CLSI EP07, Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition
CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline-Second Edition
IEC 61010-1:2010, Safety Requirement for Electrical Equipment for Measurement, Control, and Laboratory Use-Part 1: General Requirements
IEC 61010-2-010:2014, Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use-Part 2-010: Particular Requirements for Laboratory Equipment for the Heating of Materials
IEC 61010-2-101:2015, Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use-Part 2-101: Particular Requirements for In Vitro Diagnostic (IVD) Medical Equipment
IEC/EN 61326-1:2012- Electrical Equipment for Measurement, Control and Laboratory Use-EMC requirements-Part 1: General Requirements
IEC/EN 61326-2-6:2013- Electrical Equipment for Measurement, Control and Laboratory Use-EMC requirements-Part 2-6: Particular Requirements- In Vitro Diagnostic (IVD) Medical Equipment
# **L. Test Principle:**
VITROS CRBM Slides (Multiple-point colorimetric/immuno-rate assay):
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Carbamazepine in the sample competes with the carbamazepine-peroxidase conjugate for a limited number of antibody binding sites during incubation. The subsequent addition of 12 μL of VITROS Immuno-Wash Fluid to the slide removes unbound carbamazepine-peroxidase conjugate from the read area, while also providing a substrate for the enzyme mediated oxidation of leuco dye. The rate of dye formation, as monitored by reflectance spectrophotometry during incubation, is inversely proportional to the carbamazepine concentration in the sample. To determine if an adequate wash has occurred, a wash detection dye is read at 540 nm during incubation.
7
{7}
VITROS CREA Slides (Colorimetric, two-point rate assay):
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. Creatinine diffuses to the reagent layer, where it is hydrolyzed to creatine in the rate-determining step. The creatine is converted to sarcosine and urea by creatinine amidinohydrolase. The sarcosine, in the presence of sarcosine oxidase, is oxidized to glycine, formaldehyde, and hydrogen peroxide. The final reaction involves the peroxidase-catalyzed oxidation of a leuco dye to produce a colored product. Following addition of the sample, the slide is incubated. During the initial reaction phase, endogenous creatine in the sample is oxidized. The resulting change in reflection density is measured at 2-time points. The difference in reflection density is proportional to the concentration of creatinine present in the sample
VITROS TBIL Slides (Colorimetric, dual-wavelength endpoint):
A drop of patient sample is deposited on the slide and is evenly distributed by the spreading layer to the underlying layers. This layer provides a reflective background for measuring the diazo products of bilirubin and contains all reagents necessary to determine total bilirubin. The method uses dyphylline to dissociate unconjugated bilirubin from albumin. Unconjugated bilirubin, conjugated bilirubin, and albumin-linked bilirubin (delta) subsequently react with the diazonium salt 4-(N-carboxymethylsulfonyl)benzenediazonium hexafluorophosphate to produce azobilirubin chromophores that have similar molar absorptivity and absorbance maxima around 520 nm. The concentration of total bilirubin is determined by measuring the azobilirubin chromophores at two wavelengths through the transparent support. The reflectance measurement at 460 nm corrects for spectral interferences.
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
##### a. Precision/Reproducibility:
Precision studies were conducted following EP05-A3 guideline. The study was performed by testing a minimum of two quality control fluids and three human serum pools using the Carbamazepine (CRBM), Creatinine (CREA), and Total Bilirubin (TBIL) assays. Samples were analyzed using one VITROS XT 7600 Integrated System over 20 days, with 2 runs per day and 2 replicates per specimen (n=80). The results of the precision studies are shown in the tables below:
8
{8}
Carbamazepine Serum:
| Sample | Mean (μg/mL) | Repeatability (Within Run) | | Within Day | | Within Lab (Total) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| Serum Pool 1 | 3.9 | 0.13 | 3.30 | 0.13 | 3.30 | 0.16 | 3.98 |
| QC -1 | 4.7 | 0.14 | 2.97 | 0.14 | 3.02 | 0.16 | 3.32 |
| QC -2 | 10.1 | 0.24 | 2.38 | 0.25 | 2.46 | 0.28 | 2.76 |
| Serum Pool 3 | 11.6 | 0.30 | 2.57 | 0.35 | 2.98 | 0.39 | 3.37 |
| QC -3 | 13.1 | 0.29 | 0.35 | 0.35 | 2.67 | 0.41 | 3.14 |
| Serum Pool 5 | 17.6 | 0.33 | 0.37 | 0.37 | 2.10 | 0.42 | 2.41 |
Creatinine Serum:
| Sample | Mean (mg/dL) | Repeatability (Within Run) | | Within Day | | Within Lab (Total) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| QC -1 | 0.82 | 0.006 | 0.713 | 0.009 | 1.050 | 0.013 | 1.579 |
| Serum Pool 1 | 0.88 | 0.006 | 0.642 | 0.007 | 0.743 | 0.016 | 1.850 |
| S Pool (native) | 0.99 | 0.007 | 0.697 | 0.008 | 0.829 | 0.17 | 1.720 |
| QC -2 | 5.39 | 0.039 | 0.716 | 0.049 | 0.914 | 0.088 | 1.624 |
| Serum Pool 2 | 9.63 | 0.057 | 0.594 | 0.070 | 0.729 | 0.135 | 1.400 |
| Serum Pool 5 | 12.65 | 0.109 | 0.865 | 0.109 | 0.071 | 0.169 | 1.337 |
Creatinine Urine:
| Sample | Mean (mg/dL) | Repeatability (Within Run) | | Within Day | | Within Lab (Total) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| U QC -1 | 55.6 | 0.58 | 1.05 | 0.67 | 1.20 | 1.24 | 2.23 |
| Urine Pool 1 | 78.4 | 1.10 | 1.40 | 1.13 | 1.44 | 1.36 | 1.73 |
| U Pool (native) | 88.0 | 0.69 | 0.79 | 0.98 | 1.11 | 1.41 | 1.60 |
| U QC -2 | 131.2 | 1.67 | 1.27 | 1.89 | 1.44 | 2.71 | 2.06 |
| Urine Pool 4 | 251.8 | 1.99 | 0.79 | 2.12 | 0.84 | 3.90 | 1.55 |
| Urine Pool 5 | 320.9 | 2.99 | 0.93 | 3.49 | 1.09 | 5.51 | 1.72 |
Total Bilirubin serum:
| Sample | Mean (mg/dL) | Repeatability (Within Run) | | Within Day | | Within Lab (Total) | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | SD | %CV | SD | %CV | SD | %CV |
| Serum Pool 1 | 0.3 | 0.01 | 4.57 | 0.02 | 5.20 | 0.02 | 6.72 |
| QC-1 | 1.6 | 0.03 | 1.68 | 0.04 | 2.88 | 0.05 | 3.45 |
| Serum Pool 3 | 6.5 | 0.03 | 0.52 | 0.06 | 0.87 | 0.09 | 1.40 |
| QC -2 | 15.3 | 0.10 | 0.67 | 0.16 | 1.05 | 0.24 | 1.60 |
| Serum Pool 5 | 21.6 | 0.17 | 0.77 | 0.20 | 0.92 | 0.34 | 1.58 |
9
{9}
# b. Linearity/assay reportable range:
Linearity studies were performed according to CLSI EP06-A guideline. A series of eleven proportionally related admixtures of low and high levels samples were tested to verify linearity; each sample was tested in duplicate.
The results of the linearity studies support the following claimed measuring ranges for the VITROS CRBM, VITROS CREA, and VITROS TBIL assays:
| Analyte | Slope | Intercept | Correlation Coefficient | Range Tested | Claimed Measuring range |
| --- | --- | --- | --- | --- | --- |
| Carbamazepine | 1.00 | 0.27 | 1.00 | 2.1 - 22.9 | 3.0 - 20.0 μg/mL |
| Serum Creatinine | 1.00 | -0.04 | 1.00 | 0.08 - 15.8 | 0.15 - 14.0 mg/dL |
| Urine Creatinine | 1.01 | -0.16 | 1.00 | 1.6 - 384.7 | 3.2 - 346.5 mg/dL |
| Total bilirubin | 1.00 | 0.29 | 1.00 | 0.00 - 27.26 | 0.10 - 27.00 mg/dL |
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
VITROS Chemistry Products CRBM Slides: Values assigned to the VITROS Chemistry Products Calibrator Kit 9 for carbamazepine are traceable to the Certified NIST (National Institute of Standards and Technology) Reference Material, SRM (Standard Reference Material) 1599.
VITROS Chemistry Products CREA Slides: The values assigned to the VITROS Chemistry Products Calibrator Kit 1 for Creatinine are traceable to a Gas Chromatography Isotope Dilution Mass Spectrometry (GC/IDMS) method 8 and National Institute of Standards and Technology (NIST) SRM 914 creatinine standard reference material.
VITROS Chemistry Products TBIL Slides: Values assigned to the VITROS Chemistry Products Calibrator Kit 4 for total bilirubin are traceable to the Certified NIST (National Institute of Standards and Technology) Reference Material, SRM (Standard Reference Material) 916.
# d. Detection limit:
Detection capability studies for each analyte were evaluated based upon CLSI EP17-A2 guideline.
Limit of blank (LoB) studies were performed by testing 4 blank samples. Samples were tested in replicates of 6 over 3 days, using 3 lots of reagents, 4 samples every day, for a total of 216 observations (72 results per reagent lot). The LoB value for each assay was defined as the highest value achieved using blank samples with the stated probability (i.e. α = 5%). Since the data for all assays were non-gaussian, a
10
{10}
non-parametric approach was applied that estimates the LoB using the calculated rank position corresponding to the 95th percentile of the distribution of blank values observed.
Limit of detection (LoD) studies were performed by testing 4 pools of human samples with analyte concentrations close to the expected detection limit for each analyte. Samples were tested in replicates of 6 over 3 days, using 3 lots of reagents, with the 4 human sample pools every day, for a total of 216 observations (72 results per reagent lot). The LoD was calculated using a pooled SD from the low level sample results and the input LoB value for the assay, determined as described above. The LoD value for the assay was defined as the highest resultant value achieved among the combinations of reagent lots and human pools evaluated, with the stated probability (i.e. β= 5%).
Limit of quantitation (LoQ) studies were performed using 4 pools of low level samples with analyte concentrations close to the expected LoQ of the corresponding assay. Samples were tested in replicates of 4 over 3 days, using 3 lots of reagents, 4 samples every day, for a total of 144 observations (48 results per reagent lot). Ortho defines LoQ as the lowest concentration with a percent total allowable error ≤19% for carbamazepine; percent total allowable error ≤30% for creatinine in serum and urine, and total allowable error ≤0.09 mg/dL for total bilirubin in serum.
The results of the detection capability studies for each assay are presented in the table below.
| | CARB (μg/mL) | TBIL (mg/dL) | Creatinine (mg/dL) | |
| --- | --- | --- | --- | --- |
| | | | Serum/plasma | Urine |
| LoB | 0.6108 | 0.0378 | 0.0933 | 1.9973 |
| LoD | 0.6821 | 0.0722 | 0.0991 | 2.1986 |
| LoQ | 2.6860 | 0.0616 | 0.1119 | 2.0060 |
| Claimed LoQ | 3.0 | 0.10 | 0.15 | 3.2 |
| Assay Claimed Range | 3.0-20.0 | 0.10-27.00 | 0.15-14.0 | 3.2-346.5 |
# e. Analytical specificity:
Interference studies were performed in accordance with CLSI EP07-A2 guideline, by using pool of human serum, plasma or urine (samples may have been spiked or diluted to achieve the appropriate target analyte concentration). Each representative assay was tested at a minimum of two analyte concentrations in 6 replicates on one lot of reagent or multiple lots if needed. Bias was evaluated between interferent test and control samples and compared to the Maximum Allowable Interference (MAI) for each analyte. The substances that were tested included known chemical interferents, common chemical substances identified with potential to interfere based upon risk assessment.
11
{11}
### Carbamazepine:
The study was conducted using samples with carbamazepine concentrations of 3 and \(12\mu \mathrm{g / mL}\). The sponsor defined non-interference as a bias \(< 1.4~\mu \mathrm{g / mL}\) (vs. control condition). The compounds and the highest concentration that don't interfere with the carbamazepine assay are listed in the table below:
| Compound | Concentration |
| --- | --- |
| Acetaminophen | 20 mg/dL |
| Amitriptyline | 1 μg/mL |
| Amobarbital | 100 μg/mL |
| Amoxicillin | 1500 μg/mL |
| Ascorbic acid | 6 mg/dL |
| Atenolol | 20 μg/mL |
| Caffeine | 10 mg/dL |
| Cephalexin | 400 μg/mL |
| Chlordiazepoxide | 2 mg/dL |
| Chlorpromazine | 10 μg/mL |
| Cholesterol | 450 mg/dL |
| Cimetidine | 100 μg/mL |
| Ciprofloxacin | 5 mg/dL |
| Clomipramine | 3 μg/mL |
| Clonazepam | 1 μg/mL |
| Clorazepate | 7 μg/mL |
| Codeine | 4 μg/mL |
| Creatinine | 30 mg/dL |
| Desipramine | 2.5 μg/mL |
| Dextromethorphan | 3.8 μg/mL |
| Diazepam | 20 μg/mL |
| Digoxin | 30 ng/mL |
| Diltiazem | 5 μg/mL |
| Diphenhydramine | 10 μg/mL |
| Doxepin | 1.5 μg/mL |
| Enalapril | 1.2 μg/mL |
| Ethanol | 200 mg/dL |
| Ethosuximide | 500 μg/mL |
| Felbatol | 30 mg/dL |
| Fluoxetine | 8 μg/mL |
| Furosemide | 10 mg/dL |
| Gentamicin | 120 μg/mL |
| Glucose | 1200 mg/dL |
| Glutethimide | 6 mg/dL |
| Compound | Concentration |
| --- | --- |
| HPPH | 200 μg/mL |
| Hydrochlorothiazide | 2 mg/dL |
| Ibuprofen | 70 mg/dL |
| Imipramine | 0.1 mg/dL |
| Intralipid | 800 mg/dL |
| Lithium | 15 mmol/L |
| Lorazepam | 1 μg/mL |
| Mephobarbital | 200 μg/mL |
| Methsuximide | 40 mg/dL |
| Metoprolol | 3.4 μg/mL |
| Naproxen | 900 μg/mL |
| Nifedipine | 2 μg/mL |
| Nortriptyline | 5 mg/dL |
| PEMA | 40 mg/dL |
| Pentobarbital | 100 μg/mL |
| Phenobarbital | 250 μg/mL |
| Phenytoin | 100 μg/mL |
| Prednisone | 1 μg/mL |
| Primidone | 200 μg/mL |
| Promazine | 0.09 mg/dL |
| Promethazine | 10 μg/mL |
| Protriptyline | 2.5 μg/mL |
| Pseudoephedrine | 20 μg/mL |
| Ranitidine | 20 μg/mL |
| Salicylate, Na salt | 50 mg/dL |
| Secobarbital | 100 μg/mL |
| Sodium bromide | 120 mg/dL |
| Theophylline | 25 mg/dL |
| Tobramycin | 12 mg/dL |
| Total Protein | 4.0–9.0 g/dL |
| Triglycerides | 1100 mg/dL |
| Valproic acid | 500 μg/mL |
| Vancomycin | 300 μg/mL |
| Verapamil | 90 μg/mL |
12
{12}
| Compound | Concentration |
| --- | --- |
| Glyburide | 6.4 μg/mL |
| Guaifenesin | 100 mg/dL |
| Compound | Concentration |
| --- | --- |
| Warfarin | 100 μg/mL |
Cross-Reactivity:
Carbamazepine 10,11-epoxide, the major active metabolite of carbamazepine, was tested for cross-reactivity at several concentrations as listed in the table.
| Epoxide Conc. | Carbamazepine Concentration (μg/mL) | % Cross Reactivity |
| --- | --- | --- |
| 2.5 | 0.0 | 11.2 |
| | 5.2 | 10.5 |
| | 10.1 | 7.1 |
| 10.0 | 0.0 | 9.3 |
| | 5.2 | 8.1 |
| | 10.1 | 7.7 |
The substances that interfere with carbamazepine determinations in serum are summarized in the table below:
| Interferent | The highest concentrations of known interfering substances |
| --- | --- |
| Ethamsylate | 3.0 mg/dL |
| Gentisic acid | 5.0 mg/dL |
| N-acetylcysteine | 100.0 mg/dL |
| Bilirubin(conjugated) | 20.0 mg/dL |
| Hemoglobin | 200.0 mg/dL |
| Ethanol | 394.0 mg/dL |
The sponsor included in the labeling the following limitations for carbamazepine:
- Specimens with low total protein <4.0 g/dL (<40.0 g/L) may give a negative bias greater than -1.4 μg/mL (-5.9 μmol/L).
- Specimens with an elevated total protein >9.0 g/dL (>90.0 g/L) may give a positive bias greater than +1.4 μg/mL (+5.9 μmol/L).
# Creatinine:
The study was conducted using a minimum of 2 samples with creatinine concentrations of approximately 1.50 and 5.00 mg/dL. The sponsor defined non-interference as a bias <0.13 mg/dL (vs. control condition). The compounds and the highest concentration that don't interfere with the creatinine assay are listed in the table below:
13
{13}
| Compound | Concentration |
| --- | --- |
| Acetaminophen | 400 μg/mL |
| Acetoacetate | 30 mmol/L |
| Ampicillin | 1.5 mg/dL |
| Amikacin | 1.5 mg/dL |
| Ammonium Chloride | 1 mmol/L |
| Amphotericin B | 1.5 mg/dL |
| Ascorbic Acid | 3 mg/dL |
| Bacitracin | 1.5 mg/dL |
| Bicarbonate | 40 mmol/L |
| Bilirubin | 20 mg/dL |
| Bleomycin Sulfate | 1.5 mg/dL |
| Carbenicillin | 1.5 mg/dL |
| Cefazolin | 1.5 mg/dL |
| Cephalothin | 1.5 mg/dL |
| Cephaloridine | 1.5 mg/dL |
| Cephaloglycin | 1.5 mg/dL |
| Cephalexin | 1.5 mg/dL |
| Cephardine | 1.5 mg/dL |
| Cleocin | 1.5 mg/dL |
| Cloxacillin | 1.5 mg/dL |
| Demeclocycline | 1.5 mg/dL |
| Dextran | 1000 mg/dL |
| Dicloxacillin | 1.5 mg/dL |
| Doxycycline | 1.5 mg/dL |
| Di-cycloserine | 1.5 mg/dL |
| Dilantin | 2 mg/dL |
| Ethambutol | 1.5 mg/dL |
| Ethanol | 300 mg/dL |
| Furazolidone | 1.5 mg/dL |
| 5-Flurocytosine | 5 mg/dL |
| Gentamicin | 1.5 mg/dL |
| Glucose | 600 mg/dL |
| Glutathione | 1 mg/dL |
| Hypaque | 500 mg/dL |
| Intralipid | 800 mg/dL |
| Kanamycin | 1.5 mg/dL |
| Compound | Concentration |
| --- | --- |
| Isoniazid | 1.5 mg/dL |
| Limcomycin | 1.5 mg/dL |
| Methicillin | 1.5 mg/dL |
| 6-Mercaptopurine | 1.5 mg/dL |
| Minocycline | 1.5 mg/dL |
| Nalidixic Acid | 1.5 mg/dL |
| Nafcillin | 1.5 mg/dL |
| Neomycin | 1.5 mg/dL |
| Nitrofurantoin | 1.5 mg/dL |
| Oxacillin | 1.5 mg/dL |
| Oxytetracycline | 1.5 mg/dL |
| Penicillin-g | 1.5 mg/dL |
| Phenobarbital | 3 mg/dL |
| Phenoxymethyl-penicillinic acid | 1.5 mg/dL |
| pH | 6.8 |
| pH | 8.8 |
| Polymyxin B sulfate | 1.5 mg/dL |
| Polymyxin E | 1.5 mg/dL |
| Potassium | 8 mEq/L |
| Rifampicin | 1.5 mg/dL |
| Spectinomycin | 1.5 mg/dL |
| Streptomycin sulfate | 1.5 mg/dL |
| Sulfachloropyridazine | 1.5 mg/dL |
| Sulfamethoxypyridazine | 1.5 mg/dL |
| Sulfamethoxazole | 1.5 mg/dL |
| Sulfisoxazole | 1.5 mg/dL |
| Sulfadiazine | 1.5 mg/dL |
| Sulfathiazole | 6 mg/dL |
| Tetracycline | 1.5 mg/dL |
| Ticarcillin | 1.5 mg/dL |
| Tolbutamide | 22 mg/dL |
| Triglycerides | 800 mg/dL |
| Vancomycin | 1.5 mg/dL |
| Urea Nitrogen | 100 mg/dL |
| Uric Acid | 15 mg/dL |
The results of substances that interfere with creatinine determinations in serum and plasma are summarized in the table below:
14
{14}
| Interferent | Interferent concentrations (mg/dL) | Comments | Creatinine concentrations tested(mg/dL) | Bias |
| --- | --- | --- | --- | --- |
| Dipyrone (Metamizol) | 18.0 | 3X IV Therapeutic (Based on 1000 mg intravenous dose) | 1.2 | -0.24 |
| | | | 4.2 | -0.73 |
| Tolazamides | 4.5 mg/d | 3X oral Therapeutic | 3.5 | -0.39 |
Urine Creatinine:
The following preservatives have been tested and demonstrated an effect of less than 2% on creatinine results:
- Thymol
- Toluene
- Boric acid
- Glacial acetic acid (0.5 to 1.0 v/v)
- 12N HCl
- NH4OH
- Bromide
- Iodide
- 5% NaOH
The sponsor included in the labeling the following limitations for creatinine determinations in serum:
- Creatine: At a creatinine concentration of 1.5 mg/dL (133 \( \mu \) mol/L), creatine greater than 8 mg/dL (707 \( \mu \) mol/L) will be flagged with a DP code (because highly elevated creatine concentrations may cause excessive background density). For unflagged samples, residual bias because of creatine will be less than 0.15 mg/dL (13 \( \mu \) mol/L). At a creatinine concentration of 14 mg/dL (1237 \( \mu \) mol/L), creatine greater than 1 mg/dL (88 \( \mu \) mol/L) will be flagged with a DP code.
- Proline: Patients receiving hyperalimentation fluids containing proline may show an increase of 0.2 mg/dL (18 \( \mu \) mol/L). Do not collect specimens from intravenous fluid lines contaminated with hyperalimentation fluid.
- Lidocaine: Patients on long-term lidocaine therapy may show an increase of up to 1.0 mg/dL (88 \( \mu \) mol/L) due to a metabolite of lidocaine, N-ethyl glycine (NEG).
### Total Bilirubin:
The study was conducted using a minimum of 2 samples with total bilirubin concentrations of approximately 0.3 and \(15\mathrm{mg / dL}\). The sponsor defined non-interference as a bias \(< 0.2\mathrm{mg / dL}\) (vs. control condition). The compounds and the highest concentration that don't interfere with the total bilirubin assay are listed in the table below:
15
{15}
| Compound | Concentration |
| --- | --- |
| Acetaminophen | 5 mg/dL |
| Acetylsalicylic acid | 30 mg/dL |
| 5-Aminosalicylic acid | 40 μg/mL |
| Ascorbic acid | 3 mg/dL |
| Beta-carotene | 0.6 mg/dL |
| Bile acids | 6 mg/dL |
| Calcium | 4 mmol/L |
| Calcitrol | 80 μg/mL |
| Cefamandole nafate | 533 μg/mL |
| Cyclosporin-a | 238 μg/mL |
| Chloride | 120 meq/L |
| Dextran | 1000 mg/dL |
| Doxycycline | 9 μg/mL |
| Ethanol | 300 mg/dL |
| Free fatty acids | 3 mmol/L |
| Gentamicin | 5 μg/mL |
| Gentisic acid | 0.5 mg/dL |
| Glucose | 600 mg/dL |
| Hypaque | 500 mg/dL |
| Intralipid | 800 mg/dL |
| Compound | Concentration |
| --- | --- |
| L-dopa | 6 μg/mL |
| Liposyn | 10% |
| Magnesium | 1.85 mmol/L |
| Minoxidil | 2 μg/mL |
| Nafcillin | 30 μg/mL |
| Piroxicam | 10 μg/mL |
| Propanolol | 2 μg/mL |
| Rifampin | 14 μg/mL |
| Sulfapyridine | 40 μg/mL |
| Sulfasalazine | 38 μg/mL |
| Sulfathiazole | 60 μg/mL |
| Sulfobromophthalein | 150 μg/mL |
| Sulfisoxazole | 60 μg/mL |
| Tetracycline | 30 μg/mL |
| Tobramycin | 5 μg/mL |
| Total protein | 10 g/dL |
| Triglycerides | 800 mg/dL |
| Urea Nitrogen | 100 mg/dL |
| Vitamin A | 50 IU/mL |
| Vitamin K1 | 8.3 μg/mL |
The substances listed below, when tested at the concentrations indicated, caused the bias shown.
| Interferent | Interferent concentrations | Bilirubin concentrations tested (mg/dL) | Bias |
| --- | --- | --- | --- |
| Levodopa | 300 μg/mL | 0.3 | +0.9 |
| | | 17.1 | -9.8 |
| 4-Aminosalicylic acid | 8 mg/dL | 0.3 | -0.4 |
| | | 16 | +2.5 |
| Phenazopyridine | 8 mg/dL | 2.5-6.3 | +4.4 |
| Biliverdin | 4 mg/dL | 15 | +0.6 |
| Hemoglobin | 150.0 mg/dL | 1.1 | +0.3 |
The sponsor included in the labeling the following limitations for total bilirubin determinations:
- Cefotiam (Pansporin) has been reported to show very large positive biases on TBIL results. This drug is normally cleared through the kidney. Biases will be largest in specimens from patients with renal insufficiency and may be as large as 5 mg/dL (86 \( \mu \) mol/L). Because the VITROS BuBc Slide is not affected by Cefotiam, patients known to be receiving Cefotiam therapy should be monitored only with the BuBc Slide.
- Drugs and other compounds that are diazo-reactive or that absorb light in the vicinity of 540 and 460 nm may interfere.
16
{16}
- Certain drugs and clinical conditions are known to alter total bilirubin concentration in vivo.
f. Assay cut-off:
Not applicable.
# 2. Comparison studies:
a. Method comparison with predicate device:
Method comparison studies were conducted by testing a minimum of 116 human serum samples with analyte concentrations across the analytical ranges of carbamazepine, creatinine and total bilirubin assays on the VITROS XT 7600 Integrated System and the VITROS 5600 Integrated System (predicate device). In addition, 125 human urine samples were tested for creatinine on the candidate and predicate test systems. The results of the regression analyses for each of the assays are summarized below:
| Assay | N | Regression Analysis | Slope | Intercept | Correlation Coefficient | Tested range | Claimed Measuring Range |
| --- | --- | --- | --- | --- | --- | --- | --- |
| CRBM Serum (μg/mL) | 118 | Deming | 1.00 | 0.12 | 1.00 | 3.1 - 17.8 | 3.0 - 20.0 |
| CREA Serum (mg/dL) | 116 | Passing Bablok | 0.99 | 0.00 | 1.00 | 0.25 - 13.4 | 0.15 - 14.0 |
| CREA Urine (mg/dL) | 122 | Passing Bablok | 0.99 | -0.45 | 1.00 | 3.7 - 331.0 | 3.2 - 346.5 |
| TBIL Serum (mg/dL) | 125 | Passing Bablok | 0.99 | 0.01 | 1.00 | 0.14 - 23.65 | 0.10 - 27.00 |
b. Matrix comparison:
Based on the sponsor's risk assessment for adding the VITROS Chemistry Products CRBM Slides, VITROS Chemistry Products CREA Slides and VITROS Chemistry Products TBIL Slides to the VITROS 7600 System and the results of the analytical testing conducted that demonstrated there was no impact on the assays' performance characteristics, new matrix comparison studies were not conducted. FDA found this justification to be acceptable.
The VITROS Chemistry Products CRBM Slides can be used with serum and plasma
17
{17}
samples collected in lithium heparin, sodium heparin and sodium citrate.
The VITROS Chemistry Products CREA Slides can be used with serum, lithium heparin plasma, and urine.
The VITROS Chemistry Products TBIL Slides can be used with serum and lithium heparin plasma.
### 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
### 4. Clinical cut-off:
Not applicable.
### 5. Expected values/Reference range:
| Assay | Range |
| --- | --- |
| Carbamazepine^{1} | 4.0–12.0 (μg/mL) |
| Serum Creatinine^{2} | Male 0.66–1.25 mg/dL |
| | Female 0.52–1.04 mg/dL |
| Urine Creatinine^{3} | Male 1000–2000 mg/day^{4} |
| | Female 800–1800 mg/day^{4} |
| Total Bilirubin^{5} | 0.2-1.3 mg/dL |
1. Tietz NW (ed). Fundamentals of Clinical Chemistry. ed. 5. Philadelphia: WB Saunders; 1019; 2001.
2. The serum reference intervals for creatinine were established as the central 95% of results from an external study of apparently healthy adults (serum: 180 males and 180 females).
3. McPherson R, Pincus M (eds.). Henry's Clinical Diagnostics and Management by Laboratory Methods, 21st edition. Philadelphia: Saunders Elsevier [ISBN-13: 978-1-4160-0287-1; 1410; 2006.
4. Urine Creatinine calculation: Creatinine concentration (mg/dL) x 24-hour volume (dL) = mg/day.
18
{18}
5. The reference interval for total bilirubin is the central 95% of results from a study of 110 apparently healthy adults with normal liver enzymes (85 females and 25 males).
# N. Instrument Name:
VITROS XT 7600 Integrated System
# O. System Descriptions:
1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes __X__ or No ______
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ______ or No __X__
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes __X__ or No ______
3. Specimen Identification:
Specimens that are labeled with a barcode will be scanned in automatically when placed into the Reagent and Sample Manager area or manually using the bar code scanner.
4. Specimen Sampling and Handling:
The Sample Handler (SAHA) uses the universal sample tray to house patient sample containers. Once samples are loaded onto the System, they are advanced into a metering zone, where a bar code reader scans the bar code labels to identify tray and samples.
5. Calibration:
Calibration is an automatic process: bar-coded calibrators or manually programmed trays are placed on the system for assay calibration, the calibrators are processed in a random-access format (bar-coded) or in the order they were defined (manual), and a new calibration curve is calculated. The new curve is saved on the system and used for
19
{19}
subsequent assays to determine the analyte concentrations in the patient and control samples.
# 6. Quality Control:
To perform Quality Control, QC materials are run with either known, or unknown values along with patient samples to determine whether the system is functioning within the established ranges.
# **P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above:**
Not applicable.
# **Q. Proposed Labeling:**
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
# **R. Conclusion:**
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.